RT55 DIAGNOSTIC ULTRASOUND SYSTEM
Device Facts
| Record ID | K953752 |
|---|---|
| Device Name | RT55 DIAGNOSTIC ULTRASOUND SYSTEM |
| Applicant | General Electric Co. |
| Product Code | IYO · Radiology |
| Decision Date | Mar 15, 1996 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.1560 |
| Device Class | Class 2 |
| Attributes | Pediatric |
Indications for Use
The RT55 portable ultrasound diagnostic system is intended for use in obstetrics, gynecology and general radiology examinations by a qualified physician to aid in the diagnosis and evaluation of soft tissues by generating two dimensional images, time-motion images and biometric data. The specific intended uses of this system include: abdominal, small parts, pediatric and fetal imaging.
Device Story
Portable ultrasound system (<10 kg); integrated monitor/keyboard. Supports linear, convex, and micro-convex probes. Used by physicians in clinical settings for abdominal, small parts, pediatric, and fetal imaging. Generates 2D images, time-motion images, and biometric data to aid soft tissue diagnosis. Physician interprets images/data to inform clinical decision-making.
Clinical Evidence
Bench testing only. Evaluated for acoustic output, biocompatibility, and thermal, electrical, and mechanical safety. Conforms to applicable medical device safety standards.
Technological Characteristics
Portable ultrasound system; <10 kg; dimensions ~28x41x29 cm. Supports linear, convex, and micro-convex probes. Operating modes: 2D, time-motion. Design conforms to GMP standards.
Indications for Use
Indicated for use in obstetrics, gynecology, and general radiology (abdominal, small parts, pediatric, fetal) to aid soft tissue diagnosis and evaluation via 2D, time-motion, and biometric data generation. For use by qualified physicians.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
- GE Medical Systems RT2800 diagnostic ultrasound system (K873700)
Related Devices
- K081873 — EMP-2100 FULL DIGITAL ULTRASOUND DIAGNOSTIC DEVICE · Shenzhen Emperor Electronic Technology Co. , Ltd. · Jul 16, 2008
- K992208 — GE LOGIQ 200 MD DIAGNOSTIC ULTRASOUND SYSTEM · Ge Medical Systems, Inc. · Jul 12, 1999
- K052113 — DP-6600 ULTRASONIC DIAGNOSTIC IMAGING SYSTEM · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Aug 9, 2005
- K171579 — M6/M6T/M6 EXP/M6S/M6 PRO/M5 EXP/M55/M58 Diagnostic Ultrasound System · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Oct 18, 2017
- K960700 — LOGIQ A200 DIAGNOSTIC ULTRASOUND SYSTEM · GE Medical Systems · Aug 28, 1996