← Product Code [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO) · K143493

# SignosRT Bladder Ultrasound System (K143493)

_Signostics, Ltd. · IYO · Jan 14, 2015 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K143493

## Device Facts

- **Applicant:** Signostics, Ltd.
- **Product Code:** [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO.md)
- **Decision Date:** Jan 14, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

Signos RT Bladder scanner projects ultrasound energy through the lower abdomen of a patient to obtain an image of the bladder that is used to non-invasively determine bladder volume. Users must have ultrasound training before using the device.

## Device Story

Hand-held ultrasound system; projects 3MHz ultrasonic energy through lower abdomen; receives echoes to generate B-mode images on integrated LCD. Used by trained clinicians for non-invasive bladder volume measurement. Device calculates volume from single or dual scans using embedded algorithms. Output displayed on-screen; supports external USB PictBridge printing or image transfer to PC via SigViewer software. Benefits include non-invasive assessment of bladder volume, aiding clinical decision-making regarding urinary retention or catheterization. Operates on Li-Ion battery; portable form factor.

## Clinical Evidence

Bench testing only. Phantom studies demonstrated bladder volume measurement accuracy within ±9.6% (specification ±15%). Imaging performance (lateral/axial resolution) verified as identical to predicate. Compliance with IEC 60601-1, IEC 60601-2-37, IEC 60601-1-2, NEMA UD-2, NEMA UD-3, and ISO 10993 biocompatibility standards confirmed.

## Technological Characteristics

Hand-held B-mode ultrasound; 3MHz annular sector transducer. Materials: Sabic Cycoloy HC1204HF, Mitsui TPX-MED18, Sabic Versollen OMX1255NX-1. Dimensions: 6.8 x 11.5 x 2.0 cm; 0.13 kg. Power: Li-Ion battery or 100-240 VAC external. Connectivity: USB for printing/PC data transfer. Safety: Class II, Type BF, IPX7 transducer lens. Software developed per IEC 62304.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Portascan Bladder Scanner ([K033906](/device/K033906.md))
- SignosRT Ultrasound System ([K130659](/device/K130659.md))

## Reference Devices

- SpeqRT Personal Ultrasound System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

January 14, 2015

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Signostics Limited % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street. NW BUFFALO MN 55313

Re: K143493

Trade/Device Name: SignosRT Bladder Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO ITX Dated: December 28, 2014 Received: January 7, 2015

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A. Ochs

Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known)

#### K143493

Device Name Signos RT Bladder

#### Indications for Use (Describe)

Signos RT Bladder scanner projects ultrasound energy through the lower abdomen of a patient to obtain an image of the bladder that is used to non-invasively determine bladder volume. Users must have ultrasound training before using the device.

#### Contraindications

The SignosRT Bladder Scanner is designed for percutaneous scanning only. Do not attempt intracavity imaging; in particular, trans-esophageal, trans-rectal scans are contraindicated.

The SignosRT Bladder Scanner is contraindicated for pregnant patients.

Type of Use (Select one or both, as applicable)

🇿 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwoork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

#### 510(k) Indications for Use Form

#### TABLE 1 - SIGNOSRT BLADDER INDICATIONS FOR USE FORM

System: SignosRT Bladder Intended Use: Diagnostic ultrasound imaging of the human body as follows: Clinical Application Mode of Operation PWD General Specific B M CWD Color Combined Other* (Track I Only) (Tracks I & III) Doppler (Specify) (Specify) Ophthalmic Ophthalmic Fetal Abdominal N Intra-operative (Specify) Intra-operative(Neuro) Laparoscopic Pediatric Small Organ (Specify) Neonatal Cephalic Adult Cephalic Fetal Imaging Trans-rectal & Other Trans-vaginal Trans-urethral Trans-esoph. (non-Card.) Musculo-skeletal (Conventional) Musculo-skeletal (Superficial) Intravascular Other (Specify) Cardiac Adult Cardiac Pediatric Intravascular (Cardiac) Cardiac Trans-esoph. (Cardiac) Intra-cardiac Other (Specify) Peripheral Peripheral vessel Vessel Other (Specify)

N= new indication; P=previously cleared by FDA; E=added under AppendixE Additional Comments:

Type of Use (Select one or both, as applicable)

区 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

{4}------------------------------------------------

## TABLE 2 - SIGNOSRT BLADDER SCANNER INDICATIONS FOR USE FORM

System: SignosRT Bladder

Transducer: S3 (P03479)

Diagnostic ultrasound imaging of the human body as follows : Intended Use:

| Clinical Application      |                                    | Mode of Operation |   |     |     |                  |                       |                     |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|---------------------|
| General<br>(Track I Only) | Specific<br>(Tracks I & III)       | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | Other*<br>(Specify) |
| Ophthalmic                | Ophthalmic                         |                   |   |     |     |                  |                       |                     |
| Fetal Imaging<br>& Other  | Fetal                              |                   |   |     |     |                  |                       |                     |
|                           | Abdominal                          | N                 |   |     |     |                  |                       |                     |
|                           | Intra-operative (Specify)          |                   |   |     |     |                  |                       |                     |
|                           | Intra-operative(Neuro)             |                   |   |     |     |                  |                       |                     |
|                           | Laparoscopic                       |                   |   |     |     |                  |                       |                     |
|                           | Pediatric                          |                   |   |     |     |                  |                       |                     |
|                           | Small Organ (Specify)              |                   |   |     |     |                  |                       |                     |
|                           | Neonatal Cephalic                  |                   |   |     |     |                  |                       |                     |
|                           | Adult Cephalic                     |                   |   |     |     |                  |                       |                     |
|                           | Trans-rectal                       |                   |   |     |     |                  |                       |                     |
|                           | Trans-vaginal                      |                   |   |     |     |                  |                       |                     |
|                           | Trans-urethral                     |                   |   |     |     |                  |                       |                     |
|                           | Trans-esoph. (non-Card.)           |                   |   |     |     |                  |                       |                     |
|                           | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                       |                     |
|                           | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                       |                     |
|                           | Intravascular                      |                   |   |     |     |                  |                       |                     |
|                           | Other (Specify)                    |                   |   |     |     |                  |                       |                     |
| Cardiac                   | Cardiac Adult                      |                   |   |     |     |                  |                       |                     |
|                           | Cardiac Pediatric                  |                   |   |     |     |                  |                       |                     |
|                           | Intravascular (Cardiac)            |                   |   |     |     |                  |                       |                     |
|                           | Trans-esoph. (Cardiac)             |                   |   |     |     |                  |                       |                     |
|                           | Intra-cardiac                      |                   |   |     |     |                  |                       |                     |
|                           | Other (Specify)                    |                   |   |     |     |                  |                       |                     |
| Peripheral                | Peripheral vessel                  |                   |   |     |     |                  |                       |                     |
| Vessel                    | Other (Specify)                    |                   |   |     |     |                  |                       |                     |

N= new indication; P=previously cleared by FDA; E=added under AppendixE Additional Comments:

Type of Use (Select one or both, as applicable)

区 Prescription Use (Part 21 CFR 801 Subpart D)

□ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

{5}------------------------------------------------

## 510(k) Summary

#### 1. Sponsor:

Signostics Ltd PO Box 1048, Pasadena 1284 South Road Clovelly Park, SA 5042 Australia

## 2. Contact Person: Stewart Bartlett Quality and Regulatory Manager Telephone: +61 (8) 7424 0600

- 3. Date Prepared: December 28, 2014
- 4. De vice Name : SignosRT Bladder
- 5. Proprietary/Marketed Names: SignosRT Bladder

## 6. Common/Usual Name:

Diagnostic ultrasound system and transducer

### 7. Classification

Regulatory Class: II

Review Category: Tier II

Ultrasonic Pulsed Echo Imaging System (21 CFR 892.1560, 90-IYO)

(21 CFR 892.1570, 90-ITX)

Diagnostic Ultrasound Transducer

Classification Panel: Radiology

### 8. Predicate Devices

Portascan Bladder Scanner (K033906)

SignosRT Ultrasound System (K130659)

{6}------------------------------------------------

## 9. Risk Analysis Method Used

Signostics Ltd applied ISO-14971 to the design and development of the SignosRT Bladder. The conclusion from the risk analysis was the device was safe for its intended use and does not pose any unacceptable risks.

#### 10. Basis for Substantial Equivalence - Technological Characteristics

Signostics Ltd believes the SignosRT Bladder described in this Submission is substantially equivalent to the predicate devices as follows:

- a. Portascan Bladder Scanner (K033906)
- b. SignosRT Ultrasound System (K130659)

The SignosRT Bladder is substantially equivalent to the predicate devices listed above. All systems transmit ultrasonic energy into patients, then process received echoes to produce on-screen images of anatomy. All systems allow for B-mode imaging and measurements of volume of a bladder. The indication for use statement of the SignosRT Bladder is identical to the Portascan. The SignosRT Personal Ultrasound system (K130659) is a general ultrasound imaging device with an intended use statement covering a number of areas, including the abdomen which includes the bladder.

All systems enable an estimated bladder volume to be calculated from a single scan or from two scans for more accurate results. The algorithms used to calculate bladder volume are identical for the SignosRT Bladder and the predicate SpeqRT Personal Ultrasound System.

The SignosRT Bladder is identical in construction, materials, and controls to the SignosRT Personal Ultrasound (K130659), with the only difference being the indications for use statement and the software configuration. The patient contact materials are identical. The transducer and operating frequency (3.5MHz) are also identical. Both systems contain sector transducers with annular (circular) acoustic elements generating pie shaped images, have identical operating frequency, identical outer diameter of acoustic crystal, identical acoustic output, and both maintain MI and TI to be <1.0 at all times.

The predicate SignosRT Personal Ultrasound has a remote display mode to allow streaming of the video to a larger display. The SignosRT Bladder also has support to allow streaming of the video to a larger display.

The predicate Portascan has printer as part of the device. The SignosRT Bladder scanner and predicate SignosRT Ultrasound System both can be connected directly to a USB PictBridge printer for full-size printouts, or transfer images to a PC running SigViewer accessory software to print images.

{7}------------------------------------------------

# 11. Device Description

The Signostics Ltd SignosRT Bladder is a hand-held, diagnostic ultrasound system with an on-screen display. Its purpose is to acquire ultrasound echo data and display it in B-Mode on an LCD display to enable non-invasive volume measurement of a patients bladder.

Technical specifications for the Signostics SignosRT Bladder are as follows:

| System                     |        |                                                                                                |                          |         |            |
|----------------------------|--------|------------------------------------------------------------------------------------------------|--------------------------|---------|------------|
| Transducer frequencies:    |        | 3MHz (S3 transducer)                                                                           |                          |         |            |
| Frame rate:                |        | 8Fps or 16Fps (Imaging only)                                                                   |                          |         |            |
| Ultrasound lines/frame:    |        | 128 lines for 90° frame at 10cm                                                                |                          |         |            |
| Fields of View:            |        | 1-18 cm for 3MHz                                                                               |                          |         |            |
| External Video Output:     |        | No                                                                                             |                          |         |            |
| Liquid-Crystal Display:    |        | 18 bit, 262,000 Color, Active Matrix TFT LCD                                                   |                          |         |            |
| Materials                  |        | Sabic Cycoloy HC1204HF, Mitsui TPX-MED18, Sabic Versollen<br>OMX1255NX-1                       |                          |         |            |
| Size: -                    |        |                                                                                                |                          |         |            |
| Width:                     |        | 6.8 cm                                                                                         |                          |         |            |
| Height:                    |        | 11.5 cm                                                                                        |                          |         |            |
| Depth:                     |        | 2.0 cm                                                                                         |                          |         |            |
| Weight:                    |        | 0.13 kg                                                                                        |                          |         |            |
| Electrical                 |        |                                                                                                |                          |         |            |
| External Power:            | Input: | 100-240 VAC,<br>50-60Hz                                                                        |                          | Output: | 5 VDC @ 2A |
| Battery:                   |        | Li-Ion battery pack (2 Whr)                                                                    |                          |         |            |
| Leakage Current:           |        | 10 $          μ  $ A maximum                                                                   |                          |         |            |
| Primary Breakdown Voltage: |        | 3000VAC                                                                                        |                          |         |            |
| Safety Standards:          |        | IEC 60601-1:2005, ES60601-1:2005, IEC 60601-2-37:2007, IEC<br>60601-1-2:2007, ISO 10993-1:2003 |                          |         |            |
| Protection Class:          |        | Class II: per IEC 60601-1                                                                      |                          |         |            |
| Degree of Protection:      |        |                                                                                                | Type BF: per IEC 60601-1 |         |            |

| Environmental            |                                                                     |
|--------------------------|---------------------------------------------------------------------|
| Mechanical Shock:        | Drop and push testing per ES60601-1                                 |
| Mechanical Vibration:    | Random Acceleration Profile per JIS Z 0232:2004 5Hz-300Hz           |
| Drop Test (to concrete): | 1 meter                                                             |
| Operating Temperature:   | 0 to 40 C                                                           |
| Humidity:                | 20 to 80% RH, non-condensing                                        |
| Water Resistance:        | Transducer IPX7 lens, IPX1 probe degree of protection against water |
| Altitude:                | 0.7 – 1.05 standard atmospheres (2500m or 8200 feet) operating      |
| Storage                  |                                                                     |
| Temperature:             | -20 to 45 C                                                         |
| Humidity:                | 10 to 90% RH, non-condensing                                        |
| Altitude:                | 0.5 – 1.05 standard atmospheres storage                             |

{8}------------------------------------------------

## 12. Non-clinical Performance Data

The SignosRT Bladder has been bench tested for imaging performance and measurement accuracy, with tests showing the SignosRT Bladder imaging performance and measurement accuracy to be substantially equivalent to the predicate devices. The measured lateral and axial resolution for the SignosRT Bladder is identical to the predicate SpeqRT Personal Ultrasound System. The measured volume accuracy on a phantom was a maximum of +-9.6%. well within the specified +-15%.

The SignosRT Bladder Scanner device has been tested by independent laboratories to IEC 60601-1:2005, IEC 60601-2-37:2007, IEC 60601-1-2:2007, ISO 10993-5:2009, ISO 10993-10:2002, ISO 10993-10:2010, ISO 10993-12:2007, NEMA UD-2-2004 (R2009), NEMA UD-3-2004(R2009) and found to comply with all standards.

The software and firmware in the SignosRT Bladder has been developed and verified according to IEC 62304:2006. The verification reports (Appendix R and V), traceability (Appendix Q), and risk analysis (Appendix A) demonstrate the SignosRT Bladder operates as intended and risks mitigated in firmware have been verified.

The conclusion from the testing is the device is safe and effective for its intended use, and performs as well or better than the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K143493](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K143493)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
