K180084 · Konica Minolta, Inc. · IYO · Feb 8, 2018 · Radiology
Device Facts
Record ID
K180084
Device Name
Ultrasound System SONIMAGE MX1
Applicant
Konica Minolta, Inc.
Product Code
IYO · Radiology
Decision Date
Feb 8, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
The Ultrasound System SONIMAGE MX1 and its transducers are products designed to collect ultrasonic image data of the human body for diagnostic purposes. The system employs the ultrasonic pulse-echo method to visualize the anatomic structures, characteristics, and dynamics of the human body, and using an image display, Doppler display and/or Doppler sound, offers a procedure applied to the human body for medical diagnosis or examination. The range of intended clinical applications is same as other conventional ultrasound imaging systems for general purpose, such as small parts, abdomen, musculoskeletal, and peripheral vascular.
Device Story
Portable ultrasound system; collects ultrasonic image data via pulse-echo method; visualizes anatomic structures/dynamics. Operates in B-mode, Color Doppler, and PW Doppler. Used by qualified medical personnel in clinical settings. System includes transducers, cradle, battery, and foot switch. Connects via wired Ethernet or wireless LAN (USB-WiFi). Output displayed on-screen for medical diagnosis. Benefits include non-invasive visualization of internal geometry and fluid flow.
Clinical Evidence
Bench testing only. No clinical studies required for this conventional ultrasound diagnostic equipment. Verification activities confirmed geometric accuracy in B-mode and accuracy of position/velocity in Color/PW Doppler modes.
Technological Characteristics
Portable ultrasound system; B-mode, Color Doppler, PW Doppler. Transducers contact body surface; materials identical to predicate. Connectivity: Ethernet, wireless LAN (WPA/WPA2/WEP). Standards: AAMI/ANSI ES 60601-1, IEC 60601-1-6, IEC 60601-1-2, IEC 60601-2-37, IEC 62304, NEMA UD 2, NEMA UD 3. Real-time thermal/mechanical output indices (Track 3).
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis of the human body, including small parts (thyroid, breast), abdomen, musculoskeletal, and peripheral vascular applications. Contraindicated for ophthalmological exams, exams where waves pass through eyeballs, and direct contact with central nerves or cardiac systems.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
K162065 — Ultrasound System SONIMAGE HS1 · Konica Minolta, Inc. · Oct 19, 2016
K151060 — Ultrasound System SONIMAGE HS1 · Konica Minolta, Inc. · Jun 5, 2015
K152577 — ULTRASOUND SYSTEM SONIMAGE HS 1 · Konica Minolta, Inc. · Oct 8, 2015
K171579 — M6/M6T/M6 EXP/M6S/M6 PRO/M5 EXP/M55/M58 Diagnostic Ultrasound System · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Oct 18, 2017
K100598 — PENRITH ELETTRA ULTRASOUND SYSTEM · Penrith Corporation · Nov 19, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration."
February 8, 2018
Konica Minolta, Inc. % Russell D. Munves US Agent STORCH AMINI PC 140 East 45th Street. 25th Floor NEW YORK NY 10017
Re: K180084
Trade/Device Name: Ultrasound System SONIMAGE MX1 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: January 10, 2018 Received: January 11, 2018
Dear Mr. Munves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K180084
Device Name Ultrasound System SONIMAGE MX1
Indications for Use (Describe)
The Ultrasound System SONIMAGE MX1 and its transducers are products designed to collect ultrasonic image data of the human body for diagnostic purposes. The system employs the ultrasonic pulse-echo method to visualize the anatomic structures, characteristics, and dynamics of the human body, and using an image display, Doppler display and/or Doppler sound, offers a procedure applied to the human body for medical diagnosis or examination. The range of intended clinical applications is same as other conventional ultrasound imaging systems for general purpose, such as small parts, abdomen, musculoskeletal, and peripheral vascular.
**Type of Use (Select one or both, as applicable)**
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
#### Ultrasound System SONIMAGE MX1 System:
Transducer:
| Clinical Application | | | Mode of Operation | | | | | | |
|-----------------------------|---------------------------------|---|-------------------|-----|-----|------------------|-----------------------|---------------------|--|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetal | | | | | | | | |
| | Abdominal | N | | N | | N | Note1 | | |
| | Intra-operative (Specify) | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | | | | | | | | |
| | Small Organ (Note2) | N | | N | | N | Note1 | | |
| | Neonatal Cephalic | | | | | | | | |
| | Adult Cephalic | | | | | | | | |
| | Trans-rectal | | | | | | | | |
| | Trans-vaginal | | | | | | | | |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | N | | N | | N | Note1 | | |
| | Musculo-skeletal (Superficial) | N | | N | | N | Note1 | | |
| | Intravascular | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | |
| | Intra-cardiac | | | | | | | | |
| | Other (Specify) | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | | N | | N | Note1 | | |
| | Other (Specify) | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Combined mode BD (B + Pulse Doppler) and BcD (B Color Doppler + Pulse Doppler) Note 2: General small organ includes thyroid and breast
{4}------------------------------------------------
### Ultrasound System SONIMAGE MX1 System: Transducer: C5-2
| Clinical Application | | Mode of Operation | | | | | | |
|-----------------------------|---------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | N | | N | | N | Note1 | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Note2) | N | | N | | N | Note1 | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | | | | | | | |
| | Other (Specify) | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Combined mode BD (B + Pulse Doppler) and BcD (B Color Doppler + Pulse Doppler) Note 2: General small organ
{5}------------------------------------------------
### Ultrasound System SONIMAGE MX1 System: Transducer: L14-4
| Clinical Application | | | | Mode of Operation | | | | | | |
|-----------------------------|---------------------------------|---|---|-------------------|-----|------------------|-----------------------|---------------------|--|--|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) | | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetal | | | | | | | | | |
| | Abdominal | | | | | | | | | |
| | Intra-operative (Specify) | | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | | | | | | | | | |
| | Small Organ (Note2) | N | | N | | N | Note1 | | | |
| | Neonatal Cephalic | | | | | | | | | |
| | Adult Cephalic | | | | | | | | | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | | |
| | Musculo-skeletal (Conventional) | N | | N | | N | Note1 | | | |
| | Musculo-skeletal (Superficial) | N | | N | | N | Note1 | | | |
| | Intravascular | | | | | | | | | |
| | Other (Specify) | | | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | | |
| | Intra-cardiac | | | | | | | | | |
| | Other (Specify) | | | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | | N | | N | Note1 | | | |
| | Other (Specify) | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Combined mode BD (B + Pulse Doppler) and BcD (B Color Doppler + Pulse Doppler) Note 2: General small organ includes thyroid and breast
{6}------------------------------------------------
### Ultrasound System SONIMAGE MX1 System: Transducer: L11-3
| Clinical Application | Mode of Operation | | | | | | | |
|-----------------------------|---------------------------------|---|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Note2) | N | | N | | N | Note1 | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | N | | N | | N | Note1 | |
| | Musculo-skeletal (Superficial) | N | | N | | N | Note1 | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | | N | | N | Note1 | |
| | Other (Specify) | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Combined mode BD (B + Pulse Doppler) and BcD (B Color Doppler + Pulse Doppler) Note 2: General small organ includes thyroid and breast
{7}------------------------------------------------
### Ultrasound System SONIMAGE MX1 System: Transducer: MC10-3
| Clinical Application | Mode of Operation | | | | | | | |
|-----------------------------|---------------------------------|---|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic | Ophthalmic | | | | | | | |
| Fetal<br>Imaging<br>& Other | Fetal | | | | | | | |
| | Abdominal | | | | | | | |
| | Intra-operative (Specify) | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | |
| | Laparoscopic | | | | | | | |
| | Pediatric | | | | | | | |
| | Small Organ (Note2) | N | | N | | N | Note1 | |
| | Neonatal Cephalic | | | | | | | |
| | Adult Cephalic | | | | | | | |
| | Trans-rectal | | | | | | | |
| | Trans-vaginal | | | | | | | |
| | Trans-urethral | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | |
| | Musculo-skeletal (Conventional) | N | | N | | N | Note1 | |
| | Musculo-skeletal (Superficial) | N | | N | | N | Note1 | |
| | Intravascular | | | | | | | |
| | Other (Specify) | | | | | | | |
| Cardiac | Cardiac Adult | | | | | | | |
| | Cardiac Pediatric | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | |
| | Intra-cardiac | | | | | | | |
| | Other (Specify) | | | | | | | |
| Peripheral<br>Vessel | Peripheral vessel | N | | N | | N | Note1 | |
| | Other (Specify) | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Combined mode BD (B + Pulse Doppler) and BcD (B Color Doppler + Pulse Doppler) Note 2: General small organ
{8}------------------------------------------------
# 510(k) Summary
| Submitter's Name: | KONICA MINOLTA, INC. |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 1 Sakura-machi,<br>Hino-shi, 191-8511 Japan |
| Contact: | Tsutomu Fukui |
| Telephone: | +81 42 589 8429 |
| Date: | January 10, 2018 |
| Trade Name: | Ultrasound System SONIMAGE MX1 |
| Model No: | SONIMAGE MX1 |
| Common Name: | Ultrasound Imaging System |
| Classification Name(s): | Ultrasonic Pulsed Doppler Imaging System (21 CFR 892.1550)<br>Ultrasonic Pulsed Echo Imaging System (21 CFR 892.1560)<br>Diagnostic Ultrasound Transducer (21 CFR 892.1570) |
| Classification Number(s): | 90-IYN; 90-IYO; 90-ITX |
| Regulatory Class: | Class II |
| Predicate Device(s): | K162065 - Ultrasound System SONIMAGE HS1, Konica Minolta, Inc |
Device Description:
The Ultrasound System SONIMAGE MX1 is a portable ultrasound system for general purposes. The system provides ultrasound imaging information such as used for the purpose of diagnosing the human body, which visually represents the internal geometry, characteristics and dynamics of the human body, and transmits / receives ultrasound waves to obtain image data of the visual representation.
This system provides ultrasound images in conventional modes of Bmode, Color Doppler-mode, and PW Doppler-mode.
{9}------------------------------------------------
The optional items are available, such as a Cradle, an Additional Battery, a Three port probe unit, and a Foot Switch with dual/triple pedals.
The system can be connected to LAN through the wired Ethernet and, is also capable of wireless LAN with the OTS USB-WiFi adapter supporting security of WPA/WPA2 and WEP.
This system conforms to Real Time Display of Thermal and Mechanical Output Indices on Diagnostic Ultrasound Equipment (Track 3). Transducers have their own characteristic applications, and are brought into contact with the body surface.
The Ultrasound System SONIMAGE MX1 is designed to comply with the following standards:
AAMI / ANSI ES 60601-1: 2005 and A1: 2012, C1:2009 and A2:2010; IEC 60601-1-6: 2010; IEC 60601-1-2: 2007; IEC 60601-2-37: 2007; IEC 62304: 2006; NEMA UD 2-2004; NEMA UD 3-2004;
# Intended Use:
The Ultrasound System SONIMAGE MX1 and its transducers are products designed to collect ultrasonic image data of the human body for diagnostic purposes.
The system employs the ultrasonic pulse-echo method to visualize the anatomic structures, characteristics, and dynamics of the human body, and using an image display, Doppler display and/or Doppler sound, offers a procedure applied to the human body for medical diagnosis or examination.
The range of intended clinical applications is same as the predicate device, such as small parts, abdomen, musculoskeletal, and peripheral vascular.
This system is contraindicated for ophthalmological exams or exams in which ultrasonic waves may pass through the eyeballs.
This system is NOT designed for use in direct contact to central nerves and cardiac systems.
{10}------------------------------------------------
The system should not be used by persons other than fully qualified and certified medical personnel.
# Summary of Technological Characteristics Compared to Predicate Device:
The proposed Ultrasound System SONIMAGE MX1 employs the same fundamental scientific technologies as the primary predicate device (K162065). The summary of comparisons of technological characteristics for the proposed device and the predicate systems are provided as follows;
# Intended Use
The intended use of the proposed device is essentially the same as the conventional ultrasound diagnostic system for general purpose, the same as the indication for use of the predicate device (K162065).
# Operating Principle and designing
The proposed and predicate systems transmit ultrasonic energy into patients by using same transducers, then perform post processing of received echoes to generate on-screen display of anatomic structures and dynamics of the human body. The form of the system is different but both systems are designed to be transportable with battery operation. These operating principle and designing are the same as the predicate device.
The proposed system doesn't support a conventional M mode but supports the same operating modes and display format as those supported by the predicate system. The both systems support same fundamental measurement functions for anatomic structures, internal geometry, characteristics and dynamics of the human body.
# Non-clinical test
The geometric accuracy verification in B-mode and the accuracy verification of the position and the velocity for Color Doppler-mode and PW Doppler-mode, and the time for PW Doppler-mode are comprehensively tested as Bench-tests with successful results.
The proposed system and the predicate systems do not have a quantitative claims of sensitivity regarding Color Doppler-mode and PW Doppler-mode.
{11}------------------------------------------------
All of the verification activities, including as required by the risk analysis, for the proposed system was performed and the results demonstrated that the predetermined acceptance criteria were met.
### Safety
The system is in conformance with the standards described above, which are same standards as those of the primary predicate device (K162065). The both systems conform to real time display of thermal and mechanical output indices under Track 3.
# Biocompatibility
No transducer is newly introduced in this application. The all patient contact materials of human body surface are identical to those of the predicate, and then both systems achieve same acceptance level for biocompatibility.
# Conclusion:
The clinical studies are not required to support substantial equivalence for these conventional ultrasound diagnostic equipment. In addition to that, as discussed in the above technological comparison, the technological characteristics of the Ultrasound System SONIMAGE MX1 are deemed to be substantially equivalent to the aforementioned predicate devices that have already been cleared for USA distribution with 510(k) premarket notification.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.