ALOKA MODEL SSD-4000 DIAGNOSTIC ULTRASOUND SYSTEM W/OPTIONAL SONOREAL 3D SYSTEM ACCESSORY

K040719 · Aloka Co., Ltd. · IYO · Mar 26, 2004 · Radiology

Device Facts

Record IDK040719
Device NameALOKA MODEL SSD-4000 DIAGNOSTIC ULTRASOUND SYSTEM W/OPTIONAL SONOREAL 3D SYSTEM ACCESSORY
ApplicantAloka Co., Ltd.
Product CodeIYO · Radiology
Decision DateMar 26, 2004
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1560
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Indications for Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: [Table of clinical applications including Fetal, Abdominal, Intraoperative, Pediatric, Small Organ, Neonatal/Adult Cephalic, Cardiac, Transesophageal, Transrectal, Transvaginal, Peripheral Vascular, Laparoscopic, and Musculo-skeletal (Conventional/Superficial) using A, B, M, PWD, Color Doppler, and Amplitude Doppler modes].

Device Story

Aloka SSD-4000 Ver. 3.1 is a diagnostic ultrasound system with optional SONOReal 3D accessory. It uses piezo-electric transducers to transmit high-frequency sound waves into the body; reflected energy (based on acoustic impedance differences) is processed into images. Doppler shift detection is used for blood flow analysis. Used in clinical settings by trained professionals to aid in diagnosis of disease or abnormality. Output is displayed on the system console for clinician review to inform clinical decision-making. Benefits include non-invasive visualization of internal structures and hemodynamics.

Clinical Evidence

Bench testing only. The device complies with NEMA-UD2: 1992, AIUM 1994, IEC-60601-1, and ISO10993-1:1997 standards. Acoustic output levels are within FDA-allowed limits. No clinical data was required for this 510(k) submission.

Technological Characteristics

Piezo-electric transducer-based ultrasound system. Supports A, B, M, PWD, Color Doppler, and Amplitude Doppler modes. Complies with NEMA-UD2: 1992, AIUM 1994, IEC-60601-1 (2001-09), UL 2601-1 (1997), and ISO10993-1:1997. Patient contact materials evaluated for biocompatibility. Includes optional SONOReal 3D accessory.

Indications for Use

Indicated for diagnostic ultrasound imaging or fluid flow analysis of the human body, including fetal, abdominal, intraoperative, pediatric, small organ, cephalic, cardiac, transesophageal, transrectal, transvaginal, peripheral vascular, laparoscopic, and musculo-skeletal applications. Prescription use only.

Regulatory Classification

Identification

An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # MAR 2 6 2004 ## 510(k) Summary: Aloka Model SSD-4000 Diagnostic Ultrasound System w/ optional SONOReal 3D system accessory. This summary statement complies with 21 CFR, section 807.92 as amended March 14, 1995. This premarket notification has been submitted by Aloka Co., Ltd. and covers the Aloka SSD-4000 Ver. 3.1 diagnostic ultrasound system and transducers w/optional SONOReal 3D system accessory. The address is: > 10 Fairfield Boulevard Wallingford. CT 06492 (203) 269-5088 The contact person is: Richard J. Cehovsky, RA/QA Coordinator The proprietary name is the Aloka SSD-4000 diagnostic ultrasound system and transducers. The common name for this type of device is a diagnostic ultrasound system and transducers. The item in this submission is covered under the following classification: | 90 IYN | Ultrasonic Pulsed Doppler Imaging System | 21 CFR 892.1550 | |--------|------------------------------------------|-----------------| | 90 ITX | Diagnostic Ultrasound Transducer | 21 CFR 892.1570 | | 90 IYO | Ultrasonic Pulsed Echo Imaging System. | 21 CFR 892.1560 | The above as stated in 21 CFR, part 892.1570, 1560 and 1550, has been classified as regulatory Class II. The Aloka SSD-4000 Ver. 3.1 and its transducers are substantially equivalent to its predicates; the Aloka SSD-4000 Ver.1.0 (K003739) and SSD-5500 -(K992663), (K002784), (K011457), (K011315) & (K032875) and their transducers. The Aloka SSD-4000 Ver.3.1 functions in the same manner as its predicates and other diagnostic ultrasound devices. High frequency sound waves are transmitted into the body by a piezo-electric transducer. In the body, differences in the acoustic impedance of different tissues reflect a certain amount of the ultrasound energy back to the transducer, where it is transmitted via the probe cable to the system console and processed into an image. The Aloka SSD-4000 transducers can also use the Doppler shift of sound reflected from moving tissues (blood) to detect and display flow. The Aloka SSD-4000, like other Aloka marketed diagnostic ultrasound systems and transducers is indicated for imaging body structures to aid in the diagnosis of disease or abnormality. The Aloka SSD-4000 Ver.3.1 diagnostic ultrasound system and transducers are similar in technological characteristics to Aloka's initial predicate ultrasound system: SSD-4000 Ver.1.0 (K003739) as well as Aloka's SSD-5500 Ver.3.1 (K992663), SSD-5500 Ver.4.2 (K002784/ K011457), SSD-5500 Ver. 4.2.2 (K011315) and SSD-5500 Ver. 6.0 (K032875). - The SSD-4000 Ver. 3.1 is indicated for the same diagnostic ultrasound applications to Aloka's . ultrasound systems: SSD-4000 (K003739), SSD-5500- V3.1, V4.2, V4.2.2 and V6.0. - . The SSD-4000 Ver. 3.1 has the same gray-scale and Doppler abilities to Aloka's ultrasound systems as mentioned above. {1}------------------------------------------------ ## 510(k) Summary: Aloka Model SSD-4000 Diagnostic Ultrasound System w/ optional SONOReal 3D system. - The SSD-4000 Ver.3.1 uses the same technologies for imaging, Doppler functions and signal . processing as the following products currently marketed by Aloka : SSD-4000 Ver.1.0 -(K003739) and SSD-5500- V3.1, V4.2 , V4.2.2 and V6.0 - The SSD-4000 Ver. 3.1 has the same method of use as the following products currently marketed . by Aloka: SSD-4000 Ver. 1.0 - (K003739) and SSD-5500-V3.1, V4.2, V4.2.2 and V6.0. - The SSD-4000 Ver. 3.1 acoustic power output levels are below the maximum levels allowed by the . FDA. - The SSD-4000 Ver. 3.1 is subjected to the same Quality Assurance systems in development and . production as other products currently marketed by Aloka such as the: SSD-4000 Ver. 1.0 -(K003739) and SSD-5500- V3.1,V4.2 ,V4.2.2 and V6.0. - The patient contact materials used in the SSD-4000 Ver. 3.1 have been evaluated for safety via the . same standards and methods as the above mentioned products marketed by Aloka. These materials have been found to be safe for the intended uses. - The SSD-4000 Ver. 3.1 complies with electrical and physical safety standards as other products ● currently marketed by Aloka such as the: SSD-4000 Ver.1.0 - (K003739) and SSD-5500- V3.1, V4.2, V4.2.2 and V6.0. - The SONOReal accessory (K023473) is an add on device that has met UL-2601-1, EN 60601-1-. A 1 & A2, and EN-60601-1-1 standards. The device is also CE marked to Directive 93/42/EEC per Annex II, as a Class IIa device. - Aloka Co., Ltd. Certifies that the SSD-4000 Ver. 3.1 complies with NEMA-UD2: 1992, AIUM . 1994 "Acoustic Output Labeling Standard for Diagnostic Ultrasound Equipment", IEC-60601-1 (2001-09 Class A), UL 2601-1, 2th edition (1997), Part 1, 2nd edition including Amendments 1822 and ISO10993-1:1997. All testing was completed and the results meet the requirements of the standards above. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 2 6 2004 Aloka Co., Ltd. % Mr. Derwyn Reuber Vice President, Compliance Intertek Testing Services NA, Inc. 70 Codman Hill Road BOXBOROUGH MA 01719 Re: K040719 Trade Name: SSD-4000 Diagnostic Ultrasound System (Version 3.1) with Optional SONOReal 3D Digital System Accessory Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: 90 IYN, IYO, and ITX Dated: March 17, 2004 Received: March 19, 2004 Dear Mr. Reuber: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. This determination of substantial equivalence applies to the following transducers intended for use with the SSD-4000 Diagnostic Ultrasound System (Version 3.1) with Optional SONOReal 3D Digital System Accessory, as described in your premarket notification: {3}------------------------------------------------ #### Transducer Model Number | ASU-67 | UST-1003 | UST-5546 | |-------------------------|---------------|--------------| | UST-579T-7.5 | UST-2265-2 | UST-5548 | | UST-670P-5 | UST-5268P-5 | UST-5710-7.5 | | UST-672-5/7.5 (C) & (L) | UST-5293-5 | UST-9101-7.5 | | UST-676P-5 | UST-5298 | UST-9102-3.5 | | UST-978-3.5 | UST-5299 | UST-9103-5 | | UST-979-3.5 | UST-5524-5 | UST-9104-5 | | UST-984-5 | UST-5524-7.5 | UST-9112-5 | | UST-987-7.5 | UST-5526L-7.5 | UST-9121 | | UST-990-5 | UST-5531 | UST-9123 | | UST-995-7.5 | UST-5534T-7.5 | UST-9124 | | ASU-1000C-3.5 | UST-5536-7.5 | UST-52109 | | ASU-1001 | UST-5542 | | If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded. The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to: > Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850 {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801, please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html". If you have any questions regarding the content of this letter, please contact Rodrigo C. Perez at (301) 594-1212. Sincerely yours. David A. Lynn Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure(s) {5}------------------------------------------------ 4.3.1 #### Diagnostic Ultrasound Indications for Use Form SSD-4000 Ver.3.1 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Modes of operation | | | | | | | | | | |----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | Opthalmic | | | | | | | | | | | | Fetal | | P | P | P | | P | P | | See Below | | | Abdominal | | P | P | P | | P | P | | See Below | | | Intraoperative (specify) | | P | P | P | | P | P | | See Below | | | Intraoperative Neurological | | P | P | P | | P | P | | See Below | | | Pediatric | | P | P | P | | P | P | | See Below | | | Small Organ (specify) | | P | P | P | | P | P | | See Below | | | Neonatal Cephalic | | P | P | P | | P | P | | See Below | | | Adult Cephalic | | P | P | P | | P | P | | See Below | | | Cardiac | | P | P | P | | P | P | | See Below | | | Transesophageal | | P | P | P | | P | P | | See Below | | | Transrectal | | P | P | P | | P | P | | See Below | | | Transvaginal | | P | P | P | | P | P | | See Below | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | P | P | P | | P | P | | See Below | | | Laparoscopic | | P | P | P | | P | P | | See Below | | | Musculo-skeletal<br>Conventional | | P | P | P | | P | P | | See Below | | | Musculo-skeletal<br>Superficial | | P | P | P | | P | P | | See Below | | | Other | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Mixed mode operation includes B/M, B/PWD, B/Bflow/PWD. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801/ FR 801(09) Daniel G. Seum (Division Sign-Off) Division of Reproductive, Al and Radiological Devices 510(k) Numberg 19 {6}------------------------------------------------ દરભાવ) #### Diagnostic Ultrasound Indications for Use Form ASU-67 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Modes of operation | | | | | | | | | | |---------------------------------------------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | Opthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative (specify) | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ (specify) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | P | P | P | | P | P | | See Below | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional<br>Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Mixed mode operation includes B/M, B/PWD, B/Bflow/PWD. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) 801(09) Omilla. byrom (Division Sign-Off) Division of Reproductive, Abdomin and Radiological Device 510(k) Number {7}------------------------------------------------ UST-579T-7.5 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Modes of operation | | | | | | | | | | |----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | Opthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative (specify) | | P | P | P | | P | P | | See Below | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ (specify) | | P | P | P | | P | P | | See Below | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | P | P | P | | P | P | | See Below | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Mixed mode operation includes B/M, B/PWD, B/Bflow/PWD. Intraoperative applications: liver, pancreas, and gall bladder. Small parts applications: breast, testes and thyroid. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801) (Pinsion Sign Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________ {8}------------------------------------------------ Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Modes of operation | | | | | | | | | | |----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | Opthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative (specify) | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ (specify) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | P | P | P | | P | P | | See Below | | | Transvaginal | | P | P | P | | P | P | | See Below | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Mixed mode operation includes B/M, B/PWD, B/Bflow/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) David A. bynum (Division Sign-Off Division of Reproductive, Abdom and Radiological Devices 510(k) Number {9}------------------------------------------------ ## Diagnostic Ultrasound Indications for Use Form UST-672-5/7.5 (C) & (L) Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Modes of operation | | | | | | | | | | |----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | Opthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative (specify) | | P | P | P | | P | P | | See Below | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ (specify) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | P | P | P | | P | P | | See Below | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Mixed mode operation includes B/M, B/PWD,B/Bflow?PWD Intraoperative applications: liver, pancreas and gall bladder. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 80).109 David A. Symm (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number __ {10}------------------------------------------------ UST-676P-5 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Modes of operation | | | | | | | | | | |----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | Opthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative (specify) | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ (specify) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | P | P | P | | P | P | | See Below | | | Transvaginal | | P | P | P | | P | P | | See Below | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Mixed mode operation includes B/M, B/PWD, B/Bflow/PWD. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) David A. Lynn (Chuision Sign-Off) Civision of Reproductive, Abdominal and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________ {11}------------------------------------------------ ## Diagnostic Ultrasound Indications for Use Form UST-978-3.5 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Modes of operation | | | | | | | | | | |----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | Opthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | P | P | P | | P | P | | See Below | | | Intraoperative (specify) | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | P | P | P | | P | P | | See Below | | | Small Organ (specify) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Mixed mode operation includes B/M, B/PWD, B/Bflow/PWD. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) Donald A. Jessop (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________ {12}------------------------------------------------ ## Diagnostic Ultrasound Indications for Use Form UST-979-3.5 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Modes of operation | | | | | | | | | | |----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | Opthalmic | | | | | | | | | | | | Fetal | | P | P | P | | P | P | | See Below | | | Abdominal | | P | P | P | | P | P | | See Below | | | Intraoperative (specify) | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ (specify) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Mixed mode operation includes B/M, B/PWD, B/Bflow/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Prescription Use (Per 21 CFR 801.109) David A. Symms (Division Sign-Off) Olvision of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________ {13}------------------------------------------------ UST-984-5 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Modes of operation | | | | | | | | | | | |----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | | Opthalmic | | | | | | | | | | | | | Fetal | | P | P | P | | P | P | | See Below | | | | Abdominal | | | | | | | | | | | | | Intraoperative (specify) | | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | Small Organ (specify) | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | P | P | P | | P | P | | See Below | | | | Transurethral | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | | Other | | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Mixed mode operation includes B/M, B/PWD, B/Bflow/PWD. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) David A. Lynn (Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices 510(k) Number _ {14}------------------------------------------------ UST-987-7.5 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Modes of operation | | | | | | | | | | |----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | Opthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative (specify) | | P | P | P | | P | P | | See Below | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ (specify) | | | | | | | | | | | | Neonatal Cephalic | | P | P | P | | P | P | | See Below | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Mixed mode operation includes B/M, B/PWD, B/Bflow/PWD Intraoperative applications: include liver, pancreas and gall bladder. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Eyaluation (ODE) Prescription Use (Per 21 CFR 801.109 (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________ {15}------------------------------------------------ Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Modes of operation | | | | | | | | | | |----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | Ophthalmic | | | | | | | | | | | | Fetal | | P | P | P | | P | P | | See Below | | | Abdominal | | P | P | P | | P | P | | See Below | | | Intraoperative (specify) | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ (specify) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Mixed mode operation includes B/M, B/PWD, B/Bflow/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Janita Lynn (Division Sign-Off) Division of Reproductive, Abdomina and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________ {16}------------------------------------------------ Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Modes of operation | | | | | | | | | | | |----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | | Ophthalmic | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | Abdominal | | | | | | | | | | | | | Intraoperative (specify) | | P | P | P | | P | P | | See Below | | | | Intraoperative Neurological | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | Small Organ (specify) | | P | P | P | | P | P | | See Below | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac | | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | | Peripheral Vascular | | P | P | P | | P | P | | See Below | | | | Laparoscopic | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | | Other | | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Mixed mode operation includes B/M, B/PWD, B/Bflow/PWD Intraoperative applications: liver, pancreas, and gall bladder. Small parts applications: breast, testes and thyroid. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 80 R 891.109 David A. Lynn (Division Sign-Off) Division of Reproductive, Abdo and Radiological Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________ {17}------------------------------------------------ ASU-1000C-3.5 Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Modes of operation | | | | | | | | | | |----------------------------------|--------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------| | Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | | Opthalmic | | | | | | | | | | | | Fetal | | P | P | P | | P | P | | See Below | | | Abdominal | | P | P | P | | P | P | | See Below | | | Intraoperative (specify) | | | | | | | | | | | | Intraoperative Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ (specify) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Peripheral Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other | | | | | | | | | | | N= new indication; P= previously cleared by FDA; E= added under Appendix E Additional Comments: Mixed mode operation includes B/M, B/PWD, B/Bflow/PWD (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) David A. Syverson Division Sign Off (Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number _ {18}------------------------------------------------ Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Modes of operation | | | | | |…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...