HITACHI PRESTO

K040902 · Hitachi Medical Systems America, Inc. · JAK · Apr 21, 2004 · Radiology

Device Facts

Record IDK040902
Device NameHITACHI PRESTO
ApplicantHitachi Medical Systems America, Inc.
Product CodeJAK · Radiology
Decision DateApr 21, 2004
DecisionSESE
Submission TypeAbbreviated
Regulation21 CFR 892.1750
Device ClassClass 2
Attributes3rd-Party Reviewed

Indications for Use

The Presto Computed Tomography system is an x-ray imaging device that produces cross-sectional images of the body at different angles. The system reconstructs, processes, displays, and stores the collected images. The device output can provide an aid to diagnosis when used by a qualified physician.

Device Story

Presto is a multi-slice CT system; uses third-generation technology with rotating x-ray tube and detector assembly; collects up to 4 slices of data simultaneously; operates in helical (spiral) scan mode. Input: x-ray data; Output: cross-sectional images. Used in clinical settings by qualified physicians. System includes gantry, operator workstation, patient table, and high-frequency generator. Workstation uses PC technology with Windows OS for image reconstruction, display, storage, and archiving. Benefits include reduced scan times compared to single-slice systems due to simultaneous 4-slice acquisition.

Clinical Evidence

Bench testing only. Evaluations included dose profile, image noise, modulation transfer function (MTF), slice thickness, sensitivity profile, slice plane location, and CT dose index per 21 CFR 1020.33(c). Results were comparable to the predicate device.

Technological Characteristics

Third-generation CT; solid-state detector assembly; high-frequency x-ray generator; fan beam output; helical scan mode. PC-based workstation running Windows OS. 4-slice simultaneous acquisition. Gantry with slip-ring technology. Complies with 21 CFR 1020.33(c) performance standards.

Indications for Use

Indicated for use as an x-ray imaging device to produce cross-sectional images of the body for diagnostic aid by a qualified physician.

Regulatory Classification

Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ APR 2 1 2004 K040902 # HITACHI #### HITACHI MEDICAL SYSTEMS AMERICA, INC. 1959 Summit Commerce Park Twinsburg, Ohio 44087-2371 Tel.: 330.425.1313 Fax: 330.425.1410 # 510(k) Summary # Submitter Information | Submitter: | Hitachi Medical Systems America, Inc.<br>1959 Summit Commerce Park<br>Twinsburg, Ohio 44080-2371<br>ph: (330) 425-1313<br>fax: (330) 425-1410 | |------------|-----------------------------------------------------------------------------------------------------------------------------------------------| | Contact: | Douglas J. Thistlethwaite | | Date: | February 24, 2004 | ## Device Name | Device Name: | Computed tomography x-ray system | |-------------------------|-------------------------------------| | Trade/Proprietary Name: | Presto | | Common Name: | Computed Tomography X-ray System | | Classification Name: | System, X-Ray, Tomography, Computed | | Classification Number: | Sec. 892.1750 | ### Predicate Device Predicate Device: Hitachi Pronto CT, 510(k) K014240 {1}------------------------------------------------ # Device Description #### Function The Presto is a multi-slice computed tomography system that uses x-ray data to produce cross-sectional images of the body at various angles. #### Scientific Concepts The Presto system uses "third generation" CT technology, where the x-ray tube and detector assemblies are mounted on a frame that rotates continuously around the patient using slip ring technology. The solid-state detector assembly design collects up to 4 slices of data simultaneously. The x-ray sub-system features a high frequency generator, xray tube, and collimation system that produces a fan beam x-ray output. The system can operate in a helical (spiral) scan mode where the patient table moves during scanning. As the x-ray tube/detector assembly rotates around the patient, data is collected at multiple angles. The collected data is then reconstructed into cross-sectional images by a high-specd reconstruction sub-system. The images are displayed on a computer workstation, stored, printed, and archived as required. The workstation is based on current PC technology using the Windows™ operating system. #### Physical and Performance Characteristics The Presto system consisting of a gantry, operator's workstation, patient table, high-frequency x-ray generator, and accessories. The system performance is similar to the predicate device. #### Performance Comparison Because the Presto and the predicate device are both Hitachi designs, they were subjected to the same non-clinical cvaluations as stipulated in 21 CFR 1020.33(c). Evaluations include: dose profile, image noise, modulation transfer function (MTF), slice thickness and sensitivity profile, slice plane location, and CT dose index. The evaluation results of the Presto were comparable to the predicate device and support our conclusion that the Presto CT system is substantially equivalent. # Device Intended Use The Presto Computed Tomography system is an x-ray imaging device that produces cross-sectional images of the body at different angles. The system reconstructs, processes, displays, and stores the collected images. The device output can provide an aid to diagnosis when used by a qualified physician. {2}------------------------------------------------ # Device Technological Characteristics The Presto CT system acquires data in the same manner as the predicate device. Physically, the Presto is very similar to the predicate device. The key differences are the ability to collect 4 slices in a single scan as well as improvements in overall technology. The ability to collect 4 slices in a single scan allows overall scan time to be decreased but does not change the essential characteristics of the finished images. The predicate is a single slice design, meaning that if an area of 16 mm is to be examined in 2 mm increments, the x-ray tube must scan the patient 8 times, collecting 2 mm of data for each scan, to produce 8 total images. In the Presto's 4 slice design, the system need only to scan 2 times, collecting 8 mm of data during each scan. The x-ray beam is collimated to allow the exposure of 4 slices simultaneously, and the data collection system collects all 4 slices. Since the data collection system processes the data in 2 mm increments, the system produces 8 images as before, but during a shorter time. The operation of the system is virtually identical to the predicate both systems were produced using the same essential design concepts. The Presto operating system software is essentially the same, as well as the user interface. The patient table design is the same, with the exception that the weight limit was increased somewhat. Gantry controls provide the same features as the predicate, but the control layout was updated. Despite these differences, the Presto CT system is technologically equivalent in concept, function, and performance to the predicate device. # Conclusions The Presto CT system has been developed and validated according to applicable standards. Testing has proven that the system is safe and effective for the indicated use. Risk and hazard analysis shows that there are no new safety issues associated with this system as compared with the predicate device. {3}------------------------------------------------ Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines representing the wings and body. The logo is rendered in black and white. APR 2 1 2004 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Hitachi Medical Systems America, Inc. c/o Mr. Daniel Lehtonen Staff Engineer - Medical Devices Intertek Testing Services NA, Inc. 70 Codman Hill Road BOXBOROUGH MA 01779 Re: K040902 Trade/Device Name: Presto Computed Tomography X-ray System Regulation Number: 21 CFR §892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 90 JAK Dated: April 6, 2004 Received: April 7, 2004 Dear Mr. Lehtonen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (sec above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to ocgin matters of substantial equivalence of your device to a legally premarket notification. THCPDA intellight of the more and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice tor your dones as one of the following numbers, based on the regulation number at the top of the letter: | 8xx. 1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Additionally, for questions on the prosses note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general by releveloe to promation to the Act may be obtained from the Division of Strail Information on your responsional and Consumer Assistance at its toll-free number (800) 638-2041 or Mandracturers, International and Solasmiller Schools. Ida.gov/cdrh/dsmaldsmamain.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure Page 2 {5}------------------------------------------------ K040902 510(k) Number (if known): Presto Computed Tomography X-ray System Device Name: Indications for Use: The Presto Computed Tomography system is an x-ray imaging device that produces cross-sectional images of the The I resto Computed Tonography System a art, processes, displays, and stores the collected images. The device output can provide an aid to diagnosis when used by a qualified physician. (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) 6:10 1 (Division Sign-Off) Division of Reproductive, Abdominati and Radiological Devices 510(k) Number Over-the-Counter Use ાર Prescription Use
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...