← Product Code [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK) · K123500

# AQUILION LB TRITON V4.91 (K123500)

_Toshibamedical Systems Corporation · JAK · May 2, 2013 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K123500

## Device Facts

- **Applicant:** Toshibamedical Systems Corporation
- **Product Code:** [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK.md)
- **Decision Date:** May 2, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head. The Aquilion has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.

## Device Story

Whole body CT scanner; captures cross-sectional volume data sets. System comprises gantry, patient couch, and peripheral cabinets for data processing/display. Operated by trained/qualified physicians in clinical settings. Input: X-ray signals processed into volume sets. Output: Diagnostic cross-sectional images. Modifications from predicate include increased image reconstruction speed (22 images/sec), higher voltage generator (72kW), upgraded console (faster CPU, increased memory), and inclusion of AIDR 3D dose reduction and QDS noise reduction software. Device assists physicians in diagnostic imaging; provides diagnostic quality images for clinical decision-making.

## Clinical Evidence

No clinical data. Bench testing only. Image quality metrics study using anthropomorphic phantoms (brain, chest, abdomen, pelvis) compared subject device to predicate. Evaluated spatial resolution, CT number, contrast-to-noise ratio, uniformity, noise properties, and low-contrast resolution. Results demonstrated substantial equivalence to predicate.

## Technological Characteristics

Whole body CT scanner; 72kW X-ray generator. Console: 2.53 GHz CPU, 12 GB RAM. Software: AIDR 3D (dose reduction), QDS (noise reduction). Standards: IEC 60601-1 series, IEC 62304, IEC 62366, NEMA PS 3.1-3.18, NEMA XR-25. Connectivity: DICOM (NEMA PS 3.1-3.18).

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Aquilion LB CT Scanner TSX-201A ([K050458](/device/K050458.md))
- Aquilion ONE TSX-301A/2, w/4.74ER ([K113466](/device/K113466.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K123500
Page 1 of 3

SYSTEMS, INC. TOSHIBA AMERICA MEDICA 2441 Michelle Drive Tustin CA 92780 Phone: (714) 730-5000

#### 510(k) SUMMARY

- 1. SUBMITTER'S NAME: Toshiba America Medical Systems, Inc.
- 2. ADDRESS: 2441 Michelle Drive Tustin, CA. 92780-2068
- 3. ESTABLISHMENT REGISTRATION: 2020563

## 4. CONTACT PERSON: Paul Biggins Director, Regulatory Affairs (714) 730-5000

- 5. Date Prepared: November 8, 2012 (Updated: March 22, 2013)
## 6. TRADE NAME(S): Aquilion LB Triton, TSX-201A/2, v4.91

- 7. COMMON NAME: . System, X-ray, Computed Tomography
- 8. DEVICE CLASSIFICATION: Class II (per 21 CFR 892.1750)
- 9. PRODUCT CODE / DESCRIPTION: JAK - System, Computed Tomography

## 10. PERFORMANCE STANDARD:

This device conforms to applicable Performance Standards for lonizing Radiation Emitting Products [21 CFR, Subchapter J, Part 1020]

### 11. PREDICATE DEVICE:

| Product                                                          | Marketed by                        | 510(k) Number | Clearance Date |
|------------------------------------------------------------------|------------------------------------|---------------|----------------|
| Aquilion LB CT Scanner TSX-201A                                  | Toshiba America<br>Medical Systems | K050458       | March 10, 2005 |
| Aquilion ONE TSX-301A/2, w/4.74ER<br>(Dose Reduction Technology) | Toshiba America<br>Medical Systems | K113466       | April 10, 2012 |

2 2013 MAY

{1}------------------------------------------------

K123500
Page 2 of 3

## 12. REASON FOR SUBMISSION:

Modification of a cleared device

## 13. DEVICE DESCRIPTION:

The Aquilion LB Triton, TSX-201A/2, v4.91 is a whole body CT scanner. This device captures cross sectional volume data sets. The device consists of a gantry, patient couch (table) and peripheral cabinets used for data processing and display.

#### 14. SUMMARY OF INTENDED USES:

This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head.

The Aquilion has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.

#### 15. SUBSTANTIAL EQUIVALENCE:

This device is substantially equivalent to the Aquilion LB CT Scanner, TSX-201A, K050458, marketed by Toshiba America Medical Systems. The Aquillion LB Triton, TSX-201A/2, v4.91, includes modifications to the cleared device which improves image reconstruction speed and includes the addition of optional software features. There is no change to the imaging chain in the gantry. The method of operation, base software and manufacturing process remain unchanged from the cleared device.

Additionally, this device is substantially equivalent to the Aquilion ONE TSX-301A/2, w/4.74ER, K113466, in that it includes the previously cleared dose reduction technology, AIDR 3D.

| Item                                    | Aquilion LB Triton<br>TSX-201A/2, v4.91                                           | Aquilion LB CT<br>Scanner TSX-201A<br>(K050458)                              | Aquilion ONE TSX-<br>301A/2, w/4.74ER<br>(K113466)                              |
|-----------------------------------------|-----------------------------------------------------------------------------------|------------------------------------------------------------------------------|---------------------------------------------------------------------------------|
| Image reconstruction<br>(maximum speed) | 22 images per second                                                              | 12 images per second                                                         | 30 images per second                                                            |
| X-ray Generator<br>Output Power         | 72kW Maximum                                                                      | 60kW Maximum                                                                 | 70kW Maximum                                                                    |
| Dose Reduction                          | SureExposure 3D                                                                   | Real EC                                                                      | SureExposure 3D                                                                 |
| Noise Reduction<br>Processing           | QDS<br>AIDR 3D                                                                    | N/A                                                                          | QDS<br>AIDR 3D                                                                  |
| Console                                 | • 2.53 GHz CPU Clock<br>Frequency<br>• 12 GB or more CPU<br>Memory<br>• DVD Drive | • 2 GHz CPU Clock<br>Frequency<br>• 2 GB or more CPU<br>Memory<br>• CD Drive | • 3.0 GHz CPU Clock<br>Frequency<br>• 8 GB or more CPU<br>Memory<br>• DVD Drive |

Summary of Changes:

Page 2 of 3

{2}------------------------------------------------

K123500
page 3 of 3

- 1. Increased image reconstruction speed
- 2. High voltage generator is increased
- ని. Noise reduction processing software is now a standard feature
- 4. The computer console has been upgraded, which includes a faster CPU, additional CPU memory, a DVD Drive.
- ട. An optional Display Console Kit is now available.
- 6. Previously cleared software options are being added

#### 16. SAFETY:

The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the following standards: IEC60601-1, IEC60601-1-1, IEC60601-1-2, IEC60601-1-3, IEC60601-1-4, IEC60601-1-6, IEC60601-1-8, IEC60601-1-9, IEC60601-2-28. IEC60601-2-32. IEC60601-2-44. IEC60825-1. IEC62304. IEC62366. NEMA PS 3.1-3.18 and NEMA XR-25.

Additionally, this device complies with all applicable requirements of the radiation safety performance standards, as outlined in 21 CFR §1010 and §1020.

#### 17. TESTING

There is no change to the imaging chain in the gantry (i.e. geometry, detector/DAS, tube, beam limiting device). Image comparison data using brain, chest, abdomen and pelvis anthropomorphic phantoms, utilizing clinical scan conditions and reconstruction parameters, demonstrated that the subject device produces substantially equivalent images of representative structures as compared to the predicate device, Aquilion LB CT Scanner TSX-201A.

An image quality metrics study utilizing phantoms validated that the subject device is substantially equivalent to the predicate device (Aquilion LB CT Scanner TSX-201A) with regard to spatial resolution, CT number and contrast-to-noise ratio, uniformity, noise properties and low contrast resolution.

Additionally, representative diagnostic images including brain, chest, abdomen and peripheral exams were obtained using the subject device which demonstrates that the device produces images of diagnostic quality and; therefore, performs as intended.

Testing of the modified system was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices and CT Systems with satisfactory results.

#### 18. CONCLUSION

The modifications incorporated into the Aquilion LB Triton, TSX-201A/2, v4.91 do not change the indications for use or the intended use of the device. Based upon this information, conformance to standards, application of design controls and the performance data presented in this submission, Toshiba America Medical Systems, believes that the Aquilion LB Triton, TSX-201A/2, v4.91, is substantially equivalent in safety and effectiveness to the predicate devices.

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing protection and care.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 2. 2013

TOSHIBA AMERICA MEDICAL SYSTEMS, INC. % Mr. Paul Biggins Director, Regulatory Affairs 2441 Michelle Drive TUSTIN CA 92780

Re: K123500

Trade/Device Name: Aquilion LB Triton, TSX-201A/2, v4.91 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: March 22, 2013 Received: March 28, 2013

Dear Mr. Biggins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Biggins

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Sm-7)

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

## 510(k) Number (if known): K123500

Device Name: Aquilion LB Triton, TSX-201A/2, v4.91

### Indications for Use:

This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head.

The Aquilion has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician.

Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Sm. 7)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health

> K123500 510(k)

> > Page 1 of 1

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K123500](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K123500)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
