K150342 · Carestream Health, Inc. · IYO · May 8, 2015 · Radiology
Device Facts
Record ID
K150342
Device Name
Touch Ultrasound
Applicant
Carestream Health, Inc.
Product Code
IYO · Radiology
Decision Date
May 8, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: The system is intended for use by qualified physicians for ultrasound evaluation. Specific clinical applications and exam types include: Fetal (including Obstetrics) Abdominal Intraoperative Intraoperative Neuro (also known as Neurosugery) Pediatric Small Organ (also known as Small Parts) Neonatal Cephalic (also known as Neonatal Transcranial) Adult Cephalic (also known as Adult Transcranial) Transrectal Transvaginal Transurethral Musculoskeletal (Conventional and Superficial) Cardiac Adult Peripheral Vessel (also known as Peripheral Vascular) Indicated uses are different for different transducers. The Product Data sheet for the system contains a table listing the indicated uses for each transducer that can be used with the system. The bk3000 ultrasound system is not intended for ophthalmic use or any use causing the acoustic beam to pass through the eye. The Cardiac Adult application is not intended for direct use on the heart.
Device Story
Touch Ultrasound is a diagnostic ultrasound system for imaging and fluid flow analysis. It processes ultrasonic pulsed echo signals to generate B-mode, M-mode, Color Flow Mapping (CFM), and Amplitude (Power) Doppler images. It supports optional 3D reconstruction of 2D images. Operated by physicians in clinical settings, the device provides real-time visualization and geometric measurements for vascular, urological, cardiac, and OB/GYN applications. It also supports needle guidance for biopsies and punctures. The system uses linear, convex, and phased array transducers. Output is displayed on a screen for clinical assessment, aiding in diagnosis and procedural guidance. The device is a rebranded version of the bk 2300 system.
Clinical Evidence
Bench testing only. No clinical data provided. The device was evaluated for acoustic output, thermal, electrical, electromagnetic, and mechanical safety, and biocompatibility in accordance with recognized standards (e.g., IEC 60601-1, IEC 60601-2-37, ISO 10993-1).
Technological Characteristics
Ultrasonic pulsed echo imaging system using linear, convex, and phased array transducers. Modes: B-mode (with Tissue Harmonic Imaging), M-mode, PWD, CFM (Color Doppler and Amplitude Doppler). Complies with AIUM/NEMA UD-2 and UD-3 for acoustic output. Biocompatible patient contact materials. Software lifecycle follows IEC 62304. Safety standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62359.
Indications for Use
Indicated for qualified physicians performing ultrasound evaluation in fetal, abdominal, intraoperative, neurosurgical, pediatric, small organ, neonatal/adult cephalic, transrectal, transvaginal, transurethral, musculoskeletal, cardiac (non-direct), and peripheral vascular applications. Contraindicated for ophthalmic use or any use where the acoustic beam passes through the eye; cardiac application is not for direct use on the heart.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
B-K Medical Ultrasound Scanner bk 2300
Reference Devices
Pro Focus 2202
Related Devices
K152467 — Touch Ultrasound · Carestream Health, Inc. · Dec 9, 2015
K151910 — Ultrasound Scanner System BK2300 · Bk Medical Aps · Oct 5, 2015
K024236 — ULTRASOUND SCANNER, TYPE 2400 · B-K Medical A/S · Jan 17, 2003
K180737 — Ultrasound Scanner System bk2300 · Bk Medical Aps · Aug 1, 2018
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 8, 2015
Carestream Health, Inc. % Ms. Carolyn Wagner Senior Manager, Regulatory Affairs & Quality Systems 150 Verona Street ROCHESTER NY 14608
Re: K150342
Trade/Device Name: Touch Ultrasound Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, IYN, ITX Dated: March 19, 2015 Received: March 24, 2015
Dear Ms. Wagner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150342
Device Name Touch Ultrasound
| Indications for Use (Describe) |
|------------------------------------------------------------------------------------------|
| The system is intended for use by qualified physicians for ultrasound evaluation. |
| Specific clinical applications and exam types include: |
| Fetal (including Obstetrics) |
| Abdominal |
| Intraoperative |
| Intraoperative Neuro (also known as Neurosugery) |
| Pediatric |
| Small Organ (also known as Small Parts) |
| Neonatal Cephalic (also known as Neonatal Transcranial) |
| Adult Cephalic (also known as Adult Transcranial) |
| Transrectal |
| Transvaginal |
| Transurethral |
| Musculoskeletal (Conventional and Superficial) |
| Cardiac Adult |
| Peripheral Vessel (also known as Peripheral Vascular) |
| Indicated uses are different for different transducers. The Product Data sheet for the |
| system contains a table listing the indicated uses for each transducer that can be used |
| with the system. |
| The bk3000 ultrasound system is not intended for ophthalmic use or any use causing |
| the acoustic beam to pass through the eye. The Cardiac Adult application is not intended |
| for direct use on the heart. |
| Type of Use (Select one or both, as applicable) |
|X | Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### System: Touch Ultrasound
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Mode of Operation | | | | | | | | |
|---------------------------------|-------------------|---|----|---------------------|---------|-----------|---------------------|-------------|-------|
| | | | PW | Tissue | Color | Amplitude | Color | Combined | Other |
| Specific<br>(Tracks I & III) | B | M | D | Harmonic<br>Imaging | Doppler | Doppler | Velocity<br>Imaging | (Specify 1) | |
| Ophthalmic | | | | | | | | | |
| Fetal 2) | N | N | N | N | N | N | N | N | |
| Abdominal | N | N | N | N | N | N | N | N | |
| Intra-operative (Specify) | N | N | N | N | N | N | N | N | |
| Intra-operative (Neuro) 6) | N | N | N | N | N | N | N | N | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | N | N | N | N | N | |
| Small Organ (Specify) 3) | N | N | N | N | N | N | N | N | |
| Neonatal Cephalic 4) | N | N | N | N | N | N | N | N | |
| Adult Cephalic 4) | N | N | N | N | N | N | N | N | |
| Trans-rectal | N | N | N | N | N | N | N | N | |
| Trans-vaginal | N | N | N | N | N | N | N | N | |
| Trans-urethral | N | N | N | N | N | N | N | N | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skeletal (Conventional) | N | N | N | N | N | N | N | N | |
| Musculo-skeletal (Superficial) | N | N | N | N | N | N | N | N | |
| Intra-luminal | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
| Cardiac Adult 5) | N | N | N | N | N | N | N | N | |
| Cardiac Pediatric | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
| Peripheral vessel | N | N | N | N | N | N | N | N | |
| Other (Specify) | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging
Additional Comments:
==============================================================================================================================================================================
- 1) Mode combinations:
B+M, B+D, B+C, B+D+C. B mode includes Tissue Harmonic Imaging
(D is PWD, C is Color Flow mapping Doppler including Amplitude(power)Doppler)
- 2) "Fetal" (including Obstetrics).
- 3) "Small organ" also known as "Small parts".
- 4) "Cephalic" is sometimes called "Trans-cranial".
- 5) Not for direct use on the heart.
- 6) Intra-operative (Neuro) also known as "Neurosurgery"
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR)
{4}------------------------------------------------
System: Touch Ultrasound______________ Transducer: 9018
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | | | | | | | |
|------------------------------|---|---|-----|-------------------------------|------------------|----------------------|------------------------------|-------------------------|-------|
| Specific<br>(Tracks I & III) | B | M | PWD | Tissue<br>Harmonic<br>Imaging | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify 1) | Other |
| Ophthalmic | | | | | | | | | |
| Fetal | | | | | | | | | |
| Abdominal | | | | | | | | | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | | | | | | | | | |
| Small Organ (Specify) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal 2) | N | N | N | N | N | N | N | N | |
| Trans-vaginal 3) | N | N | N | N | N | N | N | N | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skel. (Conventional) | | | | | | | | | |
| Musculo-skel. (Superficial) | | | | | | | | | |
| Intra-luminal | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
| Peripheral vessel | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging
Additional Comments:
- 1) Mode combinations:
B+M, B+D, B+C, B+D+C. (D is PWD, C is Color Flow mapping Doppler including Amplitude (power) Doppler). Tissue Harmonic Imaging.
- 2) Trans-rectal: Prostate.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
- 3) Trans-vaginal: Ovaries, Uterus.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR)
{5}------------------------------------------------
System: Touch Ultrasound Transducer: 9023
| Clinical Application<br>Specific<br>(Tracks I & III) | B | M | PWD | Tissue<br>Harmonic<br>Imaging | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify 1) | Other |
|------------------------------------------------------|---|---|-----|-------------------------------|------------------|----------------------|------------------------------|-------------------------|-------|
| Ophthalmic | | | | | | | | | |
| Fetal 4) | N | N | N | N | N | N | N | | |
| Abdominal 2) | N | N | N | N | N | N | N | | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric 3) | N | N | N | N | N | N | N | | |
| Small Organ (Specify) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skel. (Conventional) | | | | | | | | | |
| Musculo-skel. (Superficial) | | | | | | | | | |
| Intra-luminal | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
| Peripheral vessel | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.
Additional Comments:
- 1) Mode combinations: B+M, B+D, B+C, B+D+C. (D is PWD, C is Color Flow mapping Doppler including Amplitude(power)Doppler)
- 2) Abdominal: Kidney, Liver, Pancreas, Spleen, Lymph Nodes, Gastrointestinal tract, Adrenal glands, Bladder and Aorta:
- 3) Pediatric: Kidney, Liver, Pancreas, Spleen, Lymph Nodes, Gastrointestinal tract, Adrenal glands, Bladder and Aorta
- 4) Fetal (including obstetrics): Gastrointestinal tract, Bladder, Spleen, Head, Liver, Umbilical cord, Kidneys, Vasculature, Aorta, Brain, Limbs, Heart, Adrenal glands, Diaphragm, Facial features, Soft tissue and Spine
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR) Prescription Use (Per 21 CFR 801.109)
II
{6}------------------------------------------------
System: Touch Ultrasound Transducer: 9051
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| ഗ്നിശ്വീ പുഴുത്ത് പ്രസ്സി | | | | | | | | | |
|---------------------------------|---|---|-----|----------|---------|-----------|----------|-------------|-------|
| Specific | B | M | PWD | Tissue | Color | Amplitude | Color | Combined | Other |
| (Tracks I & III) | | | | Harmonic | Doppler | Doppler | Velocity | (Specify 1) | |
| | | | | Imaging | | | Imaging | | |
| Ophthalmic | | | | | | | | | |
| Fetal 4) | | | | | | | | | |
| Abdominal 2) | | | | | | | | | |
| Intra-operative (Specify | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric 2) | N | N | N | N | N | N | | N | |
| Small Organ (Specify 3) | N | N | N | N | N | N | | N | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skel. (Conventional) 4) | N | N | N | N | N | N | | N | |
| Musculo-skel. (Superficial) 5) | N | N | N | N | N | N | | N | |
| Intra-luminal | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
| Peripheral vessel 6) | N | N | N | N | N | N | | N | |
| Other (Specify) | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.
Additional Comments:
- 1) Mode combinations: B+M, B+D, B+C, B+D+C. (D is PWD, C is Color Flow mapping Doppler including Amplitude(power)Doppler)
- 2) Pediatric: hip, spine (except neonatal), renal
- 3) Small Organs: breast, thyroids, testes
- 4) Musculoskeletal ultrasound (conventional), shoulder, hip, spine, knee
- 5) Musculoskeletal ultrasound (superficial), wrist, finger, toes
- 6) Peripheral vessel neck, carotid, vertebra, extremities, venous, arterial
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR)
{7}------------------------------------------------
System: Touch Ultrasound Transducer: 9062
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application<br>Specific<br>(Tracks I & III) | B | M | PWD | Tissue<br>Harmonic<br>Imaging | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify 1) | Other |
|------------------------------------------------------|---|---|-----|-------------------------------|------------------|----------------------|------------------------------|-------------------------|-------|
| Ophthalmic | | | | | | | | | |
| Fetal | | | | | | | | | |
| Abdominal | | | | | | | | | |
| Intra-operative (Specify 2 ) | N | N | N | N | N | N | | N | |
| Intra-operative (Neuro) 3) | N | N | N | N | N | N | | N | |
| Laparoscopic | | | | | | | | | |
| Pediatric | N | N | N | N | N | N | | N | |
| Small Organ (Specify) | | | | | | | | | |
| Neonatal Cephalic | N | N | N | N | N | N | | N | |
| Adult Cephalic | | | | | | | | | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skel. (Conventional) | | | | | | | | | |
| Musculo-skel. (Superficial) | | | | | | | | | |
| Intra-luminal | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
| Cardiac Adult | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
| Peripheral vessel | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.
Additional Comments:
- 1) Mode combinations: B+M, B+D, B+C, B+D+C. (D is PWD, C is Color Flow mapping Doppler including Amplitude(power)Doppler)
- 2) Intraoperative: Gall bladder
- 3) Intra-operative (Neuro) also known as "Neurosurgery" Includes spinal cord
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR)
{8}------------------------------------------------
| System: Touch Ultrasound | |
|--------------------------|------|
| Transducer: | 9077 |
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application<br>Specific<br>(Tracks I & III) | B | M | PWD | Tissue<br>Harmonic<br>Imaging | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify 1) | Other |
|------------------------------------------------------|---|---|-----|-------------------------------|------------------|----------------------|------------------------------|-------------------------|-------|
| Ophthalmic | | | | | | | | | |
| Fetal | | | | | | | | | |
| Abdominal 2) | N | N | N | N | N | N | | N | |
| Intra-operative (Specify) | | | | | | | | | |
| Intra-operative (Neuro) | | | | | | | | | |
| Laparoscopic | | | | | | | | | |
| Pediatric | | | | | | | | | |
| Small Organ (Specify) | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | |
| Adult Cephalic | N | N | N | N | N | N | | N | |
| Trans-rectal | | | | | | | | | |
| Trans-vaginal | | | | | | | | | |
| Trans-urethral | | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | | |
| Musculo-skel. (Conventional) | | | | | | | | | |
| Musculo-skel. (Superficial) | | | | | | | | | |
| Intra-luminal | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
| Cardiac Adult 3) | N | N | N | N | N | N | | N | |
| Cardiac Pediatric | | | | | | | | | |
| Trans-esoph. (Cardiac) | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
| Peripheral vessel) | | | | | | | | | |
| Other (Specify) | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E
*Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.
Additional Comments:
onlin
- 1) Mode combinations:
B+M, B+D, B+C, B+D+C. (D is PWD, C is Color Flow mapping Doppler including Amplitude (power) Doppler,
- 2) Abdominal:
==============================================================================================================================================================================
- Liver, kidney, vessels, spleen etc. and FAST
- 3) Cardiac Adult:
Not for direct use on the heart.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR)
{9}------------------------------------------------
## 510(k) Summary:
This summary is provided as part of this Premarket Notification in compliance with 21CRF, Section 807.92.
Submitters name: Carestream Health, Inc.
Address: 150 Verona Street, Rochester, NY 14608, USA
Phone: 585-627-6588
Fax: 585-323-7643
Contact person: Carolyn L Wagner, Senior Manager, Regulatory Affairs & Quality Systems
Date prepared: March 14, 2015
Trade name: Touch Ultrasound Common name: Diagnostic Ultrasound System Classification names: (90 IYO, CFR 892.1560) Ultrasonic Pulsed Echo Imaging System Ultrasonic Pulsed Doppler Imaging System (90 IYN, CFR 892.1560) Diagnostic Ultrasonic Transducer (90 ITX, CFR 892.1570)
Identification of predicate, legally marketed device:
B-K Medical Ultrasound Scanner bk 2300
## Device description:
Touch Ultrasound supports the following scanning modes and combinations thereof: B-mode (incl. Tissue Harmonic Imaging), M-mode, CFM mode, Amplitude (Power) Doppler mode.
The system can perform simple geometric measurements, and perform calculations in the areas of Vascular, Urology, Cardiology and OB/GYN applications.
The system can guide biopsy- and puncture needles.
An optional 3-D module can reconstruct a series of 2-D images into a single 3-D volume and display this on the screen.
Transducers Transducers are linear arrays, convex arrays, and phased arrays.
The patient contact materials are biocompatible.
All transducers used together with bk 2300 are Track 3 transducers.
## Acoustic output
The system controlling the Acoustic Output in the Touch Ultrasound is the same as the system in the predicate device Pro Focus 2202. The system will assure that the acoustic output always will stay below the pre-amendments upper limits i.e. Ispta ≤ 720 mW/cm² and MI ≤ 1.9 (Track 3. non ophthalmic).
{10}------------------------------------------------
The Thermal Index values are maximum 6.0, i.e. TI ≤ 6.0
### Clinical measurement accuracy
Clinical measurements and calculations are described and accuracies are provided in the User Information.
## Thermal, mechanical and electrical safety.
The scanner bk 2300 has been tested by a recognized Certified Body.
### Acoustic Output Reporting
The Acoustic Output Reporting is made according to the standards required by "Information for Manufacturers Seeking Clearance of Diagnostic Ultrasound Systems and Transducers, FDA, CDRH, September 9, 2008"
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body.
{11}------------------------------------------------
# Comparison of Technological Characteristics: Predicate Device Compared to New Device:
| | Predicate Device | New Device |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Name | bk 2300 | Touch Ultrasound |
| Modes of operation<br>Ref.: [1] | B, M, PWD, CFM1) and combinations | B, M, PWD, CFM1) and combinations |
| Intended Use: | Tissue harmonic imaging. Diagnostic ultrasound imaging or<br>image flow analysis of the human body. | Tissue harmonic imaging. Diagnostic ultrasound imaging or<br>image flow analysis of the human body. |
| Indications for Use: | The system is intended for use by qualified physicians for<br>ultrasound evaluation.<br>Specific clinical applications and exam types include:<br>Fetal (including Obstetrics)<br>Abdominal<br>Intraoperative<br>Intraoperative Neuro (also known as Neurosugery)<br>Pediatric<br>Small Organ (also known as Small Parts)<br>Neonatal Cephalic (also known as Neonatal Transcranial)<br>Adult Cephalic (also known as Adult Transcranial)<br>Transrectal<br>Transvaginal<br>Transurethral<br>Musculoskeletal (Conventional and Superficial)<br>Cardiac Adult<br>Peripheral Vessel (also known as Peripheral Vascular)<br>Indicated uses are different for different transducers. The<br>Product Data sheet for the<br>system contains a table listing the indicated uses for each<br>transducer that can be used<br>with the system.<br>The bk3000 ultrasound system is not intended for ophthalmic<br>use or any use causing the acoustic beam to pass through the<br>eye. The Cardiac Adult application is not intended for direct<br>use on the heart. | The system is intended for use by qualified physicians for<br>ultrasound evaluation.<br>Specific clinical applications and exam types include:<br>Fetal (including Obstetrics)<br>Abdominal<br>Intraoperative<br>Intraoperative Neuro (also known as Neurosugery)<br>Pediatric<br>Small Organ (also known as Small Parts)<br>Neonatal Cephalic (also known as Neonatal Transcranial)<br>Adult Cephalic (also known as Adult Transcranial)<br>Transrectal<br>Transvaginal<br>Transurethral<br>Musculoskeletal (Conventional and Superficial)<br>Cardiac Adult<br>Peripheral Vessel (also known as Peripheral Vascular)<br>Indicated uses are different for different transducers. The<br>Product Data sheet for the<br>system contains a table listing the indicated uses for each<br>transducer that can be used<br>with the system.<br>The bk3000 ultrasound system is not intended for ophthalmic<br>use or any use causing the acoustic beam to pass through the<br>eye. The Cardiac Adult application is not intended for direct use<br>on the heart. |
1) CFM= Color Flow Mapping=Color Doppler and Amplitude (Power) Doppler.
## A brief discussion of non-clinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence:
The device has been evaluated for acoustic output, thermal, electrical, electromagnetic and mechanical safety, biocompatibility, and has been found to conform with applicable medical device safety standards. The system complies with the following voluntary standards:
{12}------------------------------------------------
- AIUM/NEMA UD-2. Acoustic Output Measurement Standard For Diagnostic Ultrasound Equipment
- AIUM/NEMA UD-3, Real-time Display of Thermal and Mechanical Acoustic Output ● Indices on Diagnostic Ultrasound Equipment
- IEC 60601-1, Medical Electrical Equipment, Part 1: General requirements for safety ●
- IEC 60601-1-2, General requirements for safety, Collateral Standard, Electromagnetic Compatibility - Requirements and tests
- IEC 60601-2-37, Particular requirements for the safety of ultrasonic diagnostic medical and monitoring equipment
- ISO 14971, Application of Risk Management of Medical Devices
- EN ISO 10993-1, Biocompatibility
- IEC 62304, Medical Device Software - Software lifecycle processes
- . IEC 62359. Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields
Thermal, electrical, electromagnetic and mechanical safety is unchanged.
## Technological characteristics compared to the predicate device
The predicate device has the same technological characteristics as the subject device described above.
Minor differences consist of: Modified trade dress from BK/Analogic to Carestream Health.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.