THE ROBOGUIDE
Device Facts
| Record ID | K961849 |
|---|---|
| Device Name | THE ROBOGUIDE |
| Applicant | Shai Syg Motion and Innovations, Ltd. |
| Product Code | IZH · Radiology |
| Decision Date | Aug 21, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1710 |
| Device Class | Class 2 |
Indications for Use
The ROBOGUIDE is an automatic needle guidance system that provides means for accurate positioning of a biopsy needle of a mammography system, to enable accurate penetration to sites of detectable breast lesions. The Roboguide is an attachment to and is mounted on a parent mammography system. It comprises a robotic biopsy unit and a film read out and motor control evaluation unit. After creating a stereotactic exposure of a previously detected breast lesion, the Roboguide drives a needle holder along three orthogonal axes so that the needle penetrates the precise location of the lesion. The Roboguide can be used for fine needle aspiration (FNA) or for core biopsy by interchanging its aspiration needle carrier with a biopsy gun holder carrier. The Roboguide can be used with either plain film X-ray or with a digital camera.
Device Story
Automatic needle guidance system; attachment for mammography systems. Inputs: stereotactic X-ray images (plain film or digital camera). Operation: robotic unit calculates lesion coordinates; drives needle holder along three orthogonal axes to target site. Used in clinical settings by physicians/technicians. Output: precise needle positioning for biopsy. Benefits: accurate penetration of breast lesions; supports FNA or core biopsy via interchangeable carriers. Safety: includes immobilization and needle carrier locking mechanisms.
Clinical Evidence
Bench testing only; no clinical data provided.
Technological Characteristics
Robotic biopsy unit; film read out and motor control evaluation unit. Three-axis orthogonal drive system. Compatible with plain film X-ray or digital camera. Mechanical breast compression and immobilization. Safety interlocks for needle carrier movement.
Indications for Use
Indicated for patients with detectable breast lesions requiring fine needle aspiration (FNA) or core biopsy procedures.
Regulatory Classification
Identification
A mammographic x-ray system is a device intended to be used to produce radiographs of the breast. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
- Cytoguide (Phillips Medical)
- Mammoguide (Elscint Ltd.)
Related Devices
- K050663 — TRAXYLOC-1 SYSTEM · Traxyz Medical, Inc. · Apr 8, 2005
- K023864 — PERFORMA STEREO · Instrumentarium Corp. Imaging Division · Jan 9, 2003
- K203509 — SMART FINDER · Ims Giotto S.P.A. · Jul 14, 2021
- K122836 — AFFIRM BREAST BIOPSY GUIDANCE SYSTEM · Lorad, A Hologic Co. · Jan 11, 2013
- K113284 — BIOPSY POSITIONER · Fuji Film Medical Systems · Mar 7, 2012