← Product Code [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK) · K082761

# GE LIGHTSPEED 7.2 CT SCANNER SYSTEM (K082761)

_Ge Medical Systems, LLC · JAK · Oct 8, 2008 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K082761

## Device Facts

- **Applicant:** Ge Medical Systems, LLC
- **Product Code:** [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK.md)
- **Decision Date:** Oct 8, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The GE LightSpeed 7.2 Computed Tomography X-ray system is intended to produce cross-sectional images of the body by computer reconstruction of xray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These mages may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports. components and accessories. This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results. The GE LightSpeed 7.2 CT Scanner System is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. The device output is a valuable medical tool for the diagnosis of disease. trauma, or abnormality and for planning, guiding, and monitoring therapy.

## Device Story

GE LightSpeed 7.2 CT Scanner System is an evolutionary modification of the 64-slice LightSpeed 7.1 system. Device consists of gantry, patient table, operator console, computer, and PDU. System acquires X-ray transmission data at various angles and planes (Axial, Cine, Helical, Cardiac, Gated); computer reconstructs data into cross-sectional images. Images may be obtained with or without contrast. System includes signal analysis, display equipment, and image processing tools for trans-axial and reformatted planes. Used in clinical settings by healthcare professionals for diagnosis of disease, trauma, or abnormality and for therapy planning, guidance, and monitoring. Modifications from predicate include updates to hardware, applications software, firmware, reconstruction, and scan modes.

## Clinical Evidence

Bench testing only. System verification and validation were performed to ensure performance to specifications, federal regulations, and user requirements. Compliance with international safety standards (UL/CSA, IEC) and CDRH 21 CFR Subchapter J requirements was verified.

## Technological Characteristics

64-slice CT scanner system comprising gantry, patient table, operator console, computer, and PDU. Compliant with UL 60601-1, IEC 60601-1, and 21 CFR 1020.30/33. Utilizes X-ray transmission data acquisition and computer reconstruction. Materials and construction are equivalent to previously marketed GE CT systems.

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- LightSpeed 7.1 CT Scanner System ([K061817](/device/K061817.md))
- LightSpeed 7.0 CT Scanner System ([K040372](/device/K040372.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

OCT 0 8 2008

K082761/

GE Healthcare

3000 N. Grandview Blvd.
Waukesha, WI 53188

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h).

| Date Prepared:     | September 12, 2008                                                                                                                 |
|--------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Submitter:         | GE Medical Systems, LLC (GE Healthcare)<br>3000 N. Grandview Blvd., W-1140<br>Waukesha, WI 53188                                   |
| Primary Contact:   | Tracey Ortiz<br>Regulatory Affairs Leader<br>GE Healthcare<br>Tel: 262-544-3601, Fax: 262-918-1609<br>e-mail: tracey.ortiz@ge.com  |
| Secondary Contact: | John Jaeckle<br>Regulatory Affairs Manager<br>GE Healthcare<br>Tel: 262-312-7358, Fax: 262-312-7369<br>e-mail: john.jaeckle@ge.com |

### DEVICE IDENTIFICATION

| Trade Name:          | GE LightSpeed 7.2 CT Scanner System                                                      |
|----------------------|------------------------------------------------------------------------------------------|
| Common/Usual Name:   | LightSpeed VCT, LightSpeed VCT XT                                                        |
| Classification Name: | Computed Tomography X-ray System per<br>21CFR892.1750                                    |
| Product Code:        | 90-JAK                                                                                   |
| Predicate Device(s): | LightSpeed 7.1 CT Scanner System (K061817)<br>LightSpeed 7.0 CT Scanner System (K040372) |

Image /page/0/Picture/9 description: The image shows the General Electric (GE) logo, which consists of the letters 'G' and 'E' intertwined within a circular border. The letters are stylized with curved lines, giving them a classic and recognizable appearance. The logo is presented in black and white, with the letters and border appearing in black against a white background.

{1}------------------------------------------------

| Manufacturer(s):  | GE Medical Systems, LLC (GE Healthcare)<br>3000 N. Grandview Blvd.<br>Waukesha, WI 53188                                                                                                                                                                                                           |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                   | GE Yokogawa Medical Systems, LTD.<br>7-127, Asahigaoka 4 Chome<br>Hino-Shi, Tokyo, 191-8503  Japan                                                                                                                                                                                                 |
| Distributor:      | Same as Manufacturer                                                                                                                                                                                                                                                                               |
| Marketed Devices: | The LightSpeed 7.2 CT Scanner System is of<br>comparable type and substantially equivalent to GE's<br>currently marketed Computed Tomography X-ray<br>Systems that comply with the same or equivalent<br>standards and have similar intended uses, such as the<br>previous LightSpeed CT Scanners. |

# DEVICE DESCRIPTION

The GE LightSpeed 7.2 CT Scanner System is composed of a gantry, patient table, operator console, computer, and PDU and includes image acquisition hardware, image acquisition and reconstruction software, associated accessories and connections/interfaces to accessories. It is an evolutionary modification to the LightSpeed 7.1 (K061817). It is developed from the hardware platform of the 64 slice LightSpeed 7.1 system by adding new application features that involve changes in hardware, software, firmware, recon, and scan mode.

The GE LightSpeed 7.2 CT Scanner System is designed to be a head and whole body CT scanner incorporating the same basic fundamental operating principles and similar Indications for Use. Materials and construction are equivalent to our existing marketed products, which are compliant with UL 60601-1, IEC 60601-1 and associated collateral and particular standards, and 21CFR Subchapter J.

## INDICATIONS FOR USE

The GE LightSpeed 7.2 Computed Tomography X-ray system is intended to produce cross-sectional images of the body by computer reconstruction of xray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These mages may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports. components and accessories.

{2}------------------------------------------------

This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results.

The GE LightSpeed 7.2 CT Scanner System is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.

The device output is a valuable medical tool for the diagnosis of disease. trauma, or abnormality and for planning, guiding, and monitoring therapy.

#### COMPARISON WITH PREDICATE

The GE LightSpeed 7.2 CT Scanner System is developed from the hardware platform of the 64 slice LightSpeed 7.1 system (K061817). The GE LightSpeed 7.2 CT Scanner System involves changes from the LightSpeed 7.1 system to add new application features that involve changes in hardware, applications software, firmware, recon, and scan mode. The GE LightSpeed 7.2 CT Scanner System uses virtually the same materials, identical operating principles, and has similar indications for use as our existing marketed product, LightSpeed 7.1. We believe the GE LightSpeed 7.2 CT Scanner System is of comparable type and substantially equivalent to our currently marketed systems listed above, complies with the same or equivalent standards, and has the same intended uses.

The LightSpeed 7.2 CT Scanner System will be certified to comply with the X-ray requirements of 21CFR1020.30 and1020.33, as well as the safety requirements of UL 60601-1, and IEC 60601-1 and associated collateral and particular standards.

#### ADVERSE EFFECTS ON HEALTH

Potential electrical, mechanical and radiation hazards are identified in a risk management summary (hazard analysis) and controlled by:

- . System verification and validation to ensure performance to specifications, Federal Regulations, and user requirements.
- Adherence and certification to industry and international . standards. (UL/CSA and IEC).
- Compliance to applicable CDRH 21CFR subchapter J . requirements.

The device is designed and manufactured under the Quality System Regulations of 21CFR820.

{3}------------------------------------------------

# CONCLUSION

The GE LightSpeed 7.2 CT Scanner System is an evolutionary modification to the 64 slice LightSpeed 7.1 system (K061817). It does not result in any new potential safety risks and performs as well as or better than devices currently on the market. The GE LightSpeed 7.2 CT Scanner system will be certified to comply with the X-ray requirements of 21CFR 1020.30 and 1020.33, as well as the safety requirements of UL 60601-1, and IEC 60601-1 and associated collateral and particular standards. GE Healthcare considers the GE LightSpeed 7.2 CT Scanner Systems to be as safe, as effective, and to have substantially equivalent performance to the predicate devices(s).

Section 5: Page 5 of 5

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services (USA). The logo features a stylized eagle with three tail feathers, representing the department's commitment to health, human services, and well-being. The eagle is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES (USA)", which is arranged in a circular fashion around the logo.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

# OCT 0 8 2008

GE Medical Systems, LLC % Mr. Daniel W. Lehtonen Responsible Third Party Official Intertek Testing Services 2307 E. Aurora Rd., Unit B7 TWINSBURG OH 44087

Re: K082761

Trade/Device Name: GE LightSpeed 7.2 CT Scanner System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: September 19, 2008 Received: September 22, 2008

Dear Mr. Lehtonen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|----------------|----------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology)                      | 240-276-0120 |
| Other          |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

hope M. Whang

Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

#### STATEMENT OF INDICATIONS FOR USE

510(k) Number (if known):

K082761

Device Name:

GE LightSpeed 7.2 CT Scanner System

Indications for Use:

The GE LightSpeed 7.2 Computed Tomography X-ray system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.

This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results.

The GE LightSpeed 7.2 CT Scanner System is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.

The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.

Prescription Use X AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (Part 21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Abdominal an

Radiological Devices
510(k) Number K082761

Section 4: Page 2 of 2

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K082761](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K082761)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
