OLINDA/EXM

K033960 · Vanderbilt University · IYX · Jun 15, 2004 · Radiology

Device Facts

Record IDK033960
Device NameOLINDA/EXM
ApplicantVanderbilt University
Product CodeIYX · Radiology
Decision DateJun 15, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1100
Device ClassClass 1
AttributesSoftware as a Medical Device, Pediatric

Indications for Use

The purpose of Olinda EXM is to estimate the radiation dose received by internal organs as a result of administering a radiopharmaceutical.

Device Story

OLINDA/EXM is a personal computer software application for internal radiation dose assessment. Users input the specific radionuclide, select a body phantom model (e.g., adult male/female, child, pregnant woman), and provide radiopharmaceutical biokinetic data (time integral of activity in source organs). The EXM module can fit biokinetic data to models to calculate necessary integrals. The software uses the MIRD (Medical Internal Radiation Dose) system to calculate radiation doses to various organs. It operates on Windows 2000/XP. The output consists of tables of estimated absorbed doses to organs. It is used by medical physicists and physicians in hospitals and research facilities to investigate new radiopharmaceuticals and estimate patient doses. The software provides a more user-friendly interface and a wider range of body models compared to legacy systems, allowing for limited patient-specificity by varying organ masses. It aids clinical decision-making by providing quantitative dose estimates for safety and research purposes.

Clinical Evidence

No clinical data. Bench testing only. The device relies on established scientific literature for decay data (Stabin and da Luz 2002), dose factors (Siegel and Stabin 2003), and body phantom models (Cristy and Eckerman 1987).

Technological Characteristics

Software-based dose assessment tool. Operates on Windows 2000/XP. Uses MIRD system algorithm for internal dose calculation. Includes multiple body phantoms (adults, children, pregnant women). Provides dose estimates in mSv/MBq and rem/mCi. Standalone PC application.

Indications for Use

Indicated for estimating absorbed radiation doses to internal organs of reference patients (adult males, females, children, pregnant women) following administration of radiopharmaceuticals. Intended for use by medical physicists or trained physicians in research or hospital facilities.

Regulatory Classification

Identification

A scintillation (gamma) camera is a device intended to image the distribution of radionuclides in the body by means of a photon radiation detector. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 # SEP 1 0 2004 Mr. Michael G. Stabin, Ph.D., CHP Assistant Professor of Radiology and Radiological Sciences Vanderbilt University 1161 21st Avenue South NASHVILLE TN 37232-2675 Re: K033960 Trade/Device Name: OLINDA EXM Regulation Number: 21 CFR 892.1100 Regulation Name: Scintillation (gamma) camera Regulatory Class: I Product Code: 90 IYX Dated: May 19, 2004 . Received: May 20, 2004 Dear Dr. Stabin: This letter corrects our substantially equivalent letter of June 15, 2004 regarding the typo for the incorrect regulatory class. The June 15th letter stated this device as class II instead of class I. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications ) for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {1}------------------------------------------------ Page 2 - Dr. Stabin forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of the letter: | 8xx.1xxx | (301) 594-4591 | |----------------------------------|----------------| | 876.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4616 | | 884.2xxx, 3xxx, 4xxx, 5xxx, 6xxx | (301) 594-4616 | | 892.2xxx, 3xxx, 4xxx, 5xxx | (301) 594-4654 | | Other | (301) 594-4692 | Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97) you may obtain. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html. Sincerely yours, Vividh. Lyam Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ | 510(k) Number (if known): | K033960 | |---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | Olinda EXM | | Indications For Use: | The purpose of Olinda EXM is to estimate the radiation dose received by internal organs as a result of administering a radiopharmaceutical. | (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) | Prescription Use<br>(Per 21 CFR 801.109) | <div style="font-size: 2em;">✓</div> | |------------------------------------------|--------------------------------------| |------------------------------------------|--------------------------------------| OR Over-The-Counter Use (Optional Format 1-2-96) ![Signature](data:image/png;base64,iVBORw0KGgoAAAANSUhEUgAAAAEAAAABCAQAAAC1HAwCAAAAC0lEQVR42mNkYAAAAAYAAjCB0C8AAAAASUVORK5CYII=) (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices | 510(k) Number | K033960 | |---------------|---------| |---------------|---------| ۽ : : : 公司 : {3}------------------------------------------------ K033960 JUN 1 5 2004 ## 510(k) Summary Applicants name address, phone and fax: Vanderbilt University Contact persons name and address: Michael Stabin Vanderbilt Universitv 1161 21st Avenue South Nashville, TN 37232-2675 Telephone / Fax number of contact: (615) 343-0068 / (615) 322-3764 Date Summary prepared: #### Trade Name: OLINDA/EXM Common Name: Organ Level !Nternal Dose Assessment/EXponential Modeling Classification: Not yet classified Devices OLINDA/EXM is substantially equivalent to: CDI3, Computer Program for Tissue Doses in Diagnostic Radiology, distributed by the Food and Drug Administration. CAMIRD, distributed by the Biomedical Computing Technology Information Center, Oak Ridge National Laboratory, PO Box X, Oak Ridge TN 37830. ### Description of OLINDA/EXM: The personal computer code OLINDA, which is an acronym standing for Organ Level INternal Dose Assessment/EXponential Modeling, calculates radiation doses to different organs of the body from radiopharmaceuticals which are administered systemically (mostly intravenously, but sometimes by oral or inhalation intake routes). The code requires input from the user on: - 1) Which radionuclide is to be used. - 2) A choice of body model(s) to represent the subject(s) of interest models exist for adult males, adult females, children, and women at different stages of pregnancy, and a number of individual organ systems (not included in the body phantoms). - 3) Parameters which describe the biokinetics of the radiopharmaceutical within different organs of the body with time. Specifically, a potential user needs to provide the time integral of activity in all important source organs of the body. Alternatively, the user may provide biokinetic data and the EXM portion of the code will fit these data to a model, calculate the necessary integrals, and pass them to the OLINDA portion. {4}------------------------------------------------ The code works in the Windows 2000 or Windows XP Professional operating environments. Intended Use of Olinda EXM: The purpose of OLINDA/EXM is to estimate radiation doses received by internal organs as a result of administering a radiopharmaceutical. ## Comparison to CAMIRD This program was legally distributed by the Biomedical Computing Technology Information Center, Oak Ridge National Laboratory, PO Box X, Oak Ridge TN 37830 in April 1976. It was described in a paper by Feller (1976), a copy of which is attached. It has been used by others and is referred to in a paper by Bellina and Guzzardi (1980). It is a pre-amendment device. It has not been through a 510(k) process and does not have a document control number. It has not been classified. Three versions of CAMIRD were written. The comparison here is to version II. | Comparison | CAMIRD | OLINDA/EXM | Discussion | |---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for<br>use | Estimates the<br>absorbed doses to<br>several tissues of a<br>reference patient for a<br>specified<br>radiopharmaceutical<br>dosage | Estimates the absorbed<br>doses to several tissues<br>of a reference patient<br>for a specified<br>radiopharmaceutical<br>dosage. | Equivalent.<br>Both calculate<br>dose per unit<br>input. | | Target<br>population | Specific patients and<br>patient groups not<br>targeted. Calculations<br>are based on models of<br>an average individual<br>(reference adult<br>male). | Specific patients and<br>patient groups not<br>targeted. Calculations<br>are based on models of<br>average individuals. 10<br>models are available<br>e.g. adult male, female,<br>5 year-old, 6-month<br>pregnant woman. | Target<br>population<br>equivalent.<br>OLINDA/EXM<br>has a larger<br>number of<br>phantoms. | | Design | User specified<br>radiopharmaceutical<br>kinetic parameters and<br>previous Monte Carlo<br>calculated organ<br>contributions.<br>Algorithm: MIRD method<br>(Loevinger et al. 1988) | User specified<br>radiopharmaceutical<br>kinetic parameters and<br>previous Monte Carlo<br>calculated organ<br>contributions.<br>Algorithm: MIRD method<br>(Loevinger et al. 1988) | The programs<br>are essentially<br>equivalent in<br>input and<br>output. The<br>main difference<br>is in the number<br>of body models<br>available. | | | Input: | Input: | | | | 1. Radionuclide<br>2. Body model<br>3. Radiopharmace<br>utical<br>biokinetics | 1. Radionuclide<br>2. Body model<br>3. Radiopharmaceutical<br>biokinetics | | | Energy used<br>and/or delivered | The program estimates<br>doses from user-<br>entered criteria and<br>precalculated data; it<br>is not connected to an<br>energy-emitting<br>device. | The program estimates<br>doses from user-entered<br>criteria and<br>precalculated data; it is<br>not connected to an<br>energy-emitting device. | Equivalent | | Performance | Output of dose per<br>unit input.<br>Simpler system than<br>OLINDA/EXM. Fortran<br>IV, input driven<br>program. Only 1 body<br>phantom, fewer organs<br>in output. | Output of dose per unit<br>input.<br>User friendly, event<br>driven, more body<br>phantoms, more organs<br>in output. | OLINDA/EXM is<br>equivalent in<br>performance to<br>CAMIRD. | | Human Factors | A descriptive paper is<br>available in the open<br>literature. | Communication tools for<br>error prevention have<br>been systematically<br>implemented. Error<br>messages, help files.<br>user manual and<br>installation tests have<br>all been developed so as<br>to educate the user and<br>prevent mistakes. An<br>open literature<br>publication is in<br>preparation describing<br>the OLINDA/EXM code.<br>An open literature<br>publication is available<br>describing the MIRDOSE<br>code (Stabin 1996), on<br>which the OLINDA/EXM<br>code was based. | OLINDA/EXM is<br>more user-<br>friendly. | | Anatomical sites | Program Tissue doses<br>calculated: Adrenals,<br>Fat, Blood, ovaries,<br>Skin, Uterus, Lower<br> | Tissue doses calculated:<br>Adrenals, Brain, Gall<br>Bladder Wall, Gall<br>Bladder Cont, Lower<br>Large Intestine Wall,<br>Lower Large Intestine | More tissues<br>included in<br>OLINDA/EXM. | {5}------------------------------------------------ : {6}------------------------------------------------ dose factors are based have been established previously in the literature (Cristy and Eckerman 1987, Stabin et al. 1995) and have been widely accepted and and Eckennan 1707, on one one one one only . Both programs are limited in that they use average body models and thus are not representative of any chat they use areful gu new has more body models to choose from and also particular pations. Sales the mass of individual organs to provide a limited measure of patient-specificity. Both programs are intended to be used by mcasure oversicists or appropriately trained physicians or equivalent in research er hospital facilities. Both programs need trained professional personnel to operate the program, provide the appropriate input and interpret the results. #### References Bellina CR and Guzzardi R. CAMIRD/III: a revised version of the CAMIRD/II and MIRD-S packages for internal dose calculation: concise communication. Journal of Nuclear Medicine, Vol 21, Issue 4 379-383 1980. Cristy M. and Eckerman K. Specific absorbed fractions of energy at various ages from internal photons sources. ORNL/TM-8381 V1-V7. Oak Ridge National Laboratory, Oak Ridge, TN; 1987. Feller PA. Computer Software to Facilitate Absorbed Dose Calculations, , in Radiopharmaceutical Dosimetry Symposium -- Proceedings of a Conference Held at Oak Ridge, Tenn., April 26-29, 1976, ed. by R.J. Cloutier, J.L. Coffey, W.S. Snyder and E.E. Watson, pp.119-126, HEW Publication (FDA) 76-8044, June 1976. International Commission on Radiological Protection. 1990 Recommendations of the International Commission on Radiological Protection. ICRP Publication 60, Pergamon Press, New York, 1991. Loevinger R, Budinger T, Watson E: MIRD primer for absorbed dose calculations. Society of Nuclear Medicine; 1988. Peterson LE and Rosenstein M. Computer program for tissue doses in diagnostic radiology. Food and Drug Administration, Center for Devices and Radiological Health, Rockville, MD 1989. Rosenstein, M. HEW Publication FDA 76-8030, Food and Drug Administration, Rockville. Marvland. 1976. Stabin M. MIRDOSE - the personal computer software for use in internal dose assessment in nuclear medicine. J Nucl Med, 37:538-546; 1996. {7}------------------------------------------------ Stabin MG, da Luz CQPL. New decay data For internal and external dose assessment, Health Phys. 83(4):471-475, 2002. Stabin MG and Siegel JA. Physical Models and Dose Factors for Use in Internal Dose Assessment. Health Physics, 85(3):294-310, 2003. ### Comparison to CDI3 This program is distributed by the Food and Drug Administration through their web site www.fda.gov. It has been available since 1989 and is widely available in the radiation protection community. It has not been through a 510(k) process and does not have a document control number. It has not been classified. | Comparison | CDI3 | OLINDA/EXM | Discussion | |----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------| | Indications for<br>use | Estimates the absorbed<br>doses to various tissues<br>of a reference patient<br>for a number of<br>specified X-ray<br>procedures. The<br>program also calculates<br>a "cancer detriment<br>index" for the aggregate<br>of the tissue doses. | Estimates the absorbed<br>doses to several tissues of<br>a reference patient for a<br>specified<br>radiopharmaceutical<br>dosage. | First indication<br>essentially<br>equivalent.<br>OLINDA/EXM<br>does not<br>calculate<br>"cancer<br>detriment<br>index" | | Target<br>population | Specific patients and<br>patient groups not<br>targeted. Calculations<br>are based on model of<br>average human body<br>"phantom". | Specific patients and<br>patient groups not<br>targeted. Calculations are<br>based on models of<br>average individuals. 10<br>models are available e.g.<br>adult male, female, 5<br>year-old, 6-month<br>pregnant woman. | Target<br>population<br>equivalent.<br>OLINDA/EXM<br>has a larger<br>number of<br>phantoms. | | Design | X-ray examination input<br>parameters are<br>combined with<br>previously established<br>Monte Carlo calculations<br>of dose per unit input to<br>give dose estimates.<br>Algorithm given in<br>section 4 of "Organ<br>doses in diagnostic<br>radiology" (Rosenstein<br>1976). | User specified<br>radiopharmaceutical<br>kinetic parameters and<br>previous Monte Carlo<br>calculated organ<br>contributions. Algorithm:<br>MIRD system (Loevinger et<br>al. 1988)<br>Input:<br>Radionuclide<br>1. | Different<br>algorithms,<br>similar inputs,<br>same output. | | | Input: | | | | | 1. X-ray Spectra data | 8. Body model | | | | 2. Exposure parameters<br>(entrance exposure, R,<br>source/image distance,<br>receptor size) | 9. Radiopharmaceutic<br>al biokinetics | | | | 3. Projection<br>parameters | Output: Dose to organs in<br>mSv/MBq and rem/mCi. | | | | Out put: Dose to organs<br>in mrad | | | | Energy used<br>and/or<br>delivered | The program estimates<br>doses from user-entered<br>criteria and<br>precalculated data; it is<br>not connected to an<br>energy-emitting device. | The program estimates<br>doses from user-entered<br>criteria and precalculated<br>data; it is not connected<br>to an energy-emitting<br>device. | Equivalent | | | | | | | Performance | Output of dose per unit<br>input.<br>DOS-based, input driven<br>program. Fewer body<br>phantoms.<br>Fewer organs in output. | Output of dose per unit<br>input.<br>User friendly,<br>Event driven<br>More body phantoms<br>More organs in output.<br>Communication tools for<br>error prevention have<br>been systematically<br>implemented. Error<br>messages, help files, user<br>manual and installation<br>tests have all been<br>developed so as to | OLINDA/EXM is<br>equivalent in<br>performance to<br>CDI3. | | | | | educate the user and<br>prevent mistakes. | | Human Factors | A user's manual was<br>published by the FDA<br>(Peterson and<br>Rosenstein 1989).<br>Current availability is<br>uncertain. | An open<br>literature publication is in<br>preparation describing the<br>OLINDA/EXM code. An<br>open literature<br>publication is available<br>describing the MIRDOSE<br>code (Stabin 1996), on<br>which the OLINDA/EXM<br>code was based. | OLINDA/EXM is<br>equivalent to<br>CDI3 given the<br>current state of<br>computer<br>technology. | | | | | | | Anatomical<br>sites | Program Tissue doses<br>calculated: lungs,<br>active bone marrow,<br>ovaries, testes, thyroid,<br>uterus, total trunk<br>(excluding skeletal and<br>lung tissues), and female | Tissue doses calculated:<br>Adrenals, Brain, Gall<br>Bladder Wall, Gall Bladder<br>Cont, Lower Large<br>Intestine Wall, Lower<br>Large Intestine Cont,<br>Small Intestine, Stomach | More tissues<br>are included in<br>OLINDA/EXM. | | | | | | | | breasts. | Wall, Stomach Cont, | | | | | Upper Large Intestine | | | | | Wall, Upper Large | | | | | Intestine Cont, Heart | | | | | Wall, Heart Cont, Kidneys, | | | | | Liver, Lungs, Spleen, | | | | | Pancreas, Prostate, | | | | | Skeleton, Active Marrow, | | | | | Skin, Thyroid, Thymus, | | | | | Testes, Urin.Bl. Wall, | | | | | Urin.Bl. Cont, Whole Body | | | Compatibility<br>with other<br>devices | The only device that the<br>program interacts with is<br>the PC on which it is<br>run. | The only devices that the<br>program interacts with is<br>the PC on which it is run<br>and the printer associated<br>with the PC. | Equivalent | | Where used | Used by physicists to<br>investigate new<br>radiopharmaceuticals,<br>and estimate patient<br>doses. Used in<br>universities,<br>pharmaceutical<br>manufacturing firms,<br>government agencies,<br>hospitals and research<br>facilities. | Used by physicists to<br>investigate new<br>radiopharmaceuticals, and<br>estimate patient doses.<br>Used in universities,<br>pharmaceutical<br>manufacturing firms,<br>government agencies<br>hospitals and research<br>facilities. | Equivalent | | Standards met | None | None | Equivalent | {8}------------------------------------------------ {9}------------------------------------------------ ### Summary of comparison CDI3 and OLINDA/EXM are both computer software programs that estimate the radiation dose received by tissues in average body models. OLINDA/EXM does not estimate a "cancer detriment index", as this was deemed to be outside the scope of the program's application. OLINDA/EXM does, however, calculate equivalent dose via application of radiation weighting factors currently recommended by the ICRP (ICRP 1991). These radiation weighting factors may be modified by the user if desired. Evaluation of risk is left to the user, through application of risk models. This is not treated in OLINDA/EXM in any way. They both use Monte Carlo based calculations and require input related to the radiation delivery method. Both programs produce tables of tissue doses. The calculation undertaken is different because the radiation source is different. The utility of OLINDA/EXM is approximately equivalent to that of the CDI3 program because the calculations for both programs are based on current scientific best practice as determined by respectively the MIRD Committee (Medical Internal Radiation Dose Committee) and the FDA. The decay data and {10}------------------------------------------------ dose factors used in the OLINDA/EXM code have been extensively peer reviewed. The decay data were published in the Health Physics Journal in 2002 (Stabin and da Luz 2002), and currently may be viewed through the Health (Stable Society web site. The dose factors were then published in the Health Physics Journal in 2003 (Siegel and Stabin 2003). The phantoms on which the r nyales our are based have been established previously in the literature (Cristy dose ractors and base base al. 1995) and have been widely accepted and une ablishe international dosimetry community. Both programs are limited in that they use average body models and thus are not representative of any charticular patient. OLINDA/EXM has more body models to choose from and also permits the user to vary the mass of individual organs to provide a limited measure of patient-specificity. Both programs are intended to be used by medial c or speropriately trained physicians or equivalent in research or hospital facilities. Both programs need trained professional personnel to of nospitale program, provide the appropriate input and interpret the results. ### References Cristy M. and Eckerman K. Specific absorbed fractions of energy at various ages from internal photons sources. ORNL/TM-8381 V1-V7. Oak Ridge National Laboratory, Oak Ridge, TN; 1987. International Commission on Radiological Protection. 1990 Recommendations of the International Commission on Radiological Protection. ICRP Publication 60, Pergamon Press, New York, 1991. Loevinger R, Budinger T, Watson E: MIRD primer for absorbed dose calculations. Society of Nuclear Medicine; 1988. Peterson LE and Rosenstein M. Computer program for tissue doses in diagnostic radiology. Food and Drug Administration, Center for Devices and Radiological Health, Rockville, MD 1989. Rosenstein, M. HEW Publication FDA 76-8030, Food and Drug Administration, Rockville, Maryland. 1976. Stabin M. MIRDOSE - the personal computer software for use in internal dose assessment in nuclear medicine. J Nucl Med, 37:538-546; 1996. Stabin MG, da Luz CQPL. New decay data For internal and external dose assessment, Health Phys. 83(4):471-475, 2002. Stabin MG and Siegel JA. Physical Models and Dose Factors for Use in Internal Dose Assessment. Health Physics, 85(3):294-310, 2003.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...