← Product Code [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO) · K070982

# CTS-7700 DIGITAL ULTRASOUND IMAGING SYSTEM (K070982)

_Shantou Institute of Ultrasonic Instuments · IYO · May 2, 2007 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K070982

## Device Facts

- **Applicant:** Shantou Institute of Ultrasonic Instuments
- **Product Code:** [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO.md)
- **Decision Date:** May 2, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

Ultrasonic pulsed echo imaging and measurement for abdominal, pediatric, small organ, cardiac, peripheral vascular applications

## Device Story

Digital diagnostic ultrasound system; utilizes convex and linear transducers to perform B-mode and B/M-mode scanning. Inputs: ultrasonic pulsed echo signals from patient anatomy. Processing: 32-channel digital beam-former; dynamic receive focusing; pre-processing (gain, TGC, edge enhancement, persistence); post-processing (gray maps). Outputs: real-time 2D images on monitor; cine loops; measurements (distance, area, circumference, angle, time, slope, heart rate). Used in clinical settings by healthcare professionals. Benefits: provides diagnostic visualization and quantitative measurements for various clinical applications, including muscular-skeletal, abdominal, and small organ assessment.

## Clinical Evidence

Bench testing only. No clinical data provided.

## Technological Characteristics

Digital ultrasound system; 32-channel digital beam-former; 256 gray levels. Transducers: C3L60B (2.5-5.0 MHz convex) and L7L38B (5.0-9.0 MHz linear). Modes: B, 2B, B/M. Connectivity: USB, parallel port, B&W video printer. Storage: 60GB HDD, BMP format. Software: embedded system for image processing, measurement, and cine review.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- SonoAce 6000 ([K981510](/device/K981510.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K070982

SIUI CTS-7700 Digital Ultrasound System - Muscular-Skeletal Conventional and Superficial

MAY - 2 2007

### Tab 9 PREMARKET NOTIFICATION [510(k)] Summary

The CTS-7700 can be used for with Muscular-Skeletal Conventional and Superficial Superficial applications without any system or transducer modifications.

Nothing described in our previous submission, K061083 has been changed or modified for this submission.

All of the sections in this Tab 9 Premarket Notification [510(k)] Summary remain unchanged in this submission.

{1}------------------------------------------------

# Tab 9 PREMARKET NOTIFICATION [510(k)] Summary

| Trade Name:             | CTS-7700 with C3L60B and L7L38B Transducers                                    |                       |
|-------------------------|--------------------------------------------------------------------------------|-----------------------|
| Common Name:            | Digital Ultrasound Imaging System                                              |                       |
| Classification Name:    | Ultrasonic Pulsed Echo Imaging System, 90 IYO<br>(per 21 CFR section 892.1560) |                       |
| Manufacturer's Name:    | Shantou Institute of Ultrasonic Instruments                                    |                       |
| Address:                | #77, Jinsha Road,                                                              |                       |
|                         | Shantou SEZ, 515041, China                                                     |                       |
| Corresponding Official: | Li Delai                                                                       |                       |
| Title:                  | President                                                                      |                       |
| Telephone:              | (86) 754-8250150                                                               | Fax: (86) 754-8251499 |
| US Agent:               | Bob Leiker/ Consultant /QRS                                                    |                       |
|                         | 7263 Cronin Circle, Dublin, CA 94568                                           |                       |
|                         | Telephone: 1-925-556-1302                                                      | Fax: 1-866-718-3819   |

SonoAce 6000, K981510 Predicate Device:

#### Device Description:

The SIUI CTS-7700 is a digital diagnostic ultrasound system capable of the following operating modes: 2D (B mode) and B/M. The system is designed for use in linear and convex scanning modes and supports linear, and convex transducers. The system has cine review, image zoom, measurements and calculations, image storage and review, printing and recording capabilities. The system consists of probes, main unit, control panel and monitor.

Intended Use: Ultrasonic pulsed echo imaging and measurement for abdominal, pediatric, small organ, cardiac, peripheral vascular applications

{2}------------------------------------------------

#### Technological Characteristics:

- 1) Scanning modes: convex and linear scanning
- 2) Display modes:
	- a) B-Mode (B, 2B)
	- b) B/M-Mode
- 3) Supporting transducers:
	- C3L60B: 2.5-5.0 MHz 60R 128e convex transducer a)
	- b) L7L38B: 5.0-9.0 MHz 38mm 128e linear transducer
- 4) Focus mode:
	- a) Transmit focus mode: 1-4 selectable, focus depth: variable
	- Receive focus mode: dynamic focus ﻘ
- 5) Grey scale: 256
- 6) Pre-processing:
	- a) 32-channel digital beam-former;
	- Receive gain (include TGC): 70dB b)
	- C) Dynamic range: 35-66dB
	- Edge enhancement: 4 steps d)
	- Image persistence: 7 steps e)
	- Line density: normal, high ਿ
- 7) Post-processing
	- 10 types of gray maps, among which 4 types are user-definable
- 8) Image manipulation:
	- a) Real-time zoom in x4.0 max.
	- b) Frozen image
- 9) B/M-mode speed:

Time for full screen scroll: 1.2, 2.5, 5.0, 10.0 sec

- 10) Cine: Max. 256 frames
- 11) Image store and recall: 32 frames
- 12) Image orientation:
	- a) Left/right flip
	- b) Up/down flip
	- 90-degree rotation (selectable steps: 0, 90, 180, 270 degrees) C)
- 13) Documentation and storage:
	- a) 60GB HDD, images stored in BMP file format:
	- USB interface memory, images stored in BMP file format b)
	- Documentation devices: C)
	- B&W video printer ರ)
	- e) Parallel port printer (Inkjet or LaserJet)
- 14) Measurements and calculations
	- a) General measurements and calculations 2D: Distance, Area, Circumference, and Angle M-Mode: Distance, Time, Slope, Heart rate
	- b) Specific measurements and calculations Abdomen, Obstetrics, Gynecology, Cardiology, Small parts, Peripheral Vascular

{3}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Shantou Institute of Ultrasonic Instruments % Mr. Bob Leiker Consultant QRS 7263 Cronin Circle DUBLIN CA 94568

MAY - 2 2007

Re: K070982

Trade Name: CTS-7700 Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: IYO and ITX Dated: April 4, 2007 Received: April 6, 2007

Dear Mr. Leiker:

We have reviewed your Section 510/k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the CTS-7700, as described in your premarket notification:

Transducer Model Number

Convex Array C3L60B Linear Array L7L38B

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any

{4}------------------------------------------------

Page 2 - Mr. Leiker

Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded.

The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to:

> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

If you have any questions regarding the content of this letter, please contact Ewa Czerska at (240) 276-3666.

Sincerely vours.

Nancy L. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health 

{5}------------------------------------------------

SIUI CTS-7700 Digital Ultrasound System - Muscular-Skeletal, Conventional and Superficial Imaging

#### Diagnostic Ultrasound Indications for Use Form

# 3.1 System Indications for Use Form

Device Name: CTS-7700

| Mode of Operation                 |   |   |   |     |     |                  |                                 |                              |                       |                    |
|-----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|--------------------|
| Clinical Application              | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Opthalmic                         |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Fetal                             |   | P |   |     |     |                  |                                 |                              | P                     |                    |
| Abdominal                         |   | P |   |     |     |                  |                                 |                              | P                     |                    |
| IntraOperative<br>(Cardiac)       |   |   |   |     |     |                  |                                 |                              |                       |                    |
| IntraOperative<br>Neurological    |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Pediatric                         |   | P |   |     |     |                  |                                 |                              | P                     |                    |
| Small Organ<br>(Specify)          |   | P |   |     |     |                  |                                 |                              | P                     |                    |
| Neonatal Cephalic                 |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Adult Cephalic                    |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Cardiac (Adult)                   |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Cardiac (Pediatric)               |   | P |   |     |     |                  |                                 |                              | P                     |                    |
| Transesophageal                   |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Trans-Rectal                      |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Trans-Vaginal                     |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Trans-Urethral                    |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Intra-Vascular                    |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Peripheral Vascular               |   | P |   |     |     |                  |                                 |                              | P                     |                    |
| Laparascopic                      |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Muscular-Skeletal<br>Conventional |   | N |   |     |     |                  |                                 |                              | N                     |                    |
| Muscular-Skeletal<br>Superficial  |   | N |   |     |     |                  |                                 |                              | N                     |                    |
| Others (Specify)                  |   | P |   |     |     |                  |                                 |                              | P                     |                    |

Moda of Oneration

N = new indication P = previously cleared by FDA (K061083) E = added under Appendix E Additional Comments: Small organs include: thyroid, testes, breast Combined: B/M Mode Other uses include: Uterus, Ovary, and Prostate

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancye Brogdon
Division Sign-Off

Division of Reproductive and Radiological Device 510(k) Number

Prescription Use (Per 21 CFR 801.109)

Indications For Use

{6}------------------------------------------------

SIUI CTS-7700 Digital Ultrasound System - Muscular-Skeletal, Conventional and Superficial Imaging

# Diagnostic Ultrasound Indications for Use Form

### 3.2 Transducer Indications for Use Form

Device Name: Convex Array C3L60B

| Device Name/Center                |   |   |   |     |     |                   |                                 |                              |                       |                    |
|-----------------------------------|---|---|---|-----|-----|-------------------|---------------------------------|------------------------------|-----------------------|--------------------|
|                                   |   |   |   |     |     | Mode of Operation |                                 |                              |                       |                    |
| Clinical Application              | A | B | M | PWD | CWD | Color<br>Doppler  | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Opthalmic                         |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Fetal                             |   | P |   |     |     |                   |                                 |                              | P                     |                    |
| Abdominal                         |   | P |   |     |     |                   |                                 |                              | P                     |                    |
| IntraOperative<br>(Cardiac)       |   |   |   |     |     |                   |                                 |                              |                       |                    |
| IntraOperative<br>Neurological    |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Pediatric                         |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Small Organ<br>(Specify)          |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Neonatal Cephalic                 |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Adult Cephalic                    |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Cardiac (Adult)                   |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Cardiac (Pediatric)               |   | P |   |     |     |                   |                                 |                              | P                     |                    |
| Transesophageal                   |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Trans-Rectal                      |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Trans-Vaginal                     |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Trans-Urethral                    |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Intra-Vascular                    |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Peripheral Vascular               |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Laparascopic                      |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Muscular-Skeletal<br>Conventional |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Muscular-Skeletal<br>Superficial  |   |   |   |     |     |                   |                                 |                              |                       |                    |
| Others (Specify)                  |   | P |   |     |     |                   |                                 |                              | P                     |                    |

P = previously cleared by FDA (K061083) E = added under Appendix E N = new indication Additional Comments: Uterus, Ovary, and Prostate, Combined: B/M Mode

Concurrence of CDRH, Office of Device Evaluation (ODE)

uation (ODE)

Nancy Brogdon
(Division Sign-Off)

Division of Reprod and Radiological Devices 510(k) Number

Prescription Use (Per 21 CFR 801.109)

Indications For Use

{7}------------------------------------------------

# SIUI CTS-7700 Digital Ultrasound System - Muscular-Skeletal, Conventional and Superficial Imaging Diagnostic Ultrasound Indications for Use Form

# 3.3 Transducer Indications for Use Form

Device Name: Linear Array L7L38B

#### Mode of Operation

| Mode of Operation                 |   |   |   |     |     |                  |                                 |                              |                       |                    |
|-----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|--------------------|
| Clinical Application              | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other<br>(Specify) |
| Opthalmic                         |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Fetal                             |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Abdominal                         |   |   |   |     |     |                  |                                 |                              |                       |                    |
| IntraOperative<br>(Cardiac)       |   |   |   |     |     |                  |                                 |                              |                       |                    |
| IntraOperative<br>Neurological    |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Pediatric                         |   | P |   |     |     |                  |                                 |                              | P                     |                    |
| Small Organ<br>(Specify)          |   | P |   |     |     |                  |                                 |                              | P                     |                    |
| Neonatal Cephalic                 |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Adult Cephalic                    |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Cardiac (Adult)                   |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Cardiac (Pediatric)               |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Transesophageal                   |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Trans-Rectal                      |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Trans-Vaginal                     |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Trans-Urethral                    |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Intra-Vascular                    |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Peripheral Vascular               |   | P |   |     |     |                  |                                 |                              | P                     |                    |
| Laparascopic                      |   |   |   |     |     |                  |                                 |                              |                       |                    |
| Muscular-Skeletal<br>Conventional |   | N |   |     |     |                  |                                 |                              | N                     |                    |
| Muscular-Skeletal<br>Superficial  |   | N |   |     |     |                  |                                 |                              | N                     |                    |
| Others (Specify)                  |   |   |   |     |     |                  |                                 |                              |                       |                    |

P = previously cleared by FDA (K061083) E = added under Appendix E N = new indication Additional Comments: Small organs include: thyroid, testes, breast Combined: B/M Mode

Concurrence of CDRH, Office of Device Evaluation (ODE)

Nancy C Brogdon

(Division Sig Division of Reproductive, Ab and Radiological Devices 510(k) Number

Prescription Use (Per 21 CFR 801.109)

Indications For Use

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K070982](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K070982)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
