← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K993939

# NICOLET EME LEGEND TC22 (K993939)

_Nicolet Biomedical · IYN · Jun 29, 2000 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K993939

## Device Facts

- **Applicant:** Nicolet Biomedical
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Jun 29, 2000
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

## Device Story

Nicolet EME Legend TC22 is a diagnostic ultrasound system designed for blood flow analysis. It utilizes various transducers (2 MHz, 4 MHz, and 8 MHz) operating in Pulsed Wave (PWD) and Continuous Wave (CWD) Doppler modes. The device is intended for use by clinicians in a clinical setting to perform non-invasive vascular assessments. By transmitting ultrasound waves into the body and processing the reflected signals, the system provides real-time data on fluid flow, which assists healthcare providers in diagnosing vascular conditions. The output allows for the evaluation of hemodynamic parameters, supporting clinical decision-making regarding patient vascular health.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and intended use.

## Technological Characteristics

Diagnostic ultrasound system utilizing 2 MHz, 4 MHz, and 8 MHz transducers. Supports Pulsed Wave (PWD) and Continuous Wave (CWD) Doppler modes. System is designed for neonatal cephalic, adult cephalic, intravascular, and peripheral vascular applications.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle. The text is in all capital letters and is arranged to follow the curve of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

### JUN 2 9 2000

David W. Wagner Director Quality Assurance and Regulatory Affairs Nicolet Vascular. 6355 Joyce Drive Golden, Colorado 80403

K993939 Re: Nicolet EME Legend TC22 Regulatory class: II 21 CFR 892.1556/Procode: 90 IYN 21 CFR 892.1570/Procode: 90 ITX Dated: March 30, 2000 Received: April 4, 2000

Dear Mr. Wagner:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls rowisions of the Act. The general controls provisions of the Act include requirements for annual registration, provisions of the research gractice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the Nicolet EME Legend TC22, as described in your premarket notification:

Transducer Model Number

2 MHz 10 mm Pulsed Wave 2 MHz 15 mm Pulsed Wave 4 MHz Pulsed Wave 4 MHz Continuous Wave 8 MHz Pulsed Wave 8 MHz Continuous Wave

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{1}------------------------------------------------

#### Page -2- Mr. Wagner

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

If you have any questions regarding the content of this letter, please contact Paul M. Gammell, Ph.D. at (301) 594-1212.

Sincerely yours,

Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### K 993939

## SYSTEM LEGEND TC 22 Tiran's CRANINL FADULT CEPHALIC

Appendix F

# Poriphern Vascular Diagnstic Dev

### Diagnostic Ultrasound Indications for Use Form

## Fill out one form for each ultrasound system and each transducer.

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  |   |   |   | Mode of Operation |     |               |                   |                        |                      |                 |  |  |
|----------------------------------|---|---|---|-------------------|-----|---------------|-------------------|------------------------|----------------------|-----------------|--|--|
| Clinical Application             | A | B | M | PWD               | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined - (specify) | Other (specify) |  |  |
| Ophthalmic                       |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Fetal                            |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Abdominal                        |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Intraoperative (specify)         |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Intraoperative Neurological      |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Pediatric                        |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Small Organ (specify)            |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Neonatal Cephalic                |   |   |   | ◼                 |     |               |                   |                        |                      |                 |  |  |
| Adult Cephalic                   |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Cardiac                          |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Transesophageal                  |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Transrectal                      |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Transvaginal                     |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Transurethral                    |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Intravascular                    |   |   |   | ◼                 | ◼   |               |                   |                        |                      |                 |  |  |
| Peripheral Vascular              |   |   |   | ◼                 | ◼   |               |                   |                        |                      |                 |  |  |
| Laparoscopic                     |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Musculo-skeletal<br>Conventional |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Musculo-skeletal Superficial     |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |
| Other (specify)                  |   |   |   |                   |     |               |                   |                        |                      |                 |  |  |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

| Additional Comments: | SYSTEM UTILIZES 2EA 2MHZ PULSED WAVE PROBES, 1 EA 4MHZ PULSED AND CW PROBE AND 1 EA 8MHZ PULSED AND CW PROBE. |
|----------------------|---------------------------------------------------------------------------------------------------------------|
|----------------------|---------------------------------------------------------------------------------------------------------------|

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

(Division Sign-Off)

Division of Reproductive, Cadomanal. ONT

and Radiological Devices

| 510(k) Number | K993939 |
|---------------|---------|
|---------------|---------|

{3}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

#### Fill out one form for each ultrasound system and each transducer.

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                                          | Mode of Operation        |   |   |             |     |                  |                      |                              |                       |                    |
|--------------------------------------------------------------------------|--------------------------|---|---|-------------|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application                                                     | A                        | B | M | PWD         | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic                                                               |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Fetal                                                                    |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Abdominal                                                                |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Intraoperative (specify)                                                 |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Intraoperative Neurological                                              |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Pediatric                                                                |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Small Organ (specify)                                                    |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Neonatal Cephalic                                                        |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Adult Cephalic                                                           |                          |   |   | <div></div> |     |                  |                      |                              |                       |                    |
| Cardiac                                                                  |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Transesophageal                                                          |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Transrectal                                                              |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Transvaginal                                                             |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Transurethral                                                            |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Intravascular                                                            |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Peripheral Vascular                                                      |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Laparoscopic                                                             |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Conventional                                         |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Musculo-skeletal Superficial                                             |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Other (specify)                                                          |                          |   |   |             |     |                  |                      |                              |                       |                    |
| N= new indication; P= previously cleared by FDA; E= added under Appendix |                          |   |   |             |     |                  |                      |                              |                       |                    |
| Additional Comments:                                                     | E- 2MHZ 10MM PULSED WAVE |   |   |             |     |                  |                      |                              |                       |                    |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Prescription Use (Per 21 CFR 801.109)

F-3 Division of Reproductive, Abdominal, ENT,  
and Radiological Devices

F-3

510(k) Number J

{4}------------------------------------------------

#### Diagnostic Ultrasound Indications for Use Form

### Fill out one form for each ultrasound system and each transducer.

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                                                                                                                                                                                                                            | Mode of Operation |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|---|------|-----|-----|------------------|--------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|------------------------------------------|--------------------|--|
| Clinical Application                                                                                                                                                                                                                                       | A                 | B | M    | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler                                                                                                                 | Color<br>Velocity<br>Imaging                   | Combined<br>(specify)                    | Other<br>(specify) |  |
| Ophthalmic                                                                                                                                                                                                                                                 |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Fetal                                                                                                                                                                                                                                                      |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Abdominal                                                                                                                                                                                                                                                  |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Intraoperative (specify)                                                                                                                                                                                                                                   |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Intraoperative Neurological                                                                                                                                                                                                                                |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Pediatric                                                                                                                                                                                                                                                  |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Small Organ (specify)                                                                                                                                                                                                                                      |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Neonatal Cephalic                                                                                                                                                                                                                                          |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Adult Cephalic                                                                                                                                                                                                                                             |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Cardiac                                                                                                                                                                                                                                                    |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Transesophageal                                                                                                                                                                                                                                            |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Transrectal                                                                                                                                                                                                                                                |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Transvaginal                                                                                                                                                                                                                                               |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Transurethral                                                                                                                                                                                                                                              |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Intravascular                                                                                                                                                                                                                                              |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Peripheral Vascular                                                                                                                                                                                                                                        |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Laparoscopic                                                                                                                                                                                                                                               |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Musculo-skeletal<br>Conventional                                                                                                                                                                                                                           |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Musculo-skeletal Superficial                                                                                                                                                                                                                               |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| Other (specify)                                                                                                                                                                                                                                            |                   |   |      |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
| N= new indication; P= previously cleared by FDA; E= added under Appendix<br>Additional Comments:_E________________________________________________________________________________________________________________________________________________________ |                   |   | 2MHZ |     |     |                  |                                                                                                                                      |                                                |                                          |                    |  |
|                                                                                                                                                                                                                                                            |                   |   |      |     |     |                  | (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)<br>Concurrence of CDRH, Office of Device Evaluation (ODE) |                                                |                                          |                    |  |
| Prescription Use (Per 21 CFR 801.109)<br>ﺎ                                                                                                                                                                                                                 |                   |   |      |     |     |                  |                                                                                                                                      | (Division Sign-Ja)<br>and Radiological Devices | Division of Reproductive, Abdominal, ENT |                    |  |

510(k) Number _

{5}------------------------------------------------

PROBE FOR LEGEND TC 22 SYSTEM
(TRANSDUCER) Diagnostic Ultrasound

Appendix F

K993939

#### Diagnostic Ultrasound Indications for Use Form

#### Fill out one form for each ultrasound system and each transducer.

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                                                                    | Mode of Operation    |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
|----------------------------------------------------------------------------------------------------|----------------------|---|---|----------------------------------------------------------------|-----|------------------|----------------------|----------------------------------------------------------------------------------------------|-----------------------|--------------------|
| Clinical Application                                                                               | A                    | B | M | PWD                                                            | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging                                                                 | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic                                                                                         |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Fetal                                                                                              |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Abdominal                                                                                          |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Intraoperative (specify)                                                                           |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Intraoperative Neurological                                                                        |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Pediatric                                                                                          |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Small Organ (specify)                                                                              |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Neonatal Cephalic                                                                                  |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Adult Cephalic                                                                                     |                      |   |   | <div> <img alt="Checkmark" src="checkbox_checked.png"/> </div> |     |                  |                      |                                                                                              |                       |                    |
| Cardiac                                                                                            |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Transesophageal                                                                                    |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Transrectal                                                                                        |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Transvaginal                                                                                       |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Transurethral                                                                                      |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Intravascular                                                                                      |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Peripheral Vascular                                                                                |                      |   |   | <div> <img alt="Checkmark" src="checkbox_checked.png"/> </div> |     |                  |                      |                                                                                              |                       |                    |
| Laparoscopic                                                                                       |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Musculo-skeletal<br>Conventional                                                                   |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Musculo-skeletal Superficial                                                                       |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Other (specify)                                                                                    |                      |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| N= new indication; P= previously cleared by FDA; E= added under Appendix E<br>Additional Comments: | E- 4 MHZ PULSED WAVE |   |   |                                                                |     |                  |                      |                                                                                              |                       |                    |
| Prescription Use (Per 21 CFR 801.109)                                                              |                      |   |   |                                                                |     |                  |                      | (Division Sign-Off)<br>Division of Reproductive, Abdominal, ENT,<br>and Radiological Devices |                       |                    |

510(k) Number _

199

{6}------------------------------------------------

Appendix F ドゥツアクラク

PROBE FOR LEGEND TC-22 SYSTEM

(TRANSSICE)

Diagnostic Ultrasound Indications for Use Form

## Fill out one form for each ultrasound system and each transducer.

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Mode of Operation |   |   |     |     |               |                   |                        |                    |                 |
|----------------------------------|-------------------|---|---|-----|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| Clinical Application             | A                 | B | M | PWD | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic                       |                   |   |   |     |     |               |                   |                        |                    |                 |
| Fetal                            |                   |   |   |     |     |               |                   |                        |                    |                 |
| Abdominal                        |                   |   |   |     |     |               |                   |                        |                    |                 |
| Intraoperative (specify)         |                   |   |   |     |     |               |                   |                        |                    |                 |
| Intraoperative Neurological      |                   |   |   |     |     |               |                   |                        |                    |                 |
| Pediatric                        |                   |   |   |     |     |               |                   |                        |                    |                 |
| Small Organ (specify)            |                   |   |   |     |     |               |                   |                        |                    |                 |
| Neonatal Cephalic                |                   |   |   |     |     |               |                   |                        |                    |                 |
| Adult Cephalic                   |                   |   |   |     |     |               |                   |                        |                    |                 |
| Cardiac                          |                   |   |   |     |     |               |                   |                        |                    |                 |
| Transesophageal                  |                   |   |   |     |     |               |                   |                        |                    |                 |
| Transrectal                      |                   |   |   |     |     |               |                   |                        |                    |                 |
| Transvaginal                     |                   |   |   |     |     |               |                   |                        |                    |                 |
| Transurethral                    |                   |   |   |     |     |               |                   |                        |                    |                 |
| Intravascular                    |                   |   |   |     |     |               |                   |                        |                    |                 |
| Peripheral Vascular              |                   |   |   |     | X   |               |                   |                        |                    |                 |
| Laparoscopic                     |                   |   |   |     |     |               |                   |                        |                    |                 |
| Musculo-skeletal<br>Conventional |                   |   |   |     |     |               |                   |                        |                    |                 |
| Musculo-skeletal Superficial     |                   |   |   |     |     |               |                   |                        |                    |                 |
| Other (specify)                  |                   |   |   |     |     |               |                   |                        |                    |                 |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments: E- 4 MHZ CONTINUOUS WAVE

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

F-3

(Division Sign-off)

Division of Reproductive, Abdominal, ENT, and Radiological Devices

and Radiological Devices

510(k) Number .

{7}------------------------------------------------

PROBE FOR LEGEND TC22 SYSTEM
(TRANSDUCER)
Diagnostic Ultrasound Inc.

Appendix F

ドゥッスリ39

# Diagnostic Ultrasound Indications for Use Form

## Fill out one form for each ultrasound system and each transducer.

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                               | Mode of Operation |   |   |                                             |     |               |                   |                        |                    |                 |
|-------------------------------|-------------------|---|---|---------------------------------------------|-----|---------------|-------------------|------------------------|--------------------|-----------------|
| Clinical Application          | A                 | B | M | PWD                                         | CWD | Color Doppler | Amplitude Doppler | Color Velocity Imaging | Combined (specify) | Other (specify) |
| Ophthalmic                    |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Fetal                         |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Abdominal                     |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Intraoperative (specify)      |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Intraoperative Neurological   |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Pediatric                     |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Small Organ (specify)         |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Neonatal Cephalic             |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Adult Cephalic                |                   |   |   | <div style="background-color: #ddd;"></div> |     |               |                   |                        |                    |                 |
| Cardiac                       |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Transesophageal               |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Transrectal                   |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Transvaginal                  |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Transurethral                 |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Intravascular                 |                   |   |   | <div style="background-color: #ddd;"></div> |     |               |                   |                        |                    |                 |
| Peripheral Vascular           |                   |   |   | <div style="background-color: #ddd;"></div> |     |               |                   |                        |                    |                 |
| Laparoscopic                  |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Musculo-skeletal Conventional |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Musculo-skeletal Superficial  |                   |   |   |                                             |     |               |                   |                        |                    |                 |
| Other (specify)               |                   |   |   |                                             |     |               |                   |                        |                    |                 |

 N= new indication; P= previously cleared by FDA; E= added under Appendix E

 Additional Comments: E-8 MHZ PULSED WAVE

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Reproductive, Abdominal, ENT, and Radiological Devices

F-3

Prescription Use (Per 21 CFR 801.109)

510(k) Number _

99

39

{8}------------------------------------------------

PLEASE FOR LEGEND TC-22 SYSTEM

Appendix F ドゥッショタ

#### (THINSDIXER

## Diagnostic Ultrasound Indications for Use Form

## Fill out one form for each ultrasound system and each transducer.

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                  | Mode of Operation |   |   |     |     |                  |                      |                              |                       |                    |
|----------------------------------|-------------------|---|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|
| Clinical Application             | A                 | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic                       |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Fetal                            |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Abdominal                        |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Intraoperative (specify)         |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Intraoperative Neurological      |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Pediatric                        |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Small Organ (specify)            |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Neonatal Cephalic                |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Adult Cephalic                   |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Cardiac                          |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Transesophageal                  |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Transrectal                      |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Transvaginal                     |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Transurethral                    |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Intravascular                    |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Peripheral Vascular              |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Laparoscopic                     |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal<br>Conventional |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Musculo-skeletal Superficial     |                   |   |   |     |     |                  |                      |                              |                       |                    |
| Other (specify)                  |                   |   |   |     |     |                  |                      |                              |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Additional Comments: E - 8 MHZ CONTINUOUS WAVE

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)

Division of Reproductive, Abdominal, ENT,

and Radiological Devices

Prescription Use (Per 21 CFR 801.109)

F-3

510(k) Number ________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K993939](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K993939)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
