CS-SERIES-FP WITH 3030+OPTION

K121293 · Omega Medical Imaging, LLC · JAA · Jul 26, 2012 · Radiology

Device Facts

Record IDK121293
Device NameCS-SERIES-FP WITH 3030+OPTION
ApplicantOmega Medical Imaging, LLC
Product CodeJAA · Radiology
Decision DateJul 26, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1650
Device ClassClass 2

Indications for Use

The Omega Medical Imaging, LLC. CS-series-FP with 3030+ Option systems are intended for use in radiographic/fluoroscopic applications including cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging.

Device Story

CS-series-FP with 3030+ Option is a floor-mounted C-arm x-ray system (dual-plane systems include ceiling-suspended C-arm) for radiographic/fluoroscopic imaging. System utilizes a 29.8cm x 29.8cm solid-state flat-panel detector (FPD) featuring a cesium iodide scintillator coupled to an amorphous silicon TFT panel. Input x-ray signals are converted to digital images, processed by an acquisition system, and displayed for clinician review. Used in clinical settings (e.g., interventional suites) by physicians to guide diagnostic and interventional procedures. Output supports clinical decision-making by providing real-time or static visualization of anatomy. System includes local storage and DICOM compatibility for integration into hospital workflows.

Clinical Evidence

Bench testing only. Performance compared to predicate using commercially available test objects for low-contrast, spatial resolution, temporal resolution, and dynamic range. No clinical data presented.

Technological Characteristics

Solid-state flat-panel detector (FPD) with cesium iodide scintillator and amorphous silicon TFT panel. Dimensions: 29.8cm x 29.8cm. Complies with UL2601-1, IEC60601-1, CSA 22.2 No. 601.1-M90, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32, and IEC 60601-2-43. Connectivity: DICOM compatible. Energy source: X-ray.

Indications for Use

Indicated for patients requiring radiographic or fluoroscopic imaging, including cardiac, vascular, general diagnostic, and interventional x-ray procedures.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K121293 ### Omega Medical Imaging, LLC JUL 26 2012 ### 510(k) SUMMARY Company Name: Omega Medical Imaging, LLC > 675 Hickman Circle Address: Sanford, FL 32771 407-323-9400 Telephone No: Registration No .: 1052701 Contact person: Brian J. Fleming Date Prepared: 20 April 2012 CS-series-FP with 3030+ Option radiographic/fluoroscopy Device (trade) name: svstem Common/usual name: Image Intensified Fluoroscopic X-ray System Classification Name: Solid State X-ray Imager, Class II, 90 MQB Angiographic X-ray System, Class II, 90 IZI Classification Panel: Radiology > CFR Section: 892.1650 and 892.1600 Device Class: Class II 90 MQB, 90 IZI Device Code: #### Predicate device(s): Omega Medical Imaging CS-series-FP (K100102) . #### Device description: The Omega Medical Imaging, LLC. CS-series-FP systems currently incorporate a 19.8cm x 19.8cm solid-state flat-panel detector (FPD) as an option. This 510(k) submission adds a larger format (29.8cm x 29.8cm) flat-panel detector as an additional option. The CS-series-FP fluoroscopy single and dual plane x-ray imaging systems are configured with a floor mounted C-arm and a patient table. The dual plane systems incorporate a ceiling suspended C-arm into the system. The flat-panel image detector utilizes a cesium iodide scintillator coupled to an amorphous silicon TFT panel. The captured digital image is processed by the acquisition system which includes image processing, viewing functions, local storage, and DICOM compatibility. Materials: The solid-state x-ray imager (FPD) construction and materials comply with UL2601-1, IEC60601-1, CSA 22.2 No. 601.1-M90, and is CE Marked. The imager is classified by Underwriters Laboratories. {1}------------------------------------------------ # Omega Medical Imaging, LLC #### Intended use: - The Omega Medical Imaging, LLC. CS-series-FP with 3030+ Option systems are . intended for use in radiographic/fluoroscopic applications including cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging. #### Comparison with Predicate Devices: - The Omega Medical Imaging large format option to the CS-series-FP system utilizes . identical technology of the above mentioned predicate device. The input scintillator is Cesium lodide coupled to a Thin-Film-Transistor (TFT) panel. The image processing is achieved with conventional computer based image processing systems. It is the opinion of Omega Medical Imaging that the CS-series-FP with 3030+ Option utilizing the larger format FPD detector is essentially equivalent to the existing Omega Medical Imaging cleared CS-series-FP system utilizing the smaller format FPD. #### Substantial Equivalence: - SE was determined on non-clinical performance testing. Included in this report is detailed . data comparing performance with the existing Omega Medical Imaging CS-series-FP system utilizing the 19.8cm x 19.8cm format FPD image acquisition system. The tests that were performed utilized commercially available Test Objects that include low-contrast objects with varying absorbers simulating different patient sizes, spatial and temporal resolution test objects, and dynamic range test objects. Also included is the "Non-clinical Considerations" of the flat-panel detector. It is the opinion of Omega Medical Imaging that the CS-series-FP with 3030+ Option with the larger format FPD (29.8cm x 29.8cm) is essentially equivalent to the cleared Omega Medical Imaging CS-series-FP (K100102) smaller format detector systems. #### Safety information: - The Omega CS-series-FP with 3030+ Option systems comply with the applicable ● requirements of 21 CFR 1020.30, 21 CFR 1020.31, and 21 CFR 1020.32. - The Omega CS-series-FP with 3030+ Option systems comply with the international safety . standards IEC 60601-1, IEC 60101-1-2, IEC 60601-1-3, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32, and IEC 60601-2-43. - The Omega CS-series-FP with 3030+ Option systems comply with UL 60601-1 and . CAN/USA C22.2 No.601.1-M90 #### Conclusion: The Omeqa CS-series-FP with 3030+ Option systems with the larger format (29.9cm x 29.9cm) FPD do not introduce any new indications for use, nor does the use of the device result in any new potential hazard. Omega considers the CS-series-FP with 3030+ Option systems with this larger format flat-panel detector to be substantially equivalent with the predicate device. 5-2 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an image of an eagle with its wings spread. The eagle is facing to the right and has three lines emanating from its body. ## DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 JUL 26 2012 Mr. Brian J. Fleming President & CEO Omega Medical Imaging, LLC 675 Hickman Circle SANFORD FL 32771 Re: - K121293 Trade/Device Name: CS-series-FP with 3030+ Option Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA and OWB Dated: April 20, 2012 Received: May 1, 2012 ### Dear Mr. Fleming: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of {3}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter requirences as begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed nonication. The results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely Yours, Janine M. Morris Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): CS-series-FP with 3030+ Option Device Name: Indications for Use: The Omega Medical Imaging, LLC CS-series-FP with 3030+ Option systems are intended for use in radiographic/fluoroscopic application including cardiac, vascular, general radiographic/fluoroscopic diagnostic, and interventional x-ray imaging. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![image.png](https://i.imgur.com/9499999.png) Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) Page 1 of 1.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...