← Product Code [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK) · K143401

# SOMATOM Definition Edge (K143401)

_Siemens Medical Solutions USA, Inc. · JAK · Apr 6, 2015 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K143401

## Device Facts

- **Applicant:** Siemens Medical Solutions USA, Inc.
- **Product Code:** [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK.md)
- **Decision Date:** Apr 6, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Siemens SOMATOM Definition Edge (Project P46F) systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles. (*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)

## Device Story

The SOMATOM Definition Edge is a computed tomography (CT) scanner that acquires x-ray transmission data through continuous rotation of detectors and x-ray tube with simultaneous patient translation. The system uses the syngo VA48 (SOMARIS/7 VA48) software platform to reconstruct cross-sectional images. It is operated by trained healthcare professionals in clinical settings. The device includes advanced features such as TwinBeam scanning, adaptive spiral scanning, iterative reconstruction (ADMIRE, iMAR, MARIS), and dose reduction tools (HandCARE, CARE Dose4D). Output images are used by clinicians for diagnostic evaluation. The device benefits patients by providing high-quality diagnostic imaging while incorporating dose optimization and artifact reduction technologies to improve image clarity and safety.

## Clinical Evidence

Bench testing only. Performance was validated through verification and validation testing of software features and adherence to recognized standards (IEC 60601-2-44, IEC 61223-3-5, NEMA XR-25, NEMA XR-29). No clinical study data was required or provided.

## Technological Characteristics

Computed tomography x-ray system. Software version syngo VA48 (SOMARIS/7 VA48). Features iterative reconstruction (ADMIRE, iMAR, MARIS), dose reduction (HandCARE, CARE Dose4D), and various scanning modes (TwinBeam, Adaptive Spiral). Complies with IEC 60601-1, IEC 60601-2-44, IEC 62304, and NEMA XR-29 standards. Connectivity via DICOM. Operates as a networked medical imaging system.

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- SOMATOM Definition Edge ([K120579](/device/K120579.md))
- SOMATOM Definition AS Open ([K130901](/device/K130901.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three stylized, curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 6, 2015

Siemens Medical Solutions USA, Inc. % Ms. Kimberly Mangum Regulatory Affairs Specialist 51 Valley Stream Parkway MALVERN PA 19355

Re: K143401

Trade/Device Name: SOMATOM Definition Edge Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: February 10, 2015 Received: February 13, 2015

Dear Ms. Mangum:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are a light blue color. The word is displayed horizontally and is the main focus of the image. 

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K143401

Device Name

SOMATOM Definition Edge

Indications for Use (Describe)

The Siemens SOMATOM Definition Edge (Project P46F) systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.

(*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultancous translation of the patient.)

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a light blue or teal. The letters are evenly spaced and appear to be a solid color with no gradients or shadows.

### 510(K) SUMMARY FOR

## SOMATOM Definition Edge Computed Tomography Systems

Submitted by: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355 Date Prepared: January 29, 2015

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.

- 1. General Information: Importer/Distributor Establishment: Registration Number: 2240869 Siemens Medical Solutions, Inc. 51 Valley Stream Pkwy Malvern, PA 19355
Manufacturing Facility: Siemens AG: Medical Solutions Siemensstrasse 1

91301 Forchheim, GERMANY Establishment Registration Number: 3004977335

### 2. Contact Person:

Ms. Kimberly Mangum Technical Specialist, Regulatory Affairs Submissions Siemens Medical Solutions, Inc. USA 51 Valley Stream Parkway D02 Malvern, PA 19355-1406 Phone: (610) 448-4912 Fax: (610) 448-1787 Email: kimberly.mangum@siemens.com

### 3. Device Name and Classification

Product Name: SOMATOM Definition Edge Propriety Trade Name: SOMATOM Definition Edge Classification Name: Computed Tomography X-ray System Classification Panel: Radiology CFR Section: 21 CFR §892.1750 Device Class: Class II Product Code: 90JAK

- 4. Legally Marketed Primary Predicate Device: Product Name: SOMATOM Definition Edge

{4}------------------------------------------------

## SIEMENS

Propriety Trade Name: SOMATOM Definition Edge Classification Name: Computed Tomography X-ray System Classification Panel: Radiology CFR Section: 21 CFR §892.1750 Device Class: Class II Product Code: 90 JAK

Legally Marketed Secondary Predicate Device: Name: SOMATOM Definition AS Open Propriety Trade Name: SOMATOM Definition AS Open Classification Name: Computed Tomography X-ray System Classification Panel: Radiology CFR Section: 21 CFR §892.1750 Device Class: Class II Product Code: 90 JAK

## 5. Indications for Use

The Siemens SOMATOM Definition Edge (Project P46F) systems are intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from either the same axial plane taken at different angles or spiral planes* taken at different angles.

(*spiral planes: the axial planes resulted from the continuous rotation of detectors and x-ray tube, and the simultaneous translation of the patient.)

## 6. Substantial Equivalence:

Siemens SOMATOM Definition Edge Computed Tomography Systems with software version syngo® VA48 (SOMARIS/7 VA48) is substantially equivalent to the following medical devices in commercial distribution as listed in Table 1:

| Manufacturer | Predicate Device           | 510(k) # | Clearance Date  |
|--------------|----------------------------|----------|-----------------|
| Siemens      | SOMATOM Definition Edge    | K120579  | May 23, 2012    |
| Manufacturer | Secondary Device           | 510(k) # | Clearance Date  |
| Siemens      | SOMATOM Definition AS Open | K130901  | January 2, 2014 |

## Table 1: Predicate Devices

### 7. Device Description:

Siemens intends to market a new software version, syngo® VA48 (SOMARIS/7 VA48) for its SOMATOM Definition Edge Computed Tomography X-ray systems. The subject device SOMATOM Definition Edge will be delivered with software version syngo® VA48 (SOMARIS/7 VA48). Additionally software version synqo® VA48 (SOMARIS/7 VA48 will be offered as an optional upgrade for existing SOMATOM Definition Edge systems. syngo® VA48 (SOMARIS/7 VA48) is a further development to the SOMARIS/7 operating software cleared as part of the predicate devices. A listing of device modifications is as follows:

{5}------------------------------------------------

# SIEMENS

- 1. New software version syngo® VA48 (SOMARIS/7 VA48) to support the following additional features:
	- TwinBeam Scanning ●
	- Adaptive Spiral (zig-zag scan) for Sliding Gantry Configuration .
	- Respiratory Analysis of Respiratory Rate & Pitch Adjustment ●
	- FAST 3D Reconstruction (Auto 3D Recon FAST 3D Align) .
	- Multiphase Reconstruction with extended Field of View ●
	- FAST DE Results (Dual Energy PACS ready images) ●
	- FAST Contact ●
	- Iterative Reconstruction with Extended Field of View ●
	- OEM Varian RGSc Online Mode
	- Full 4D Lung Scan ●
	- Applications at CT syngo.via client ●
	- Temporal MIP (t-MIP) .
	- TrueD 4D Viewer .
	- . HD Field of View Pro (HD FoV 2.0)
- 2. ADMIRE Iterative Reconstruction (option)
- 3. iMAR Improved Metal Artifact Correction (option)
- 4. MARIS (Metal Artifact Reduction in Image Space) Option
- 5. HandCARE Quantitative Dose Reduction Option
- 6. CARE Dose4D Dose Reduction Option

### 8. Summary of Technical Characteristics of the Subject Device as Compared with the Predicate Device:

SOMATOM Definition Edge configured with software version syngo® VA48 does not have significant changes in materials, energy source, or technological characteristics when compared to the predicate devices. Both the subject device and predicate devices are computed tomography scanners that support various visualization and evaluation tools. The intended use and fundamental scientific technology are similar to the predicate devices; therefore Siemens believes that they are substantially equivalent to the predicate devices. Table 2 below provides a comparison of the primary features of the subject device in comparison to the predicate device.

| Subject Device Feature                                                                  | Predicate Device Comparable Feature                                                        |
|-----------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|
| New software version syngo® VA48<br>(SOMARIS/7 VA48) with supported<br>software options | System scanner software syngo®<br>VA44 (SOMARIS/7 VA44) with<br>supported software options |
| ADMIRE Iterative Reconstruction<br>(option)                                             | Optional Iterative Reconstruction<br>(option)                                              |
| iMAR Iterative Reconstruction (option)                                                  | Optional Iterative Reconstruction<br>(option)                                              |
| MARIS (option)                                                                          | MARIS (option)                                                                             |
| HandCARE with Quantitative Dose                                                         | HandCARE with Quantitative Dose                                                            |

### Table 2: Predicate Device Comparison

{6}------------------------------------------------

| Subject Device Feature                            | Predicate Device Comparable Feature               |
|---------------------------------------------------|---------------------------------------------------|
| Reduction Option                                  | Reduction Option                                  |
| CARE Dose4D Quantitative Dose<br>Reduction Option | CARE Dose4D Quantitative Dose<br>Reduction Option |

### 9. Nonclinical Testing:

SOMATOM Definition Edge configured with software version syngo® VA48 is designed to fulfill the requirements of the following standards:

- IEC 60601-2-44: Medical electrical equipment Part 2-44: Particular . requirements for the safety of X-ray equipment for computed tomography - Ed. 2.1
- IEC 61223-3-5: Evaluation and routine testing Evaluation and routine ● testing in medical imaging departments - Part 3-5: Acceptance tests -Imaging performance of computed tomography X-ray equipment CORRIGENDUM 1
- NEMA XR-25: Computed Tomography Dose Check ●
- IEC 61223-2-6: Evaluation and routine testing in medical imaging departments - Part 2-6: Constancy tests - Imaging performance of computed tomography X-ray equipment
- NEMA PS 3.1 3.18: Digital Imaging and Communications in Medicine ● (DICOM) Set
- IEC 62304 Ed. 1.0: Medical device software software life cycle ● processes
- . IEC 60601-1: Medical electrical equipment - Part 1: General requirements for Safety, 1988, Amendment 1, 1991-11, Amendment 2, 1995
- ISO 14971: Medical devices Application of risk management to medical ● devices
- NEMA XR-29: Standard Attributes on CT Equipment Related to Dose ● Optimization and Management

This submission contains performance data to demonstrate continued conformance with special controls for medical devices containing software. Non clinical tests were conducted for the SOMATOM Definition Edge configured with software version syngo® VA48 during product development. The modifications described in this Premarket Notification were supported with verification/validation testing.

The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results supports that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.

{7}------------------------------------------------

## SIEMENS

## Software Verification and Validation

Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission.

## Summary

Performance tests were conducted to test the functionality of the SOMATOM Definition Edge configured with software version syngo® VA48. These tests have been performed to test the ability of the included features of the subject device. The results of these tests demonstrate that the subject device performs as intended. The result of all conducted testing was found acceptable to support the claim of substantial equivalence.

### General Safety and Effectiveness Concerns: 10.

The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device. Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development. verification and validation testing. To minimize electrical, mechanical, and radiation hazards. Siemens adheres to recognized and established industry practice and standards.

### 11. Conclusion as to Substantial Equivalence:

The SOMATOM Definition Edge configured with software version syngo® VA48 has the same intended use and comparable indication for use as the predicate devices. The technological characteristics such as image acquisition, operating platform, and image manipulation are similar to the predicate devices

The predicate devices were cleared based on non-clinical supportive information and clinical images. The results of these tests demonstrate that the SOMATOM Definition Edge (K120579) and SOMATOM Definition AS Open (K130901) are adequate for the intended use. The comparison of technological characteristics, non-clinical performance data, and software validation demonstrates that the subject device is as safe and effective when compared to the predicate devices that are currently marketed for the same intended use.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K143401](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K143401)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
