← Product Code [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO) · K201316

# Bladder Scanner (Model: M3, M3-HD, M4, M4-HD) (K201316)

_Suzhou Peaksonic Medical Technology Co., Ltd. · IYO · Feb 26, 2021 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K201316

## Device Facts

- **Applicant:** Suzhou Peaksonic Medical Technology Co., Ltd.
- **Product Code:** [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO.md)
- **Decision Date:** Feb 26, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Bladder Scanner (Models: M3, M3-HD, M4, M4-HD) is B-mode pulsed-echo ultrasound device. It intended as a handheld battery-operated device. The M3, M3-HD, M4, M4-HD Bladder Scanner projects ultrasound energy through the lower abdomen of the patient to obtain images of the bladder which is used to calculate bladder Volume noninvasively. The M3, M3-HD, M4, M4-HD Bladder Scanner is intended to be used only by qualified medical professionals.

## Device Story

Handheld, battery-operated B-mode ultrasound device; consists of main unit, 3D (M4/M4-HD) or 2D (M3/M3-HD) probe, and app software on smartphone or tablet. Operator places probe on lower abdomen; device acquires multiple 2D ultrasound images; sophisticated image processing restores stereo image; algorithm calculates bladder volume. Results displayed on mobile device screen. Used by qualified medical professionals in clinical settings. Features include expert/easy/intelligence modes, wireless printing, and Wi-Fi/Bluetooth data transmission. Patient information is encrypted for privacy. Benefits include non-invasive, fast, and reliable bladder volume assessment to aid clinical decision-making.

## Clinical Evidence

No clinical data. Substantial equivalence supported by bench testing, including electrical safety (ES60601-1), electromagnetic compatibility (IEC 60601-1-2), acoustic output (NEMA UD 2), and wireless connectivity (FCC Part 15).

## Technological Characteristics

Handheld B-mode ultrasound; mechanical probe (2.5MHz, 120° sector angle); PC material skin contact; battery-powered; Wi-Fi/Bluetooth connectivity; software-based image processing and volume calculation; complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, and NEMA UD 2.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Bladder Scanner (Models: M2, M2-W, M1, M1-W) ([K190769](/device/K190769.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Suzhou Peaksonic Medical Technology Co., Ltd. % Calvin Ma General Manager 2A, West Side of Building G4 Kunshan Hi-Tech Medical Device Industrial Park Qiandeng, Kunshan, Suzhou, Jiangsu 215341 CHINA

February 26, 2021

### Re: K201316

Trade/Device Name: Bladder Scanner (Model: M3, M3-HD, M4, M4-HD) Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: Class II Product Code: IYO, ITX Dated: January 1, 2021 Received: January 25, 2021

### Dear Calvin Ma:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 

{1}------------------------------------------------

801) medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known)

K201316

Device Name Bladder Scanner (Model:M3, M3-HD, M4, M4-HD)

Indications for Use (Describe)

The Bladder Scanner (Models: M3, M4, M4-HD) is B-mode pulsed-echo ultrasound device. It intended as a handheld battery-operated device. The M3, M3-HD, M4, M4-HD Bladder Scanner projects ultrasound energy through the lower abdomen of the patient to obtain images of the bladder which is used to calculate bladder Volume noninvasively. The M3, M3-HD, M4, M4-HD Bladder Scanner is intended to be used only by qualified medical professionals.

| Type of Use (Select one or both, as applicable)                                     |                                                                                    |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(K) Summary

Prepared in accordance with the requirements of 21 CFR Part 807.92

Prepared Date: January 1, 2021

K201316

## 1. Submitter's Information

The submitter of this pre-market notification is:

| Name:           | SUZHOU PEAKSONIC MEDICAL TECHNOLOGY CO., LTD                                                                                                  |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Address:        | 2A, West Side of Building G4, Kunshan<br>Hi-Tech Medical Device Industrial Park ,<br>Qiandeng Town, Kunshan<br>Suzhou, JIANGSU, 215341, CHINA |
| Contact person: | Calvin Ma                                                                                                                                     |
| Title:          | General manager                                                                                                                               |
| E-mail:         | service@peaksonic.com.cn                                                                                                                      |
| Tel:            | +86- 512-36693388                                                                                                                             |

## 2. Device Identification

| Trade/Device Name: | Bladder Scanner                                                        |
|--------------------|------------------------------------------------------------------------|
| Models:            | M3, M3-HD, M4, M4-HD                                                   |
| Regulation Number: | 21 CFR 892.1560<br>21 CFR 892.1570                                     |
| Common Name:       | Ultrasonic, Pulsed echo, Imaging<br>Transducer, Ultrasonic, Diagnostic |
| Regulation Class:  | Class II                                                               |
| Product Code:      | IYO, ITX                                                               |

## 3. Predicate Device

| 510(K) number:     | K190769                                                                |
|--------------------|------------------------------------------------------------------------|
| Device Name:       | Bladder Scanner (Models:M2, M2-W, M1, M1-W)                            |
| Manufacturer:      | Suzhou Lischka Medtech Co., Ltd.                                       |
| Regulation Number: | 21 CFR 892.1560<br>21 CFR 892.1570                                     |
| Common Name:       | Ultrasonic, Pulsed echo, Imaging<br>Transducer, Ultrasonic, Diagnostic |
| Regulation Class:  | Class II                                                               |
| Product Code:      | IYO, ITX                                                               |

{4}------------------------------------------------

### 4. Device Description

The M Series Bladder Scanner is a handheld battery-operated device, which is developed by SUZHOU PEAKSONIC MEDICAL TECHNOLOGY CO., LTD., and manufactured by Suzhou Lischka Medtech Co., Ltd., it provides non-invasive bladder volume measurement utilizing real-time ultrasound imaging. The device consists of the main unit, 3D probe (M4, M4-HD)/2D probe (M3, M3-HD), Data processing and storage modules, APP software, battery and Charger.

- It features:
- . M3 and M3-HD has two Operation Modes: Expert Mode and Easy Mode, M4 and M4-HD has three Operation Modes: Expert Mode, Easy Mode and intelligence mode
- . Non-invasive, comfortable, correct, reliable, fast and simple operation. When the operator releases the button on the probe, multiple 2D plane ultrasound images are acquired in a few seconds. The equipment adopts sophisticated image processing techniques to restore the stereo image, and adopts a sophisticated algorithm to measure bladder volume and displays the measurement results on the screen.
- Use touch screen keyboard operation on tablet computer (or mobile phone)
- . Multilinqual choice
- Bluetooth Wireless Printing Report
- . Volume preset reminder
- . Information management, storage and printing
- Wifi wireless bidirectional transmission
- Leak-proof management of patient information: information is stored and transmitted after encryption, and the patient information can be viewed and uploaded only after password landing.
- . The instrument consists of a handheld wireless scanner and a tablet computer (or mobile phone)
- Built-in battery power supply .

The difference between these models is that the model of the probe is different. M4. M4-HD is 3D probe .M3, M3-HD is 2D probe. M3-HD, M4-HD will be operated by tablet computer, M3, M4 will be operated by smart mobile phone.

#### 5. Indication for use

The Bladder Scanner (Models: M3, M3-HD, M4, M4-HD) is B-mode pulsed-echo ultrasound device. It intended as a handheld battery-operated device. The M3, M3-HD, M4, M4-HD Bladder Scanner projects ultrasound energy through the lower abdomen of the patient to obtain images of the bladder which is used to calculate bladder Volume noninvasively. The M3, M3-HD, M4, M4-HD Bladder Scanner is intended to be used only by qualified medical professionals.

### 6. Comparison to Predicate Device

Comparison to the predicate devices, the subject device has same intended use, similar product design, same performance effectiveness, performance safety as the predicate device as summarized in the following table:

{5}------------------------------------------------

| SE Comparisons                                                    | Predicate device                                                                                                                                                                                                                                                                                                                         | Subject device                                                                                                                                                                                                                                                                                                                             | Comments                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|-------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer/K#                                                   | Suzhou Lischka<br>Medtech Co., Ltd./<br>K190769                                                                                                                                                                                                                                                                                          | SUZHOU PEAKSONIC<br>MEDICAL<br>TECHNOLOGY CO.,<br>LTD.                                                                                                                                                                                                                                                                                     | /                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Trade name and<br>model                                           | Bladder Scanner<br>(Models: M2, M2-W,<br>M1, M1-W)                                                                                                                                                                                                                                                                                       | Bladder Scanner<br>(Models: M3, M3-HD,<br>M4, M4-HD)                                                                                                                                                                                                                                                                                       | /                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Classifications<br>name<br>and<br>Regulation Name                 | Regulation Number:<br>21 CFR 892.1560<br>21 CFR 892.1570<br>Regulation Name:<br>Ultrasonic, Pulsed<br>echo, Imaging<br>Transducer,<br>Ultrasonic, Diagnostic<br>Product Code: IYO,<br>ITX                                                                                                                                                | Regulation Number:<br>21 CFR 892.1560<br>21 CFR 892.1570<br>Regulation Name:<br>Ultrasonic, Pulsed<br>echo, Imaging<br>Transducer, Ultrasonic,<br>Diagnostic<br>Product Code: IYO, ITX                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Indication for use                                                | The Bladder Scanner<br>(Models: M2, M2-W, M1, M1-W) projects<br>ultrasound energy<br>through the lower<br>abdomen of the<br>patient to obtain<br>images of the bladder<br>which is used to<br>calculate bladder<br>volume noninvasively.<br>The Bladder Scanner<br>is intended to be used<br>only by qualified<br>medical professionals. | The Bladder Scanner<br>(Models: M3, M3-HD,<br>M4, M4-HD) projects<br>ultrasound energy<br>through the lower<br>abdomen of the patient<br>to obtain images of the<br>bladder which is used to<br>calculate bladder<br>volume noninvasively.<br>The Bladder Scanner is<br>intended to be used only<br>by qualified medical<br>professionals. | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Contraindications                                                 | Do not use the Bladder<br>Scanner on following<br>cases:<br>a) Fetal use or<br>pregnant patients<br>b) Patients with<br>ascites<br>c) Patients with open<br>or damaged skin.<br>d) Wounds in the<br>suprapubic region                                                                                                                    | Do not use the Bladder<br>Scanner on following<br>cases:<br>a) Fetal use or pregnant<br>patients<br>b) Patients with ascites<br>c) Patients with open or<br>damaged skin.<br>d) Wounds in the<br>suprapubic region                                                                                                                         | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Modes<br>of<br>operation                                          | B mode                                                                                                                                                                                                                                                                                                                                   | B mode                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| System<br>Characteristics<br>and structure                        | Portable,<br>LCD Display,<br>Thermal Printer,<br>Power source: Battery                                                                                                                                                                                                                                                                   | Portable,<br>Smartphone or tablet<br>computer display,<br>Thermal Printer,<br>Power source: Battery                                                                                                                                                                                                                                        | Different,<br>The design of<br>predicate device is built-in<br>software and ultrasound<br>image and data is displayed<br>on the device's display.<br>Subject device has no<br>display itself and ultrasound<br>image and data is displayed<br>on the smartphone and<br>tablet computer through APP<br>software. The principles,<br>algorithms, and probes of<br>both devices are the same.<br>The subject device has<br>performed<br>software<br>verification according to the<br>FDA guidance and IEC<br>62304, the difference does<br>not affect the safety and<br>performance. |
| Display                                                           | M2: 2.4" TFT-LCD<br>M1: 2.4" TFT-LCD                                                                                                                                                                                                                                                                                                     | M3-HD, M4-HD will be<br>provided customer with<br>a Samsung SM-T590<br>10.5 inch.<br>The display of M3, M4<br>will be according to the<br>customer's smart<br>mobile<br>phone<br>specifications                                                                                                                                            | Different,<br>The size and specification of<br>the display is different and<br>does not affect product<br>safety and performance.                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Controls<br>for<br>Change<br>of<br>acoustic output<br>during scan | No                                                                                                                                                                                                                                                                                                                                       | No                                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Transducer Type                                                   | Mechanical<br>Probe                                                                                                                                                                                                                                                                                                                      | Mechanical<br>Probe                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Measurement<br>place                                              | Abdomen                                                                                                                                                                                                                                                                                                                                  | Abdomen                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Transducer<br>Resonant<br>Frequency                               | 2.5Mhz                                                                                                                                                                                                                                                                                                                                   | 2.5Mhz                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Number<br>of<br>elements                                          | 1                                                                                                                                                                                                                                                                                                                                        | 1                                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Sector Angle                                                      | 120°                                                                                                                                                                                                                                                                                                                                     | 120°                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Number of Scan<br>Planes                                          | M2, M2-W:12 ,<br>M1, M1-W:1                                                                                                                                                                                                                                                                                                              | M4, M4-HD:12 ,<br>M3, M3-HD:1                                                                                                                                                                                                                                                                                                              | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|                                                                   |                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Patient<br>Contacting<br>Material                                 | PC (skin contact)                                                                                                                                                                                                                                                                                                                        | PC (skin contact)                                                                                                                                                                                                                                                                                                                          | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Volume<br>measurement<br>range                                    | 0ml-999ml                                                                                                                                                                                                                                                                                                                                | 0ml-999ml                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Volume<br>measurement<br>accuracy                                 | M2/M2-W: $\pm$ 7%,<br>$\pm$ 7ml;<br>M1/M1-W: $\pm$ 14%,<br>$\pm$ 14ml                                                                                                                                                                                                                                                                    | M4/M4-HD:<br>under 100 mL: $\pm$ 7mL;<br>100 to 999 mL: $\pm$ 7%,<br>M3/M3-HD:<br>under 100 mL: $\pm$ 14mL;<br>100 to 999 mL: $\pm$ 14%                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Classification<br>of<br>protection against<br>electric shock      | Class II equipment                                                                                                                                                                                                                                                                                                                       | Class II equipment                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Applied part type                                                 | B type                                                                                                                                                                                                                                                                                                                                   | B type                                                                                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Real-time<br>scanning                                             | Yes (Pre-scan)                                                                                                                                                                                                                                                                                                                           | Yes (Pre-scan)                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| PC Data Upload                                                    | USB connection                                                                                                                                                                                                                                                                                                                           | USB connection                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Power                                                             | Lithium<br>battery:<br>URR18650ZY-2600<br>mAh(SNLB-435<br>7.4Vd.c. 2600mAh<br>Charger:<br>HXY-<br>084V1500A-UL<br>AC100-240Va                                                                                                                                                                                                            | Lithium<br>battery:<br>NCA653864SA-2400<br>mAh<br>(PC015-2S1P)<br>7.4Vd.c. 2400mAh<br>Charger: HXY-<br>084V1500A-UL<br>AC100-240Va                                                                                                                                                                                                         | Different, The lithium battery<br>used in the predicate device<br>and subject device is<br>different.<br>The lithium battery has been<br>performed<br>safety<br>test<br>according to the IEC62133,<br>and the subject device has<br>been performed safety test<br>according to the IEC 60601-<br>1, so the difference in lithium<br>battery specifications will not<br>raise<br>new<br>safety<br>and<br>performance risks.                                                                                                                                                        |
| WIFI                                                              | M1, M2 does not<br>contain WIFI<br>connection.<br>M1-W, M2-W contain<br>WIFI connection                                                                                                                                                                                                                                                  | WIFI connection.                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Bluetooth                                                         | Connect to the printer<br>using Bluetooth to<br>print a test image.                                                                                                                                                                                                                                                                      | Connect to the printer<br>using Bluetooth to print<br>a test image.                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Safety and EMC<br>Standards<br>compliance                         | ES60601-<br>1:2005+A1:2012<br>IEC 60601-1-2:2014<br>IEC 60601-2-37:2015                                                                                                                                                                                                                                                                  | ES60601-<br>1:2005+A1:2012<br>IEC 60601-1-2:2014<br>IEC 60601-2-37:2015                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| FDA limit                                                         | Track 1                                                                                                                                                                                                                                                                                                                                  | Track 1                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |

{6}------------------------------------------------

## 007_510(k) Summary

{7}------------------------------------------------

## 007_510(k) Summary

{8}------------------------------------------------

## 007 510(k) Summary

All the differences don't affect the safety and effectiveness which is concluded after all the required testing, so no safety and effectiveness issues relating to the system come into conclusion.

## 7. Performance Data

#### Clinical test:

Clinical testing is not required.

#### Non-clinical data

The M3,M3-HD,M4,M4-HD Bladder Scanner comply with:

Safety:

- 1. ES60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)
- 2. IEC 60601-1-2 Medical electrical equipment -- Part 1-2: General requirements for basic safety and essential performance -- Collateral Standard: Electromagnetic disturbances -- Requirements and tests

Performance:

- 3. IEC 60601-2-37 Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment
- 4. NEMA UD 2 Standard for real-time display of thermal and mechanical acoustic output indies on diagnostic ultrasound equipment.
- 5. Acoustic output testing as per the quideline "Information for Manufacturers Seeking Marketing

Clearance of Diagnostic Ultrasound Systems and Transducers" dated September 9, 2008.

WIFI and Bluetooth connection:

6. FCC CFR TITLE 47 PART 15 SUBPART C SECTION 15.247

The tests were selected to show substantial equivalence between the subject device and the predicate.

### 8. Conclusion

The M3,M3-HD,M4,M4-HD Bladder Scanner was evaluated with safety, EMC and Acoustic Output. The conclusions drawn from testing of the M3,M3-HD,M4,M4-HD Bladder Scanner demonstrate that the device is as safe and effective as the legally marketed predicate devices(K190769).

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K201316](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K201316)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
