← Product Code [IZH](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IZH) · K152038

# MammoGRIP (K152038)

_Women'S Imaging Solutions Enterprises, LLC · IZH · Dec 22, 2015 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IZH/K152038

## Device Facts

- **Applicant:** Women'S Imaging Solutions Enterprises, LLC
- **Product Code:** [IZH](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IZH.md)
- **Decision Date:** Dec 22, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1710
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

MammoGRIP aids in positioning during radiologic visualization of the breast.

## Device Story

MammoGRIP is a non-medicated, 0.1% benzalkonium chloride foam solution applied to a technician's hands. The solution creates a slightly tacky surface while damp, improving the technician's grip on dry breast tissue to facilitate optimal positioning within the mammography machine's field of view. It is used in clinical settings by radiology technicians during standard mammogram positioning. The device benefits the patient by assisting in accurate breast positioning for radiologic visualization.

## Clinical Evidence

Bench testing only. Biocompatibility testing performed per ISO 10993-10, including Intracutaneous Injection Test (Irritation) and Kligman Maximization Test (Sensitivity). Radiolucency demonstrated. Lot-specific verification includes pH, appearance, odor, and benzalkonium chloride content.

## Technological Characteristics

Non-medicated 0.1% benzalkonium chloride foam solution. Radiolucent. Non-sterile. Biocompatibility verified per ISO 10993-10.

## Regulatory Identification

A mammographic x-ray system is a device intended to be used to produce radiographs of the breast. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- MammoPad Radiolucent Cushion ([K062141](/device/K062141.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 22, 2015

Women's Imaging Solutions Enterprises, LLC Erika Huffman Prin. Medical Research Manager 4050 OLSON MEMORIAL HIGHWAY MINNEAPOLIS MN 55422

Re: K152038

Trade/Device Name: MammoGRIP Regulation Number: 21 CFR 892.1710 Regulation Name: Mammographic X-Ray System (accessory) Regulatory Class: II Product Code: IZH Dated: December 14, 2015 Received: December 16, 2015

Dear Ms. Huffman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert Ocks

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K152038

Device Name MammoGRIP

Indications for Use (Describe)

MammoGRIP aids in positioning during radiologic visualization of the breast.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 6.0 510(k) Summary

| Submitter:                                                                             | MammoGRIP LLC                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                                                                        | Erika Huffman<br>Medical Research Manager, Regulatory<br>NAMSA<br>4050 Olson Memorial Hwy.<br>Suite 450<br>Minneapolis, MN 55422<br>Phone: +1-763-588-9857<br>Fax: +1-763-287-3836<br>ehuffman@namsa.com                                                                                                                                                                                                                                                                                                                                     |
| Date Prepared:                                                                         | July 17, 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Trade Name:                                                                            | MammoGRIP™                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Common Name:                                                                           | Mammography positioning aid                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Classification Name:                                                                   | Accessory to X-Ray Mammographic System,<br>21 CFR Part 892.1710                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Regulatory Class:                                                                      | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Product Code:                                                                          | IZH                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Predicate Device:                                                                      | K062141-MammoPad Radiolucent Cushion<br>This predicate device has not been subject to a design-related<br>recall.                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Device Description:                                                                    | MammoGRIP™ is a non-medicated, 0.1% benzalkonium<br>chloride foam solution intended to be used during<br>mammography to facilitate breast positioning. When<br>MammoGRIP is applied to the technician's hands, it imparts a<br>slightly tacky or sticky surface while it is still damp, thereby<br>allowing the technician to have a better grip of the dry breast<br>tissue for optimal positioning in the field of view of the<br>mammography machine. MammoGRIP is intended to be used<br>with standard mammogram positioning techniques. |
| Indications for Use:                                                                   | MammoGRIP aids in positioning during radiologic<br>visualization of the breast.                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Comparison of the<br>Technological<br>Characteristics with<br>the Predicate<br>Device: | The MammoGRIP device is similar to the MammoPad device<br>in the following ways:<br>• Each of the devices is intended to be used as an aid to<br>positioning during radiologic visualization of the<br>breast.<br>• Each of the devices is radiolucent.<br>• Each of the devices is provided non-sterile.                                                                                                                                                                                                                                    |
|                                                                                        | The MammoGRIP device is different from the predicate<br>device in the following ways:<br>• Physical Form                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Performance Data:                                                                      | The MammoGRIP device meets the following<br>biocompatibility standards:<br>• Intracutaneous Injection Test (Irritation) per ISO 10993-10<br>• Kligman Maximization Test (Sensitivity) per ISO 10993-10<br>The following characteristics are verified for each lot: pH,<br>appearance, odor, and benzylkonium chloride content. In<br>addition, radiolucency of the MammoGRIP device has been<br>demonstrated.                                                                                                                                |
| Conclusion:                                                                            | The data provided in this submission support the safety of the<br>MammoGRIP device and demonstrate that the MammoGRIP<br>device is substantially equivalent to the predicate device<br>which is currently marketed for the same intended use.                                                                                                                                                                                                                                                                                                |

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IZH/K152038](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IZH/K152038)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
