← Product Code [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO) · K990248

# DICOM IMAGE MANAGER, MODEL 3-D VISUALIZATION OPTION (K990248)

_Dicomit Imaging Systems Corp. · IYO · Apr 16, 1999 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K990248

## Device Facts

- **Applicant:** Dicomit Imaging Systems Corp.
- **Product Code:** [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO.md)
- **Decision Date:** Apr 16, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

The DICOMIT Dicom Image Manager medical image management system is intended for acceptance, transfer, display, storage and digital processing of medical images. The software component provides functions for performing operations related to image manipulation. The "3D ROI" software option enables the device to display and manipulate an adjunctive 3D volume image created with data from a diagnostic ultrasound scanning procedure.

## Device Story

Dicom Image Manager "3D ROI" is a software module for medical image management systems. It accepts sequential 2D ultrasound images acquired via standard transducers during freehand scanning procedures. The software processes these sequential images—stored in system memory (Cine Loop)—to render a 3D volume image. This volume is displayed and manipulated in windows on a monitor. Used in clinical settings, the device assists healthcare providers by providing adjunctive 3D visualization of ultrasound data, aiding in image review and clinical assessment. It does not require specialized positioning devices or sensors for data acquisition.

## Clinical Evidence

No clinical data. The submission relies on a comparison of technological characteristics and intended use with legally marketed predicate devices.

## Technological Characteristics

Software-based 3D visualization module for medical image management systems. Operates on standard ultrasound system hardware. Utilizes freehand scanning acquisition without specialized positioning sensors. Rendering process uses sequential B-mode ultrasound images stored in Cine memory. Connectivity via DICOM standard.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Dicom Image Manager ([K951925](/device/K951925.md))
- ATL HDI 3000/5000 Ultrasound systems ([K961459](/device/K961459.md))
- GE LOGIQ 700 Ultrasound system ([K964617](/device/K964617.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

4/16/99

### Page A - 4 510(k) Notification DICOM IMAGE MANAGER - Addition of 3D Visualization Software Module 510 (k) SUMMARY

K990248

This following summary is provided as part of this Premarket Notification in compliance with and based on the format set forth in the Final Rule as published in the Federal Register, December 14, 1994. (See 21 CFR § 807.92)

## (1) Submitters Name / Contact Person:

Dicomit Imaging Systems Corp. 75 East Beaver Creek Road, Unit 9 Richmond Hill, Ontario Canada L4B 1K6

| Contact Person: | Terry Callahan                |
|-----------------|-------------------------------|
|                 | Tel.: (905) 886-9496          |
|                 | Fax: (905) 886-2109           |
|                 | E-mail: tcallahan@dicomit.com |

January 22, 1999 Date prepared:

(2) Name of device:

| Trade Name:          | Dicom Image Manager™ "3D ROI™"        |
|----------------------|---------------------------------------|
| Common Name:         | 3D Visualization Module               |
| Classification Name: | Ultrasonic Pulsed Echo Imaging System |

# (3) Identification of predicate devices:

| Manufacturer                     | Device                               | 510(k) Number |
|----------------------------------|--------------------------------------|---------------|
| Dicomit Imaging Systems Corp.    | Dicom Image Manager                  | K951925       |
| Advanced Technology Labs         | ATL HDI 3000/5000 Ultrasound systems | K961459       |
| General Electric Medical Systems | GE LOGIQ 700 Ultrasound system       | K964617       |

# (4) Description of the device:

The software changes described in this submission enable the DICOMIT Dicom Image Manager - with 3D-visualization software module to store in memory, multiple sequential ultrasound images when using a standard ultrasound transducer during any standard ultrasound scanning procedure. These images can then be rendered as a 3D image that can be displayed and manipulated in "windows" on the monitor.

{1}------------------------------------------------

#### Page A - 5 510(k) Notification DICOM IMAGE MANAGER - Addition of 3D Visualization Software Module

## (5) A statement of the intended use of the device:

The DICOMIT Dicom Image Manager medical image management system is intended for acceptance, transfer, display, storage and digital processing of medical images. The software component provides functions for performing operations related to image manipulation.

The "3D ROI" software option enables the device to display and manipulate an adjunctive 3D volume image created with data from a diagnostic ultrasound scanning procedure.

## (6) Predicate Device Comparison:

The DICOMIT Dicom Image Manager - with 3D Visualization Software Module is substantially equivalent to similar features in the predicate devices and has the same intended uses and technological characteristics as the predicate devices.

The DICOMIT system and the predicate devices all use standard ultrasound system probes. All systems acquire multiple sequential images scanned in a "freehand" manner by the user without special positioning devices or position sensors.

The resulting sequential images are stored in the ultrasound system memory commonly called "Cine" or "Cine Loop". On all three systems the 3D volume images are rendered using the individual B-mode scan images acquired from the "freehand" scan.

The DICOMIT Dicom Image Manager - with 3D Visualization Software Module creates a 3D volume image in the same manner as the predicate devices described in this submission.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an image of an abstract eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 1 6 1999

Terry Callahan President Dicomit Imaging Systems Corp. 75 East Beaver Creek Rd., Unit 9 Richmond Hill. Ontario Canada L4B 1K6

Re:

K990248 DICOM Image Manager, 3D Visualization Option "3D ROI" Dated: January 22, 1999 Received: January 26, 1999 Regulatory class: II 21 CFR 892_1560/Procode: 90 IYO

Dear Mr. Callahan:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marked predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

APT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

### Page 5 510(k) Notification DICOM IMAGE MANAGER Addition of 3D Visualization Software Module

510(k) Number (if Known): K990248

### Device Name: DICOMIT Dicom Image Manager "3D ROI" Option

Indications For Use:

The DICOMIT Dicom Image Manager medical image management system is intended for acceptance, transfer, display, storage and digital processing of medical images. The software component provides functions for performing operations related to image manipulation.

The "3D ROI" software option enables the device to display and manipulate an adjunctive 3D volume image created with data from a diagnostic ultrasound scanning procedure.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

Dmeld W. Slu for DDS

vision Sion Off ivision of Reproductive, Abdominal, EN and Radiological C 510(k) Number

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K990248](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K990248)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
