← Product Code [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO) · K081110

# SIUI CTS-900 DIGITAL ULTRASOUND IMAGING SYSTEM (K081110)

_Shantou Institute of Ultrasonic Instuments · IYO · May 16, 2008 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K081110

## Device Facts

- **Applicant:** Shantou Institute of Ultrasonic Instuments
- **Product Code:** [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO.md)
- **Decision Date:** May 16, 2008
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

Ultrasonic pulsed echo imaging and measurement for abdominal, pediatric, small organ, cardiac, peripheral vascular and muscular-skeletal applications.

## Device Story

Digital diagnostic ultrasound system; utilizes convex (C3L60B) and linear (L7L38B) transducers to acquire acoustic signals; processes signals via 32-channel digital beam-former; supports B-mode and B/M-mode imaging. System features include cine review, image zoom, measurements, and calculations. Operated by clinicians in clinical settings; output displayed on integrated monitor for real-time visualization and diagnostic assessment. Facilitates clinical decision-making through anatomical imaging and quantitative measurements (distance, area, circumference, angle, heart rate). Benefits patients by providing non-invasive diagnostic imaging for various organ systems and vascular structures.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance verified through technical specifications, acoustic output measurements, and comparison of imaging capabilities to predicate devices.

## Technological Characteristics

Digital ultrasound system; 32-channel digital beam-former; 256 gray scale; convex (2.5-5.0 MHz) and linear (5.0-9.0 MHz) transducers; dynamic receive focus; 60GB HDD storage; USB interface; B&W video printer output; software-based image processing (zoom, orientation, cine loop).

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- SIUI CTS-7700 ([K061083](/device/K061083.md))
- SIUI CTS-7700 ([K070982](/device/K070982.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K081110

SIUI CTS-900 Digital Ultrasound Imaging System

MAY 1 6 2008

# Section 7 510(k) Summary of Safety and Effectiveness

| Trade Name:             | CTS-900 with C3L60B and L7L38B Transducers                                                                                                            |
|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name:            | Digital Ultrasound Imaging System                                                                                                                     |
| Classification Name:    | Ultrasonic Pulsed Echo Imaging System, IYO<br>(Per 21 CFR section 892.1560)<br>Diagnostic Ultrasonic Transducer, ITX<br>(Per 21 CFR section 892.1570) |
| Manufacturer's Name:    | Shantou Institute of Ultrasonic Instruments                                                                                                           |
| Address:                | #77, Jinsha Road,<br>Shantou, Guangdong 515041, China                                                                                                 |
| Corresponding Official: | Flower Cai                                                                                                                                            |
| Title:                  | Overseas Cooperation Supervisor                                                                                                                       |
| Telephone:              | (86) 754- 8250150 Fax: (86) 754-8251499                                                                                                               |
| US Agent:               | Bob Leiker/ Consultant /QRS<br>Dublin, CA 94568<br>Telephone: 1-925-556-1302 Fax: 1-866-718-3819                                                      |
| Predicate Device:       | SIUI CTS-7700 , K061083 and K070982                                                                                                                   |

Device Description: The SIUI CTS-900 is a digital diagnostic ultrasound system capable of the following operating modes: 2D (B mode) and B/M. The system is designed for use in linear and convex scanning modes and supports linear, and convex transducers. The system has cine review, image zoom, measurements and calculations, image storage and review, printing and recording capabilities. The system consists of probes, main unit, control panel and monitor.

Intended Use: Ultrasonic pulsed echo imaging and measurement for abdominal, pediatric, small organ, cardiac, peripheral vascular and muscular-skeletal applications.

## Technological Characteristics:

- 1) Scanning modes: convex and linear scanning
- 2) Display modes:
	- a) B-Mode (B, 2B)
	- b) B/M-Mode

{1}------------------------------------------------

- 3) Supporting transducers:
	- a) C3L60B: 2.5-5.0 MHz 60R 128e convex transducer
	- b) L7L38B: 5.0-9.0 MHz 38mm 128e linear transducer
- 4) Focus mode:
	- a) Transmit focus mode:
		- 1-4 selectable, focus depth: variable
	- b) Receive focus mode: dynamic focus
- 5) Grey scale: 256
- 6) Pre-processing:
	- 32-channel digital beam-former: a)
	- Receive gain (include TGC): 70dB b)
	- Dynamic range: 35-66dB c)
	- Edge enhancement: 4 steps d)
	- Image persistence: 7 steps e)
	- Line density: normal, high f)
- 7) Post-processing

10 types of gray maps, among which 4 types are user-definable

- 8) Image manipulation:
	- a) Real-time zoom in x4.0 max.
	- b) Frozen image
- 9) B/M-mode speed:

Time for full screen scroll: 1.2, 2.5, 5.0, 10.0 sec

- 10)Cine: Max. 256 frames
- 11)Image store and recall: 32 frames
- 12)Image orientation:
	- c) Left/right flip
	- d) Up/down flip
	- e) 90-degree rotation (selectable steps: 0, 90, 180. 270 degrees)

13)Documentation and storage:

- a) 60GB HDD, images stored in BMP file format;
- b) USB interface memory, images stored in BMP file format
- c) Documentation devices:
- d) B&W video printer

14)Measurements and calculations

- a) General measurements and calculations 2D: Distance, Area, Circumference, and Angle M-Mode: Distance, Time, Slope, Heart rate
- b) Specific measurements and calculations Abdomen, Obstetrics, Gynecology, Cardiology, Small parts, Peripheral Vascular.

- 510(k) Summary of Safety and Effectiveness
Page 2 of 2

{2}------------------------------------------------

Public Health Service

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Shantou Institute of Ultrasonic Instruments (SIUI) % Mr. Bob Leiker Consultant Quality & Regulatory Services 7263 Cronin Circle DUBLIN CA 94568

Re: K081110

Trade/Device Name: SIUI CTS-900 Digital Ultrasound Imaging System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO and ITX Dated: April 15, 2008 Received: April 18, 2008

Dear Mr. Leiker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

MAY 1 6 2008

This determination of substantial equivalence applies to the following transducers intended for use with the SISU CTS-900 Digital Ultrasound Imaging System as described in your premarket notification:

#### Transducer Model Number

#### Convex Array C3L60B Linear Array L7L38B

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 

{3}------------------------------------------------

CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (cnclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

If you have any questions regarding the content of this letter, please contact Sundar Rajan at (240) 276-3666.

Sincerely yours.

Vorui M. Whag

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{4}------------------------------------------------

### SIUI CTS-900 Digital Ultrasound Imaging System Diagnostic Ultrasound Indications for Use Form

# 1. System Indications for Use Form

Device Name: CTS-900

| Mode of Operation    |   |    |   |     |     |         |             |          |           |           |
|----------------------|---|----|---|-----|-----|---------|-------------|----------|-----------|-----------|
| Clinical Application | A | B  | M | PWD | CWD | Color   | Power       | Color    | Combined  | Other     |
|                      |   |    |   |     |     | Doppler | (Amplitude) | Velocity | (Specify) | (Specify) |
|                      |   |    |   |     |     |         | Doppler     | Imaging  |           |           |
| Opthalmic            |   |    |   |     |     |         |             |          |           |           |
| Fetal                |   | P1 |   |     |     |         |             |          | P1        |           |
| Abdominal            |   | P1 |   |     |     |         |             |          | P1        |           |
| IntraOperative       |   |    |   |     |     |         |             |          |           |           |
| (Cardiac)            |   |    |   |     |     |         |             |          |           |           |
| IntraOperative       |   |    |   |     |     |         |             |          |           |           |
| Neurological         |   |    |   |     |     |         |             |          |           |           |
| Pediatric            |   | P1 |   |     |     |         |             |          | P1        |           |
| Small Organ          |   | P1 |   |     |     |         |             |          | P1        |           |
| (Specify)            |   |    |   |     |     |         |             |          |           |           |
| Neonatal Cephalic    |   |    |   |     |     |         |             |          |           |           |
| Adult Cephalic       |   |    |   |     |     |         |             |          |           |           |
| Cardiac              |   | P1 |   |     |     |         |             |          | P1        |           |
| Transesophageal      |   |    |   |     |     |         |             |          |           |           |
| Trans-Rectal         |   |    |   |     |     |         |             |          |           |           |
| Trans-Vaginal        |   |    |   |     |     |         |             |          |           |           |
| Trans-Urethral       |   |    |   |     |     |         |             |          |           |           |
| Intra-Vascular       |   |    |   |     |     |         |             |          |           |           |
| Peripheral Vascular  |   | P1 |   |     |     |         |             |          | P1        |           |
| Laparascopic         |   |    |   |     |     |         |             |          |           |           |
| Muscular-Skeletal    |   | P2 |   |     |     |         |             |          | P2        |           |
| Conventional         |   |    |   |     |     |         |             |          |           |           |
| Muscular-Skeletal    |   | P2 |   |     |     |         |             |          | P2        |           |
| Superficial          |   |    |   |     |     |         |             |          |           |           |
| Others (Specify)     |   | P1 |   |     |     |         |             |          | P1        |           |

N=new indication

:

E=added under Appendix E

P1=previously cleared by FDA (K061083)

E=added under Appendix E

P2=previously cleared by FDA (K070982)

Additional Comments: Small organs include: thyroid, testes, breast Combined: B/M Mode

Other uses include: Uterus, Ovary, and Prostate

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, office of Device Evaluation(((Division Sign-Off)

y

Division of Reproductive, Abdominal and Radiological Devices 510(k) Number _______________________________________________________________________________________________________________________________________________________________ KO81110

:ription Use (Per 21 CFR 801.109)

ગા 8

Indications For Use

{5}------------------------------------------------

### SIUI CTS-900 Digital Ultrasound Imaging System

#### Diagnostic Ultrasound Indications for Use Form

## 2. Transducer Indications for Use Form

Device Name: Convex Array C3L60B

## Mode of Operation

| Clinical Application | A | B  | M | PWD | CWD | Color   | Power       | Color    | Combined  | Other     |
|----------------------|---|----|---|-----|-----|---------|-------------|----------|-----------|-----------|
|                      |   |    |   |     |     | Doppler | (Amplitude) | Velocity | (Specify) | (Specify) |
|                      |   |    |   |     |     |         | Doppler     | Imaging  |           |           |
| Opthalmic            |   |    |   |     |     |         |             |          |           |           |
| Fetal                |   | P1 |   |     |     |         |             |          | P1        |           |
| Abdominal            |   | P1 |   |     |     |         |             |          | P1        |           |
| IntraOperative       |   |    |   |     |     |         |             |          |           |           |
| (Cardiac)            |   |    |   |     |     |         |             |          |           |           |
| IntraOperative       |   |    |   |     |     |         |             |          |           |           |
| Neurological         |   |    |   |     |     |         |             |          |           |           |
| Pediatric            |   |    |   |     |     |         |             |          |           |           |
| Small Organ          |   |    |   |     |     |         |             |          |           |           |
| (Specify)            |   |    |   |     |     |         |             |          |           |           |
| Neonatal Cephalic    |   |    |   |     |     |         |             |          |           |           |
| Adult Cephalic       |   |    |   |     |     |         |             |          |           |           |
| Cardiac              |   | P1 |   |     |     |         |             |          | P1        |           |
| Transesophageal      |   |    |   |     |     |         |             |          |           |           |
| Trans-Rectal         |   |    |   |     |     |         |             |          |           |           |
| Trans-Vaginal        |   |    |   |     |     |         |             |          |           |           |
| Trans-Urethrai       |   |    |   |     |     |         |             |          |           |           |
| Intra-Vascular       |   |    |   |     |     |         |             |          |           |           |
| Peripheral Vascular  |   |    |   |     |     |         |             |          |           |           |
| Laparascopic         |   |    |   |     |     |         |             |          |           |           |
| Muscular-Skeletal    |   |    |   |     |     |         |             |          |           |           |
| Conventional         |   |    |   |     |     |         |             |          |           |           |
| Muscular-Skeletal    |   |    |   |     |     |         |             |          |           |           |
| Superficial          |   |    |   |     |     |         |             |          |           |           |
| Others (Specify)     |   | P1 |   |     |     |         |             |          | P1        |           |

N=new indication

E=added under Appendix E

P1=previously cleared by FDA (K061083) P2=previously cleared by FDA (K070982)

Additional Comments: other uses include: Uterus, Ovary and Prostate Combined: BIM Mode

PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, office of Device Evaluation (ODE)

Jorge M. Ortiz

Division of Reproductive, Abdominal and Radiological Devices Kog 1110 510(k) Number ________________________________________________________________________________________________________________________________________________________________

Prescription Use (Per 21 CFR 801.109) Section 8

Indications For Use

Page 3 of 4

{6}------------------------------------------------

# SIUI CTS-900 Digital Ultrasound Imaging System

### Diagnostic Ultrasound Indications for Use Form

## 3. Transducer Indications for Use Form

Device Name: Linear Array L7L38B

| Clinical Application              | A | B  | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
|-----------------------------------|---|----|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|--------------------|
| Opthalmic                         |   |    |   |     |     |                  |                                 |                              |                       |                    |
| Fetal                             |   |    |   |     |     |                  |                                 |                              |                       |                    |
| Abdominal                         |   |    |   |     |     |                  |                                 |                              |                       |                    |
| IntraOperative<br>(Cardiac)       |   |    |   |     |     |                  |                                 |                              |                       |                    |
| IntraOperative<br>Neurological    |   |    |   |     |     |                  |                                 |                              |                       |                    |
| Pediatric                         |   | P1 |   |     |     |                  |                                 |                              | P1                    |                    |
| Small Organ<br>(Specify)          |   | P1 |   |     |     |                  |                                 |                              | P1                    |                    |
| Neonatal Cephalic                 |   |    |   |     |     |                  |                                 |                              |                       |                    |
| Adult Cephalic                    |   |    |   |     |     |                  |                                 |                              |                       |                    |
| Cardiac                           |   |    |   |     |     |                  |                                 |                              |                       |                    |
| Transesophageal                   |   |    |   |     |     |                  |                                 |                              |                       |                    |
| Trans-Rectal                      |   |    |   |     |     |                  |                                 |                              |                       |                    |
| Trans-Vaginal                     |   |    |   |     |     |                  |                                 |                              |                       |                    |
| Trans-Urethral                    |   |    |   |     |     |                  |                                 |                              |                       |                    |
| Intra-Vascular                    |   |    |   |     |     |                  |                                 |                              |                       |                    |
| Peripheral Vascular               |   | P1 |   |     |     |                  |                                 |                              | P1                    |                    |
| Laparascopic                      |   |    |   |     |     |                  |                                 |                              |                       |                    |
| Muscular-Skeletal<br>Conventional |   | P2 |   |     |     |                  |                                 |                              | P2                    |                    |
| Muscular-Skeletal<br>Superficial  |   | P2 |   |     |     |                  |                                 |                              | P2                    |                    |
| Others (Specify)                  |   |    |   |     |     |                  |                                 |                              |                       |                    |

#### Mode of Operation

N=new indication

E=added under Appendix E

P1=previously cleared by FDA (K061083) P2=previously cleared by FDA (K070982) Additional Comments: Small organs include: thyroid, testes, breast Combined: B/M Mode

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, office of Device Evaluation (ODE)

Ibrú Mthz

(Division Sian-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number Page 4 of 4

Prescription Use (Per 21 CFR 801.109) Section 8

Indications For Use

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K081110](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K081110)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
