← Product Code [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK) · K171738

# Supria True64 (K171738)

_Hitachi Healthcare Americas Corporation · JAK · Aug 18, 2017 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K171738

## Device Facts

- **Applicant:** Hitachi Healthcare Americas Corporation
- **Product Code:** [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK.md)
- **Decision Date:** Aug 18, 2017
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The Supria True64 system is indicated for head, whole body, and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, or dynamic modes. The volume datasets acquired by the Supria can be post processed by the system to provide additional information. Post processing capabilities included in the Supria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering. Volume datasets acquired by the Supria can be transferred to external devices via a DICOM standard interface. The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.

## Device Story

Multi-slice CT system using 64-detector technology; rotates X-ray tube and detector assembly around patient via slip ring. Collects data at multiple angles; high-frequency generator produces fan beam X-ray. High-speed reconstruction sub-system processes data into cross-sectional images. Operated by qualified physicians in clinical settings. Includes guideShot option for interventional imaging via remote in-room display. Post-processing includes CTA, MPR, and volume rendering. Output aids diagnosis; images transferred via DICOM. Benefits include high-quality imaging with potential for low-dose acquisition via iterative reconstruction.

## Clinical Evidence

Bench testing only. Performance metrics (dose profile, noise, CT number, uniformity, spatial resolution, MTF) compared to predicate and found equivalent. Clinical evidence limited to 3 clinical image examples for usability assessment and comparative evaluation of FBP vs. Iterative Reconstruction showing noise/artifact reduction.

## Technological Characteristics

64-slice CT; solid-state detector; 750mm gantry aperture; 5 MHU anode heat storage; 48kW system max power. PC-based workstation (Windows OS). Connectivity via DICOM. Iterative reconstruction (Intelli IP Advanced/Quick). Automatic Exposure Control (IntelliEC). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-44, NEMA XR 25, NEMA XR 26, IEC 62304.

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- HITACHI SUPRIA Whole-body X-ray CT System Phase 3 ([K163528](/device/K163528.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or movement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 18, 2017

Hitachi Healthcare Americas Corporation % Mr. Douglas J. Thistlethwaite Manager of Regulatory Affairs 1959 Summit Commerce Park TWINSBURG OH 44087

Re: K171738

Trade/Device Name: Supria True64 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: July 26, 2017 Received: July 28, 2017

Dear Mr. Thistlethwaite:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely,

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

#### K171738

Device Name Supria True64 Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

Indications for Use (Describe)

The Supria True64 system is indicated for head, whole body, and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, or dynamic modes.

The volume datasets acquired by the Supria can be post processed by the system to provide additional information. Post processing capabilities included in the Supria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.

Volume datasets acquired by the Supria can be transferred to external devices via a DICOM standard interface. The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary

{4}------------------------------------------------

## Submitter Information

| Submitter:        | Hitachi Healthcare Americas<br>1959 Summit Commerce Park<br>Twinsburg, Ohio 44087-2371 |
|-------------------|----------------------------------------------------------------------------------------|
| Contact:          | Douglas J. Thistlethwaite                                                              |
| Telephone number: | 330-425-1313                                                                           |
| Telephone number: | 330-963-0749                                                                           |
| E-mail:           | thistlethwaited@hitachihealthcare.com                                                  |
| Date:             | April 18, 2017                                                                         |

## Subject Device Name

| Trade/Proprietary Name: | Supria True64                            |
|-------------------------|------------------------------------------|
| Regulation Number:      | 21 CFR 892.1750                          |
| Regulation Name:        | Computed tomography x-ray system         |
| Product Code            | JAK, System, X-Ray, Tomography, Computed |
| Class                   | II                                       |
| Panel                   | Radiology                                |

## Predicate Device Name

| Predicate Device(s): | HITACHI SUPRIA Whole-body X-ray CT System Phase 3 (K163528) |
|----------------------|-------------------------------------------------------------|
| Regulation Number:   | 21 CFR 892.1750                                             |
| Regulation Name:     | Computed tomography x-ray system                            |
| Product Code         | JAK, System, X-Ray, Tomography, Computed                    |
| Class                | II                                                          |
| Panel                | Radiology                                                   |

### Device Intended Use

The Supria True64 system is indicated for head, whole body, and vascular X-ray Computed Tomography applications in patients of all ages. The images can be acquired in either axial, helical, or dvnamic modes,

The volume datasets acquired by the Supria can be post processed by the system to provide additional information. Post processing capabilities included in the Supria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.

Volume datasets acquired by the Supria can be transferred to external devices via a DICOM standard interface.

The guideShot Option adds a remote in-room display and controls to support interventional imaging. The device output can provide an aid to diagnosis when used by a qualified physician.

## Device Description

### Function

The Supria True64 is a multi-slice computed tomography system designed to perform multi-slice CT scanning supported by 64-detector technology. The system allows optimum clinical applications ranging from routine exams in response to the diversified circumstances in imaging whole body regions.

{5}------------------------------------------------

### Scientific Concepts

The Supria True64 system uses 64-slice CT technology, where the X-ray tube and detector assemblies are mounted on a frame that rotates continuously around the patient using slip ring technology. The solid-state detector assembly design collects up to 64 slices of data simultaneously. The X-ray sub-system features a high frequency generator. X-ray tube, and collimation system that produces a fan beam X-ray output. The system can operate in a helical (spiral) scan mode where the patient table moves during scanning. As the X-ray tube/detector assembly rotates around the patient, data is collected at multiple angles.

The collected data is then reconstructed into cross-sectional images by a high-speed reconstruction sub-system. The images are displayed on a Computer Workstation, stored, printed, and archived as required. The workstation is based on current PC technology using the Windows™ operating system.

### Physical and Performance Characteristics

The Supria True64 system consists of a Gantry, Operator's Workstation, Patient Table, High-Frequency X-ray Generator, and accessories. The system performance is similar to the predicate device.

### Performance Comparison

As part of our design validation performance comparison analysis was conducted to demonstrate continued conformance with a special control or recognized standard.

A rationale analysis was then conducted and the results are contained in Table 1:

| Testing Type                  | Rationale Analysis                                                                                                                                                                                                                                                                                                   |
|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Validation Testing - Bench    | There is no change about the following performance.<br>Dose Profile, Noise, Mean CT number and Uniformity, Spatial Resolution, Tomographic Section Thickness and Sensitivity<br>Profile, Tomographic Plane Location, CT dose index<br>Therefore, we judged that Supria is substantially equivalent to the predicate. |
| Validation Testing - Clinical | We provide 3 kinds of clinical image example which we judged to be sufficient to judge a clinical usability.                                                                                                                                                                                                         |

Table 1 Performance Comparison Analysis

The analysis confirms the performance characteristics of the Supria True64 are comparable to the predicate device and support our conclusion that the subject system is substantially equivalent.

# Device Technological Characteristics

The technological characteristics of the Supria True64 and the predicate device are listed in Table 2.

| ITEM                                                         | Supria True64                                                                          | HITACHI Supria PHASE 3 (K163528)                        | Difference |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------|---------------------------------------------------------|------------|
| Gantry                                                       |                                                                                        |                                                         |            |
| Geometry                                                     | Rotate-rotate with offset detector system,<br>slip ring                                | Rotate-rotate with offset detector system,<br>slip ring | No         |
| Scan Time                                                    | 0.75, 1.0, 1.5, 2.0 [s]                                                                | 0.75, 1.0, 1.5, 2.0 [s]                                 | No         |
| X-ray Fan Beam Angle                                         | 51 [deg]                                                                               | 51 [deg]                                                | No         |
| Gantry Tilt                                                  | -30 to +30 [deg]                                                                       | -30 to +30 [deg]                                        | No         |
| Gantry Aperture                                              | 750 [mm]                                                                               | 750 [mm]                                                | No         |
| Gantry Dimensions                                            | 1990 x 920 x 1842.5 [mm]                                                               | 1990 x 920 x 1842.5 [mm]                                | No         |
| Gantry Weight                                                | 1600 [kg]                                                                              | 1600 [kg]                                               | No         |
| Scan Localizer                                               | Laser                                                                                  | Laser                                                   | No         |
| ITEM                                                         | Supria True64                                                                          | HITACHI Supria PHASE 3 (K163528)                        | Difference |
| Detector                                                     |                                                                                        |                                                         |            |
| Type                                                         | Solid state                                                                            | Solid state                                             | No         |
| Number of Channels                                           | 880 [ch] (16 slice)<br>888 [ch] (64 slice)                                             | 880 [ch] (8ch reference)                                | No         |
| Number of Rows                                               | 16 (16 slice)<br>64 (64 slice)                                                         | 16                                                      | See 01     |
| Number of Slices                                             | 16 [slice/scan] (Axial) (16 slice)<br>64 [slice/scan] (Axial) (64 slice)               | 16 [slice/scan] (Axial)                                 | No         |
| X-ray Tube                                                   |                                                                                        |                                                         |            |
| Anode Heat Storage                                           | 5 [MHU]                                                                                | 5 [MHU]                                                 | No         |
| Dissipation Rate                                             | 470 [kHU/min]                                                                          | 470 [kHU/min]                                           | No         |
| Tube cooling                                                 | Oil/air                                                                                | Oil/air                                                 | No         |
| Tube focal spot                                              | Dual 0.7 x 0.8, 1.2 x 1.4 [mm]                                                         | Dual 0.7 x 0.8, 1.2 x 1.4 [mm]                          | No         |
| X-ray Generator                                              |                                                                                        |                                                         |            |
| kW Output                                                    | System Maximum 48[kW] /<br>Generator Maximum 51 [kW]                                   | System Maximum 48[kW] /<br>Generator Maximum 51 [kW]    | No         |
| Max. Power Input                                             | 75 [kVA]                                                                               | 75 [kVA]                                                | No         |
| kVp Range                                                    | 80, 100, 120, 140 [kVp]                                                                | 80, 100, 120, 140 [kVp]                                 | No         |
| mA Range                                                     | 10 to 400 [mA] @120kV, 48kW                                                            | 10 to 400 [mA] @120kV, 48kW                             | No         |
| Patient Table                                                |                                                                                        |                                                         |            |
| Range of Movement,<br>Vertical                               | 450 to 1000 [mm] (CT-WT-21)                                                            | 450 to 1000 [mm] (CT-WT-21)                             | No         |
| Range of Movement,<br>Longitudinal                           | 1910 [mm] (CT-WT-21)                                                                   | 1910 [mm] (CT-WT-21)                                    | No         |
| Range of Movement,<br>Lateral                                | N/A                                                                                    | N/A                                                     | No         |
| Scannable Range                                              | 155 cm                                                                                 | 155 cm                                                  | No         |
| Maximum Load Capacity                                        | 227 [kg]                                                                               | 227 [kg]                                                | No         |
| Display                                                      |                                                                                        |                                                         |            |
| Monitor Type                                                 | 24" LCD                                                                                | 24" LCD                                                 | No         |
| Matrices, Pixels                                             | 1920 x 1200                                                                            | 1920 x 1200                                             | No         |
| Image Enlargements                                           | Up to 9.99x                                                                            | Up to 9.99x                                             | No         |
| Max. Slices Displayed at Once                                | 25                                                                                     | 25                                                      | No         |
| Image Storage                                                |                                                                                        |                                                         |            |
| Hard Disk                                                    | 110 [GB] (images),<br>200 [GB] (raw data) (16 slice)<br>800 [GB] (raw data) (64 slice) | 110 [GB] (images),<br>200 [GB] (raw data)               | See 02     |
| Storage Images                                               | 200,000                                                                                | 200,000                                                 | No         |
| Archival Storage (Media)                                     | DVD-R/RW, CD-R/RW                                                                      | DVD-R/RW, CD-R/RW                                       | No         |
| ITEM                                                         | Supria True64                                                                          | HITACHI Supria PHASE 3 (K163528)                        | Difference |
| Scanning, Reconstruction                                     |                                                                                        |                                                         |            |
| Localization Scan                                            | Real time                                                                              | Real time                                               | No         |
| Localization Scan Length                                     | 150, 250, 350, 500, 750,<br>1000, 1250, 1500, 1750 [mm]                                | 150, 250, 350, 500, 750,<br>1000, 1250, 1500, 1750 [mm] | No         |
| Max. Scan Time                                               | 100 [s]                                                                                | 100 [s]                                                 | No         |
| Helical Beam Pitch                                           | 0.56, 0.81, 1.06, 1.31, 1.56 (16 slice)<br>0.58, 0.83, 1.08, 1.33, 1.58 (64 slice)     | 0.56, 0.81, 1.06, 1.31, 1.56                            | See 03     |
| Collimation                                                  | 1.25, 5, 10, 15, 20 [mm]                                                               | 1.25, 5, 10, 15, 20 [mm]                                | No         |
| Reconstruction Matrix                                        | 512 x 512 [pix]                                                                        | 512 x 512 [pix]                                         | No         |
| Reconstruction FOVs                                          | 20 to 500 [mm]                                                                         | 20 to 500 [mm]                                          | No         |
| Slice Thickness                                              | 0.625, 1.0, 1.25, 2.5, 3.75,<br>5.0, 7.5, 10.0 [mm]                                    | 0.625, 1.0, 1.25, 2.5, 3.75,<br>5.0, 7.5, 10.0 [mm]     | No         |
| Range of CT numbers                                          | -2000 to +4000 (13bit)<br>-32768 to +32767 (16bit)                                     | -2000 to +4000 (13bit)<br>-32768 to +32767 (16bit)      | No         |
| Reconstruction Time                                          | 0.1 seconds per image or less                                                          | 0.1 seconds per image or less                           | No         |
| Performance                                                  |                                                                                        |                                                         |            |
| High-contrast spatial resolution                             | 0.35 [mm]                                                                              | 0.35 [mm]                                               | No         |
| Low-contrast resolution mm at<br>% at <4 rads                | 2.5 [mm] @ 0.25%                                                                       | 2.5 [mm] @ 0.25%                                        | No         |
| 10% MTF                                                      | 14.7 [lp/cm]                                                                           | 14.7 [lp/cm]                                            | No         |
| 50% MTF                                                      | 12.2 [lp/cm]                                                                           | 12.2 [lp/cm]                                            | No         |
| Dose Controls                                                |                                                                                        |                                                         |            |
| Bow Tie Filter                                               | Yes. Normal                                                                            | Yes. Normal                                             | No         |
| Automatic Exposure Control                                   | Yes. IntelliEC                                                                         | Yes. IntelliEC                                          | No         |
| Longitudinal Modulation                                      | Yes                                                                                    | Yes                                                     | No         |
| Angular Modulation                                           | Yes                                                                                    | Yes                                                     | No         |
| Iterative Reconstruction                                     | Yes. Intelli IP Advanced, Intelli IP Quick                                             | Yes. Intelli IP Advanced, Intelli IP Quick              | No         |
| Maximum possible pitch with full<br>image quality            | 1.56 (16 slice)<br>1.58 (64 slice)                                                     | 1.56                                                    | See 03     |
| Dose Displays                                                |                                                                                        |                                                         |            |
| CTDIv                                                        | Yes                                                                                    | Yes                                                     | No         |
| DLP                                                          | Yes                                                                                    | Yes                                                     | No         |
| Features                                                     |                                                                                        |                                                         |            |
| Axial Scan                                                   | Yes                                                                                    | Yes                                                     | No         |
| Helical Scan                                                 | Yes                                                                                    | Yes                                                     | No         |
| Dynamic Scan                                                 | Yes                                                                                    | Yes                                                     | No         |
| Predict Scan                                                 | Yes                                                                                    | Yes                                                     | No         |
| ECG Retrospective Scan<br>(Helical)                          | No                                                                                     | No                                                      | No         |
| ECG Prospective Scan<br>(Axial)                              | No                                                                                     | No                                                      | No         |
| guideShot Scan                                               | Yes                                                                                    | Yes                                                     | No         |
| ITEM                                                         | Supria True64                                                                          | HITACHI Supria PHASE 3 (K163528)                        | Difference |
| Automatic Exposure Control                                   | Yes                                                                                    | Yes                                                     | No         |
| Automatic Exposure Control<br>using Iterative Reconstruction | No                                                                                     | No                                                      | No         |
| ECG Dose Modulation                                          | No                                                                                     | No                                                      | No         |
| Adaptive Filter                                              | No                                                                                     | No                                                      | No         |
| Iterative Reconstruction                                     | Yes. Intelli IP Advanced, Intelli IP Quick                                             | Yes. Intelli IP Advanced, Intelli IP Quick              | No         |
| Injector Synchronization                                     | Yes                                                                                    | Yes                                                     | No         |
| Dose Check                                                   | Yes                                                                                    | Yes                                                     | No         |
| Access Control                                               | Yes                                                                                    | Yes                                                     | No         |
| Automatic Cardiac Phase<br>Search                            | No                                                                                     | No                                                      | No         |
| Preview Scan                                                 | No                                                                                     | No                                                      | No         |
| Double Slice at Axial Scan                                   | No                                                                                     | No                                                      | No         |
| Priority Recon.                                              | No                                                                                     | No                                                      | No         |
| Dose Report                                                  | Yes. Simple Dose Report                                                                | Yes. Simple Dose Report                                 | No         |
| DICOM                                                        | Yes                                                                                    | Yes                                                     | No         |
| ID Reader                                                    | Yes                                                                                    | Yes                                                     | No         |
| Exam Split                                                   | Yes                                                                                    | Yes                                                     | No         |
| Multi-Planar Reconstruction<br>(MPR)                         | Yes                                                                                    | Yes                                                     | No         |
| Volume Rendering                                             | Yes                                                                                    | Yes                                                     | No         |
| CT Angiography (CTA)                                         | Yes                                                                                    | Yes                                                     | No         |
| Segmentation                                                 | Yes                                                                                    | Yes                                                     | No         |
| Retouch                                                      | Yes                                                                                    | Yes                                                     | No         |
| Quality Exam                                                 | Yes                                                                                    | Yes                                                     | No         |
| HiMAR                                                        | Yes                                                                                    | Yes                                                     | No         |
| Orbit synchronization scan                                   | Yes                                                                                    | Yes                                                     | No         |
| Off-time mode                                                | Yes                                                                                    | Yes                                                     | No         |
| On-time standby                                              | Yes                                                                                    | Yes                                                     | No         |
| Shutter Scan Reduction                                       | Yes                                                                                    | Yes                                                     | No         |

#### Table 2 Technological Characteristic Differences

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

The differences from the predicate device to Supria 64 are explained below table.

| Detector      |                                                                                                                        |
|---------------|------------------------------------------------------------------------------------------------------------------------|
| 01            | The 64 lines detector is added.                                                                                        |
| Image Storage |                                                                                                                        |
| 02            | Raw data disk for 64 slices is added because the data amount for each raw data of 64 slices is .bigger than 16 slices. |
| Dose Controls |                                                                                                                        |
| 03            | Helical Beam Pitch is fine-tuned for 64 lines.                                                                         |

Therefore, based on a thorough analysis and comparison of Supra 64 and the predicate device, the technological characteristics do not impact safety and effectiveness.

{9}------------------------------------------------

# Substantial Equivalence

A summary decision was based on analysis of Table 3.

| ITEM                        | Overall Rationale Analysis                                                                                                                          |
|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Gantry                      | There are no significant changes in technology characteristics, hardware, and software from the predicate device, HITACHI Supria PHASE 3 (K163528). |
| Detector                    |                                                                                                                                                     |
| X-ray Tube                  |                                                                                                                                                     |
| X-ray Generator             |                                                                                                                                                     |
| Patient Table               |                                                                                                                                                     |
| Display                     |                                                                                                                                                     |
| Image Storage               |                                                                                                                                                     |
| Scanning,<br>Reconstruction |                                                                                                                                                     |
| Performance                 |                                                                                                                                                     |
| Dose Controls               |                                                                                                                                                     |
| Dose Displays               |                                                                                                                                                     |
| Features                    |                                                                                                                                                     |

Table 3 Rationale Analysis: Supria vs. Predicate

Therefore, based on a thorough analysis and comparison of the functions, scientific concepts, physical and performance characteristics, performance comparison and technological characteristics, the proposed Supria True64 Whole-body X-ray CT System is considered substantially equivalent to the currently marketed predicate device (SUPRIA Whole-body X-ray CT System (K163528)) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.

{10}------------------------------------------------

# Summary of Non-Clinical Testing

This device complies with all applicable requirements for Dose Profile. Noise. Mean CT number and Uniformity, Spatial Resolution, Tomographic Section Thickness and Sensitivity Profile, Tomographic Plane Location, and CT dose index.

Noise, Mean CT number, Uniformity and Spatial Resolution test results for the Supria True64 were compared to results from the Supria PHASE 3 (K163528. Our findings determined the systems to be equivalent. The summary report on these findings is provided in this submission.

In addition, the Supria True64 system is in conformance with the applicable parts of the following standards:

- AAMI ANSI ES60601-1:2005/(R) 2012 and A1:2012, C1:2009/(R)2012 and . A2:2010/(R)2012
Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)

- IEC 60601-1-2 Edition 3: 2007 . Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- IEC 60601-1-3 Edition 2.0 2008-01

Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment

- IEC 60601-2-44 Edition 3.0 2009-02 Medical electrical equipment Part 2-44: particular requirements for the basic safety and essential performance of x-ray equipment for computed tomography.
- NEMA XR 25 Computed Tomography Dose Check
- NEMA XR26

Access Controls for Computer Tomography: Identification, Interlocks, and Logs

- IEC 62304 First edition 2006-05. Medical device software Software life cvcle processes .
# Summary of Clinical Testing

As part of design validation, Hitachi has provided 3 kinds of clinical image examples which we judged to be sufficient to judge a clinical usability.

In addition, a comparative evaluation was conducted of via FBP and Iterative Reconstruction. It resulted in finding the images from the Supria True64 realized both low dose and high quality through reduction of image noise and artifacts.

# Conclusions

Hitachi believes that, based on the information included in the submission, Supria True64 Whole-body X-ray CT System is substantially equivalent with respect to hardware, base elements of the software, safety, effectiveness, and functionality to the SUPRIA Whole-body Xray System Phase 3 (K163528).

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K171738](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K171738)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
