QUANTITATIVE ANALYSIS OF LEFT VENTRICULAR ANGIOGRAMS (QLV)
K993765 · Medis Medical Imaging Systems BV · IZI · Nov 26, 1999 · Radiology
Device Facts
Record ID
K993765
Device Name
QUANTITATIVE ANALYSIS OF LEFT VENTRICULAR ANGIOGRAMS (QLV)
Applicant
Medis Medical Imaging Systems BV
Product Code
IZI · Radiology
Decision Date
Nov 26, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1600
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
OLV-CMS allows the user to work productively in the increasingly digital world of images that are archived in DICOM formats as a result of angiography procedures. The QLV-CMS product is used to select images of the left ventricle for quantitative analysis, resulting in measurements of the selected images. The quantitative information is provided for use by physicians and scientists in both research and clinical settings. Image archiving and administration purposes are supported with the QLV-CMS product as well.
Device Story
QLV-CMS software performs quantitative analysis of left ventricular angiograms; inputs include digitized 35mm cinefilm or DICOM-standard files. System selects left ventricle images; calculates dimensions and performance characteristics; outputs results for printing or database archiving. Used in clinical and research settings by physicians and scientists. Operates on Windows NT as a modular software package; integrates with other software like CMS-View for image handling. Provides objective measurements to assist clinical decision-making and research assessment.
Clinical Evidence
No clinical data provided; substantial equivalence based on software verification and validation testing.
Technological Characteristics
Modular software package; runs on Windows NT; supports DICOM image formats. Utilizes established image analysis algorithms for left ventricular quantification. Software level of concern is minor.
Indications for Use
Indicated for physicians and scientists to obtain quantitative measurements of left ventricular dimensions and performance characteristics from angiographic x-ray images. Used in cardiac cath labs for research studies, assessment of angiographic patient condition, evaluation of device/drug therapy results, and review/analysis of patient imaging records.
Regulatory Classification
Identification
An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Cardiovascular Measurement System - (CMS) (K940172)
Reference Devices
CMS-View
Related Devices
K993763 — QUANTITATIVE ANALYSIS OF CORONARY ANGIOGRAMS (QCA) · Medis Medical Imaging Systems BV · Nov 26, 1999
K983821 — PHILIPS INTURIS CARDIO IMAGE VIEWING PACKAGE · Philips Medical Systems North America, Inc. · Dec 11, 1998
K982203 — CAAS II LVA BIPLANE OPTION · Pie Medical Equipment B.V. · Jul 14, 1998
K082628 — CMR42 CARDIAC MRI IMAGING SYSTEM, 2.1.23 · Circle Cardiovascular Imaging, Inc. · Nov 20, 2008
K023100 — ACCUSKETCH CARDIAC QUANTITATIVE SYSTEM W/ ADVANCED ANALYSIS COMPONENTS · General Electric Medical Systems Information Techn · Oct 11, 2002
Submission Summary (Full Text)
{0}------------------------------------------------
OLV-CMS Special 510(k) Notification "Device Modification" SUMMARY 510(k) Summary of safe and effectiveness information conform 21CFR. 807.87(h).
| 1) Submitter | : MEDIS Medical Imaging Systems, Inc., |
|--------------------|--------------------------------------------------------------|
| Address | : 109 Danbury Road<br>Ridgefield, CT 06877 USA |
| Telephone | : 203.438.5588 |
| Fax | : 203.438.5393 |
| Contact Person | : Douglas Orr, President |
| Prepared | : July 30, 1999 |
| 2) Device Name | : Quantitative Analysis of Left Ventricular Angiograms (QLV) |
| Common Name | : QLV-CMS |
| Device Class. Name | : System, X-ray, Angiographic |
| Regulation Number | : 21 CFR 892.1600 (90 IZI; Class II) |
3) Predicate Device(s) : Cardiovascular Measurement System - (CMS) 510(k) Number: K940172
4) Description of the device:
QLV-CMS performs quantitative analysis of angiograms (specific to the left ventricle) that are input to the system in formats that include digitized video (from 35 mm cinefilm) and DICOM-file standard formats. The analysis results of the OLV-CMS operation may be printed or archived in files for export to a general-purpose database.
## 5) Intended use:
OLV-CMS allows the user to work productively in the increasingly digital world of images that are archived in DICOM formats as a result of angiography procedures. The QLV-CMS product is used to select images of the left ventricle for quantitative analysis, resulting in measurements of the selected images. The quantitative information is provided for use by physicians and scientists in both research and clinical settings. Image archiving and administration purposes are supported with the QLV-CMS product as well.
## 6) Substantial equivalence information:
The OLV-CMS software, as a software package, is substantially equivalent to the predicate device "Cardiovascular Measurement System - CMS" by utilizing the same technical standards and image analysis algorithms for efficient, accurate and reproducible results. The differences in the devices are noted as:
a. QLV-CMS now runs on Windows NT, as compared with the predicate device operation on MS-DOS, and
b. QLV-CMS is a modular software package, as compared with the fully-integrated predicate device. Image handling (review, playback, selection) may now accomplished with other modular software packages, including the CMS-View from MEDIS.
Conclusion respecting safety and effectiveness:
It is the opinion of MEDIS medical imaging systems that QLV-CMS is safe and potential hazards are controlled by a risk management plan for the software development process, including hazard analysis. verification and validation tests and evaluations by hospitals. In our opinion the level of concern for stand-alone software to analyze images is "minor" and that the use of QLV-CMS software does not change the intended use of the angiographic systems in practice.
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top and bottom. Inside the circle is a stylized image of an eagle or bird with three curved lines representing its wings or feathers.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 26 1999
Douglass Orr President MEDIS Medical Imaging Systems, Inc. 109 Danbury Road Ridgefield, CT 06877
Re: K993765
Quantitative Analysis of Left Ventricular Angiograms - Cardiovascular Dated: July 30, 1999 Received: November 8, 1999 Regulatory class: II 21 CFR 892.1600/Procode: 90 IZI
Dear Mr. Orr:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Page__________________________________________________________________________________________________________________________________________________________________________ ુર્ભ
510(k) Number (if known): K993765
Device Name: _
Indications For Use:
510(k) Number: Device Name:
K993765 Quantitative Analysis of Left Ventricular Angiograms (QLV-CMS)
Indications for Use:
QLV-CMS provides quantitative information regarding the calculated dimensions and calculated performance characteristics of the left ventricle of the heart, imaged during angiographic x-ray procedures typically performed in cardiac cath labs. This information is suitable for use by physicians and scientists in the following applications:
1. Scientific and research studies, assessing the angiographic condition of patients and the results of device and drug therapies,
2. Review and analysis of patient angiographic imaging records, providing additional information to physicians and administrators.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| | <div>(Division Sign-Off)</div> <div>Division of Reproductive, Abdominal, ENT,</div> <div>and Radiological Devices</div> |
| 510(k) Number | K993765 |
| Prescription Use | <div>✓</div> <div>(Per 21 CFR 801.109)</div> |
|------------------|----------------------------------------------|
|------------------|----------------------------------------------|
OR
| Over-The-Counter Use | <div>(Optional Format 2-16)</div> |
|----------------------|-----------------------------------|
|----------------------|-----------------------------------|
1-2-96)
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.