← Product Code [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO) · K171766

# Acuson SC2000 Diagnostic Ultrasound System (K171766)

_Siemens Medi Cal Solutions, Inc. · IYO · Jul 12, 2017 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K171766

## Device Facts

- **Applicant:** Siemens Medi Cal Solutions, Inc.
- **Product Code:** [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO.md)
- **Decision Date:** Jul 12, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Indications for Use

The SC2000 ultrasound imaging system is intended for the following applications: Cardiac. Neo-natal and Fetal Cardiac. Pediatric, Transesophageal, Adult Cephalic, Peripheral Vessel, Abdominal, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are: Cardiac Imaging Applications and Analysis The system transmits ultrasound energy into adult, pediativ, neonatal, and fetal cardiac patients creating 2D (B). 3D. M-Mode (M), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. The system also supports catheters which are intended for intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The system has Cardiac Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. Vascular Imaging Applications and Analysis The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the carotid arteries or juggler veins in the neck; superficial and arteries in the arms and legs and abdomen; and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. The system has Vascular Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. Superficial Imaging Applications The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of conventional or superficial musculoskeletal structures and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. Intraoperative Imaging Applications The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), and Pulsed Wave Doppler (PWD) to obtain images and blood flow velocity that provide guidance during intraoperative procedures. Transcranial Imaging Applications The system transmits ultrasound energy into the cranium of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the brain and surrounding anatomical structures to evaluate the presence of pathology. The system provides Measurement Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

## Device Story

Mobile, software-controlled diagnostic ultrasound system; acquires primary/secondary harmonic ultrasound echo data. Inputs: ultrasound signals via various transducers (matrix, phased, linear, curved arrays) and optional ECG for synchronization. Transforms inputs into 2D (B), 3D, 4D (real-time 3D), M-Mode, Pulsed/Continuous Wave Doppler, Color Doppler, Amplitude Doppler, and Color Velocity Imaging. Used in clinical settings by physicians for diagnostic imaging and intraoperative guidance. Output: visual display of anatomical structures and blood flow velocity; measurement/calculation packages provide adjunctive data for clinical diagnosis. Benefits: non-invasive visualization of pathology in heart, vessels, brain, and musculoskeletal structures; supports intraluminal visualization via compatible catheters.

## Clinical Evidence

No clinical data required; substantial equivalence established via bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal/electromagnetic/mechanical safety testing.

## Technological Characteristics

Mobile ultrasound system; supports matrix, phased, linear, and curved array transducers. Modes: B, M, PWD, CWD, Color Doppler, Amplitude Doppler, Color Velocity, 3D/4D. Connectivity: Wireless, DICOM. Standards: AAMI/ANSI 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62366, IEC 62304, NEMA UD-2, NEMA UD-3, ISO 10993-1.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- ACUSON SC2000 Diagnostic Ultrasound System ([K170315](/device/K170315.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of three human profiles facing right, with flowing lines suggesting movement or progress. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 12, 2017

SIEMENS MEDICAL SOLUTIONS INC. 0 Mr. MARK JOB RESPONSIBLE THIRD PARTY OFFICIAL REGULATORY TECHNOLOGY SERVICES, LLC 1394 25TH STREET, NW BUFFALO MN 55313

Re: K171766

Trade/Device Name: Acuson SC2000 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYO, IYN, ITX, OBJ Dated: June 2, 2017 Received: June 14, 2017

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D. O'Hara For

Robert Ochs Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

K171766

#### Device Name

Acuson SC2000TM Diagnostic Ultrasound System

#### Indications for Use (Describe)

The SC2000 ultrasound imaging system is intended for the following applications: Cardiac. Neo-natal and Fetal Cardiac. Pediatric, Transesophageal, Adult Cephalic, Peripheral Vessel, Abdominal, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are:

Cardiac Imaging Applications and Analysis

The system transmits ultrasound energy into adult, pediativ, neonatal, and fetal cardiac patients creating 2D (B). 3D. M-Mode (M), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

The system also supports catheters which are intended for intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The system has Cardiac Measurements and Calculation Packages that provide information that may be used adjunctively

with other medical data obtained by a physician for clinical diagnosis purposes.

Vascular Imaging Applications and Analysis

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the carotid arteries or juggler veins in the neck; superficial and arteries in the arms and legs and abdomen; and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

The system has Vascular Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Superficial Imaging Applications

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of conventional or superficial musculoskeletal structures and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images.

Intraoperative Imaging Applications

The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), and Pulsed Wave Doppler (PWD) to obtain images and blood flow velocity that provide guidance during intraoperative procedures.

Transcranial Imaging Applications

The system transmits ultrasound energy into the cranium of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the brain and surrounding anatomical structures to evaluate the presence of pathology. The system provides Measurement Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

{3}------------------------------------------------

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Number (if known):

#### Device Name: SC2000 Diagnostic Ultrasound System

Ultrasound imaging or fluid flow analysis of the human body as follows: Intended Use:

| Clinical<br>Application          | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>3D | Other:<br>Real<br>Time<br>3D |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|--------------|------------------------------|
| Ophthalmic                       |   |   |   |     |     |                  |                                 |                              |                       |                               |              |                              |
| Fetal                            | P | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |              | P                            |
| Abdominal                        | P | P | P | P   | P   | P                | P                               |                              | P*                    | P                             | P            |                              |
| Intraoperative<br>Abdominal      | P | P | P | P   | P   |                  |                                 |                              | P*                    | P                             |              |                              |
| Intraoperative<br>Neurological   |   |   |   |     |     |                  |                                 |                              |                       |                               |              |                              |
| Pediatric                        | P | P | P | P   | P   | P                | P                               |                              | P*                    | P                             | P            | P                            |
| Small Organ<br>(specify)         |   |   |   |     |     |                  |                                 |                              |                       |                               |              |                              |
| Neonatal<br>Cephalic             |   |   |   |     |     |                  |                                 |                              |                       |                               |              |                              |
| Adult Cephalic                   | P | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |              |                              |
| Cardiac                          | P | P | P | P   | P   | P                | P                               |                              | P*                    | P                             | P            | P                            |
| Trans-esophageal                 | P | P | P | P   | P   |                  |                                 |                              | P*                    |                               | P            |                              |
| Transrectal                      |   |   |   |     |     |                  |                                 |                              |                       |                               |              |                              |
| Transvaginal                     |   |   |   |     |     |                  |                                 |                              |                       |                               |              |                              |
| Transurethral                    |   |   |   |     |     |                  |                                 |                              |                       |                               |              |                              |
| Intra-Luminal                    | P | P | P | P   | P   | P                | P                               |                              | P*                    |                               |              | P                            |
| Peripheral Vessel                | P | P | P | P   | P   | P                | P                               | P                            | P*                    | P                             |              |                              |
| Laparoscopic                     |   |   |   |     |     |                  |                                 |                              |                       |                               |              |                              |
| Musculo-skeletal<br>Conventional | P | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |              |                              |
| Musculo-skeletal<br>Superficial  | P | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |              |                              |
| Other (Neonatal<br>Cardiac)      | P | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |              |                              |
| Other (Intra-<br>Cardiac)        | P | P | P | P   | P   | P                | P                               |                              | P*                    |                               |              | P                            |
| Other<br>(Abdominal<br>Vascular) | P | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |              |                              |

N=new indication. P = Previously Cleared in 510(k) K170315

#### Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{5}------------------------------------------------

510(k) Number (if known):

CW2 Device Name:

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical<br>Application          | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude) | Color<br>Velocity | Combined<br>(Specify) | Other:<br>Harmonic |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|-------------------|-----------------------|--------------------|
|                                  |   |   |   |     |     |                  | Doppler              | Imaging           |                       | Imaging            |
| Ophthalmic                       |   |   |   |     |     |                  |                      |                   |                       |                    |
| Fetal                            |   |   |   |     |     |                  |                      |                   |                       |                    |
| Abdominal                        |   |   |   |     |     |                  |                      |                   |                       |                    |
| Intraoperative<br>Abdominal      |   |   |   |     |     |                  |                      |                   |                       |                    |
| Intraoperative<br>Neurological   |   |   |   |     |     |                  |                      |                   |                       |                    |
| Pediatric                        |   |   |   |     | P   |                  |                      |                   |                       |                    |
| Small Organ<br>(specify)         |   |   |   |     |     |                  |                      |                   |                       |                    |
| Neonatal<br>Cephalic             |   |   |   |     |     |                  |                      |                   |                       |                    |
| Adult Cephalic                   |   |   |   |     |     |                  |                      |                   |                       |                    |
| Cardiac                          |   |   |   |     | P   |                  |                      |                   |                       |                    |
| Trans-esophageal                 |   |   |   |     |     |                  |                      |                   |                       |                    |
| Transrectal                      |   |   |   |     |     |                  |                      |                   |                       |                    |
| Transvaginal                     |   |   |   |     |     |                  |                      |                   |                       |                    |
| Transurethral                    |   |   |   |     |     |                  |                      |                   |                       |                    |
| Intra-Luminal                    |   |   |   |     |     |                  |                      |                   |                       |                    |
| Peripheral Vessel                |   |   |   |     | P   |                  |                      |                   |                       |                    |
| Laparoscopic                     |   |   |   |     |     |                  |                      |                   |                       |                    |
| Musculo-skeletal<br>Conventional |   |   |   |     |     |                  |                      |                   |                       |                    |
| Musculo-skeletal<br>Superficial  |   |   |   |     |     |                  |                      |                   |                       |                    |
| Other (specify)                  |   |   |   |     |     |                  |                      |                   |                       |                    |

N=new indication. Previously Cleared in 510(k) K170315

#### Additional Comments:

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{6}------------------------------------------------

### 510(k) Number (if known):

Device Name:

4Z1c

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical<br>Application          | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------|
| Ophthalmic                       |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Fetal                            |   | P | P | P   | P   | P                |                                 |                              | P*                    | P                             | P                            |
| Abdominal                        |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Intraoperative<br>Abdominal      |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Intraoperative<br>Neurological   |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Pediatric                        |   | P | P | P   | P   | P                |                                 |                              | P*                    | P                             | P                            |
| Small Organ<br>(specify) **      |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Neonatal<br>Cephalic             |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Adult Cephalic                   |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Cardiac                          |   | P | P | P   | P   | P                |                                 |                              | P*                    | P                             | P                            |
| Trans-esophageal                 |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Transrectal                      |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Transvaginal                     |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Transurethral                    |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Intra-Luminal                    |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Peripheral Vessel                |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Laparoscopic                     |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Musculo-skeletal<br>Conventional |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Musculo-skeletal<br>Superficial  |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Other (specify)                  |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |

N=new indication. P = Previously Cleared in 510(k) K170315

#### Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{7}------------------------------------------------

510(k) Number (if known):

Device Name:

4V1c

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical<br>Application          | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|
| Ophthalmic                       |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Fetal                            |   | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |
| Abdominal                        |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Intraoperative<br>Abdominal      |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Intraoperative<br>Neurological   |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Pediatric                        |   | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |
| Small Organ<br>(specify)         |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Neonatal<br>Cephalic             |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Adult Cephalic                   |   | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |
| Cardiac                          |   | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |
| Trans-esophageal                 |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Transrectal                      |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Transvaginal                     |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Transurethral                    |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Intra-Luminal                    |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Peripheral Vessel                |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Laparoscopic                     |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Musculo-skeletal<br>Conventional |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Musculo-skeletal<br>Superficial  |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Other (Neonatal<br>Cardiac)      |   | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |

N=new indication. Previously Cleared in 510(k) K170315

#### Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{8}------------------------------------------------

### 510(k) Number (if known):

Device Name:

V5Ms

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical<br>Application          | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude) | Color<br>Velocity | Combined<br>(Specify) | Other:<br>Harmonic | Other:<br>3D |
|----------------------------------|---|---|---|-----|-----|------------------|----------------------|-------------------|-----------------------|--------------------|--------------|
|                                  |   |   |   |     |     |                  | Doppler              | Imaging           |                       | Imaging            |              |
| Ophthalmic                       |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Fetal                            |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Abdominal                        |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Intraoperative<br>Abdominal      |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Intraoperative<br>Neurological   |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Pediatric                        |   | P | P | P   | P   | P                |                      |                   | P*                    |                    | P            |
| Small Organ<br>(specify)         |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Neonatal                         |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Cephalic                         |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Adult Cephalic                   |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Cardiac                          |   | P | P | P   | P   | P                |                      |                   | P*                    |                    | P            |
| Trans-esophageal                 |   | P | P | P   | P   | P                |                      |                   | P*                    |                    | P            |
| Transrectal                      |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Transvaginal                     |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Transurethral                    |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Intra-Luminal                    |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Peripheral Vessel                |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Laparoscopic                     |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Musculo-skeletal<br>Conventional |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Musculo-skeletal<br>Superficial  |   |   |   |     |     |                  |                      |                   |                       |                    |              |
| Other (specify)                  |   |   |   |     |     |                  |                      |                   |                       |                    |              |

N=new indication. P = Previously Cleared in 510(k) K170315

#### Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{9}------------------------------------------------

### 510(k) Number (if known):

Device Name:

### ୧୯.IHD

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical<br>Application          | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|
| Ophthalmic                       |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Fetal                            |   | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |
| Abdominal                        |   | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |
| Intraoperative<br>Abdominal      |   | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |
| Intraoperative                   |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Pediatric                        |   | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |
| Small Organ<br>(specify)         |   | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |
| Neonatal<br>Cephalic             |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Adult Cephalic                   |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Cardiac                          |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Trans-esophageal                 |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Transrectal                      |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Transvaginal                     |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Transurethral                    |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Intra-Luminal                    |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Peripheral Vessel                |   | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |
| Laparoscopic                     |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Musculo-skeletal<br>Conventional |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Musculo-skeletal<br>Superficial  |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Other<br>(Abdominal<br>Vascular) |   | P | P | P   | P   | P                | P                               |                              | P*                    | P                             |

N=new indication. P = Previously Cleared in 510(k) K170315

#### Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{10}------------------------------------------------

### 510(k) Number (if known):

Device Name:

#### Z6Ms

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical<br>Application          | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------|
| Ophthalmic                       |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Fetal                            |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Abdominal                        |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Intraoperative<br>Abdominal      |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Intraoperative<br>Neurological   |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Pediatric                        |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Small Organ<br>(specify)         |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Neonatal<br>Cephalic             |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Adult Cephalic                   |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Cardiac                          | P | P | P | P   | P   | P                |                                 | P                            | P*                    | P                             | P                            |
| Trans-esophageal                 | P | P | P | P   | P   | P                |                                 | P                            | P*                    | P                             | P                            |
| Transrectal                      |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Transvaginal                     |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Transurethral                    |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Intra-Luminal                    |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Peripheral Vessel                |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Laparoscopic                     |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Musculo-skeletal<br>Conventional |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Musculo-skeletal<br>Superficial  |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Other (specify)                  |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |

N=new indication. P = Previously Cleared in 510(k) K170315

#### Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{11}------------------------------------------------

### 510(k) Number (if known):

Device Name:

V7M

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical<br>Application          | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>3D |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|--------------|
| Ophthalmic                       |   |   |   |     |     |                  |                                 |                              |                       |                               |              |
| Fetal                            |   |   |   |     |     |                  |                                 |                              |                       |                               |              |
| Abdominal                        | P | P | P | P   | P   | P                | P                               |                              | P*                    | P                             | P            |
| Intraoperative<br>Abdominal      |   |   |   |     |     |                  |                                 |                              |                       |                               |              |
| Intraoperative<br>Neurological   |   |   |   |     |     |                  |                                 |                              |                       |                               |              |
| Pediatric                        | P | P | P | P   | P   | P                |                                 |                              | P*                    |                               | P            |
| Small Organ<br>(specify)         |   |   |   |     |     |                  |                                 |                              |                       |                               |              |
| Neonatal<br>Cephalic             |   |   |   |     |     |                  |                                 |                              |                       |                               |              |
| Adult Cephalic                   |   |   |   |     |     |                  |                                 |                              |                       |                               |              |
| Cardiac                          | P | P | P | P   | P   | P                |                                 |                              | P*                    |                               | P            |
| Trans-esophageal                 | P | P | P | P   | P   | P                |                                 |                              | P*                    |                               | P            |
| Transrectal                      |   |   |   |     |     |                  |                                 |                              |                       |                               |              |
| Transvaginal                     |   |   |   |     |     |                  |                                 |                              |                       |                               |              |
| Transurethral                    |   |   |   |     |     |                  |                                 |                              |                       |                               |              |
| Intra-Luminal                    |   |   |   |     |     |                  |                                 |                              |                       |                               |              |
| Peripheral Vessel                |   |   |   |     |     |                  |                                 |                              |                       |                               |              |
| Laparoscopic                     |   |   |   |     |     |                  |                                 |                              |                       |                               |              |
| Musculo-skeletal<br>Conventional |   |   |   |     |     |                  |                                 |                              |                       |                               |              |
| Musculo-skeletal<br>Superficial  |   |   |   |     |     |                  |                                 |                              |                       |                               |              |
| Other (specify)                  |   |   |   |     |     |                  |                                 |                              |                       |                               |              |

N=new indication. P = Previously Cleared in 510(k) K170315

#### Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{12}------------------------------------------------

### 510(k) Number (if known):

Device Name:

8V3

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical<br>Application          | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|
| Ophthalmic                       |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Fetal                            | P | P | P | P   | P   | P                | P                               |                              | P *                   | P                             |
| Abdominal                        |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Intraoperative<br>Abdominal      |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Intraoperative<br>Neurological   |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Pediatric                        | P | P | P | P   | P   | P                | P                               |                              | P *                   | P                             |
| Small Organ<br>(specify)         |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Neonatal<br>Cephalic             |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Adult Cephalic                   |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Cardiac                          | P | P | P | P   | P   | P                | P                               |                              | P *                   | P                             |
| Trans-esophageal                 |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Transrectal                      |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Transvaginal                     |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Transurethral                    |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Intra-Luminal                    |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Peripheral Vessel                |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Laparoscopic                     |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Musculo-skeletal<br>Conventional |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Musculo-skeletal<br>Superficial  |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Other (Neonatal<br>Cardiac)      | P | P | P | P   | P   | P                | P                               |                              | P *                   | P                             |

N=new indication. Previously Cleared in 510(k) K170315

#### Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{13}------------------------------------------------

510(k) Number (if known):

Device Name:

9L4

Indications for Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical<br>Application          | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|
| Ophthalmic                       |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Fetal                            |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Abdominal                        |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Intraoperative<br>Abdominal      |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Intraoperative<br>Neurological   |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Pediatric                        |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Small Organ<br>(specify)         |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Neonatal<br>Cephalic             |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Adult Cephalic                   |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Cardiac                          |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Trans-esophageal                 |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Transrectal                      |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Transvaginal                     |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Transurethral                    |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Intra-Luminal                    |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Peripheral Vessel                | P | P | P | P   |     | P                | P                               | P                            | P*                    | P                             |
| Laparoscopic                     |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Musculo-skeletal<br>Conventional | P | P | P | P   |     | P                | P                               | P                            | P*                    | P                             |
| Musculo-skeletal<br>Superficial  | P | P | P | P   |     | P                | P                               | P                            | P*                    | P                             |
| Other (specify)                  |   |   |   |     |     |                  |                                 |                              |                       |                               |

N=new indication. P = Previously Cleared in 510(k) K170315

#### Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{14}------------------------------------------------

### 510(k) Number (if known):

Device Name:

10V4

Intended Use:

Ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical<br>Application          | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|
| Ophthalmic                       |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Fetal                            |   | P | P | P   | P   | P                | P                               |                              | P *                   | P                             |
| Abdominal                        |   | P | P | P   | P   | P                | P                               |                              | P *                   | P                             |
| Intraoperative<br>Abdominal      |   | P | P | P   | P   | P                | P                               |                              | P *                   | P                             |
| Intraoperative<br>Neurological   |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Pediatric                        |   | P | P | P   | P   | P                | P                               |                              | P *                   | P                             |
| Small Organ<br>(specify)         |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Neonatal<br>Cephalic             |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Adult Cephalic                   |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Cardiac                          |   | P | P | P   | P   | P                | P                               |                              | P *                   | P                             |
| Trans-esophageal                 |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Transrectal                      |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Transvaginal                     |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Transurethral                    |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Intra-Luminal                    |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Peripheral Vessel                |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Laparoscopic                     |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Musculo-skeletal<br>Conventional |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Musculo-skeletal<br>Superficial  |   |   |   |     |     |                  |                                 |                              |                       |                               |
| Other (Neonatal<br>Cardiac)      |   | P | P | P   | P   | P                | P                               |                              | P *                   | P                             |

N=new indication. Previously Cleared in 510(k) K170315

#### Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{15}------------------------------------------------

510(k) Number (if known):

Device Name:

### AcuNav 8F and 10F Ultrasound Catheter

Intended Use:

Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients as follows:

| Clinical<br>Application          | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------|
| Ophthalmic                       |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Fetal                            |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Abdominal                        |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Intraoperative<br>Abdominal      |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Intraoperative<br>Neurological   |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Pediatric                        | P | P | P | P   | P   | P                | P                               |                              | P*                    |                               |                              |
| Small Organ<br>(specify) **      |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Neonatal<br>Cephalic             |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Adult Cephalic                   |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Cardiac                          | P | P | P | P   | P   | P                | P                               |                              | P*                    |                               |                              |
| Trans-esophageal                 |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Transrectal                      |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Transvaginal                     |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Transurethral                    |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Intra-Luminal                    | P | P | P | P   | P   | P                | P                               |                              | P*                    |                               |                              |
| Peripheral Vessel                |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Laparoscopic                     |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Musculo-skeletal<br>Conventional |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Musculo-skeletal<br>Superficial  |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Other (Intra-<br>Cardiac)        | P | P | P | P   | P   | P                | P                               |                              | P*                    |                               |                              |

N=new indication. P = Previously Cleared in 510(k) K170315

#### Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{16}------------------------------------------------

510(k) Number (if known):

Device Name:

Intended Use:

ACUSON AcuNav™ V 10F Ultrasound Catheter

Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients as follows:

| Clinical<br>Application          | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------|
| Ophthalmic                       |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Fetal                            |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Abdominal                        |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Intraoperative<br>Abdominal      |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Intraoperative<br>Neurological   |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Pediatric                        | P | P | P | P   | P   | P                | P                               |                              | P*                    |                               | P                            |
| Small Organ<br>(specify) **      |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Neonatal<br>Cephalic             |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Adult Cephalic                   |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Cardiac                          | P | P | P | P   | P   | P                | P                               |                              | P*                    |                               | P                            |
| Trans-esophageal                 |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Transrectal                      |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Transvaginal                     |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Transurethral                    |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Intra-Luminal                    | P | P | P | P   | P   | P                | P                               |                              | P*                    |                               | P                            |
| Peripheral Vessel                |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Laparoscopic                     |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Musculo-skeletal<br>Conventional |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Musculo-skeletal<br>Superficial  |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |
| Other (Intra-<br>Cardiac)        | P | P | P | P   | P   | P                | P                               |                              | P*                    |                               | P                            |

N=new indication. P = Previously Cleared in 510(k) K170315

#### Additional Comments:

*Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON A SEPARATE PAGE IF NEEDED)

{17}------------------------------------------------

510(k) Number (if known):

Device Name:

SoundStar 10F Ultrasound Catheter

Intended Use:

Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart:

| Clinical<br>Application          | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D |  |
|----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------|--|
| Ophthalmic                       |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |  |
| Fetal                            |   |   |   |     |     |                  |                                 |                              |                       |                               |                              |  |
| Abdominal                        |   |   |   |     |…

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K171766](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K171766)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
