← Product Code [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO) · K150580

# UltraVision 2 Diagnostic Ultrasound System (K150580)

_Winprobe Corporation · IYO · Apr 7, 2015 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K150580

## Device Facts

- **Applicant:** Winprobe Corporation
- **Product Code:** [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO.md)
- **Decision Date:** Apr 7, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

This diagnostic ultrasound system (UltraVision) and its transducers are intended for use in clinical examinations to evaluate differences in the echogenicity of soft tissue of small parts in adult patients. (thyroid, breast, and testicles) and for superficial muscular skeletal diagnosis. The system is for prescription use only by a trained sonographer under the direction of a qualified physician or directly by a qualified physician.

## Device Story

UltraVision 2 is a portable diagnostic ultrasound system; utilizes hand-held linear transducers to transmit ultrasonic energy into patients; receives and processes echoes to generate B-mode images of anatomic structures on a 15-inch LCD display. Operated by trained sonographers or physicians in clinical settings. Provides visual output for evaluation of soft tissue echogenicity; assists clinicians in diagnostic decision-making for small parts and superficial musculoskeletal conditions. Portable form factor (4.5 kg) distinguishes it from larger, stationary predicate systems.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), diagnostic ultrasound safety (IEC 60601-2-37), acoustic output testing per FDA guidance, and biocompatibility (ISO 10993).

## Technological Characteristics

Portable ultrasound system; 15-inch LCD display; 128-element linear transducer (5-12 MHz). Operates in B-mode. Materials evaluated per ISO 10993. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37. Dimensions: 400mm x 50mm x 250mm; Weight: 4.5 kg.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Siemens Acuson S2000 Ultrasound System ([K081148](/device/K081148.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 7, 2015

Winprobe Corporation % Mr. Mark Job Responsible Third Party Official 1394 25th Street, NW BUFFALO MN 55313

Re: K150580

Trade/Device Name: Ultravision 2 Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, ITX Dated: March 6, 2015 Received: March 9, 2015

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration            |
| Indications for Use                     |

#### Indications for Use

510(k) Number (if known)

#### K150580

#### Device Name

UltraVision 2 Diagnostic Ultrasound System

#### Indications for Use (Describe)

This diagnostic ultrasound system (UltraVision) and its transducers are intended for use in clinical examinations to evaluate differences in the echogenicity of soft issue of small patients, the in clincal examinations to
superticial museular skelets dissue of small parts in adult patients, superficial muscular skeletal diagnosis.

The system is for prescription use only by a trained sonographer under the direction of a qualified physician or directly by
a qualified physician a qualified physician.

Type of Use (Select one or both, as applicable)

[X] Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017

See PRA Statement below.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this hurden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (1/14)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

# Indications for Use Table (system)

| Clinical Application      |                                 | Mode of Operation |   |     |     |                  |                       |  |
|---------------------------|---------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--|
| General<br>(Track 1 only) | Specific<br>(Tracks 1 & 3)      | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(specify) |  |
| Ophthalmic                | Ophthalmic                      |                   |   |     |     |                  |                       |  |
| Fetal Imaging<br>& Other  | Fetal                           |                   |   |     |     |                  |                       |  |
|                           | Abdominal                       |                   |   |     |     |                  |                       |  |
|                           | Intra-operative (specify)       |                   |   |     |     |                  |                       |  |
|                           | Intra-operative (neuro)         |                   |   |     |     |                  |                       |  |
|                           | Laparoscopic                    |                   |   |     |     |                  |                       |  |
|                           | Pediatric                       |                   |   |     |     |                  |                       |  |
|                           | Small Organ (specify)           | N                 |   |     |     |                  |                       |  |
|                           | Breast, Thyroid, Testes         |                   |   |     |     |                  |                       |  |
|                           | Neonatal Cephalic               |                   |   |     |     |                  |                       |  |
|                           | Adult Cephalic                  |                   |   |     |     |                  |                       |  |
|                           | Trans-rectal                    |                   |   |     |     |                  |                       |  |
|                           | Trans-vaginal                   |                   |   |     |     |                  |                       |  |
|                           | Trans-urethral                  |                   |   |     |     |                  |                       |  |
|                           | Trans-esoph. (non-cardiac)      |                   |   |     |     |                  |                       |  |
|                           | Musculo-skeletal (conventional) |                   |   |     |     |                  |                       |  |
|                           | Musculo-skeletal (superficial)  | N                 |   |     |     |                  |                       |  |
|                           | Intravascular                   |                   |   |     |     |                  |                       |  |
|                           | Other (specify)                 |                   |   |     |     |                  |                       |  |
| Cardiac                   | Cardiac Adult                   |                   |   |     |     |                  |                       |  |
|                           | Cardiac Pediatric               |                   |   |     |     |                  |                       |  |
|                           | Intravascular (Cardiac)         |                   |   |     |     |                  |                       |  |
|                           | Trans-esoph. (Cardiac)          |                   |   |     |     |                  |                       |  |
|                           | Intra-cardiac                   |                   |   |     |     |                  |                       |  |
|                           | Other (specify)                 |                   |   |     |     |                  |                       |  |
| Peripheral<br>Vessel      | Peripheral vessel               |                   |   |     |     |                  |                       |  |
|                           | Other (specify)                 |                   |   |     |     |                  |                       |  |

System: UltraVision 2 Diagnostic Ultrasound System

N = new indication; P= previously cleared by FDA

{4}------------------------------------------------

# Indication for Use Table (transducer)

| Clinical Application      |                                 | Mode of Operation |   |     |     |                  |                       |                    |
|---------------------------|---------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------------------|
| General<br>(Track 1 only) | Specific<br>(Tracks 1 & 3)      | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic                | Ophthalmic                      |                   |   |     |     |                  |                       |                    |
| Fetal Imaging<br>& Other  | Fetal                           |                   |   |     |     |                  |                       |                    |
|                           | Abdominal                       |                   |   |     |     |                  |                       |                    |
|                           | Intra-operative (specify)       |                   |   |     |     |                  |                       |                    |
|                           | Intra-operative (neuro)         |                   |   |     |     |                  |                       |                    |
|                           | Laparoscopic                    |                   |   |     |     |                  |                       |                    |
|                           | Pediatric                       |                   |   |     |     |                  |                       |                    |
|                           | Small Organ (specify)           | N                 |   |     |     |                  |                       |                    |
|                           | Breast, Thyroid, Testes         |                   |   |     |     |                  |                       |                    |
|                           | Neonatal Cephalic               |                   |   |     |     |                  |                       |                    |
|                           | Adult Cephalic                  |                   |   |     |     |                  |                       |                    |
|                           | Trans-rectal                    |                   |   |     |     |                  |                       |                    |
|                           | Trans-vaginal                   |                   |   |     |     |                  |                       |                    |
|                           | Trans-urethral                  |                   |   |     |     |                  |                       |                    |
|                           | Trans-esoph. (non-cardiac)      |                   |   |     |     |                  |                       |                    |
|                           | Musculo-skeletal (conventional) |                   |   |     |     |                  |                       |                    |
|                           | Musculo-skeletal (superficial)  | N                 |   |     |     |                  |                       |                    |
|                           | Intravascular                   |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                 |                   |   |     |     |                  |                       |                    |
| Cardiac                   | Cardiac Adult                   |                   |   |     |     |                  |                       |                    |
|                           | Cardiac Pediatric               |                   |   |     |     |                  |                       |                    |
|                           | Intravascular (Cardiac)         |                   |   |     |     |                  |                       |                    |
|                           | Trans-esoph. (Cardiac)          |                   |   |     |     |                  |                       |                    |
|                           | Intra-cardiac                   |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                 |                   |   |     |     |                  |                       |                    |
| Peripheral<br>Vessel      | Peripheral vessel               |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                 |                   |   |     |     |                  |                       |                    |

Transducer: HL5-10EPN transducer for use with UltraVision 2

N = new indication; P= previously cleared by FDA

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a coat of arms with a white background and a black border. In the center of the coat of arms is a red griffin holding a spear. Above the griffin is a banner with the word "imagine" written on it.

# 510(k) Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.

The assigned 510(k) number is:

#### Submitter

WinProbe Corporation 11770 US Highway 1, suite 405E Palm Beach Gardens, Florida 33408 Telephone: (561)318-8367 Fax: (561)624-0914 Registration #

#### Official Correspondent:

Guy Scott WinProbe Corp. President Email: qscott@winprobe.com Telephone: (561)318-8367 11662 Lake Shore Place North Palm Beach, Florida 33408

#### Or

Rose Malchow WinProbe Corp. Quality Manager Email: rmalchow@winprobe.com Telephone: (561) 318-8367 11770 US Highway 1, suite 405E Palm Beach Gardens, Florida 33408

# Date Prepared:

February 19, 2015

# Device Name and Classification:

Common/Usual Name:

Diagnostic Ultrasound System

Proprietary Name:

UltraVision 2 Diagnostic Ultrasound System

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows a coat of arms with a white background and a black border. In the center of the coat of arms is a red winged lion holding a spear. Above the lion is a white banner with the word "imagine" written on it in red. The lion is depicted in a dynamic pose, giving the impression of movement and strength.

| Classification Names:                 | FR Number | Product Code |
|---------------------------------------|-----------|--------------|
| Ultrasound Pulsed Echo Imaging System | 892.1560  | IYO          |
| Diagnostic Ultrasound Transducer      | 892.1570  | ITX          |

## Predicate Device:

Siemens Acuson S2000 Ultrasound System, K081148.

#### Device Description:

The UltraVision 2 is a portable Diagnostic Ultrasound System which applies the latest technologies to produce optimal images. Various image parameter adjustments, a 15 inch high resolution display, and custom probes are configured to provide clear and stable images. It operates in B-mode and.

## Intended Use:

This diagnostic ultrasound system (UltraVision) and its transducers are intended for use in clinical examinations to evaluate differences in the echogenicity of soft tissue of small parts in adult patients. (thyroid, breast, and testicles) and for superficial muscular skeletal diagnosis.

The system is for prescription use only by a trained sonographer under the direction of a qualified physician or directly by a qualified physician.

#### Effectiveness and Safety Contraindications:

Clinical Test: Clinical testing not required

#### Non-clinical Test:

Testing according to the following safety standards are conducted on the subject device:

- 1. IEC 60601-1 Safety Requirements for Medical Equipment
- 2. IEC 60601-1-2 EMC Requirements for Medical Equipment
- 3. IEC 60601-2-37 Diagnostic Ultrasound Safety Standards
- 4. Acoustic output testing as per the FDA guidance "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers", dated September 9, 2008
- 5. ISO 10993-1 Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process, ISO 10993-5 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity, ISO 10993-10 Biological evaluation of medical devices – Part 10: Tests for irritation and delayed-type hypersensitivity

#### Comparison to the predicate device:

The UltraVision 2 diagnostic ultrasound system uses the same fundamental scientific technologies as the predicate device (Siemens Acuson S2000, K081148). Table 1 compares the UltraVision 2 to the predicate device with respect to indications for use, intended use, and principles of operation for the determination of substantial equivalence.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows a coat of arms with a white background. In the center of the coat of arms is a red griffin holding a spear. Above the griffin is a white banner with the word "IMAGINE" written on it in red letters. The coat of arms has a black border.

| Manufacturer            | WinProbe                                                                                                                                                                                                                                                                                                                                                                                                                                                | Siemens                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name              | UltraVision 2                                                                                                                                                                                                                                                                                                                                                                                                                                           | Acuson S2000                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| 510(k) Number           |                                                                                                                                                                                                                                                                                                                                                                                                                                                         | K081148                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Product Code            | IYO, ITX                                                                                                                                                                                                                                                                                                                                                                                                                                                | IYN, IYO, ITX                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Regulation Number       | 892.1560<br>892.1570                                                                                                                                                                                                                                                                                                                                                                                                                                    | 892.1550<br>892.1560<br>892.1570                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Regulation Name         | Ultrasonic pulsed echo imaging system                                                                                                                                                                                                                                                                                                                                                                                                                   | Ultrasonic pulsed Doppler imaging system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Indications for Use     | This diagnostic ultrasound system (UltraVision) and its transducers are intended for use in clinical examinations to evaluate differences in the echogenicity of soft tissue of small parts in adult patients, (thyroid, breast, and testicles) and for superficial muscular skeletal diagnosis.<br>The system is for prescription use only by a trained sonographer under the direction of a qualified physician or directly by a qualified physician. | <b>Indications for Use:</b><br>The Modified S2000, S2000 ABVS Ultrasound System, is intended for the following applications: General Radiology, Abdominal, Fetal, Small Parts, Transcranial, OB/GYN, Cardiac, Pelvic, Neonatal/Adult Cephalic, Pediatric, Urology, Vascular, Musculoskeletal, Superficial Musculoskeletal, and Peripheral Vascular applications. The system supports the transducers listed in the Ultrasound Indications for Use tables, including the 14L5BV for B-mode imaging of a patient's breast using an optional automatic scanning function. The device is not intended to be used as a replacement for screening mammography.<br>The system also provides for the measurement of anatomical structures and analysis as provided in the original S2000 Ultrasound System, which provides information that is used by medical health care professionals for clinical diagnosis purposes. |
| Significant Differences | The main difference is the size<br>Dimensions: 400 mm (W) x 50 mm (H) x 250 mm (D)<br>Weight: 4.5 kg                                                                                                                                                                                                                                                                                                                                                    | Dimensions: 623 mm (W) x 1300 mm (H) x 1103 mm (D)<br>Weight: 166 kg                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |

#### Table 1. Predicate Device Comparison

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image is a coat of arms with a white background and a black border. In the center of the coat of arms is a red griffin holding a spear. Above the griffin is a banner with the word "IMAGISTER" written on it. The griffin is standing on its hind legs and facing to the left.

## Intended Use:

The intended use and clinical applications of the UltraVision 2 system are narrowed, but still in the scope of the predicate device. Both systems are intended to be used with a conventional extracorporeal transducer. The type of transducer specified for use with the UltraVision 2 system is linear which is also used with the predicate system. A comparison of the transducers is provided in Table 2:

|                        | HL5-10EPN<br>(UltraVision 2)          | 14L5<br>(Acuson S2000)                                  |
|------------------------|---------------------------------------|---------------------------------------------------------|
| Type                   | Linear                                | Linear                                                  |
| Frequency<br>Bandwidth | 5 - 12 MHz                            | 5 - 14 MHz                                              |
| Applications           | Breast Testes Thyroid Musculoskeletal | Breast Testes Thyroid Musculoskeletal Peripheral Vessel |
| Number of<br>Elements  | 128                                   | 128                                                     |
| Modes of Operation     | B                                     | B, C, D, M                                              |
| Array Footprint        | 38.1 mm                               | 39 mm                                                   |

Table 2. Probe Comparison

# Operating Principle and Design:

Both the UltraVision 2 and the predicate system transmit ultrasonic energy into patients then perform postprocessing of received echoes to generate on-screen displays of anatomic structures of the human body. Both systems are designed with an LCD display screen and hand-held transducers. Both systems support the same operating mode (B) and the same measurement functions for anatomic structures. The operation characteristics (installation and use, mode of operation) and the power supply are the same in both systems.

# Non-clinical Test and Safety:

The UltraVision 2 system is in conformance with the standards described above which are the same or equivalent to those of the predicate device.

# Biocompatibility:

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image is a coat of arms featuring a red winged lion holding a spear on a white background. The lion is standing on its hind legs and facing left. A banner with the word "Imagine" is draped behind the lion. The coat of arms has a simple shield shape with a black outline.

The patient contact materials of human body surface are evaluated under ISO 10993 and determined acceptable for the specified usage of the system. Both systems have the same acceptance level for biocompatibility.

## Conclusion:

Clinical studies are not required to support substantial equivalence for these conventional ultrasound systems. In addition, as discussed in the above technological comparison, the technological characteristics of the Ultra\lision 2 system are substantially equivalent to the referenced predicate device that has been previously cleared for USA distribution.

# Substantially Equivalent Determination:

By definition, a device is substantially equivalent to a predicate device has the same intended use and the same operational characteristics as the previously cleared predicate device, or the device has the same intended use and different operational characteristics in which substantial equivalency can be demonstrated in the device in comparison to the predicate device. In addition, the new device does not raise new questions regarding its safety and effectiveness as compared to the predicate device.

It is shown in this 510(k) submission that the difference between the UltraVision 2 and the predicate device does not raise any questions regarding its safety and effectiveness. The UltraVision 2, as designed and produced, is determined to be substantially equivalent to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K150580](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K150580)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
