SCANDIA-45 GAMMA CAMERA SYSTEM

K101391 · Scandia Corporation · IYX · Aug 4, 2010 · Radiology

Device Facts

Record IDK101391
Device NameSCANDIA-45 GAMMA CAMERA SYSTEM
ApplicantScandia Corporation
Product CodeIYX · Radiology
Decision DateAug 4, 2010
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1100
Device ClassClass 1

Indications for Use

For use in the acquisition, processing, display, and analysis of planar images of the thyroid and other small organs.

Device Story

Scandia-45 is a high-resolution, small-field-of-view (SFOV) digital gamma camera system. It utilizes a single detector assembly with 45 high-quantum-efficiency photomultipliers, magnetic shielding, and integrated electronics to capture gamma radiation signals. The system includes a lightweight, adjustable gantry and exchangeable microcast collimators. Operated by clinical staff in a medical setting, the device acquires, processes, and displays planar images of the thyroid and small organs. The system uses a fast PCI bus acquisition interface for data handling. Clinicians use the displayed images to analyze organ structure and function, aiding in diagnostic decision-making. The device provides high-resolution imaging capabilities for small-organ assessment.

Clinical Evidence

Clinical images were obtained from human subjects using the Scandia-45. Image quality was compared to reference predicate devices and found to be at least equal. Bench testing was conducted using NEMA NU1 phantoms under standard protocols, confirming performance met or exceeded predicate specifications. Electrical safety testing was also performed and met applicable standards.

Technological Characteristics

Small, mobile scintillation gamma camera. Features a single SFOV detector with 45 high-quantum-efficiency photomultipliers. Includes a gantry, acquisition station, and microcast collimators. Employs full-digital electronics with a fast PCI bus acquisition interface. Construction is integrated, one-board, without tuning potentiometers. Designed for thyroid and small organ imaging.

Indications for Use

Indicated for acquisition, processing, display, and analysis of planar images of the thyroid and other small organs in human subjects.

Regulatory Classification

Identification

A scintillation (gamma) camera is a device intended to image the distribution of radionuclides in the body by means of a photon radiation detector. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Scandia Corporation 10200 Dennis Drive, Suite 5 Urbandale, Iowa 50322 Phone: 515-334-5131 Image /page/0/Picture/1 description: The image shows a sequence of handwritten characters, which appear to be alphanumeric. The sequence reads as 'K101391'. The characters are written in a simple, slightly irregular style, typical of handwriting. The image is in black and white, with the characters standing out against the background. AUG 0 4 2010 # SECTION G . 510(K) SUMMARY ### In accordance with 21CFR 807.92 #### 1.0 Submitter Information Scandia Corporation Name: Address: 10200 Dennis Drive, Suite 5 Urbandale, Iowa 50322 Phone: 515-334-5131 Fax: 515-334-5145 Contact Person: David Askew, Chief Operations Officer Date of Submission: April 16, 2010 #### 2.0 Device Identification Name of Device: Scandia-45 Common Name: Scintillation (Gamma) Camera System Classification Name: Emission Computed Tomography System #### 3.0 Identification of Equivalent Device 1. T-Quest - MEDX Corporation - [K073456] #### 4.0 Description of Device The Scandia-45 is a high resolution, SFOV digital gamma camera for thyroid and small organs imaging. The Scandia-45, is made up of a single, small-field detector assembled with high optical and mechanical quality. The systems has 45 high quantum efficiency photomultipliers characterized by improved energy resolution, magnetic shielding and long-term stability and utilizes a compact, highly integrated, one board, easily serviceable construction without tuning potentiometers. The acquisition console has full-digital electronics, including fast PCI bus acquisition interface. The system also has a light-weight, easilyadjustable gantry and light, easily-exchangeable microcast collimators. G-1 {1}------------------------------------------------ #### Intended Use / Indications for Use 5.0 For use in the acquisition, processing, display, and analysis of planar images of the thyroid and other small organs. #### 6.0 Technological Characteristics The Scandia-45 Gamma Camera System is a small, mobile scintillation camera with a gantry, a single small-field detector, an acquisition station, and related hardware and software components. The characteristics of the Scandia-45 compare substantially with the predicate device in materials used, technology applied, and functional methodology. Thus, the Scandia-45 Gamma Camera System raises no new issues of safety or efficacy. #### 7.0 Performance Testing and Data Performance testing was performed using NEMA NU1 phantoms, under the NEMA Standard test protocols. In all cases, performance of the Scandia-45 device met or exceeded that of predicate device. Clinical images were obtained using the Scandia-45 in human subjects. Image quality was at least equal to images produced by reference predicate devices. Furthermore, electrical safety testing has been performed and found to meet applicable standards and defined acceptance criteria. #### 8.0 Substantial Equivalence The Scandia-45 Gamma Camera System has the same intended use, similar principles of operation, and consistent technological characteristics as the predicate devices. Thus, the Scandia-45 is substantially equivalent to the predicate devices and no new safety or effectiveness concerns are raised. G-2 {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three curved lines. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and 10;ng Administration 10903 News ampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002 Mr. David Askew Chief Operations Officer Scandia Corporation 10200 Dennis Drive, Suite 5 URBANDALE IA 50322 AUG 0 4 2010 Re: K101391 Trade/Device Name: Scandia-45 Regulation Number: 21 CFR 892.1100 Regulation Name: Scintillation (gamma) camera Regulatory Class: II Product Code: IYX Dated: May 1, 2010 Received: May 18, 2010 ### Dear Mr. Askew: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting of {3}------------------------------------------------ medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(V), remarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Donald J. Trump Donald J. St.Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use ## 510(k) Number (if known): K101391 ## Device Name: Scandia-45 # AUG 0 4 2010 Indications for Use: The Scandia-45 is indicated for use in the acquisition, processing, display, and analysis of planar images of the thyroid and other small organs Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use , (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD) *(Division Sign-Off)* Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety 510K. Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...