← Product Code [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK) · K032915

# ADVANTAGE 4D OPTION (K032915)

_Ge Medical Systems, LLC · JAK · Oct 30, 2003 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K032915

## Device Facts

- **Applicant:** Ge Medical Systems, LLC
- **Product Code:** [JAK](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK.md)
- **Decision Date:** Oct 30, 2003
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1750
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

Advantage 4D is a non-invasive software / hardware option that can be used to provide and display CT images of all phases of a breathing cycle for the evaluation of respiration-induced motion. The software will allow the user to retrospectively define the best respiratory phase from an image quality standpoint, and group images by the phase selected. Advantage 4D can also be used for target or treatment volume (DICOM Radiation Therapy Structure Sets) verification

## Device Story

Advantage 4D is a CT software/hardware option for respiratory motion management. It synchronizes CT data acquisition with an external respiratory signal to capture organ/tissue motion. The system reconstructs images across all breathing phases, allowing clinicians to retrospectively select the optimal phase for image quality and group images accordingly. It also displays DICOM Radiation Therapy Structure Sets for target/treatment volume verification. Used in clinical settings by radiologists or radiation oncologists, the device enables visualization of motion-free images, facilitating more accurate treatment planning and target verification, ultimately benefiting patients by reducing motion-related artifacts in CT scans.

## Clinical Evidence

No clinical data provided. Safety and effectiveness are supported by software development, validation, and verification processes, and adherence to industry and international standards.

## Technological Characteristics

Software/hardware option for CT systems. Features include respiratory signal synchronization, phase-based image reconstruction, Cine Phase Movie display, and DICOM RTSS overlay. Operates as an accessory to CT systems (21 CFR 892.1750).

## Regulatory Identification

A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Smart Breath Respiratory Compensation ([K022919](/device/K022919.md))
- Advantage Sim 6.0 ([K021780](/device/K021780.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows handwritten text on a white background. The text includes the identifier "K032915" at the top, followed by "Page 1 of 2" below it. The handwriting appears to be cursive and slightly slanted.

Image /page/0/Picture/2 description: The image shows the General Electric (GE) logo. The logo consists of the letters "G" and "E" intertwined in a stylized, cursive font. The letters are enclosed within a circular border that has a swirling or wave-like pattern around the perimeter. The logo is black and white.

GE Medical Systems 3000 N. Grandview Blvd Waukesha, WI 53188

# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).

Larry A. Kroger, Ph.D. Submitter Senior Regulatory Program Manager Telephone: (262) 544-3894, FAX: (262) 548-4768 Date Prepared: August 8, 2002

# PRODUCT IDENTIFICATION

| Name:                     | Advantage 4D option                                                                                                                                                                                                                                                                                                  |  |  |  |  |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
| Classification Name:      | Accessory to Computed Tomography System                                                                                                                                                                                                                                                                              |  |  |  |  |
| Classification Panel      | 892 - Radiology                                                                                                                                                                                                                                                                                                      |  |  |  |  |
| Classification<br>Number: | 892.1750                                                                                                                                                                                                                                                                                                             |  |  |  |  |
| Manufacturer :            | GE Medical Systems<br>3000 N. Grandview Blvd<br>Waukesha, WI 53188, USA                                                                                                                                                                                                                                              |  |  |  |  |
| Distributor:              | General Electric Medical Systems, Milwaukee, WI                                                                                                                                                                                                                                                                      |  |  |  |  |
| Marketed Devices:         | Option is substantially equivalent to:<br>Model: Smart Breath Respiratory Compensation<br>Manufacturer: General Electric Medical Systems<br>510(k) #: K022919<br>Option also incorporates features derived from:<br>Model: Advantage Sim 6.0.<br>Manufacturer: General Electric Medical Systems<br>510(k) #: K021780 |  |  |  |  |

### Device Description:

Advantage 4D enables a patient to be scanned on a CT with normal breathing. CT data are acquired and images are reconstructed without image artifacts due to organ and tissue motion. CT Images are synchronized with respiratory signal coming from external device and organ/tissue motion within the CT data reflecting both the organ motions and the chest motion.

Page 1/2

Image /page/0/Picture/12 description: The image shows the words "ATTACHMENT A" in a bold, sans-serif font. The text is set against a textured background, which appears to be a mix of light and dark shades, possibly indicating a grainy or patterned surface. The words are all capitalized and evenly spaced, suggesting a title or heading for a document or section.

{1}------------------------------------------------

K032915
Page 2 of 2

Image /page/1/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circular shape. The letters and the circle are in black, set against a white background. The logo has a classic and recognizable design.

K∅32915

GE Medical Systems 3000 N. Grandview Blvd Waukesha, WI 53188

The user can then visually determine the optimum phase to perform volume imaging and post processing [quantification, contouring, segmentation. The respiratory phase identification is also provided.

In addition. Advantage 4D has the capability to display DICOM Radiation Therapy Structure Sets and can be used for target or treatment volume verification. This new capability comes from features derived from Advantage Sim 6.0 (K021780).

# Indications for Use:

Advantage 4D is a non-invasive software / hardware option that can be used to provide and display CT images of all phases of a breathing cycle for the evaluation of respiration-induced motion. The software will allow the user to retrospectively define the best respiratory phase from an image quality standpoint, and group images by the phase selected. Advantage 4D can also be used for target or treatment volume (DICOM Radiation Therapy Structure Sets) verification

# Comparison with Predicate:

This device will use a similar technology to that already used by GE Medical Systems in our cleared devices Smart Breath Respiratory Compensation (K022919) and Advantage Sim 6.0 (K021780). Fundamentally, this device will use the same phase selection process that is used in Smart Breath Respiratory Compensation to group images based on respiratory phase. In addition, Advantage 4D will have the ability to display reformat images through a Cine Phase Movie, and to overlay RTSS (Radiotherapy Structure Sets) contours over multiphase CT slices. These additional features are inherited from the viewer of Advantage Sim 6.0 (K021780). A more detailed comparison is included in Attachment C of this submission.

| Device Name                           | FDA Clearance Number |
|---------------------------------------|----------------------|
| Smart Breath Respiratory Compensation | K022919              |
| Advantage Sim 6.0                     | K021780              |

# Adverse Effects on Health :

The potential hazards are identified in a risk management summary (hazard analysis) and are controlled by:

- · Software Development, Validation and Verification Process to ensure performance to specifications, Federal Regulations and user requirements.
- · Adherence to industry and international standards.

### Conclusions:

Advantage 4D Option does not result in any new potential safety risks and performs as well as devices currently on the market. GE considers features and technology of the Advantage 4D Option to be equivalent to those of Smart Breath Respiratory Compensation []K022919 and Advantage Sim 6.0 [K021780.

Page 2/2

Image /page/1/Picture/18 description: The image shows the words "ATTACHMENT A" in a black box. The words are written in a sans-serif font and are centered in the box. The background of the box is a light gray color. The words are written in all capital letters.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular design with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines suggesting movement or connection.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 3 0 2003

GE Medical Systems % Mr. Juergen Welte Responsible Third Party TUV Rheinland of North America 12 Commerce Road NEWTOWN CT 06470

Re: K032915

Trade/Device Name: Advantage 4D Option Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: 90 JAK Dated: October 14, 2003 Received: October 15, 2003

Dear Mr. Welte:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter:

| 8xx. 1xxx                        | (301) 594-4591 |
|----------------------------------|----------------|
| 876.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4616 |
| 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 |
| 892.2xxx, 3xxx, 4xxx, 5xxx       | (301) 594-4654 |
| Other                            | (301) 594-4692 |

Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html.

Sincerely vours.

Nancy C. Brogdon

Nancv C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. There are four decorative flourishes, one in each quadrant around the central letters.

GE Medical Systems 3000 N. Grandview Blvd Waukesha, W1 53188

### STATEMENT OF INTENDED USE

510(k) Number (if known): Ka32915

Device Name: Advantage 4D

# Indications For Use:

Advantage 4D is a non-invasive software / hardware option that can be used to provide and display CT images of all phases of a breathing cycle for the evaluation of respiration-induced motion. The software will allow the user to retrospectively define the best respiratory phase from an image quality standpoint, and group images by the phase selected. Advantage 4D can also be used for target or treatment volume (DICOM Radiation Therapy Structure Sets) verification

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| Prescription Use     | <div style="display:inline-block;">✓</div> |
|----------------------|--------------------------------------------|
| (Per 21 CFR 801.109) |                                            |

-OR-

| Over-The-Counter Use |  |
|----------------------|--|
|----------------------|--|

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
K032915
510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K032915](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAK/K032915)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
