← Product Code [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO) · K143122

# Volcano iFR Harvest Software (Offline) (K143122)

_Volcano Corporation · IYO · Feb 20, 2015 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K143122

## Device Facts

- **Applicant:** Volcano Corporation
- **Product Code:** [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO.md)
- **Decision Date:** Feb 20, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

The Volcano iFR® Harvest Modality is an offline analysis tool of the currently marketed vasodilator-free pressure measurement module, iFR® Modality of the Volcano s5/s5i/CORE and CORE Mobile systems. It is intended to calculate iFR values from hyperemia-free FFR pressure recordings in a laptop or desktop setting only.

## Device Story

Software tool for offline analysis of hyperemia-free FFR pressure recordings; inputs existing FFR case files collected/archived on Volcano s5/s5i consoles; separates baseline (iFR) and Pd/Pa pressure runs; calculates iFR values; generates summary report. Used in laptop/desktop settings by physicians to analyze pressure measurements; facilitates clinical assessment of coronary/peripheral vasculature; provides quantitative data for decision-making.

## Clinical Evidence

Bench testing only. Performance data included software verification and software validation to demonstrate equivalence to the predicate device.

## Technological Characteristics

Software-based offline analysis tool. Operates on laptop/desktop platforms. Utilizes identical pressure measurement algorithms as the predicate Volcano iFR Modality. Connectivity via import of archived FFR case files from Volcano s5/s5i consoles.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Volcano iFR Modality ([K133323](/device/K133323.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white and has a simple, clean design. The caduceus symbol is a common symbol for healthcare and medicine.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W 066-G609 Silver Spring, MD 20993-0002

February 20, 2015

Volcano Corporation Neeta Sharma Director, Regulatory Affairs 3721 Valley Centre Drive, Suite 500 San Diego, CA 92130

Re: K143122 Trade/Device Name: Volcano Harvest iFR Modality Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic Pulsed Echo Imaging System Regulatory Class: Class II Product Code: IYO Dated: January 20, 2015 Received: January 22, 2015

Dear Ms. Sharma,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Melissa A. Torres -S

For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use Statement

5l0(k) Number (if known) K143122

Page 1 of 1

Volcano Harvest iFR® Modality Device Name

## Indications for Use

The Volcano iFR® Harvest Modality is an offline analysis tool of the currently marketed vasodilator-free pressurement module, iFR® Modality of the Volcano s5/s5i/CORE and CORE Mobile systems. It is intended to calculate iFR values from hyperemia-free FFR pressure recordings in a laptop or desktop setting only.

X_____________________________________________________________________________________________________________________________________________________________________________ Prescription Use (Per 21 CFR 801.109)

OR

Over the Counter Use _________________________________________________________________________________________________________________________________________________________

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

# 510 (k) SUMMARY

| SPONSOR:            | Volcano Corporation<br>3721 Valley Center Drive<br>San Diego, CA 92130                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT/SUBMITTER:  | Neeta Sharma<br>Director, Regulatory Affairs<br>Volcano Corporation<br>3721 Valley Center Drive<br>San Diego, CA 92130<br>Tel: (858) 720-4187<br>Fax :( 858) 720-0335                                                                                                                                                                                                                                                                                                                                                               |
| DATE PREPARED:      | February 18, 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| DEVICE:             | Volcano Harvest iFR® Modality                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| TRADE NAME:         | Volcano Harvest iFR® Modality                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| COMMON NAME:        | Ultrasonic Pulsed Echo Imaging System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| CLASSIFICATION:     | 21 CFR Part 892.1560<br>IYO: System, Imaging, Pulsed Echo, Ultrasonic<br>Class II Device                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| PREDICATE DEVICE:   | Volcano iFR Modality (K133323)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| DEVICE DESCRIPTION: | The Volcano Harvest iFR® Modality is a software tool<br>designed to calculate offline iFR values from hyperemia-<br>free FFR pressure recordings made using Volcano's s5<br>v3.2.x software. It takes an existing FFR case file, separates<br>the baseline (iFR) and Pd/Pa (FFR under resting<br>conditions) pressure runs, calculates the iFR values from<br>the baseline runs, and prepares a report summarizing the<br>values. Input case files are collected and archived on the s5<br>or s5i console from Volcano Corporation. |
| INTENDED USE:       | The Volcano iFR® Harvest Modality is an offline analysis<br>tool of the currently marketed vasodilator-free pressure<br>measurement module, iFR® Modality of the Volcano<br>s5/s5i/CORE and CORE Mobile systems. It is intended to<br>calculate iFR values from hyperemia-free FFR pressure<br>recordings in a laptop or desktop setting only.                                                                                                                                                                                      |

{4}------------------------------------------------

## COMPARISON OF CHARACTERISTICS:

The Harvest iFR® Modality is an offline version of the iFR Modality with similar in intended use that both software are designed to measure blood pressure in the coronary and peripheral vasculature. Both softwares have identical technological features. The Harvest iFR® Modality software allows for an offline analysis of the pressure measurement readings to be done if desired by the physician in a laptop or desktop setting.

## PERFORMANCE DATA:

Performance testing completed for a determination of substantial equivalence included the following:

- . Software Verification
- . Software Validation

The results of the performance data demonstrate equivalence to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K143122](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K143122)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
