← Product Code [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK) · K181414

# uCT 530, uCT 550 (K181414)

_Shanghai United Imaging Healthcare Co., Ltd. · JAK · Aug 14, 2018 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K181414

## Device Facts

- **Applicant:** Shanghai United Imaging Healthcare Co., Ltd.
- **Product Code:** [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK.md)
- **Decision Date:** Aug 14, 2018
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The uCT Computed Tomography X-ray System uCT 530/uCT 550 is a computed tomography x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

## Device Story

Multi-slice CT scanner; uses continuously rotating tube-detector system; operates on fan beam principle. Inputs: X-ray transmission data collected at multiple angles/planes during gantry rotation. Processing: Data sent to computer for reconstruction into cross-sectional images (slices) using filter back-projection (FBP) algorithm. Features KARL iterative denoising and MAC metal artifact correction. Output: DICOM-format cross-sectional images for clinical review. Used in clinical settings by trained operators (technicians/physicians). Output assists clinicians in diagnostic imaging of internal organs and tissues. Benefits: Provides cross-sectional anatomical visualization for diagnostic purposes.

## Clinical Evidence

No clinical study included. Substantial equivalence supported by non-clinical bench testing, including dosimetry, image performance tests, electrical safety (ES 60601-1), EMC (IEC 60601-1-2), and biocompatibility (ISO 10993).

## Technological Characteristics

Multi-slice CT; solid-state GOS detectors; 50kW power; 700mm aperture; 0.55mm minimum slice thickness; 40/80 slice capability. Continuous rotation up to 0.5s/360°. Connectivity: DICOM. Software: KARL 3D iterative denoising, MAC metal artifact correction. Biocompatibility: ISO 10993-5/10 compliant.

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- uCT Computed Tomography X-Ray System (uCT 760, uCT 780) ([K172135](/device/K172135.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.

Shanghai United Imaging Healthcare Co., Ltd. % Shumei Wang OM & RA Director No. 2258 Chengbei Rd., Jiading Industrial District Shanghai. 201807 CHINA

August 14th, 2018

Re: K181414

Trade/Device Name: uCT 530, uCT 550 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: May 28, 2018 Received: May 30, 2018

Dear Shumei Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jeff Rodgers

for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K181414

Device Name uCT 530, uCT 550

### Indications for Use (Describe)

The uCT Computed Tomography X-ray System uCT 530/550 is a computed tomography x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

| Type of Use (Select one or both, as applicable)                                     |                                                                                    |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The logo is simple and modern, and the use of bold fonts gives it a strong and confident look.

# 510 (k) SUMMARY

- Date of Preparation 1. May 28, 2018
#### 2. Sponsor Identification

Shanghai United Imaging Healthcare Co.,Ltd. No.2258 Chengbei Rd. Jiading District, 201807, Shanghai, China

Contact Person: Shumei Wang Position: QM&RA VP Tel: +86-021-67076888-6776 Fax: +86-021-67076889 Email: shumei.wang@united-imaging.com

#### 3. Identification of Proposed Device

Trade Name: uCT 530, uCT 550 Common Name: Computed Tomography X-ray System Model(s): uCT 530, uCT 550

Regulatory Information Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-ray System Regulatory Class: II Product Code: JAK Review Panel: Radiology

#### Identification of Predicate Device(s) 4.

Predicate Device

510(k) Number: K172135 Device Name: uCT Computed Tomography X-Ray System Model(s): uCT 760, uCT 780

### Regulatory Information

Regulation Number: 21 CFR 892.1750 Regulation Name: Computed Tomography X-ray System Regulatory Class: II Product Code: JAK Review Panel: Radiology

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED" and "IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The logo is simple and modern, and the colors are muted.

#### 5. Device Description:

The uCT 530/uCT 550 is a multi-slice X-ray computed tomography scanner which features a continuously rotating tube-detector system and functions according to the fan beam principle. The system provides the filter back-projection (FBP) algorithm to reconstruct images in DICOM format. which can be used by post-processing applications.

The system consists of the Gantry, X-ray System, Data Management System, Patient Table, Console, Power Supply Cabinet, Image Processing Computer, and Software. The system software is a program used for patient management, data management, Xray scan control, image reconstruction, and image archive.

A motorized patient table moves the patient through a circular opening in the Gantry. As the patient passes through the Gantry, a source of x rays rotates around the inside of the circular opening. Detectors on the exit side of the patient record the X rays exiting the section of the patient's body being irradiated as an X-ray "snapshot". Many different "snapshots" (angles) are collected during one complete rotation. The data are sent to a computer to reconstruct all of the individual "snapshots" into a crosssectional image (slice) of the internal organs and tissues for each complete rotation of the source of x rays.

There are two features for denoising and reduce metal artifact, which are KARL iterative denoising reconstruction algorithm and MAC Metal artifact correction algorithm.

|         | Spec.<br>Model | HV Power | Rotation speed                     | Minimum<br>slice thickness | Maximum slices<br>generated per rotation |
|---------|----------------|----------|------------------------------------|----------------------------|------------------------------------------|
| uCT 530 |                | 50kW     | Up to 0.5 sec per<br>360 °rotation | 0.55mm                     | 40                                       |
| uCT 550 |                | 50kW     | Up to 0.5 sec per<br>360 °rotation | 0.55mm                     | 80                                       |

This proposed device includes two models: uCT 530, uCT 550.The differences between the two models are as follows:

#### Indications for Use 6.

The uCT Computed Tomography X-ray System uCT 530/uCT 550 is a computed tomography X-ray system intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes and indicated for the whole body (including head, neck, vascular).

#### 7. Comparison of Technological Characteristics with the Predicate Devices

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is stacked on top of each other and is in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The logo is simple and modern, and the colors are muted.

The uCT 530/uCT 550 Computed Tomography X-ray system has the same indications for use as the predicate device uCT 760/uCT 780. The fundamental scientific technology of the proposed device is same as the predicate device.

Table 1 below provides a comparison of the technological characteristics of the proposed device in comparison to the predicate device.

| ITEM                                                                     | Proposed Device                                                                                                                                                                                                                                                                                                                                     | Predicate Device                                                                                                                                                                                                                                                                                                                                 | Remark        |
|--------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
|                                                                          | uCT 530, uCT 550                                                                                                                                                                                                                                                                                                                                    | uCT 760, uCT 780                                                                                                                                                                                                                                                                                                                                 |               |
| General                                                                  |                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                  |               |
| Product Code                                                             | JAK                                                                                                                                                                                                                                                                                                                                                 | JAK                                                                                                                                                                                                                                                                                                                                              | Same          |
| Regulation No.                                                           | 21 CFR 892.1750                                                                                                                                                                                                                                                                                                                                     | 21 CFR 892.1750                                                                                                                                                                                                                                                                                                                                  | Same          |
| Class                                                                    | II                                                                                                                                                                                                                                                                                                                                                  | II                                                                                                                                                                                                                                                                                                                                               | Same          |
| Intended Use                                                             | The uCT Computed<br>Tomography X-ray System<br>uCT530/uCT550 is a computed<br>tomography X-ray system<br>intended to produce cross-<br>sectional images of the body<br>by computer reconstruction of<br>X-ray transmission data taken<br>at different angles and planes<br>and indicated for the whole<br>body (including head, neck,<br>vascular). | The uCT Computed<br>Tomography X-ray System is<br>a computed tomography x-ray<br>system intended to produce<br>cross-sectional images of the<br>body by computer<br>reconstruction of x-ray<br>transmission data taken at<br>different angles and planes<br>and indicated for the whole<br>body (including head, neck,<br>cardiac and vascular). | Note<br>No.1  |
| Specifications                                                           |                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                  |               |
| Scan Regime                                                              | Continuous Rotation                                                                                                                                                                                                                                                                                                                                 | Continuous Rotation                                                                                                                                                                                                                                                                                                                              | Same          |
| Scan Modes                                                               | Scout<br>Axial Scan<br>Helical Scan                                                                                                                                                                                                                                                                                                                 | Scout<br>Axial Scan<br>Helical Scan                                                                                                                                                                                                                                                                                                              | Same          |
| Detector<br>Material                                                     | Solid-state GOS                                                                                                                                                                                                                                                                                                                                     | Solid-state GOS                                                                                                                                                                                                                                                                                                                                  | Same          |
| Z-plane<br>coverage                                                      | 22mm                                                                                                                                                                                                                                                                                                                                                | 40mm                                                                                                                                                                                                                                                                                                                                             | Note<br>No.2  |
| Size of detector<br>element in Z-<br>plane                               | 0.55mm                                                                                                                                                                                                                                                                                                                                              | 0.5mm                                                                                                                                                                                                                                                                                                                                            | Note<br>No.3  |
| Number of<br>element per row                                             | 864                                                                                                                                                                                                                                                                                                                                                 | 936                                                                                                                                                                                                                                                                                                                                              | Note<br>No.4  |
| Number of<br>detector row                                                | 40                                                                                                                                                                                                                                                                                                                                                  | 80                                                                                                                                                                                                                                                                                                                                               | Note<br>No.5  |
| Maximum slices<br>generated per<br>rotation (multi-<br>slice capability) | 40 for uCT 530<br>80 for uCT 550                                                                                                                                                                                                                                                                                                                    | 128 for uCT 760<br>160 for uCT 780                                                                                                                                                                                                                                                                                                               | Note<br>No.6  |
| Minimum slice<br>thickness                                               | 0.55mm                                                                                                                                                                                                                                                                                                                                              | 0.625mm for uCT 760<br>0.5mm for uCT 780                                                                                                                                                                                                                                                                                                         | Note No.7     |
| Maximum<br>sampling rate                                                 | Up to 4800 views per 360 °                                                                                                                                                                                                                                                                                                                          | Up to 4800 views per 360 °                                                                                                                                                                                                                                                                                                                       | Same          |
| Tube anode<br>storage capacity                                           | 5.3MHU                                                                                                                                                                                                                                                                                                                                              | 7.5MHU                                                                                                                                                                                                                                                                                                                                           | Note No.8     |
| Maximum<br>cooling rate                                                  | 815 kHU/min                                                                                                                                                                                                                                                                                                                                         | 1386 kHU/min                                                                                                                                                                                                                                                                                                                                     | Note No.9     |
| Focal spot size                                                          | 0.5x1.0mm<br>1.0x1.0mm                                                                                                                                                                                                                                                                                                                              | 0.7x0.7mm<br>1.0x1.0mm                                                                                                                                                                                                                                                                                                                           | Note No.10    |
| Power                                                                    | 50kW                                                                                                                                                                                                                                                                                                                                                | 80kW for uCT 760<br>100 kW for uCT 780                                                                                                                                                                                                                                                                                                           | Note No.11    |
| mA Range                                                                 | 10-420mA                                                                                                                                                                                                                                                                                                                                            | 6-667mA for uCT 760<br>6-833mA for uCT 780                                                                                                                                                                                                                                                                                                       | Note No.12    |
| kV Settings                                                              | 70, 80, 100, 120, 140                                                                                                                                                                                                                                                                                                                               | 70, 80, 100, 120, 140                                                                                                                                                                                                                                                                                                                            | Same          |
| Aperture                                                                 | 700mm                                                                                                                                                                                                                                                                                                                                               | 700mm                                                                                                                                                                                                                                                                                                                                            | Same          |
| Rotation speed                                                           | Up to 0.5 sec per 360 °rotation                                                                                                                                                                                                                                                                                                                     | Up to 0.35 sec per 360 °<br>rotation for uCT 760<br>Up to 0.3 sec per 360 °<br>rotation for uCT 780                                                                                                                                                                                                                                              | Note No.13    |
| Gantry Tilt                                                              | $\pm$ 30 °with 0.5 increment                                                                                                                                                                                                                                                                                                                        | $\pm$ 30 °with 0.5 increment                                                                                                                                                                                                                                                                                                                     | Same          |
| Scannable range                                                          | 1700 mm                                                                                                                                                                                                                                                                                                                                             | 1700 mm                                                                                                                                                                                                                                                                                                                                          | Same          |
| Horizontal<br>motion range                                               | 2180 mm                                                                                                                                                                                                                                                                                                                                             | 2180 mm                                                                                                                                                                                                                                                                                                                                          | Same          |
| Table<br>Horizontal<br>Speed                                             | Up to 200mm/sec                                                                                                                                                                                                                                                                                                                                     | Up to 200mm/sec                                                                                                                                                                                                                                                                                                                                  | Same          |
| Vertical motion<br>range                                                 | 480 mm-950 mm from the<br>floor                                                                                                                                                                                                                                                                                                                     | 480 mm-950 mm from the<br>floor                                                                                                                                                                                                                                                                                                                  | Same          |
| Vertical speed                                                           | Up to 40 mm/sec                                                                                                                                                                                                                                                                                                                                     | Up to 40 mm/sec                                                                                                                                                                                                                                                                                                                                  | Same          |
| Table<br>Horizontal<br>Position<br>accuracy                              | $\pm$ 0.25mm                                                                                                                                                                                                                                                                                                                                        | $\pm$ 0.25mm                                                                                                                                                                                                                                                                                                                                     | Same          |
| Table Maximum<br>table load                                              | 205kg                                                                                                                                                                                                                                                                                                                                               | 205kg                                                                                                                                                                                                                                                                                                                                            | Same          |
| Image Spatial<br>Resolution                                              | High mode:<br>>20 lp/cm @ MTF 0%<br>16.5 $\pm$ 1.7 lp/cm @ MTF10%<br>11.5 $\pm$ 1.2 lp/cm @ MTF50%                                                                                                                                                                                                                                                  | High mode:<br>>20 lp/cm @ MTF 0%<br>16.5 $\pm$ 1.7 lp/cm @ MTF10%<br>11.5 $\pm$ 1.2 lp/cm @ MTF50%                                                                                                                                                                                                                                               | Same          |
| Image Noise                                                              | 3.0 $\pm$ 0.5 HU at 120 kV, 5 mm<br>slice thickness, CTDIvol 28.9<br>mGy                                                                                                                                                                                                                                                                            | 3.0 $\pm$ 0.5 HU at 120 kV, 5 mm<br>slice thickness, CTDIvol<br>29.1mGy                                                                                                                                                                                                                                                                          | Note No.14    |
| CT Number<br>Display Range                                               | -1024 ~+8191 HU                                                                                                                                                                                                                                                                                                                                     | -1024 ~+8191 HU                                                                                                                                                                                                                                                                                                                                  | Same          |
| Scan Field of View                                                       | Up to 500 mm<br>600mm with extend FOV                                                                                                                                                                                                                                                                                                               | Up to 500 mm<br>600mm with extend FOV                                                                                                                                                                                                                                                                                                            | Same          |
| Reconstruction<br>Field of View                                          | 40mm-500mm<br>40mm-600mm with extend<br>FOV                                                                                                                                                                                                                                                                                                         | 40mm-500mm<br>40mm-600mm with extend<br>FOV                                                                                                                                                                                                                                                                                                      | Same          |
| Maximum<br>scannable length                                              | 1700mm                                                                                                                                                                                                                                                                                                                                              | 1700mm                                                                                                                                                                                                                                                                                                                                           | Same          |
| Image Matrix                                                             | Up to 1024 x 1024<br>0.55mm,1.1mm,2.2mm,5.5mm<br>,11mm (axial)<br>0.55-10mm(helical)                                                                                                                                                                                                                                                                | Up to 1024 x 1024                                                                                                                                                                                                                                                                                                                                | Same          |
| Reconstructed<br>slice thickness                                         |                                                                                                                                                                                                                                                                                                                                                     | uCT 760:<br>0.625mm,1.25mm,2.5mm,5mm,10mm (axial)<br>0.625-10mm(helical)<br>uCT 780:<br>0.5mm,0.625mm,1.25mm,2.5mm,5mm,10mm (axial)<br>0.5-10mm (helical)                                                                                                                                                                                        | Note<br>No.15 |
| Pitch                                                                    | 0.1~2.0                                                                                                                                                                                                                                                                                                                                             | 0.1~2.0                                                                                                                                                                                                                                                                                                                                          | Same          |
| Maximum<br>continuous<br>exposure time                                   | Up to 100seconds                                                                                                                                                                                                                                                                                                                                    | Up to 100seconds                                                                                                                                                                                                                                                                                                                                 | Same          |
| Safety                                                                   |                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                  |               |
| Electrical Safety                                                        | Comply with ES60601-1                                                                                                                                                                                                                                                                                                                               | Comply with ES60601-1                                                                                                                                                                                                                                                                                                                            | Same          |
| EMC                                                                      | Comply with IEC60601-1-2                                                                                                                                                                                                                                                                                                                            | Comply with IEC60601-1-2                                                                                                                                                                                                                                                                                                                         | Same          |
| Biocompatibility                                                         | Patient Contact Materials were<br>tested and demonstrated no<br>cytotoxicity (ISO 10993-5), no<br>evidence for irritation and<br>sensitization (ISO 10993-10).                                                                                                                                                                                      | Patient Contact Materials<br>were tested and demonstrated<br>no cytotoxicity (ISO 10993-<br>5), no evidence for irritation<br>and sensitization (ISO 10993-<br>10).                                                                                                                                                                              | Same          |
| Clinical                                                                 | Sample clinical images for proposed devices are provided in Section 36<br>Clinical Evaluation. Electronic file for each image are provide in MISC<br>Folder.                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                  |               |
| Justification                                                            |                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                  |               |
| Note ID                                                                  | Justification                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                  |               |
| Note 1                                                                   | The uCT 530 and uCT550 is indicated for computed tomography of the<br>whole body including head, neck, and vascular but not including cardiac<br>which does not affect safety and effectiveness the computed tomography<br>system.                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                  |               |
| Note 2                                                                   | Provides the shorter detector Z-plane coverage induces longer scanning<br>time for CT imaging, which does not affect safety and effectiveness.                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                  |               |
| Note 3                                                                   | Provides the bigger minimum detector element size that induces lower z-<br>plane spatial resolution for CT imaging, which does not affect safety and<br>effectiveness.                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                  |               |

Table 1 Comparison of Technological Characteristics

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the words "UNITED IMAGING" in bold, sans-serif font. To the right of the words is a stylized logo. The logo is a dark blue color and appears to be a stylized letter U.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for United Imaging. The text "UNITED IMAGING" is in bold, sans-serif font. To the right of the text is a stylized letter "U" that is split in the middle by a vertical white line. There is also a horizontal white line in the middle of the "U".

{8}------------------------------------------------

Image /page/8/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The color of the logo is a dark teal.

| Note 4                                                          | Provides the smaller detector element number per row that induces less data sampling for CT imaging but provides sufficient data sampling for reconstruction which does not affect safety and effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                |
|-----------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Note 5                                                          | The number of detector rows is determined by Z-plane coverage and size of detector element in Z-plane, which does not affect safety and effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Note 6                                                          | With a reconstruction process, 40 slices or 80 slices can be achieved from the 40 rows detector. Provides the smaller slice number in one rotation induces longer scanning time for CT imaging, which does not affect safety and effectiveness.                                                                                                                                                                                                                                                                                                                                                                             |
| Note 7                                                          | Minimum slice thickness is determined by size of detector element in Z-plane, which does not affect safety and effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Note 8                                                          | Tube anode storage capacity is a kind of measurement about the maximum throughput of CT scanner. When an independent scan is implemented, the tube is heated and if the time interval between two independent scans is long enough, the heat can be dissipated timely but for short scan time interval, the heat which has not been dissipated timely should be stored within tube device and thus for the continuous scans (for short scan time intervals), higher storage capacity means that more continuous scan numbers can be supported. However, tube anode storage capacity has no effect on each independent scan. |
| Note 9                                                          | Maximum cooling rate is a kind of measurement about the maximum throughput of CT scanner. For the continuous scans, higher maximum cooling rate means that the tube heat can be dissipated faster and thus shorter scan time interval and more continuous scan numbers can be supported. However, it has no effect on each independent scan.                                                                                                                                                                                                                                                                                |
| Note 10                                                         | Focus spot size has effect on image spatial resolution and smaller size is helpful for resolution improvement. However, the image spatial resolution between the two kinds of devices is equivalent substantially.                                                                                                                                                                                                                                                                                                                                                                                                          |
| Note 11                                                         | Provides the smaller power output that induces lower ability of x-ray penetration when scanning the object with high BMI with higher possibility of photon starvation, and the safety has been evaluated by the related testing and verification.                                                                                                                                                                                                                                                                                                                                                                           |
| Note 12                                                         | Provides the smaller mA output that induces lower ability of x-ray penetration when scanning the object with high BMI with higher possibility of photon starvation, and the safety has been evaluated by the related testing and verification.                                                                                                                                                                                                                                                                                                                                                                              |
| Note 13                                                         | Provides slower rotation speed that induces longer scan time for examination which does not affect safety and effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Note 14                                                         | With 120 kV and 5mm slice thickness, the image noise for typical head is 3HU on CTDIvol 28.9 mGy. The image noise level is equivalent substantially considering the proposed device has measured its noise based on the smaller CTDIvol than the Predicate Device.                                                                                                                                                                                                                                                                                                                                                          |
| Note 15                                                         | Provides the more slice thickness that induce more choice for various clinical scan situations and thinner slice thickness should bring higher                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| spatial resolution in z-plane, which does not affect safety and |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| effectiveness.                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the text is a stylized letter "U" that is split into two parts by a horizontal line. The logo is simple and modern, and the color is a dark teal.

The proposed device and the predicate device are the same in regard to most of application features, except for cardiac.

Table 2 below provides a comparison of the application features of the proposed device in comparison to the predicate device.

| ITEM                      | Proposed Device<br>uCT 530, uCT 550 | Predicate Device<br>uCT 760, uCT 780 | Remark |
|---------------------------|-------------------------------------|--------------------------------------|--------|
| Application Features      |                                     |                                      |        |
| Iterative noise reduction | KARL 3D                             | KARL 3D                              | Same   |
| Adaptive Filter           | Adaptive Filter                     | Same                                 | Same   |
| Metal artifact reduction  | MAC                                 | MAC                                  | Same   |

Table 2 Comparison of Application Features

#### 8. Performance Data

The following performance data were provided in support of the substantial equivalence determination.

### Non-Clinical Testing

Non-clinical testing including dosimetry and image performance tests were conducted for the uCT 530/uCT 550 during the product development.

UNITED IMAGING HEALTHCARE claims conformance to the following standards and guidance:

## Electrical Safety and Electromagnetic Compatibility (EMC)

Electrical Safety and Electromagnetic Compatibility (EMC) testing were conducted on the uCT 530/550 in accordance with the following standards:

- A ES 60601-1:2005(R)2012+A1:2012+C1:2009/(R)2012+A2:2010/(R)2012 Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- A IEC 60601-2-44 Edition 3.0 2009, Medical Electrical Equipment - Part 2-44: Particular Requirements For The Basic Safety And Essential Performance Of Xray Equipment For Computed Tomography
- A IEC 60601-1-2:2014, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests
- > IEC 60825-1 Edition 2.0 2007-03, Safety Of Laser Products - Part 1: Equipment Classification, And Requirements

## Product Particular Standards

- NEMA XR 25-2010, Computed Tomography Dose Check A
- A NEMA XR 28-2013, Supplemental Requirements For User Information And

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image shows the logo for United Imaging. The logo consists of the words "UNITED IMAGING" in bold, sans-serif font, stacked on top of each other. To the right of the words is a stylized "U" shape, which is dark gray. The logo is simple and modern.

System Function Related To Dose In CT

- A NEMA XR 29-2013, Standard Attributes on CT Equipment Related to Dose Optimization and Management
- IEC 60601-1-3 Edition 2.1 2013-04, Medical Electrical Equipment Part 1-3: A General Requirements For Basic Safety And Essential Performance - Collateral Standard: Radiation Protection In Diagnostic X-ray Equipment
- A IEC 61223-3-5 First Edition 2004-08, Evaluation And Routine Testing In Medical Imaging Departments - Part 3-5: Acceptance Tests - Imaging Performance Of Computed Tomography X-ray Equipment

# Performance Verification

- A Clinical Evaluation for sample clinical images evaluation;
- A AEC Test Report for AEC performance study.

## Software

- A NEMA PS 3.1-3.20(2011): Digital Imaging and Communications in Medicine (DICOM)
- A IEC 62304: Medical Device Software - software life cycle process
- > Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices
- > Content of Premarket Submissions for Management of Cybersecurity in Medical Devices

# Biocompatibility

- ISO 10993-5 Third Edition 2009-06-01, Biological Evaluation Of Medical A Devices - Part 5: Tests For In Vitro Cytotoxicity
- A ISO 10993-10 Third Edition 2010-08-01, Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization

## Other Standards and Guidances

- ISO 14971: Medical Devices Application of risk management to medical A devices
- Code of Federal Regulations, Title 21, Part 820 Quality System Regulation A
- > Code of Federal Regulations, Title 21, Subchapter J - Radiological Health
- > Laser Products - Conformance with IEC 60825-1 and IEC 60601-2-22; Guidance for Industry and FDA Staff (Laser Notice No. 50)
- > Provision for Alternate Measure of the Computed Tomography Dose Index (CTDI) to Assure Compliance with the Dose Information Requirements of the Federal Performance Standard for Computed Tomography

## Software Verification and Validation

Software documentation for a Moderate Level of Concern software per FDA' Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" is included as a part of this submission.

{11}------------------------------------------------

Image /page/11/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED" is stacked on top of the text "IMAGING" and both are in a bold, sans-serif font. To the right of the text is a stylized icon that appears to be a capital letter "U" with a vertical line through the center.

The risk analysis was completed and risk control was implemented to mitigate identified hazards. The testing results show that all the software specifications have met the acceptance criteria. Verification and validation testing of the proposed device was found acceptable to support the claim of substantial equivalence.

UNITED IMAGING HEALTHCARE conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modification, misuse or denial of use, or unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Cybersecurity information in accordance with guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices" is included in this submission.

### Clinical Testing

No Clinical Study is included in this submission.

### Summarv

The features described in this premarket submission are supported with the results of the testing mentioned above, the uCT 530/uCT 550 was found to have a safety and effectiveness profile that is similar to the predicate device.

#### 9. Conclusions

Based on the comparison and analysis above, the proposed device has same intended use, similar performance, equivalence safety and effeteness as the predicate device. The differences above between the proposed device and predicate device do not affect the intended use, technology characteristics, safety and effectiveness. And no issues are raised regarding to safety and effectiveness. The proposed device is determined to be Substantially Equivalent (SE) to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K181414](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K181414)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
