← Product Code [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK) · K211645

# Epione (K211645)

_Quantum Surgical Sas · JAK · Feb 24, 2022 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K211645

## Device Facts

- **Applicant:** Quantum Surgical Sas
- **Product Code:** [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK.md)
- **Decision Date:** Feb 24, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The EPIONE device is a user controlled, stereotactic accessory intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous procedures. During the planning phase, the desired instrument placement and performance is defined relative to the target anatomy. During the guidance phase, the device enables to monitor respiratory levels and verify patient position prior to instrument advancement. During the assessment phase, the achieved instrument and performance are displayed relative to the previously defined plan through an overlay of the pre- and post-treatment image data. The device is indicated for use in liver ablation procedures under general anesthesia with rigid straight instruments such as needles and probes of diameters ranging from 11G to 19G by physicians trained for CT procedures.

## Device Story

Epione is a robotic, stereotactic accessory for CT-guided percutaneous liver ablation. System components include a robot cart with a robotic arm and force sensor for hand-guided positioning, a display cart with touchscreen, a navigation cart with camera, and Epionet software. Input: CT images and optical tracking of patient markers. Operation: Physician defines trajectory on CT images; robot arm aligns needle guide; user manually advances instrument. System monitors respiratory levels via patient reference markers to ensure alignment during breath-hold. Output: Visual overlay of planned vs. achieved instrument position on CT images. Used in clinical settings by trained physicians. Benefits: Enhanced accuracy and verification of needle placement relative to target anatomy during ablation procedures.

## Clinical Evidence

Multi-center, prospective, interventional clinical study on 21 patients undergoing liver ablation under general anesthesia. Primary endpoints: safety and performance. Results: No adverse events reported; device demonstrated safety and effectiveness for CT-guided percutaneous liver ablation.

## Technological Characteristics

Robotic arm with force sensor; optical tracking system for patient/instrument registration; touchscreen interface. Materials: Biocompatible per ISO 10993. Standards: ASTM F2554-10 (positional accuracy), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 62366-1 (usability). Software: Epionet workflow assistance. Sterilization: Validated reprocessing for instruments.

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- MAXIO ([K132108](/device/K132108.md))

## Reference Devices

- CAS-One IR ([K152473](/device/K152473.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text on the right. The symbol on the left is a stylized representation of a human figure. The text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters. The words "U.S. FOOD & DRUG" are on the top line, and the word "ADMINISTRATION" is on the bottom line.

February 24, 2022

Quantum Surgical SAS % Elise Lagacherie Ouality Assurance and Regulatory Affairs Director ZAC Eureka 1000 rue du Mas de Verchant Montpellier, 34000 FRANCE

Re: K211645

Trade/Device Name: Epione Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-Ray System Regulatory Class: Class II Product Code: JAK Dated: January 19, 2022 Received: January 21, 2022

Dear Elise Lagacherie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Laurel Burk Assistant Director Diagnostic X-ray Systems Team Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K211645

Device Name EPIONE

#### Indications for Use (Describe)

The EPIONE device is a user controlled, stereotactic accessory intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous procedures.

During the planning phase, the desired instrument placement and performance is defined relative to the target anatomy. During the guidance phase, the device enables to monitor respiratory levels and verify patient position prior to instrument advancement. During the assessment phase, the achieved instrument and performance are displayed relative to the previously defined plan through an overlay of the pre- and post-treatment image data.

The device is indicated for use in liver ablation procedures under general anesthesia with rigid straight instruments such as needles and probes of diameters ranging from 11G to 19G by physicians trained for CT procedures.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Quantum Surgical. The logo consists of an orange geometric shape on the left, resembling a stylized letter "Q". To the right of the shape, the words "QUANTUM" and "surgical" are written in gray, with "QUANTUM" on top of "surgical". The logo appears to be for a medical or technology company.

# K211645 510(k) SUMMARY

Date prepared: 23, February 2022

#### Submitter

## Quantum Surgical SAS

1000 rue du mas de Verchant 34000 Montpellier France

## Contact Person:

Elise Lagacherie, Quality & Requlatory Affairs Director e.lagacherie@quantumsurgical.com

#### Device:

Name of Device: EPIONE™ Common or Usual Name: Computer assisted Surgical System. Classification Name: Computed tomography X-ray system (21 CFR 892.1750) Requlatory Class: II Product Code: JAK

#### Predicate Device

MAXIO (JAK, K132108), Perfint Healthcare Pvt. Ltd.,

#### Reference Device:

CAS-One IR (JAK, K152473), CAScination AG

#### Device Description

The EPIONE device is a user controlled, stereotactic accessory intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous procedures.

During the planning phase, the desired instrument placement and performance is defined relative to the tarqet anatomy. During the quidance phase, the device enables to monitor respiratory levels and verify patient position prior to instrument. During the assessment phase, the achieved instrument placement and performance are displayed relative to the previously defined plan through an overlay of the pre- and post-treatment image data.

The device is indicated for use in liver ablation procedures under general anesthesia with rigid straight instruments such as needles and probes of diameters ranging from 11G to 19G by physicians trained for CT procedures.

The EPIONE device consists in the following components:

| Robot cart | This mobile equipment can be moved in and out of the intervention room and is positioned next to the patient. The cart carries a robotic arm including a force sensor assembly allowing handguiding of the robotic arm by the user. The robotic arm movements are enabled by the user using a footswitch. The robot cart also embeds the electronic systems required to power and operate the robot arm. |
|------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for Quantum Surgical. The logo consists of an orange geometric shape on the left and the words "QUANTUM surgical" on the right. The word "QUANTUM" is in orange, and the word "surgical" is in gray.

| Display cart       | This mobile equipment can be moved in and out of the intervention room and is positioned next to the operator. The cart carries a touchscreen to operate the system.                          |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Navigation cart    | This mobile equipment can be moved in and out of the intervention room and is positioned next to the table. The cart carries a navigation camera.                                             |
| EPIONET™ software  | A software provides the step-by-step workflow assistance for intervention planning and intraoperative positioning of the robotic arm for instruments placement and post-operative assessment. |
| EPIONE instruments | Needle guide<br>Patient reference<br>Navigation probe                                                                                                                                         |

#### Indications for Use

The EPIONE device is a user controlled, stereotactic accessory intended to assist in the planning and manual advancement of one or more instruments, as well as in verification of instrument position during Computed Tomography (CT) guided percutaneous procedures.

During the planning phase, the desired instrument placement and performance is defined relative to the target anatomy. During the guidance phase, the device enables to monitor respiratory levels and verify patient position prior to instrument advancement. During the assessment phase, the achieved instrument placement and performance are displayed relative to the previously defined plan through an overlay of the pre- and post-treatment image data.

The device is indicated for use in liver ablation procedures under general anesthesia with rigid straight instruments such as needles and probes of diameters ranging from 11G to 19G by physicians trained for CT procedures.

|                                            | EPIONE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | MAXIO<br>predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|--------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use /<br>Indication for Use       | The EPIONE device is a user controlled,<br>stereotactic accessory intended to assist in the<br>planning and manual advancement of one or<br>more instruments, as well as in verification of<br>instrument position during Computed<br>Tomography (CT) guided percutaneous<br>procedures.<br>During the planning phase, the desired<br>instrument placement and performance is<br>defined relative to the target anatomy. During<br>the guidance phase, the device enables to<br>monitor respiratory levels and verify patient<br>position prior to instrument advancement.<br>During the assessment phase, the achieved<br>instrument placement and performance are<br>displayed relative to the previously defined plan<br>through an overlay of the pre- and post-<br>treatment image data.<br>The device is indicated for use in liver ablation<br>procedures under general anesthesia with rigid<br>straight instruments such as needles and probes<br>of diameters ranging from 11G to 19G by<br>physicians trained for CT procedures. | MAXIO is a user controlled, stereotactic<br>accessory intended to assist in the planning and<br>manual advancement of one or more<br>instruments during Computed Tomography (CT)<br>guided percutaneous procedures.<br>MAXIO permits physician verification of patient<br>position prior to needle advancement and<br>monitoring of respiratory levels during the<br>procedure.<br>Image registration and overlay tools available in<br>MAXIO are intended to provide guidance to the<br>user during planning and instrument placement.<br>MAXIO is indicated for use with rigid straight<br>instruments such as needles and probes used in<br>Computed Tomography (CT) guided<br>percutaneous interventional procedures<br>performed by physicians trained for CT<br>procedures on organs and anatomical structures<br>in the thorax, abdomen and pelvis. |
| Patient Anesthesia<br>Conditions           | For use under general anesthesia conditions                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | For use under local to general anesthesia<br>conditions                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                            | EPIONE                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | MAXIO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Intra-interventional<br>Planning           | Physician defines trajectory per an entry point<br>and a target point on CT images.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Physician defines trajectory per an entry point<br>and a target point on CT images.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Interventional<br>instruments              | Rigid straight interventional instruments such as<br>needles and probes for ablation from 11G to<br>19G.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Rigid straight interventional instruments such as<br>needles, probes for biopsy, ablation and<br>drainage from 11G to 21G.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Needle<br>configuration and<br>performance | • Needle selection among a list of predefined<br>needles<br>• Manufacturer defined performances (e.g.<br>ablation zones).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | • Needle selection among a list of predefined<br>needles<br>• Manufacturer defined performances (e.g.<br>ablation zones).<br>• User-defined performances (e.g. ablation<br>zones).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Needle guidance                            | • Robotic arm with needle guide is positioned<br>on the desired trajectory as was planned on<br>the intra-interventional CT images.<br>• Instrument (e.g. needle) is then manually<br>advanced through the guide on the planned<br>trajectory.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | • Robotic arm with needle guide instrument is<br>positioned on the desired trajectory as was<br>planned on the intra-interventional CT images.<br>• Instrument (e.g. needle) is then manually<br>advanced through the guide on the planned<br>trajectory.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Patient registration                       | Automatic patient location to image registration<br>per array of markers on a patient reference<br>instrument placed on the patient's skin and<br>included in the CT-scan images and which are<br>located using an optical tracking system.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | The location of the robotic workstation is located<br>to a fixed established reference position relative<br>to the C-arm and bed per the use of a docking<br>mat and calibration process performed during<br>the initial workstation installation.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Patient<br>Registration<br>verification    | The navigation probe is pointed to pre-identified<br>points on the patient (skin markers) to verify<br>their relative locations.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Skin markers are placed on the patient and<br>included in the CT-scan and located when<br>planning. Then before needle placement, it is<br>verified using a laser system that the patient has<br>not moved by verifying that the locations of the<br>skin markers are still in the same location as<br>when scanned. In addition, table markers on the<br>C-arm are also used to verify also per a laser<br>cross-hairs that the relative locations of the C-<br>arm and robot workstation have not changed<br>from the original installation.                                                                                                                                                                                                                                                                                                             |
| Respiratory Motion<br>Management           | Respiratory motion control using breath-hold<br>Measurement of apnea reference level during<br>the intra-interventional planning CT imaging.<br>Live display of respiratory level using the markers<br>/ patient reference placed on the patient and<br>visual comparison with the reference level<br>during needle guidance.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Respiratory motion control using breath-hold.<br>Measurement of apnea reference level during<br>the intra-interventional planning CT imaging.<br>Live display of respiratory level using a<br>respiratory belt mounted on the patient<br>(Medspira Breath-Hold for Interventional<br>Radiology) and visual comparison with the<br>reference level during needle guidance.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Intra-interventional<br>verification       | Verification CT Image registration with planning<br>CT to compare achieved needle position to<br>planned needle trajectory.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Verification CT Image registration with planning<br>CT to compare achieved needle position to<br>planned needle trajectory.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Post-procedure<br>verification             | Register intra-interventional CT series with post<br>procedure CT series to display planned target<br>position on Post-procedure CT image.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Register intra-interventional CT series with post<br>procedure CT series to display planned target<br>position on Post-procedure CT image.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |

#### Comparison of Technological Characteristics with the predicate device

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains the logo for Quantum Surgical. The logo consists of an orange hexagon with a stylized "Q" shape on the left side. To the right of the hexagon is the text "QUANTUM" in a larger, bolder font, stacked above the word "surgical" in a smaller font.

The comparisons between the EPIONE and the predicate determined that:

- The EPIONE device has the same general intended use as the predicate. The indications for use of the EPIONE is for liver ablations as in the predicate except that the predicate has a wider indication to also include other procedures and anatomical areas. The EPIONE is therefore equivalent to the predicate in this respect within its indications for use in liver ablations.
- The EPIONE device is limited for use for patients under general anesthesia as in the predicate except that the predicate allows also for local anesthesia conditions. The EPIONE device is therefore equivalent to the predicate within its use under general anesthesia conditions.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Quantum Surgical. The logo consists of an orange geometric shape resembling a stylized "q" on the left, followed by the words "QUANTUM" in orange and "surgical" in gray on the right. The geometric shape is made up of connected lines and angles, creating a modern and abstract design.

- The EPIONE device is compatible for use with interventional instruments with diameters ranging from 11G to 19G as compared to the predicate with a wider range of 11G to 21G. Given that the proposed EPIONE range is within that of the predicate, the EPIONE device is therefore equivalent to the predicate within its narrower range.
- · The EPIONE device was shown to have equivalent technological characteristics to the predicate including a robotic arm to position the needle guide per the planned trajectory. An important difference is that the EPIONE additionally implements an optical tracking system to both register and track the patient location and instrumentation to ensure correct placement of the needle relative to the CT based planned trajectory. The predicate alternately utilizes a locating mat and location calibration process at installation to provide for a fixed known location of the C-arm, and a laser system to thereafter ensure that the patient location has not changed. However, this different technological approach in the EPIONE is, along with other related differences in instrumentation for use with the tracking system, is supported by an equivalent implementation of an optical tracking system in the reference device, the CAS-One IR, for the same purposes and the same intended use as the EPIONE. This should then not raise any new questions of safety or effectiveness.

#### Summary of Non-Clinical Performance Data:

The following nonclinical performance testing was conducted to establish substantial equivalence and to verify that the EPIONE device will perform safely and effectively per its intended use:

- . Overall System Accuracy Tests
- . Respiratory Monitoring Effectiveness Tests
- . Needle Bending Test Study
- Positional accuracy testing in accordance with the ASTM F2554-10 Standard Practice for Measurement of Positional Accuracy of Computer Assisted Surgical Systems
- Electrical Safety Tests, in accordance with IEC 60601-1
- . Electromagnetic Compatibility Tests, in accordance with IEC 60601-1-2
- . Biocompatibility Tests, in accordance with standard ISO 10993 for patient contacting components
- Cleaning and Sterilization Validation Tests, for the instrument reprocessing
- . Usability Testing, in accordance with IEC 62366-1:2015 and the FDA quidance document Applying Human Factors and Usability Engineering to Medical Devices.

A preclinical animal study on the liver was conducted on porcine models using the EPIONE device to demonstrate safety and effectiveness of the device. The study evaluated the accuracy, safety, and feasibility of robotically assisted CT-guided needle placement in the liver of an in-vivo model (with breathing movements) with the EPIONE device. A total of 36 needle placements were performed in the liver. The average accuracy of the system demonstrates that the EPIONE device is safe and effective for CT-guided needle placement in the liver.

Additionally, Software design and velidation testing was completed in completed in compliance with the FDA guidance "Content of Premarket Submissions for Software Contained in Medical Devices", the IEC 62304 Standard "Medical Device Software - Life Cycle Process", as well as FDA cybersecurity premarket guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices".

The performance testing of the EPIONE device demonstrated that it should be as safe and as effective as the predicate device.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image contains the logo for Quantum Surgical. The logo consists of an orange hexagon with a stylized "q" shape on the left side. To the right of the hexagon, the words "QUANTUM" are written in orange, and below that, the word "surgical" is written in gray.

## Clinical study

A multi-center, prospective and interventional clinical study, was performed using the EPIONE™ device for liver ablations on 21 patients under general anesthesia to supplement the bench and preclinical animal testing and confirm the safety and performance of the results of this study show that the EPIONE™ device is safe and effective for CT-quided percutaneous procedures for liver ablations under general anesthesia conditions with no adverse events reported for any of the subjects.

## Conclusions

Compared to the predicate, the EPIONE device has the same intended use, indications for use within that of the predicate, patient anesthesia condition within that of the predicate, and equivalent technological characteristics. The differences between the EPIONE device and the predicate do not raise different questions of safety and effectiveness. The testing including bench testing, animal, and clinical testing demonstrated that the device should be as safe and effective in achieving its intended use as in the predicate. The EPIONE™ device is therefore substantially equivalent to the identified predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K211645](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K211645)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
