← Product Code [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO) · K093371

# MODEL DGH 6000 SCANMATE A (K093371)

_Dgh Technology, Inc. · IYO · Apr 5, 2010 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K093371

## Device Facts

- **Applicant:** Dgh Technology, Inc.
- **Product Code:** [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO.md)
- **Decision Date:** Apr 5, 2010
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The intended use of the DGH 6000 is the measurement of AL, ACD, LT of the human eye. The DGH 6000 is also intended to calculate the optical power of an IOL that is to be implanted during cataract surgery. The DGH 6000 is intended to be used solely by qualified medical professionals. Clinical consideration and judgment should be applied when using the DGH 6000.

## Device Story

DGH 6000 Scanmate A is a USB-connected A-Mode pulsed echo ultrasound biometer. Device captures ultrasonic echoes from human eye structures; processes signals to measure axial length, anterior chamber depth, and lens thickness. Software integrates established IOL calculation formulas to determine required IOL power for cataract surgery. Operated by qualified medical professionals in clinical settings; output displayed on PC monitor. Healthcare providers use measurements and calculated IOL values to guide surgical planning. Benefits include accurate ocular biometry for improved IOL selection during cataract procedures.

## Clinical Evidence

No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including phantom comparative tests and adherence to electrical safety and acoustic output standards.

## Technological Characteristics

A-Mode pulsed echo ultrasound; USB-connected transducer (DGH 6006); PC-based display. Measures AL, ACD, LT. Software includes IOL calculation formulas. Complies with EN 60601-1, EN 60601-1-2, NEMA UD-2, and NEMA UD-3.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Quantel Medical AXIS II Ultrasonic Biometer ([K000554](/device/K000554.md))
- DGH 3000A Ultrasonic A-Scan ([K872726](/device/K872726.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# TECHNOLOGY, INC.

## 510(k) Summary (21 CFR 807.92(c)) 510(k) K093371

Submitter:

Contact Person:

DGH Technology, Inc. 110 Summit Drive, Suite B Exton, PA 19341

M. Luther Detweiler Vice President Regulatory Affairs and Quality Assurance

Telephone: 610-594-9100 Fax: 610-594-0390 Email: Lou@dghkoi.com

Date Summary Prepared:

March 12th, 2010

Device Trade Name:

DGH 6000 Scanmate A.

Device Common Name:

Ultrasound A-Scan biometer.

Device Classification:

Device: System, Imaging, Pulsed Echo, Ultrasonic Panel: Radiology Product Code: IYO Device Class: II Regulation Number: 21 CFR 892.1560

Device: Diagnostic Ultrasonic Transducer Panel: Radiology Product Code: ITX Device Class: II Regulation Number: 21 CFR 892.1570

110 Summit Drive, Suite B Exton, PA 19341

PA 19341

Phone: (610) 594-9100 (800) 722-3883

Fax: (610) 594-0390

APR - 5 2010

.
.
.
.
.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for DGH Technology, Inc. The letters "DGH" are in large, bold, black font on the left side of the image. To the right of the letters is a stylized graphic that appears to be a horizontal line with curved extensions above and below it. Above the graphic, the words "TECHNOLOGY, INC." are printed in a smaller font.

Legally Marketed Predicate Device(s):

Description of Device:

The DGH 6000 A-Scan is a USB plug-in device that uses A-Mode, pulsed echo ultrasound technology to measure the axial length (AL), anterior chamber depth (ACD), and lens thickness (LT) of the human eye. The device includes formulas to calculate the implanted IOL power, using the axial length measurement.

Quantel Medical AXIS II Ultrasonic Biometer (K000554)

DGH 3000A Ultrasonic A-Scan (K872726)

Intended Use of the Device:

The intended use of the DGH 6000 is the measurement of AL, ACD, LT of the human eye. The DGH 6000 is also intended to calculate the optical power of an IOL that is to be implanted during cataract surgery. The DGH 6000 is intended to be used solely by qualified medical professionals. Clinical consideration and judgment should be applied when using the DGH 6000.

#### Technological Characteristics:

Axial length (AL), anterior chamber depth (ACD), and lens thickness (LT) measurements are obtained using the same technology as the predicate device, DGH 3000A. The technology is based on ultrasonic pulse echo technology, whereby short bursts of ultrasonic energy are transmitted and the resulting echoes are captured, amplified, filtered and processed. Specific time distances in the captured echo peaks are then measured and converted into distance information. Digital signal processing algorithms validate individual peaks and proper probe alignment. The resulting measurements and waveforms are then displayed on a PC monitor for the medical (ophthalmic) professional user. In addition, the software assists the user in calculating replacement intraocular lens values based on established IOL calculation formulas.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for DGH Technology, Inc. The letters "DGH" are in a bold, sans-serif font. To the right of the letters is a stylized graphic that resembles a wing or a stylized line. Above the graphic, the words "TECHNOLOGY, INC." are printed in a smaller font.

#### Performance Tests:

The following tests were performed to demonstrate substantial equivalence:

(A) Non-Clinical Tests

- . Comparative test block (phantom) tests
- . EN 60601-1:1990 Medical Electrical Equipment Part 1 General Requirements for Electrical Safety
- . EN 60601-1-2:2002 Medical Electrical Equipment Collateral Standard: Electromagnetic Compatibility
- . NEMA Standard Publication UD-2 2004: Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment
- . NEMA Standard Publication UD-3 2004: Standard for Real-Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment.

(B) Clinical Tests

Since the DGH 6000 Scanmate-A uses the same technology as existing devices, clinical tests are not required.

#### Conclusions:

We have reviewed the results of the performance tests and have determined that the DGH 6000 is substantially equivalent, in safety and efficacy, to the legally marketed predicate device.

Phone: (610) 594-9100 (800) 722-3883

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

### APR - 5 2010

DGH Technology, Inc. % Mr. Ned Devine Senior Staff Engineer Underwriters Laboratories, Inc. 333 Pfingsten Road NORTHBROOK IL 60062

Re: K093371

Trade/Device Name: DGH 6000 Scanmate A Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed Doppler imaging system Regulatory Class: II Product Code: IYO and ITX Dated: March 23, 2010 Received: March 24, 2010

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the DGH 6000 Scanmate A, as described in your premarket notification:

#### Transducer Model Number

#### DGH 6000

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Shahram Vaezy at (301) 796-6242.

Sincerely yours,

Donald J. St. Pierre Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure(s)

{5}------------------------------------------------

## Indications for Use

510(k) Number (if known): K093371

Device Name: DGH 6000 Scanmate A

Indications for Use: The intended use of the DGH 6000 is the measurement of AL. ACD, LT of the human eye. The DGH 6000 is also intended to calculate the optical power of an IOL that is to be implanted during cataract surgery. The DGH 6000 is intended to be used solely by qualified medical professionals. Clinical consideration and judgment should be applied when using the DGH 6000.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

KMR

(Division Sign-Off) Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510(k) Number_________________________________________________________________________________________________________________________________________________________________

Page 1 of 3

{6}------------------------------------------------

## . Attachment 3

## Indications for Use Form

.. ...........................................................................................................................................................................

13

510(k) K093371 System: DGH 6000 Scanmate A Transducer: DGH 6006

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                 | Mode of Operation |   |     |     |                  |                       |        |
|------------------------------|---------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------|
| General<br>(Track 1<br>Only) | Specific<br>(Tracks 1 & 3)      | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | A-Mode |
| Ophthalmic                   | Ophthalmic                      |                   |   |     |     |                  |                       | P      |
| Fetal Imaging<br>& Other     | Fetal                           |                   |   |     |     |                  |                       |        |
|                              | Abdominal                       |                   |   |     |     |                  |                       |        |
|                              | Intra-operative (Specify)       |                   |   |     |     |                  |                       |        |
|                              | Intra-operative (Neuro)         |                   |   |     |     |                  |                       |        |
|                              | Laparoscopic                    |                   |   |     |     |                  |                       |        |
|                              | Pediatric                       |                   |   |     |     |                  |                       |        |
|                              | Small Organ (Specify)           |                   |   |     |     |                  |                       |        |
|                              | Neonatal Cephalic               |                   |   |     |     |                  |                       |        |
|                              | Adult Cephalic                  |                   |   |     |     |                  |                       |        |
|                              | Trans-rectal                    |                   |   |     |     |                  |                       |        |
|                              | Trans-vaginal                   |                   |   |     |     |                  |                       |        |
|                              | Trans-urethral                  |                   |   |     |     |                  |                       |        |
|                              | Trans-esoph. (non-Card.)        |                   |   |     |     |                  |                       |        |
|                              | Musculo-skeletal (Conventional) |                   |   |     |     |                  |                       |        |
|                              | Musculo-skeletal (Superficial)  |                   |   |     |     |                  |                       |        |
|                              | Intravascular                   |                   |   |     |     |                  |                       |        |
|                              | Other (Specify)                 |                   |   |     |     |                  |                       |        |
| Cardiac                      | Cardiac Adult                   |                   |   |     |     |                  |                       |        |
|                              | Cardiac Pediatric               |                   |   |     |     |                  |                       |        |
|                              | Intravascular (Cardiac)         |                   |   |     |     |                  |                       |        |
|                              | Trans-esoph. (Cardiac)          |                   |   |     |     |                  |                       |        |
|                              | Intra-cardiac                   |                   |   |     |     |                  |                       |        |
|                              | Other (Specify)                 |                   |   |     |     |                  |                       |        |
| Peripheral                   | Peripheral vessel               |                   |   |     |     |                  |                       |        |
| Vessel                       | Other (Specify)                 |                   |   |     |     |                  |                       |        |

N = new indication; P = previously cleared by FDA; E = added under this appendix

(Division Sign-Off)
Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510K K093371

{7}------------------------------------------------

## Indications for Use Form

#### 510(k) K093371 Transducer: DGH 6006

Intended Use: The intended use of the DGH 6006 transducer is the measurement of AL, ACD, LT of the human eye. The DGH 6006 transducer is intended to be used solely by qualified medical professionals. Clinical consideration and judgment should be applied when using the DGH 6006 transducer.

| Clinical Application         |                                 | Mode of Operation |   |     |     |                  |                       |        |  |
|------------------------------|---------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------|--|
| General<br>(Track 1<br>Only) | Specific<br>(Tracks 1 & 3)      | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify) | A-Mode |  |
| Ophthalmic                   | Ophthalmic                      |                   |   |     |     |                  |                       | P      |  |
| Fetal Imaging<br>& Other     | Fetal                           |                   |   |     |     |                  |                       |        |  |
|                              | Abdominal                       |                   |   |     |     |                  |                       |        |  |
|                              | Intra-operative (Specify)       |                   |   |     |     |                  |                       |        |  |
|                              | Intra-operative (Neuro)         |                   |   |     |     |                  |                       |        |  |
|                              | Laparoscopic                    |                   |   |     |     |                  |                       |        |  |
|                              | Pediatric                       |                   |   |     |     |                  |                       |        |  |
|                              | Small Organ (Specify)           |                   |   |     |     |                  |                       |        |  |
|                              | Neonatal Cephalic               |                   |   |     |     |                  |                       |        |  |
|                              | Adult Cephalic                  |                   |   |     |     |                  |                       |        |  |
|                              | Trans-rectal                    |                   |   |     |     |                  |                       |        |  |
|                              | Trans-vaginal                   |                   |   |     |     |                  |                       |        |  |
|                              | Trans-urethral                  |                   |   |     |     |                  |                       |        |  |
|                              | Trans-esoph. (non-Card.)        |                   |   |     |     |                  |                       |        |  |
|                              | Musculo-skeletal (Conventional) |                   |   |     |     |                  |                       |        |  |
|                              | Musculo-skeletal (Superficial)  |                   |   |     |     |                  |                       |        |  |
|                              | Intravascular                   |                   |   |     |     |                  |                       |        |  |
|                              | Other (Specify)                 |                   |   |     |     |                  |                       |        |  |
| Cardiac                      | Cardiac Adult                   |                   |   |     |     |                  |                       |        |  |
|                              | Cardiac Pediatric               |                   |   |     |     |                  |                       |        |  |
|                              | Intravascular (Cardiac)         |                   |   |     |     |                  |                       |        |  |
|                              | Trans-esoph. (Cardiac)          |                   |   |     |     |                  |                       |        |  |
|                              | Intra-cardiac                   |                   |   |     |     |                  |                       |        |  |
|                              | Other (Specify)                 |                   |   |     |     |                  |                       |        |  |
| Peripheral                   | Peripheral vessel               |                   |   |     |     |                  |                       |        |  |
| Vessel                       | Other (Specify)                 |                   |   |     |     |                  |                       |        |  |

N = new indication; P = previously cleared by FDA; E = added under this appendix

(Division Sign-Off)

Division of Radiological Office of In Vitro Diagnostic Device Ev

510K K093371

3/3

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K093371](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K093371)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
