K981626 · Siemens Medical Solutions USA, Inc. · IYO · May 27, 1998 · Radiology
Device Facts
Record ID
K981626
Device Name
SONOLINE ELEGRA DIAGNOSTIC ULTRASOUND SYSTEM
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
IYO · Radiology
Decision Date
May 27, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The SONOLINE Elegra ultrasound imaging system is intended for the following applications: General Radiology, Abdominal, Intraoperative, Small Parts, Transcranial, OB/GYN, Pelvic, Neonatal/Adult Cephalic, Urology, Vascular, Musculoskeletal. Superficial Musculoskeletal, and Peripheral Vascular applications. The addition of 3-D Imaging will not add new indications for use to the Elegra. It will allow the user to display regions of interest in 3-dimensional format, either by showing views in three orthagonal planes or allowing surface rendering of objects. The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
Device Story
The SONOLINE Elegra is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound data via various transducers and displays it in B-Mode, M-Mode, Color Mode, Pulsed (PW) Doppler, and Continuous (CW) Doppler. This submission adds 3D imaging capabilities, allowing users to visualize regions of interest through three orthogonal planes or surface rendering. The system is operated by clinicians in clinical settings. It provides anatomical measurements and analysis packages to assist in clinical diagnosis. The 3D reconstruction technique is functionally equivalent to the previously cleared SieScape panoramic imaging feature. The device benefits patients by providing enhanced visualization of anatomical structures for diagnostic purposes.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Mobile, software-controlled diagnostic ultrasound system. Supports B-Mode, M-Mode, Color, PW Doppler, and CW Doppler. Includes 3D reconstruction and surface rendering. Complies with UL 2601, CSA 22.2 No. 601-1, AIUM/NEMA 1992 (thermal/mechanical indices), and 93/42/EEC. Connectivity includes on-screen display for indices.
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis in general radiology, abdominal, intraoperative, small parts, transcranial, OB/GYN, pelvic, neonatal/adult cephalic, urology, vascular, and musculoskeletal applications. Prescription use only.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
### MAY 2 7 1998 510(K) SUMMARY
## Addition of 3 D Imaging to SONOLINE Elegra Diagnostic Ultrasound system
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with the Safe Medical Device Act of 1990 revisions to 21 CFR, Part 807.92, Content and Format of a 510(k) Summary.
#### Submitted By: 1.
Siemens Medical Systems, Inc., Ultrasound Group 22010 S.E. 51st Street Issaquah, WA 98027-7002
## Contact Person: Steve Hesler Manager of Regulatory Affairs (425) 557-1629
## Date Prepared: March 31, 1998
- 2. Proprietary Name: SONOLINE Elegra Advanced SONOLINE Elegra
Common/ Usual Name: Diagnostic Ultrasound System with Accessories
## Classification Name: Ultrasonic Pulsed Doppler Imaging System (Product Code 90 IYN, 21 CFR 892.1550)
#### 3. Predicate Device:
K945072, 11/21/95, cleared as the Q4000, marketed as the SONOLINE Elegra Advanced
K950517, 4/5/96, cleared as Q64XX, marketed as the SONOLINE Elegra K961833, 10/29/96, SieScape panoramic display for SONOLINE Elegra Advanced
#### 4. Device Description:
The SONOLINE Elegra is a general purpose, mobile, software-controlled, diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bioeffect mechanisms. Its function is to acquire ultrasound data and display it in B-Mode, M-Mode, Color Mode, Pulsed (PW) Doppler Mode, Continuous (CW) Doppler Mode, or in a combination of modes, on a CRT display.
The SONOLINE ® Elegra, has been designed to meet the following product safety standards:
- UL 2601, Safety Requirements for Medical Equipment .
- CSA 22.2 No. 601-1, Safety Requirements for Medical Equipment .
- Standard for Real Time Display of Themal and Mechanical Indices on . Diagnostic Ultrasound Equipment, AlUM/NEMA, 1992.
- 93/42/EEC Medical Devices Directive
EN60601 = (IEC 601-1-1 + IEC 601-1-2), Safety and EMC Requirements for Medical Equipment
{1}------------------------------------------------
#### 5. Intended Uses:
The SONOLINE Elegra ultrasound imaging system is intended for the following applications: General Radiology, Abdominal, Intraoperative, Small Parts, Transcranial, OB/GYN, Pelvic, Neonatal/Adult Cephalic, Urology, Vascular, Musculoskeletal. Superficial Musculoskeletal, and Peripheral Vascular applications. The addition of 3-D Imaging will not add new indications for use to the Elegra. It will allow the user to display regions of interest in 3-dimensional format, either by showing views in three orthagonal planes or allowing surface rendering of objects.
The system also provides for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.
#### Technological Comparison to Predicate Device: 6.
SONOLINE Elegra is a previously cleared device. The purpose of this submission is to receive clearance for the addition of 3 dimensional imaging to the already-cleared svstem. This real-time reconstruction technique is equivalent to the SieScape panaoramic imaging feature already cleared by 510(k) (see K961833, 10/29/96).
## End of 510(k) Summary
t with 5
{2}------------------------------------------------
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three birds in flight, arranged in a row.
# MAY 2 7 1998
Siemens Medical Systems. Inc. Ultrasound Group c/o Joseph P. Murnane Underwriters Laboratories, Inc. 1285 Walt Whitman Road Melville, New York 11747-3081 Rc:
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
K981626 Sonoline® Elegra Diagnostic Ultrasound System Dated: May 1, 1998 Received: May 7, 1998 Regulatory class: II Procodes: 90 IYN, 21 CFR 892.1550 90 IYO, 21 CFR 892.1560 90 ITX, 21 CFR 892. 1570
Dear Mr. Murnane:
We have reviewed your Section 510%) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Festeral Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions agginst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (2) CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your devise in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electrotic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmam.html".
Sincerely yours,
William Yu
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{3}------------------------------------------------
510 (k) Number (if
known) :
SONOLINE Elegra with SieScape and 3 D Imaging Device Name :
Intended Use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | P | P | P | | BMDC (P) | |
| Abdominal | | P | P | P | P | P | P | | BMDC (P) | |
| Intraoperative<br>(Specify) | | P | P | P | P | P | P | | BMDC (P) | |
| Pediatric | | P | P | P | P | P | P | | BMDC (P) | |
| Small Organ<br>(Specify) | | P | P | P | P | P | P | | BMDC (P) | |
| Neonatal Cephalic | | P | P | P | P | P | P | | BMDC (P) | |
| Adult Cephalic | | P | P | P | P | P | P | | BMDC (P) | |
| Cardiac | | P | P | P | P | P | P | | BMDC (P) | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | P | P | P | P | P | P | | BMDC (P) | |
| Transvaginal | | P | P | P | P | P | P | | BMDC (P) | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | P | P | P | | BMDC (P) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | P | P | P | | BMDC (P) | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication;
P = previously cleared by FDA;
E = added
under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David G. Seymore
(Division Sign-Off)
(Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K981626
{4}------------------------------------------------
510 (k) Number (if
known) :
2.0 CW probe Device Name :
Intended Use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
## Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|---------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | P | | | | | |
| Abdominal | | | | | P | | | | | |
| Intraoperative<br>(Specify) | | | | | P | | | | | |
| Pediatric | | | | | P | | | | | |
| Small Organ<br>(Specify) | | | | | P | | | | | |
| Neonatal Cephalic | | | | | P | | | | | |
| Adult Cephalic | | | | | P | | | | | |
| Cardiac | | | | | P | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | P | | | | | |
| Transvaginal | | | | | P | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | P | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal | | | | | P | | | | | |
| Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
under Appendix E
Additional Comments:___
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David C. Severson
(Division Sign-Off)
(Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K481626
{5}------------------------------------------------
510 (k) Number (if
known) :
## 2.5PL20 Phased Array Transducer Device Name :
Intended Use:
2.5F E20 Friased Aray andy or fluid flow analysis of the human body as follows:
## Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) | |
|----------------------------------------------------------------------------------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Ophthalmic | | | | | | | | | | | |
| Fetal | | E | E | E | E | | E | | BMDC (E) | | |
| Abdominal | | E | E | E | E | | E | | BMDC (E) | | |
| Intraoperative<br>(Specify) | | | | | | | | | | | |
| Pediatric | | E | E | E | E | E | E | | BMDC (E) | | |
| Small Organ<br>(Specify) | | | | | | | | | | | |
| Neonatal Cephalic | | E | E | E | E | E | E | | BMDC (E) | | |
| Adult Cephalic | | E | E | E | E | E | E | | BMDC (E) | | |
| Cardiac | | E | E | E | E | E | E | | BMDC (E) | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral vessel | | E | E | E | E | E | | E | | BMDC (E) | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
| N = new indication; P = previously cleared by FDA; E = added<br>under Appendix E<br>Additional Comments: | | | | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David A. Segerson
Division Sign Off
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number _ K981626
{6}------------------------------------------------
## 510 (k) Number (if
known) :
### 3.5PL28 Phased Array Transducer Device Name :
Intended Use:
5.5F E20 Thused Airay Truid flow analysis of the human body as follows:
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|--------------------------------------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | P | P | P | P | | BMDC (P) | |
| Fetal | | P | P | P | P | P | P | | BMDC (P) | |
| Abdominal | | P | P | P | P | | P | | BMDC (P) | |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | P | P | P | P | P | P | | BMDC (P) | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal Cephalic | | P | P | P | P | P | P | | BMDC (P) | |
| Adult Cephalic | | P | P | P | P | P | P | | BMDC (P) | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | P | P | P | | BMDC (P) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N = new indication; P = previously cleared by FDA; E = added | | | | | | | | | | |
| under Appendix E | | | | | | | | | | |
| Additional Comments: | | | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Jarid a. Jesson
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices
981626 510(k) Number
{7}------------------------------------------------
510 (k) Number (if
known) :
## 3.5C40 Curved Array Transducer Device Name :
Intended Use:
3.5040 Cul ved Array Transade of the data visis of the human body as follows:
Mode of Operation
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
|----------------------------------------------------------------------------------------------------|---|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | E | E | E | | E | E | | BMDC (E) | |
| Abdominal | | E | E | E | | E | E | | BMDC (E) | |
| Intraoperative (Specify) | | | | | | | | | | |
| Pediatric | | E | E | E | | E | E | | BMDC (E) | |
| Small Organ (Specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | E | E | E | | E | E | | BMDC (E) | |
| Laparoscopic | | E | E | E | | E | E | | BMDC (E) | |
| Musculo-skeletal Conventional | | E | E | E | | E | E | | | |
| Musculo-skeletal Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N = new indication; P = previously cleared by FDA; E = added under Appendix E Additional Comments: | | | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Lyons
---
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K487422
{8}------------------------------------------------
510 (k) Number (if known) : Device Name :
# 5.0HDPL40 Linear Array Transducer
Diagnostic ultrasound imaging or fluid flow analysis of the Intended Use: human body as follows:
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC (P) | |
| Abdominal | | P | P | P | | P | P | | BMDC (P) | |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | P | P | P | | P | P | | BMDC (P) | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC (P) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC (P) | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication;
P = previously cleared by FDA;
E = added
under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David G. Bergman
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K981626
{9}------------------------------------------------
510 (k) Number (if known) :
### 5.0C50 Curved Array Transducer Device Name :
Intended Use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|--------------------------------------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC (P) | |
| Abdominal | | P | P | P | | P | P | | BMDC (P) | |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | P | P | P | | P | P | | BMDC (P) | |
| Small Organ<br>(Specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC (P) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC (P) | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N = new indication: P = previously cleared by FDA: E = added | | | | | | | | | | |
## under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Hanson
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K981626
{10}------------------------------------------------
510 (k) Number (if
known) :
### 6.5EC10 Curved Array Transducer Device Name :
Diagnostic ultrasound imaging or fluid flow analysis of the Intended Use: human body as follows:
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------------------------------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) | | E | E | E | | E | E | | BMDC (E) | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | E | E | E | | E | E | | BMDC (E) | |
| Transvaginal | | E | E | E | | E | E | | BMDC (E) | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N = new indication; P = previously cleared by FDA; E = added<br>under Appendix F | | | | | | | | | | |
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Seigman
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K981626
{11}------------------------------------------------
510 (k) Number (if
known) :
### 6.5EV13 Curved Array Transducer Device Name :
Intended Use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ<br>(Specify) | | P | P | P | | P | P | | BMDC (P) | |
| Neonatal Cephalic | | P | P | P | | P | P | | BMDC (P) | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | P | P | P | | P | P | | BMDC (P) | |
| Transvaginal | | P | P | P | | P | P | | BMDC (P) | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
## under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Seymon
(Division Sign-Off) (Porision of Reproductive, Abdominal, ENT,
and Radiological Devices and Radiological Devices
510(k) Number K981626
{12}------------------------------------------------
510 (k) Number (if known) : Device Name :
# 7.5C30 Curved Array Transducer
Diagnostic ultrasound imaging or fluid flow analysis of the Intended Use: human body as follows:
Mode of Operation
| Clinical | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|--------------------------------------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Application | | | | | | | | | | |
| Ophthalmic | | | | | | | | | | |
| Fetal | | E | E | E | | E | E | | BMDC (E) | |
| Abdominal | | E | E | E | | E | E | | BMDC (E) | |
| Intraoperative | | | | | | | | | | |
| (Specify) | | | | | | | | | | |
| Pediatric | | E | E | E | | E | E | | BMDC (E) | |
| Small Organ | | E | E | E | | E | E | | BMDC (E) | |
| (Specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | E | E | E | | E | E | | BMDC (E) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | E | E | E | | E | E | | BMDC (E) | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N = new indication; P = previously cleared by FDA; E = added | | | | | | | | | | |
under Appendix E Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Thail G. Seyman
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number_K 78/626
(
{13}------------------------------------------------
510 (k) Number (if
known) :
7.5L40 Linear Array Transducer Device Name :
Intended Use:
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|---------------------------------------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | P | P | P | | P | P | | BMDC (P) | |
| Abdominal | | P | P | P | | P | P | | BMDC (P) | |
| Intraoperative<br>(Specify) | | P | P | P | | P | P | | BMDC (F) | |
| Pediatric | | P | P | P | | P | P | | BMDC (P) | |
| Small Organ<br>(Specify) | | P | P | P | | P | P | | BMDC (P) | |
| Neonatal Cephalic | | P | P | P | | P | P | | BMDC (P) | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | P | P | P | | P | P | | BMDC (P) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | BMDC (P) | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
| N = new indication. D = previously cleared by FDA. E = added. | | | | | | | | | | |
## under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
Hind G. Hyam
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number_K481626
{14}------------------------------------------------
510 (k) Number (if
known) :
### 7.5PL13 Phased Array Transducer Device Name :
Diagnostic ultrasound imaging or fluid flow analysis of the Intended Use: human body as follows:
Mode of Operation
| Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Ophthalmic | | | | | | | | | | |
| Fetal | | E | E | E | E | E | E | | BMDC (E) | |
| Abdominal | | E | E | E | E | E | E | | BMDC (E) | |
| Intraoperative<br>(Specify) | | | | | | | | | | |
| Pediatric | | E | E | E | E | E | E | | BMDC (E) | |
| Small Organ<br>(Specify) | | E | E | E | E | E | E | | BMDC (E) | |
| Neonatal Cephalic | | E | E | E | E | E | E | | BMDC (E) | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac | | | | | | | | | | |
| Transesophageal | | | | | | | | | | |
| Transrectal | | | | | | | | | | |
| Transvaginal | | | | | | | | | | |
| Transurethral | | | | | | | | | | |
| Intravascular | | | | | | | | | | |
| Peripheral vessel | | E | E | E | E | E | E | | BMDC (E) | |
| Laparoscopic | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | E | E | E | E | E | E | | BMDC (E) | |
| Musculo-skeletal<br>Superficial | | | | | | | | | | |
| Other (specify) | | | | | | | | | | |
N = new indication;
P = previously cleared by FDA;
E = added
under Appendix E
Additional Comments:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Thind le. Beynam
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K981626
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.