← Product Code [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO) · K042831

# ATP (AUTO NON-CONTACT TONOMETER/PACHYMETER) (K042831)

_Reichert, Inc. · IYO · Oct 28, 2004 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K042831

## Device Facts

- **Applicant:** Reichert, Inc.
- **Product Code:** [IYO](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO.md)
- **Decision Date:** Oct 28, 2004
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1560
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

ATP is intended to measure intra-ocular pressure and comeal thickness of the eye for the purpose of aiding in the diagnosis and monitoring of glaucoma.

## Device Story

ATP (Auto Non-Contact) Tonometer/Pachymeter measures intra-ocular pressure and corneal thickness. Device utilizes ultrasonic pulsed echo imaging technology via detachable probes. Operated by clinical staff in ophthalmic settings. Provides quantitative measurements to assist physicians in glaucoma diagnosis and monitoring. Benefits include non-invasive assessment of ocular parameters.

## Technological Characteristics

Ultrasonic pulsed echo imaging system; utilizes detachable probes (13861-401); A-mode operation for ophthalmic measurements.

## Regulatory Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three wavy lines, which is the symbol of HHS. The image is in black and white.

Public Health Service

OCT 2 8 2004

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Reichert, Inc. % Mr. N. E. Devine Responsible Third Party Official Entela, Inc. 3033 Madison Ave., SE GRAND RAPIDS MI 49548

Re: K042831

Trade/Device Name: ATP (Auto Non-Contact) Tonometer/Pachymeter) Regulation Number: 21 CFR 892.1560 and 886.1930 Regulation Name : Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: 90 IYO and 86 HKX Dated: October 12, 2004 Received: October 13, 2004

Dear Mr. Devine:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling. and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

{1}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrb/dsmaldsmamain.html

Sincerely yours,

Nancy C. Brogdon

Nancy C. srogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form

Fill out one form for each ultrasound system and each transducer

Indended Use: ATP is intended to measure intra-ocular pressure and comeal thickness of the eye for the purpose of aiding in the diagnosis and monitoring of glaucoma.

|                               | Mode of Operation |   |   |     |     |                  |                       |                              |                       |                    |
|-------------------------------|-------------------|---|---|-----|-----|------------------|-----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application          | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Dopplier | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic                    | P                 |   |   |     |     |                  |                       |                              |                       |                    |
| Fetal                         |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Abdominal                     |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Interoperative (specify)      |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Interoperative Neurological   |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Pediatric                     |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Small Organ (specify)         |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Neonatal Cephalic             |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Adult Cephalic                |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Cardiac                       |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Transesophageal               |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Transcrectal                  |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Transvaginal                  |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Transurethral                 |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Intravascular                 |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Peripheral Vascular           |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Laparoscopic                  |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Musculo-skeletol Conventional |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Musculo-skeletol Superficial  |                   |   |   |     |     |                  |                       |                              |                       |                    |
| Other (specify)               |                   |   |   |     |     |                  |                       |                              |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments: The above indications for use is for the ATP with the (13861-401)

detachable probes.

510k) N

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Elaluation (ODE)

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices K042831

A3-7

Prescription Use

.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K042831](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYO/K042831)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
