LIGHTSPEED 7.1 CT SCANNER SYSTEM

K061817 · Ge Medical Systems, LLC · JAK · Jul 12, 2006 · Radiology

Device Facts

Record IDK061817
Device NameLIGHTSPEED 7.1 CT SCANNER SYSTEM
ApplicantGe Medical Systems, LLC
Product CodeJAK · Radiology
Decision DateJul 12, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1750
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The LightSpeed 7.1 CT Scanner System is indicated for head and whole body Xray Computed Tomography applications.

Device Story

LightSpeed 7.1 CT Scanner System; evolutionary modification of LightSpeed 7.0 (K040372). System components: gantry, patient table, operator console, computer, PDU. Inputs: X-ray signals acquired during patient scanning. Processing: image acquisition and reconstruction software/firmware transforms raw X-ray data into diagnostic images. Output: CT images for clinical review. Usage: clinical environment; operated by trained radiology personnel. Healthcare providers use output for diagnostic decision-making; enables visualization of internal anatomy for clinical assessment.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing, system verification, and validation to ensure performance meets specifications, federal regulations, and user requirements.

Technological Characteristics

Computed Tomography X-ray system; gantry-based architecture. Compliant with UL 60601-1, IEC 60601-1, and 21 CFR 1020.30/1020.33. Includes image acquisition hardware and reconstruction software. Materials and construction equivalent to predicate LightSpeed 7.0. System is a modification of the 32/64 slice platform.

Indications for Use

Indicated for head and whole body X-ray Computed Tomography applications in patients requiring diagnostic imaging.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ JUL 1 2 2006 R661817 GE Healthcare 3000 N. Grandview Blvd. W-1140 Waukesha, WI 53188 # 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h). Submitter: John Jaeckle Regulatory Affairs Program Manager- MI & CT Tel: 262-312-7358, Fax: 262-312-7369 e-mail: John.Jaeckle@med.ge.com Date Prepared: June 13, 2006 ### PRODUCT IDENTIFICATION Name: LightSpeed 7.1 CT Scanner System - Classification Name: Computed Tomography X-ray System 21CFR892.1750, 90-JAK GE Medical Systems LLC (GE Healthcare) Manufacturer: 3000 N. Grandview Blvd. Waukesha, WI 53188 > GE YOKOGAWA MEDICAL SYSTEMS 7-127 Asahigaoka 4-chome Hino-shi, Tokyo, JAPAN 191 Distributor: Same as Manufacturer Marketed Devices: The LightSpeed 7.1 CT Scanner System is of comparable type and substantially equivalent to GE's currently marketed Computed Tomography X-ray Systems that comply with the same or equivalent standards and have the same intended uses, such as the previous LightSpeed CT Scanners. {1}------------------------------------------------ #### DEVICE DESCRIPTION The LightSpeed 7.1 CT Scanner System is composed of a gantry, patient table, operator console, computer, and PDU and includes image acquisition hardware, image acquisition and reconstruction software, associated accessories and connections/interfaces to accessories. The LightSpeed 7.1 Scanner System is an evolutionary modification to LightSpeed 7.0 (K00372) It is developed from the hardware platform of LightSpeed 7.0 32/64 slice system by adding new application features that involve changes in software, firmware, recon and scan mode. The LightSpeed 7.1 Scanner System is designed to be a head and whole body CT scanner incorporating the same basic fundamental operating principles and Indications for Use. Materials and construction are equivalent to our existing marketed products, which are compliant with UL 60601-1, IEC 60061-1 and associated collateral and particular standards, and 21CFR Subchapter J. #### Indications for Use: The LightSpeed 7.1 Scanner System is indicated for head and whole body X-ray Computed Tomography applications. #### Comparison with Predicate: The LightSpeed 7.1 Scanner System is developed from the hardware platform of our LightSpeed 7.0 32/64 slice system (K040372). The LightSpeed 7.1 involves changes from the LightSpeed 7.0 system to add new application features that involve changes in application software, firmware, recon and scan mode. The LightSpeed 7.1 Scanner System uses the same materials and operating principle as our existing marketed product, LightSpeed 7.0, as well as having identical indications for use. We believe the LightSpeed 7.1 Scanner System is of comparable type and substantially equivalent to currently marketed system listed above and complies with the same or equivalent standards and have the same intended uses. LightSpeed 7.1 Scanner System will be certified to comply with the X-ray requirements of 21CFR1020.30 and 1020.33, as well as the safety requirements of UL 60601-1, and IEC 60601-1 and associated collateral and particular standards. #### Adverse Effects on Health: Potential electrical, mechanical and radiation hazards are identified in a risk management summary (hazard analysis) and controlled by: - . System verification and validation to ensure performance to specifications, Federal Regulations, and user requirements. - Adherence and certification to industry and international standards. (UL/CSA and IEC). . - Compliance to applicable CDRH 21CFR subchapter J requirements. . The device is designed and manufactured under the Quality System Regulations of 21CFR820. {2}------------------------------------------------ #### CONCLUSIONS The LightSpeed 7.1 CT Scanner System is developed from the hardware platform of our LightSpeed 7.0 32/64 slice system (K040372) and does not result in any new potential safety risks and performs as well as or better than devices currently on the market. GE considers the LightSpeed 7.1 CT Scanner System to be equivalent to other marketed devices with the same indications for use and meeting similar standards. {3}------------------------------------------------ Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850 ### JUL 1 2 2006 GE Medical Systems LLC (GE Healthcare) % Mr. Neil E. Devine, Jr. Responsible Third Party Official Intertek Testing Services NA, Inc. 2307 East Aurora Road, Unit B7 TWINSBURG OH 44087 Re: K061817 Trade/Device Name: LightSpeed 7.1 CT Scanner System Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: June 27, 2006 Received: June 28, 2006 #### Dear Mr. Devine: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Image /page/3/Picture/9 description: The image shows a circular logo with the text "1906-2006" at the top and "Centennial" in the middle. The letters "FDA" are prominently displayed in the center of the logo. Three stars are arranged in a horizontal line below the word "Centennial". The text around the circle is difficult to read due to the image quality. *Protecting and Promoting Public Health* {4}------------------------------------------------ Page 2 - Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 | |----------------|---------------------------------|--------------| | 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 894.xxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Indications for Use 510(k) Number (if known): Device Name: LightSpeed 7.1 CT Scanner System Indications For Use: The LightSpeed 7.1 CT Scanner System is indicated for head and whole body Xray Computed Tomography applications. Prescription Use (Part 21 CFR 801 Subpart D) ANDIOR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BEL.OW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) David R. Ingram (Division Sian-Division of I and Radiolor 510(k) Num Page 1 of 1 (32
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...