ARTIS ZEE / ZEEGOS WITH SW VC14

K090745 · Siemens Medical Solutions USA, Inc. · IZI · Jun 18, 2009 · Radiology

Device Facts

Record IDK090745
Device NameARTIS ZEE / ZEEGOS WITH SW VC14
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeIZI · Radiology
Decision DateJun 18, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1600
Device ClassClass 2
AttributesPediatric

Indications for Use

Artis zee / zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients. Procedures that can be performed with the Artis zee / zeego family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions. Artis zee / zeego can also support the acquisition of position triggered imaging for spatial data synthesis.

Device Story

Modular angiographic X-ray system; components include C-arm, X-ray tube, flat detector, digital imaging system, generator, and examination table. Used in clinical and OR settings by healthcare professionals for diagnostic imaging and interventional procedures. System acquires X-ray data; supports image fusion and navigation systems; enables position-triggered imaging for spatial data synthesis. New SW version VC14 adds wide screen display, OR table, and laser light crosshair. System provides visual/audible warnings; monitors functions; blocks operation upon error detection. Benefits include enhanced visualization for image-guided surgery and interventional procedures.

Clinical Evidence

Bench testing only. Performance testing evaluated modifications to the Artis zee/zeego family; results met established test plans and protocols.

Technological Characteristics

Modular X-ray system; components include C-arm, X-ray tube, flat detector, digital imaging system, collimator, and generator. Features include wide screen display, OR table, and laser light crosshair. Software version VC14. System includes visual/audible safety warnings and automated error monitoring/blocking.

Indications for Use

Indicated for pediatric and obese patients requiring single or biplane diagnostic imaging and interventional procedures, including cardiac, neuro, general, rotational, and multipurpose angiography, as well as whole-body radiographic/fluoroscopic procedures, image-guided surgery, and position-triggered imaging for spatial data synthesis.

Regulatory Classification

Identification

An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K090745 Page 1 of 3 JUN 18 2009 # SECTION 5 # 510(K) SUMMARY ## FOR # SIEMENS ARTIS ZEE / ZEEGO FAMILY VC14 Submitted by: Siemens AG, Healthcare Sector, USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355 ## March 17, 2009 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92. ### 1. Contact Person: Mr. Garv Johnson Technical Specialist, Regulatory Affairs Submissions Siemens AG, Healthcare Sector, USA, Inc 51 Valley Stream Parkway E-50 Malvern, PA 19355-1406 Phone:(601) 448-1778 Fax: (610) 448-1787 ### Device Name and Classification 2. | Product Name: | Artis zee and Artis zeego - Modular Angiographic System | | |-----------------------|---------------------------------------------------------|--| | Classification Name: | Angiographic X-ray System | | | Classification Panel: | Radiology | | | CFR Section: | 21 CFR §892.1600 | | | Device Class: | Class II | | | Product Code: | 90 IZI | | #### 3. Intended Use: Artis zee / zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to pediatric and obese patients. Procedures that can be performed with the Artis zee / zeego family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, {1}------------------------------------------------ KC9C745 Page 2 of 3 Artis zee / zeego can also support the acquisition of position triggered imaging for spatial data synthesis. The intended use and indications for use of the Artis zee / zeego VC14 have minor changes from its predicate device the Artis zee / zeego VC13, K073290 #### Device Description: 4. The Artis zee / zeego Modular Angiography System is designed as a set of components that may be combined into different configurations to provide specialized angiography systems. For these models, a new SW version VC14 will be available. The new SW version VC14 will allow additional functionality of currently used applications and is extended with new features (i.e. a wide screen display, a new OR table, and laser light crosshair). SW VC14 also includes an extension of the intended use to surgical angiography, including general surgical use for patients in the OR. The Artis zee / zeego Modular Angiography System with SW VC14 is substantially equivalent to the AXIOM Artis Modular Angiography System VC13 with all its components as described in the Device Description, Section 11 and the Substantial Equivalence Section 12. ### 5. Substantial Equivalence: The Artis zee / zeego Modular Angiography System SW VC14 is a modification of a legally marketed device and substantial equivalent to Artis zee, Artis zeego SW VC13 as listed below. | 510(k) Number | Date of Clearance | Device Name | |---------------|---------------------|---------------------------------------------------| | K073290 | February 11th, 2008 | Artis zee, Artis zeego Angiographic X-ray Systems | A detailed Substantial Equivalence Comparison is provided in Section 12. ## 6. Summary of Technological Characteristics of the Principal Device as compared with the Predicate Device: Artis zee / zeego Modular Angiography System is designed as a set of components (Carm, X-ray tube and housing, flat detector, digital imaging system, collimator, generator etc.) that may be combined into different configurations to provide specialized angiography systems. Many of the components used with Artis zee / zeego are either commercially available with current Siemens systems or include minor modifications to existing components. New or modified features provided with Artis zee/zeego SW VC 14 are provided in the Executive Summary as well as in the Device Description. {2}------------------------------------------------ KC9C745. Page 3 of 3 ### 7. General Safety and Effectiveness Concerns: Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features including visual and audible warnings are incorporated into the system design. In addition the Artis zee / zeego Modular Angiography System is continually monitored, and if an error occurs, the system functions will be blocked and an error message will be displayed. Furthermore the operators are health care professionals familiar with and responsible for the X-ray examinations to be performed. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. # Performance Data: Applicable testing was performed to evaluate the modifications to the Artis zee / zeego family. The test results were found to be acceptable as required by the respective test plans and protocols. ## Conclusion: The testing reported in this 510(k) establishes the device is safe and effective for its intended use and substantially equivalent to the predicate device. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows a logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with outstretched wings. # DEPARTMENT OF HEALTH & HUMAN SERVICES JUN 18 2009 Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # Mr. Garv Johnson Technical Specialist, Regulatory Affairs, Submissions Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway MALVERN PA 19355 Re: K090745 Trade/Device Name: Artis zee / zeego Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic x-ray system Regulatory Class: II Product Code: IZI Dated: June 5, 2009 Received: June 9, 2009 Dear Mr. Johnson: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/cdrl/mdr/ for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Janine M. Morris Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # SECTION 4 # Indications For Use 510(k) Number (if known): 090 745 Device Name: Artis zee / zeego ## Indications for Use: Artis zee / zeego is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not linited to, pedication and obese patients. Procedures that can be performed with the Artis zee / zeego family include cardiac angiography, neuro-angiography, general angiography, rotational angiography, multipurpose angiography and whole body radiographic/fluoroscopic procedures as well as procedures next to the table for i.e. patient extremities. Additional procedures that can be performed include angiography in the operating room, image guided surgery by X-ray, by image fusion, and by navigation systems. The examination table as an integrated part of the system can be used for X-ray imaging, surgery and interventions. Artis zee / zeego can also support the acquisition of position triggered imaging for spatial data synthesis. Prescription Use (Part 21 CFR 801 Subpart D) C) AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart (Please do not write below this line - continue on another page if needed) Concyrrence of the CDRH, Office of Device Evaluation (ODE) HoguhkPh (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices KO90745 510(k) Number 510(k) Artis zeelzeego Page 4 of 133 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Siemens AG Healthcare Sector
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...