← Product Code [JAA](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAA) · K203087

# CombiDiagnost R90 (K203087)

_Philips Medical Systems Dmc GmbH · JAA · Dec 3, 2020 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAA/K203087

## Device Facts

- **Applicant:** Philips Medical Systems Dmc GmbH
- **Product Code:** [JAA](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAA.md)
- **Decision Date:** Dec 3, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1650
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Indications for Use

CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography.

## Device Story

CombiDiagnost R90 is a remote-controlled fluoroscopy and digital radiography system. It uses a Cesium Iodide flat-panel detector to capture X-ray images of patients in standing, seated, or lying positions. The system includes a high-voltage generator, X-ray tube, collimator, and the Eleva Workspot workstation. Operators use the workstation to preset exam data, acquire, process, store, and export images. The system incorporates advanced imaging features like Digital Subtraction Imaging, bone suppression, and table-side stitching. It is used in clinical settings by healthcare professionals. The output is digital radiographic/fluoroscopic images, which assist clinicians in diagnostic decision-making. The device benefits patients by providing high-quality diagnostic imaging for a wide range of routine and specialized examinations.

## Clinical Evidence

No clinical data. Substantial equivalence was demonstrated through design features, indications for use, fundamental scientific technology, and non-clinical performance testing (bench testing, image quality assessment, and usability engineering).

## Technological Characteristics

Multi-functional R/F system; Cesium Iodide flat-panel detector; Velara X-ray generator (65kW/80kW); SRO/SRM X-ray tube; motorized collimation; Eleva Workspot workstation. Connectivity: PACS/RIS integration. Software: Eleva system (Front-end, Back-end, Image Processing). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-54, IEC 6220-1-1, IEC 62304, ISO 14971.

## Regulatory Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- CombiDiagnost R90 ([K163210](/device/K163210.md))

## Reference Devices

- DigitalDiagnost C90 ([K202564](/device/K202564.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Philips Medical Systems DMC GmbH % Prithul Bom Most Responsible Person Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k SAINT PAUL MN 55114

Re: K203087

Trade/Device Name: CombiDiagnost R90 Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: Class II Product Code: JAA, KPR, MOB Dated: November 23, 2020 Received: November 24, 2020

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

December 3, 2020

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K203087

Device Name CombiDiagnost R90

Indications for Use (Describe)

CombiDiagnost R90 is a multi-functional general RF system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary of Safety and Effectiveness

510(K) Number : K203087

This 510(k) summary of safety and effectiveness information is prepared in accordance with 21 CFR §807.92.

| Issue Date:           | July 17th, 2020                                                                                                                                                     |                                             |  |
|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|--|
| Manufacturer:         | Philips Medical Systems DMC GmbH<br>Roentgenstrasse 24-26<br>22335 Hamburg<br>GERMANY<br>Establishment registration number: 3003768251                              |                                             |  |
| Contact Person:       | Dr. Supriya A. Dalvi<br>Regulatory Operations Specialist<br>Phone: +91 9825604544 / +91<br>8733918445<br>Fax: +49 40 5078-2425<br>E-mail: supriya.dalvi@philips.com |                                             |  |
| Device Trade<br>Name: | CombiDiagnost R90                                                                                                                                                   |                                             |  |
| Classification:       | Classification Name:                                                                                                                                                | Image-intensified fluoroscopic x-ray system |  |
|                       | Classification Regulation:                                                                                                                                          | 21CFR 892.1650                              |  |
|                       | Classification Panel:                                                                                                                                               | 90 - Radiology                              |  |
|                       | Device Class:                                                                                                                                                       | Class II                                    |  |
|                       | Primary Product Code:                                                                                                                                               | JAA                                         |  |
|                       | Secondary Product Code:                                                                                                                                             | KPR, MQB                                    |  |
| Predicate<br>Device:  | Trade Name:                                                                                                                                                         | CombiDiagnost R90                           |  |
|                       | Manufacturer:                                                                                                                                                       | Philips Medical Systems DMC                 |  |
|                       | 510(k) Clearance:                                                                                                                                                   | K163210 - January 31, 2017                  |  |
|                       | Classification Name:                                                                                                                                                | Image-intensified fluoroscopic x-ray system |  |
|                       | Classification Regulation:                                                                                                                                          | 21CFR 892.1650                              |  |
|                       | Classification Panel:                                                                                                                                               | Radiology                                   |  |

{4}------------------------------------------------

| Device Class:              | Class II                         |
|----------------------------|----------------------------------|
| Primary Product Code:      | JAA                              |
| Secondary Product Codes:   | KPR, MQB                         |
| <b>Reference Device:</b>   |                                  |
| Trade Name:                | <i>DigitalDiagnost C90</i>       |
| Manufacturer:              | Philips Medical Systems DMC GmbH |
| 510(k) Clearance:          | K202564 - September 30, 2020     |
| Classification Name:       | Stationary x-ray system          |
| Classification Regulation: | 21CFR 892.1680                   |
| Classification Panel:      | 90 -- Radiology                  |
| Device Class:              | Class II                         |
| Product code:              | MQB, KPR, LLZ                    |

#### Device Description:

The CombiDiagnost R90 is a multi-functional general Radiography/Fluoroscopy (R/F) system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography.

The CombiDiagnost R90 is a remote-controlled fluoroscopy system in combination with high-end digital radiography. The system is suitable for routine X-ray examinations and special examinations on patients in standing, seated or lying positions. The CombiDiagnost R90 retrieves images by means of a Cesium Iodide flat panel detector.

Philips fluoroscopy systems (standard configuration) consist of the Basic unit ("geometry" or "table unit"). Workstation Eleva Workspot (with integrated generator control, hand switch, keyboard, mouse, touch screen and PC), dual screen-monitor, Spot film device (digital camera or flat panel detector), Dynamic detector, Fixed detector, X-ray Generator Velara, X-ray tube assembly, Receptor (Flat panel detector). The optional component like Skyplate wireless portable detectors small and large, Ceiling Suspension (CSM3), Vertical Wall stand (VS2), Ceiling Suspension for monitors, monitor trolley, Remote control for RF viewer, accessories for "Stitching on the Table", are also available.

The Eleva software of the proposed CombiDiagnost R90 is based on a workstation i.e. Eleva Workspot (computer, keyboard, display and mouse) that is used by an operator to preset examination data and to generate, process and handle digital x-ray images. As part of the radiographic system, the Eleva software is intended to acquire, process, store, display and export digital Fluoroscopy and radiographic images.

The Eleva Software system is decomposed into software components. These components are clustered in three component collections like the image handling focused Back-end (BE), the acquisition focused Front-end (FE) and Image Processing (IP).

{5}------------------------------------------------

- The Front-end Software is intended to acquire images.

- The Back-end Software is intended to query patient data from Radiology Information System (RIS), store, display and export digital radiographic images to Picture Archiving and Communication System (PACS)

- The Image Processing Software is intended to perform the pre and post processing on the acquired raw images.

The CombiDiagnost R90 uses the same workflow from the currently marketed and predicate device, CombiDiagnost R90 (K163210) with only the following modifications:

- additional optional components (like the reference monitor, remote control), -
- -Eleva Workspot updated to incorporate new imaging features mainly from the previously approved reference device, DigitalDiagnost C90 (K202564) along with functional clusters like Digital Subtraction Imaging and stitching on the table
- updates to improve usability and serviceability. -

### Indications for Use:

CombiDiagnost R90 is a multi-functional general R/F system. It is suitable for all routine radiography and fluoroscopy exams, including specialist areas like angiography or pediatric work, excluding mammography.

### Fundamental Scientific Technology:

The CombiDiagnost R90 employs the same basic construction and fundamental scientific technology as the currently marketed and predicate device. CombiDiagnost R90 (K163210), with regards to the functionality of all its components with the same high voltage generator, X-ray tube, Collimator, Wireless portable detectors, workstation (ELEVA) for images post-processing, storage and viewing. Many of the modifications to the CombiDiagnost R90 are already FDA cleared in reference device, DigitalDiagnost C90 (K202564). All features characteristics that potentially have an influence on image quality are assessed and verified according to FDA Guidance for Industry and Food and Drug Administration Staff: Guidance for the Submission of 510/k)'s for Solid State X-ray Imaging Devices; issued on September 1, 2016. Further the Non-clinical information sufficiently supports the substantial equivalence as per chapter VII- Non clinical Considerations of FDA Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices.

These modifications have been evaluated in safety risk assessment report. The risks associated with these changes are considered in the risk management report and risk management activities show that all risks are sufficiently mitigated and that the overall residual risks are acceptable.

The outcome of this technological characteristics comparison between the CombiDiagnost R90 and currently marketed and predicate device, CombiDiagnost R90 (K163210) also demonstrate that the minor differences in the technological characteristics do not affect the safety or effectiveness of the CombiDiagnost R90 when compared to the currently marketed predicate.

{6}------------------------------------------------

### Summary of technological characteristics:

The CombiDiagnost R90 has the same indications for use and technological characteristics as the currently marketed and predicate device, CombiDiagnost R90 (K163210) demonstrating its substantial equivalence to the predicate device. Refer Table 1 below for comparisons of the technological characteristics of CombiDiagnost R90 and the predicate device and Table 2 for comparison between CombiDiagnost R90 and the reference device, DigitalDiagnost C90 (K202564)

Table 1: Comparison of the Technological Characteristics and Indications for use of the CombiDiagnost R90 and the Currently Marketed and Predicate device, CombiDiagnost R90

|                                                       | Predicate device:<br>CombiDiagnost R90<br>(K163210)                                                                  | CombiDiagnost R90                                                                                                                                                          | Discussion                                                                                                                                                                                                          |
|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulation<br>number                                  | 21CFR 892.1650                                                                                                       | Identical                                                                                                                                                                  | -                                                                                                                                                                                                                   |
| Classification<br>Panel                               | 90 - Radiology                                                                                                       | Identical                                                                                                                                                                  | -                                                                                                                                                                                                                   |
| Device Class                                          | Class II                                                                                                             | Identical                                                                                                                                                                  | -                                                                                                                                                                                                                   |
| Product codes                                         | JAA, KPR, MQB                                                                                                        | Identical                                                                                                                                                                  | -                                                                                                                                                                                                                   |
| Table Features                                        |                                                                                                                      |                                                                                                                                                                            |                                                                                                                                                                                                                     |
| Working height<br>(table top center<br>to floorplate) | 65 cm - 133 cm                                                                                                       | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Table tilt<br>movement                                | -90° to +90°                                                                                                         | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Table top<br>suspension                               | Two sides suspensions                                                                                                | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Table top<br>material                                 | Plastic laminate or carbon<br>fiber                                                                                  | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Table top<br>movement                                 | Lateral: 32 cm (12.6") (± 16<br>cm (6.3"))<br>Longitudinal: Only detector<br>movements to<br>improve patient comfort | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Table top<br>absorption                               | Plastic, with Carbon fiber:<br>0.6mm Al @ 100kV,<br>HVL = 3.6mm Al                                                   | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Maximum<br>patient weight                             | 284 kg (626 lbs)                                                                                                     | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Lateral scan<br>distance                              | 32 cm ± 16 cm                                                                                                        | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Lateral scan<br>speed                                 | 5 cm/s, soft start and stop<br>Auto centering                                                                        | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Longitudinal<br>scan distance                         | 148 cm (58.3") longitudinal, motorized                                                                               | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Longitudinal<br>scan speed /<br>Detector<br>movement  | 3 cm – 20 cm / sec                                                                                                   | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Table column<br>angulation                            | -40° to +40°                                                                                                         | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Source image<br>distance                              | 113cm – 183cm                                                                                                        | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Collimator                                            | Motorized automatic<br>collimation                                                                                   | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Preparation time<br>for exposure                      | 1 sec (approximately)                                                                                                | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Grid                                                  | Parkable                                                                                                             | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Auto Grid<br>Selection                                | Yes                                                                                                                  | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Automatic pre-<br>position of the<br>table            | Yes                                                                                                                  | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Picture archiving<br>and<br>communication<br>system   | Yes                                                                                                                  | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Image chain<br>(fluoroscopy)                          | Philips dynamic Eleva<br>Image Chain                                                                                 | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Detector                                              | Pixium FE 4343F (cleared<br>via K080859 – Villa Sistemi<br>Medicali S.p.A.)                                          | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Generator                                             | Philips Velara 65kW,<br>optional 80 kW                                                                               | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| Tube                                                  | Philips SRO 33100 ROT<br>380 or SRM 0608 ROT GS<br>505                                                               | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                |
| System Control                                        | Remote, optional nearby<br>with nearby control trolley                                                               | Identical                                                                                                                                                                  | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device                                                                                                                                 |
| Image<br>Acquisition                                  | Eleva WorkSpot                                                                                                       | Updated Eleva WorkSpot<br>to include new imaging<br>features UNIQUE 2 and<br>Bone Suppression from the<br>reference device,<br>DigitalDiagnost C90<br>(K202564) along with | The basis and functionality of<br>the Image acquisition remains<br>same. The Eleva software is<br>upgraded to incorporate the<br>new features for better<br>imaging was tested and<br>verified to have no impact on |

{7}------------------------------------------------

{8}------------------------------------------------

|                                      |                                                                                                                                                                                                                                                                 | Subtraction Imaging and<br>stitching on the table                                | the device. Thus<br>demonstrating Substantial<br>Equivalence                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Operating<br>System                  | Microsoft Windows 7                                                                                                                                                                                                                                             | Microsoft Windows 10                                                             | The upgrade to Window 10 is<br>a part of routine software<br>upgrade done by Microsoft.<br>Further verification tests<br>demonstrates no impact on the<br>safety and effectiveness of the<br>device. Thus demonstrating<br>Substantial Equivalence                                                                                                                                                                                                                                                                            |
| Image<br>Processing                  | UNIQUE (fluoroscopy +<br>radiography modality)                                                                                                                                                                                                                  | UNIQUE (fluoroscopy<br>modality only)<br>UNIQUE 2 (radiography<br>modality only) | UNIQUE 2 is only used for<br>radiography workflows for<br>the subject device. It is<br>intended to provide improved<br>image processing, reduced<br>noise, and improved contrast<br>compared to UNIQUE for<br>radiography only.<br>UNIQUE 2 does not replace<br>UNIQUE for fluoroscopy or<br>alter the clinical workflow,<br>hence there is no change to<br>the safety or effectiveness of<br>the device. Both UNIQUE<br>and UNIQUE 2 are embedded<br>in their respective workflows<br>and cannot be used cross-<br>modality. |
| Reference<br>monitor                 | No                                                                                                                                                                                                                                                              | Yes                                                                              | The additional reference<br>monitor is made available for<br>the display a reference image<br>in the Control Room and the<br>Examination Room. This does<br>not impact safety and<br>effectiveness of the device.<br>Thus, demonstrating<br>Substantial Equivalence                                                                                                                                                                                                                                                           |
| Firmware for the<br>dynamic detector | No                                                                                                                                                                                                                                                              | Yes                                                                              | This is a service tool to<br>provide functionality to<br>perform firmware updates of<br>the detector.  It doesn't<br>impact safety and<br>effectiveness of the device.<br>Thus, demonstrating<br>Substantial Equivalence                                                                                                                                                                                                                                                                                                      |
| Indications for<br>Use               | CombiDiagnost R90 is a<br>multi-functional general R/F<br>system. It is suitable for all<br>routine radiography and<br>fluoroscopy exams,<br>including specialist areas<br>like angiography or<br>pediatric work, excluding<br>mammography.                     | Identical                                                                        | Substantially Equivalent; No<br>impact to safety and<br>effectiveness of the device.                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                      | CombiDiagnost R90                                                                                                                                                                                                                                               | Reference Device<br>DigitalDiagnost C90<br>(K202564)                             | Discussion                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Image<br>Acquisition                 | Updated Eleva WorkSpot to<br>include new imaging<br>features UNIQUE 2 and<br>Bone Suppression from the<br>reference device,<br>DigitalDiagnost C90<br>(K202564) along with<br>functional clusters, Digital<br>Subtraction Imaging and<br>stitching on the table | Eleva WorkSpot                                                                   | The basis and functionality of<br>the Image acquisition remains<br>same. The Eleva software is<br>upgraded to incorporate the<br>new features for better<br>imaging was tested and<br>verified to have no impact on<br>the safety or effectiveness of<br>the device.                                                                                                                                                                                                                                                          |
| Operating<br>System                  | Microsoft Windows 10                                                                                                                                                                                                                                            | Identical                                                                        | No impact on safety and<br>effectiveness of the device.<br>Further testing has been<br>carried to confirm the safety<br>and effectiveness                                                                                                                                                                                                                                                                                                                                                                                     |
| Image<br>Processing                  | UNIQUE 2 (radiography<br>modality only)                                                                                                                                                                                                                         | UNIQUE 2 (radiography<br>modality only)                                          | No impact on safety and<br>effectiveness of the device as<br>the feature is included only in<br>the radiography workflow of<br>the subject device, which is<br>same as the reference device<br>Further testing has been<br>carried to confirm the safety<br>and effectiveness                                                                                                                                                                                                                                                 |
| Post-processing<br>application       | Bone Suppression                                                                                                                                                                                                                                                | Identical                                                                        | No impact on safety and<br>effectiveness of the device.<br>Further testing has been<br>carried to confirm the safety<br>and effectiveness                                                                                                                                                                                                                                                                                                                                                                                     |
| Scatter<br>correction                | Skyflow Plus                                                                                                                                                                                                                                                    | Identical                                                                        | No impact on safety and<br>effectiveness of the device.<br>Further testing has been<br>carried to confirm the safety<br>and effectiveness                                                                                                                                                                                                                                                                                                                                                                                     |
| Tube Head                            | Eleva Tube head with<br>optional live camera<br>package                                                                                                                                                                                                         | Identical                                                                        | No impact on safety and<br>effectiveness of the device.<br>Further testing has been<br>carried to confirm the safety<br>and effectiveness                                                                                                                                                                                                                                                                                                                                                                                     |

{9}------------------------------------------------

Table 2: Comparison of the Technological Characteristics of the CombiDiagnost R90 and the reference device, DigitalDiagnost C90

## Summary of Non-Clinical Data:

This 510(k) premarket notification contains technical documentation which includes non-clinical verification and validation tests as well as image quality testing. Tests were performed on the CombiDiagnost R90 according to the following FDA recognized standards and guidance documents as well as additional non-standard performance tests:

{10}------------------------------------------------

- . AAMI ANSI ISO 14971:2007/ (R) 2010, Medical devices. Application of risk management to medical devices
- AAMI ES60601-1:2005/(R)2012 and . C1:2009/(R)2012 ANSI and A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment Part 1-2: General ● requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60601-1-3 Edition 2.1 2013-04 Medical electrical equipment - Part 1-3: General requirements for basic safety and essential performance - Collateral Standard: Radiation protection in diagnostic X-ray equipment
- . IEC 60601-2-54 Edition 1.1 2015-04 CONSOLIDATED VERSION Medical electrical equipment - Part 2-54: Particular requirements for the basic safety and essential performance of X-ray equipment for radiography and radioscopy
- . IEC 6220-1-1 Edition 1.0 2015-03 Medical electrical equipment-Characteristics of digital X-ray imaging devices Part 1-1: Determination of the detective quantum efficiency Detectors used in radiographic imaging
- IEC 62304 Edition 1.1 2015-06 CONSOLIDATED VERSION Medical device software -● Software life cvcle processes
- Guidance for the Submission of 510(k)s for Solid State X-Ray Imaging Devices, issued ● September 1, 2016
- Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, issued May 11, 2005
- Guidance for Pediatric Information for X-ray Imaging Device Premarket Notifications, ● issued November 2017

Additional non-standard performance tests have been performed to demonstrate the safety and effectiveness as follows:

- . System Verification Test
- . Image Quality Test
- Human Factors and Usability Engineering Test

The test results of Non-Clinical testing demonstrate that the CombiDiagnost R90 meets the acceptance criteria and is adequate for its intended use. Additionally, the risk management activities show that all risks are sufficiently mitigated and overall residual risks are acceptable.

Based on the supporting data provided in this 510(k) premarket notification, the CombiDiagnost R90 is substantially equivalent to the currently marketed and predicate device, CombiDiagnost R90 (K163210) in terms of safety and effectiveness.

{11}------------------------------------------------

#### Summary of Clinical Data:

The CombiDiagnost R90 did not require a clinical study since substantial equivalence to the currently marketed and predicate device, CombiDiagnost R90 (K163210) was demonstrated with the following attributes:

- Design features;
- Indication for use;
- Fundamental scientific technology; ●
- Non-clinical performance testing and
- Safety and effectiveness. ●

#### Substantial Equivalence Conclusion:

The comparison of technological characteristics, non-clinical performance data, safety testing demonstrates that CombiDiagnost R90 is substantially equivalent to the currently marketed and predicate device, CombiDiagnost R90 (K163210) and is safe and effective.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAA/K203087](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/JAA/K203087)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
