SONOACE 6000C DIAGNOSTIC ULTRASOUND SYSTEM AND TRANSDUCERS, ALSO CALLED ULTRASOUND SYSTEM AND TRANSDUCERS

K990970 · Medison America, Inc. · IYN · Apr 7, 1999 · Radiology

Device Facts

Record IDK990970
Device NameSONOACE 6000C DIAGNOSTIC ULTRASOUND SYSTEM AND TRANSDUCERS, ALSO CALLED ULTRASOUND SYSTEM AND TRANSDUCERS
ApplicantMedison America, Inc.
Product CodeIYN · Radiology
Decision DateApr 7, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Indications for Use

SonoAce 6000C/UM 400C intended uses as defined FDA guidance documents are: Fetal (includes infertility monitoring of follicle development) Abdominal Intra-operative (abdominal organs and peripheral vessel, neurological) Pediatric Small Organ Neonatal Cephalic Adult Cephalic Cardiac Trans-Rectal Trans-Vaginal Peripheral-Vascular Muscular-Skeletal (conventional, superficial)

Device Story

Mobile, software-controlled diagnostic ultrasound system. Acquires ultrasound data via piezoelectric transducers (2.0-9.0 MHz) to display B-mode, M-mode, Color-Flow, Pulsed Doppler, Power Doppler, and 3D images. Digital beamforming and digital dynamic receive focusing enhance image quality. Operator uses control panel to adjust transmit focusing, filtering, and image enhancement. System calculates anatomical measurements (distances, areas, volumes) and obstetric parameters (BPD, HC, AC, FL, etc.). Cine loop stores up to 64 sequential images; image-filing manages patient history. DICOM 3.0 output supports remote viewing. Used by healthcare professionals in clinical settings for diagnostic imaging and biopsy guidance. Real-time acoustic output (mechanical/thermal indices) displayed for safety. Benefits include high-resolution visualization of internal structures and blood flow to assist clinical diagnosis.

Clinical Evidence

Bench testing only. No clinical data provided. System performance validated through acoustic output measurements (SPTA 720 mW/cm2, MI 1.9) and compliance with international electromechanical safety standards (IEC 601-1, EN 60601-1-2).

Technological Characteristics

Digital beamforming ultrasound system. Transducers use piezoelectric materials. Supports linear and convex probes (2.0-9.0 MHz). Connectivity via DICOM 3.0. Software-controlled image processing (filtering, enhancement, dynamic windowing). Acoustic output: SPTA 720 mW/cm2, MI 1.9. Safety standards: EN 60601-1, UL 2601-1, C22.2 No. 601.1, CEMEC 1157:1992, EN 60601-1-2.

Indications for Use

Indicated for diagnostic ultrasound imaging or fluid flow analysis of the human body, including fetal, abdominal, intra-operative, pediatric, small organ, neonatal/adult cephalic, cardiac, trans-rectal, trans-vaginal, peripheral-vascular, and muscular-skeletal applications. Includes biopsy guidance and 3D imaging. Contraindications: Not for in vitro fertilization or percutaneous umbilical blood sampling.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 4/7/99 510(k) Premarket Notification # 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92. Submitter's name, address, telephone number, contact person: 1) Medison America, Inc. 6616 Owens Drive Pleasanton, CA 94588 Bob Leiker Vice President, Regulatory and Quality Telephone: (510) 463-1830 Prepared February 25, 1999 Name of the device, including the trade or proprietary name if applicable, the common or 2) usual name, and the classification name, if known: > Common/Usual Name: Diagnostic Ultrasound System and Accessories #### Proprietary Name: SonoAce 6000C Diagnostic Ultrasound System and Transducers. Also called Ultramark® 400C Diagnostic Ultrasound System and Transducers. | Classification Names: | FR Number | Product Code | |------------------------------------------|-----------|--------------| | Ultrasound Pulsed Echo Imaging System | 892.1560 | 90-IYO | | Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN | | Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX | #### 3) Identification of the predicate or legally marketed device: Medison America, Inc. believes that SA 6000C/UM 400C Ultrasound System is substantially equivalent to the currently marketed SA 8800/HDI® 1500 and the SA 6000/UM 400 systems (K974269 and K981510). {1}------------------------------------------------ #### Device Description: 4) The SA 6000C/UM 400C System is a general purpose, mobile, software controlled, diagnostic ultrasound system. Its function is to acquire ultrasound data and to display the data as B-mode. M-mode, Color-Flow Doppler, Pulsed (PW) Doppler, Power Doppler, 3D or in a combination of these modes. M-mode uses the sweep display method which has its images flow from the left to the right on the monitor The SA 6000C/UM 400C also gives the operator the ability to measure anatomical structures and offers analysis packages that provide information that is used to make a diagnosis by competent health care professionals. The system has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed. Eight different models of transducers are available and any two may be connected at the same time. In addition to the initial operational settings for each transducer preprogrammed in the system, user-customized parameter settings for each transducer may be inserted by the operator and stored for recall as needed via the system control panel. Customization includes transmit focusing, filtering, image enhancement processing, dynamic window curve selection. Controls are also provided to select display format (single and various combinations), to activate zoom features, and to utilize the cine loop function. Patient contact materials are the same as those previously cleared in the predicate device, SA 8800/HDI 1500 and SA 6000C/UM 400C ultrasound systems and transducers (K974267 and K981510). The SA 6000C/UM 400C uses digital beamforming technology, and supports a variety of Linear and Convex probes for a wide variety of applications. It is an ultrasound scanner, which provides high resolution, high penetration performance, and various measurement functions. Probes are supported in frequencies from 2.0 MHz to 9.0 MHz. The system can be used to measure distances and calculate areas, circumferences and volumes, as well as calculate the date of delivery by using BPD (biparietal diameter), OFD (occipito-frontal diameter), HC (head circumference), AC (abdominal circumference), AD (abdominal diameter), FL (femur length), CRL (crown rump length), APTD (anteroposterior trunk diameter), TTD (transverse trunk diameter), GS (gestational sac), LMP (last menstural period.), Cardiac Analysis and Vascular Analysis. Biopsy guidelines are provided on screen to assist in the collection of tissue samples, using biopsy guide adapters offered as an optional accessory. The SA 6000C/UM 400C supports the Cine function (capable of storing up to 64 sequential images) and real-time zoom function to the region-of-interest. The system provides the ability to perform remote viewing of images, without {2}------------------------------------------------ compression, via a Dicom 3.0 compatible output. Management of patient history is possible by image-filing function. High-resolution images are provided by utilizing a technology called digital dynamic receive focusing. The SA 6000C/UM 400C has been designed to meet the following electromechanical safety standards: - EN 60601-1 (IEC 601-1,) European Norm, Medical Electrical Equipment ( - UL 2601-1, Underwriters Laboratories Standards, Medical Electrical Equipment ( - C22.2 No. 601.1, Canadian Standards Association, Medical Electrical Equipment ( - CEMEC 1157:1992, International Electrotechnical Commission, Requirements for the ( declaration of the acoustic output of medical diagnostic ultrasonic equipment - EN 60601-1-2 (IEC 601-1-2,) European Norm, Collateral Standard: Electromagnetic ( Compatibility - Compliant with the European Medical Device Directive Certificate issued by TUV. ( #### Intended Use: ર) SonoAce 6000C/UM 400C intended uses as defined FDA guidance documents are: - Fetal (includes infertility monitoring of follicle development) ( - ( Abdominal - ( Intra-operative (abdominal organs and peripheral vessel, neurological) - ( Pediatric - Small Organ ( - Neonatal Cephalic ( - Adult Cephalic ( - { Cardiac - ( Trans-Rectal - Trans-Vaginal ( - ( Peripheral-Vascular - ( Muscular-Skeletal (conventional, superficial) {3}------------------------------------------------ Typical examinations performed using the system are: - ( General abdominal and pelvic studies including organ surveys, assessment, and retro-peritoneal cavity studies. - Study of small parts including breasts, shoulders, thyroid, and the abdominal wall. く - Pediatric scans of organs and bony structures. ( - く Peripheral vascular applications including carotid arteries, legs, arms, feet, and penile artery. - Monitoring procedures for infertility studies (other than in vitro fertilization). ( - First, second and third trimester pregnancy studies. ・( - Prostate, prostate biopsy guidance, and rectal wall studies. ( - Neonatal head studies. ( - Transcranial studies of middle cerebral arteries, internal carotid artery, and vertebral arteries. ( - ( Cardiac studies in adults and children. - Biopsy guidance for tissue or fluid sampling. ( - Conventional podiatry scans. ( - ( Intraoperative application including soft tissue structures. #### Technological Characteristics: 6) This device operates identical to the predicate devices in that piezoelectric material in the transducer is used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as a 2D and M-mode, Spectral Doppler, Color Doppler, Power Doppler, or 3D images. Transducer patient contact materials are biocompatible. The device's acoustic output limits are: All Applications: | | (Maximum Range) | |--------|-----------------| | I SPTA | 720 mW/cm2 | | MI | 1.9 | The limits are the same as predicate Track 3 devices. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with its head turned to the left. The eagle is surrounded by a circular border containing the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 7 1999 APR Medison America, Inc. c/o Carol Stamp TUV Product Services 1775 Old Highway 8 New Brighton, MN 55112-1891 K990970 Re: SonoAce 6000C/Ultramark 400C Requlatory Class: II/21 CFR 892.1550, 21 CFR 892.1560 and 21 CFR 892.1570 Product Code: 90/IYN, 90 IYO and 90 ITX Dated: March 22, 1999 Received: March 23, 1999 Dear Ms. Stamp: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. This determination of substantial equivalence applies to the followinq transducers intended for use with the SonoAce 6000C/Ultramark 400C, as described in your premarket notification: Transducer Model Number HLS-9 Linear Array 7.5 MHz/40 mm L5-9/60mm Linear Array 7.5 MHz/60mm C2-4/30R Curved Linear Array 3.0MHz/30R/60deg MC2-5 Curved Linear Array 3.5MHz/40R/89deg HC3-6 Curved Linear Array 3.5MHz/60R/60deg C4-7 Curved Linear Array 6.5MHz/10R/10R/140deg C4-9 Curved Linear Array 5.0MHz/40R/60deg EC 4-9/10R Curved Linear Array 6.5MHz/10R/140deg (Endocavity) If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic QS inspections, the FDA will verify such assumptions. Failure to comply with the GMP regulation may result in {5}------------------------------------------------ Page 2 - Carol Stamps requlatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification does not affect any obligation you may have under sections 531 and 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. Please be advised that the determination above is based on the fact that no medical devices have been demonstrated to be safe and effective for in vitro fertilization or percutaneous umbilical blood sampling, nor have any devices been marketed for these uses in interstate commerce prior to May 28, 1976, or reclassified into class I (General Controls) or class II (Special Controls). FDA considers devices specifically intended for in vitro fertilization and percutaneous umbilical blood sampling to be investigational, and subject to the provision of the investigational device exemptions (IDE) regulations, 21 Therefore, your product labeling must be consistent with FDA's CFR, Part 812. position on this use. This determination of substantial equivalence is granted on the condition that prior to shipping the first device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in Appendix G, (enclosed) of the Center's September 30, 1997 "Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers." If the special report is incomplete or contains unacceptable values (e.g., acoustic output greater than approved levels), then the 510(k) clearance may not apply to the production units which as a result may be considered adulterated or misbranded. The special report should reference the manufacturer's 510(k) number. It should be clearly and prominently marked "ADD-TO-FILE" and should be submitted in duplicate to: > Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850 This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Other general information on your responsibilities under the Act may 807.97). be obtained from the Division of Small Manufacturers Assistance at its tollfree number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". {6}------------------------------------------------ Page 3 - Carol Stamps If you have any questions regarding the content of this letter, please contact II you have all quosebood at (301) 594-1212. Sincerely yours, Daniel h. Syym CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosures {7}------------------------------------------------ # Section 4.3 INDICATION FOR USE Ultrasound Device Indications Statement 510(k) Number: SonoAce 6000C / Ultramark 400C Ultrasound System Device Name: Indications for Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Mode of Operation (*includes simultaneous B-mode) | Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) | |-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------------|------------------------|--------------------| | Ophthalmic | | | | | | | | | | | | Fetal (See Note 3) | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Abdominal | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Intra-Operative<br>(See Note 5) | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Intra-Operative<br>Neurological | | N | N | N | | N | N | | Note 1 | Note 6 | | Pediatric | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Small Organ<br>(See Note 4) | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Neonatal Cephalic | | P | P | N | | N | N | | Note 1 | Note 6 | | Adult Cephalic | | N | N | N | | N | N | | Note 1 | Note 6 | | Cardiac | | P | P | N | | N | N | | Note 1 | Note 6 | | Transesophageal | | | | | | | | | | | | Trans-Rectal | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Trans-Vaginal | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Trans-Urethral | | | | | | | | | | | | Intra-Vascular | | | | | | | | | | | | Peripheral -Vascular | | P | P | N | | N | N | | Note 1 | Note 6 | | Laparoscopic | | | | | | | | | | | | Muscular-Skeletal<br>Conventional | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Muscular-Skeletal<br>Superficial | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Others(Specify) | | | | | | | | | | | N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes: Note 1: PWD/Color Doppler, PWD/Power Doppler, CWD/Color Doppler, CWD/Power Doppler Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 5: Abdominal organs and peripheral vessel Note 6: 3D Imaging Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) . David he. Sezman (Division Sign-Off) Division of Reproductive, Abdominal and Radiological 510(k) Number K990970 {8}------------------------------------------------ | 510(k) Number: | | |----------------------|------------------------------------------------------------------------------------| | Device Name: | SonoAce 6000C / Ultramark 400C Ultrasound System | | Transducer: | HL5-9, 7.5MHz/40mm, Linear Array Probe | | Indications for Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | Mode of Operation (*includes simultaneous B-mode) | | |--|---------------------------------------------------|--| |--|---------------------------------------------------|--| | Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) | |-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------------|------------------------|--------------------| | Ophthalmic | | | | | | | | | | | | Fetal (See Note 3) | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intra-Operative<br>(See Note 5) | | N | N | N | . | N | N | | Note 1 | Notes 2, 6 | | Intra-Operative<br>Neurological | | | | | | | | | | | | Pediatric | | N | N | N | | N | N | | Note 1 | Notes 2, 6 | | Small Organ<br>(See Note 4) | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Neonatal Cephalic | | P | P | N | | N | N | | Note 1 | Note 6 | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Trans-Rectal | | | | | | | | | | | | Trans-Vaginal | | | | | | | | | | | | Trans-Urethral | | | | | | | | | | | | Intra-Vascular | | | | | | | | | | | | Peripheral -Vascular | | P | P | N | | N | N | | Note 1 | Note 6 | | Laparoscopic | | | | | | | | | | | | Muscular-Skeletal<br>Conventional | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Muscular-Skeletal<br>Superficial | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Others(Specify) | | | | | | | | | | | N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes: Note 1: PWD/Color Doppler, PWD/Power Doppler Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 5: Abdominal organs and peripheral vessel Note 6: 3D Imaging Concurrence of CDRH, Office of Device Evaluation (ODE) L Prescription Use (Per 21 CFR 801.109) Daniel h. Sesom (Division Sign-Off) Reproductive, Abdominal, E and Radiological 510(k) Number K990970 {9}------------------------------------------------ 510(k) Number: SonoAce 6000C / Ultramark 400C Ultrasound System Device Name: L5-9/60mm, 7.5MHz/60mm, Linear Array Probe Transducer: Indications for Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | A | B | M* | PWD* | CWD* | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) | |-----------------------------------|---|---|----|------|------|-------------------|----------------------------|------------------------------|------------------------|--------------------| | Ophthalmic | | | | | | | | | | | | Fetal (See Note 3) | N | N | N | N | | N | N | | Note 1 | Note 6 | | Abdominal | N | N | N | N | | N | N | | Note 1 | Note 6 | | Intra-Operative<br>(See Note 5) | | | | | | | | | | | | Intra-Operative<br>Neurological | | | | | | | | | | | | Pediatric | N | N | N | N | | N | N | | Note 1 | Notes 2, 6 | | Small Organ<br>(See Note 4) | P | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Neonatal Cephalic | P | P | P | N | | N | N | | Note 1 | Note 6 | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Trans-Rectal | | | | | | | | | | | | Trans-Vaginal | | | | | | | | | | | | Trans-Urethral | | | | | | | | | | | | Intra-Vascular | | | | | | | | | | | | Peripheral -Vascular | P | P | P | N | | N | N | | Note 1 | Note 6 | | Laparoscopic | | | | | | | | | | | | Muscular-Skeletal<br>Conventional | N | N | N | N | | N | N | | Note 1 | Notes 2, 6 | | Muscular-Skeletal<br>Superficial | N | N | N | N | | N | N | | Note 1 | Notes 2, 6 | | Others(Specify) | | | | | | | | | | | Mode of Operation (*includes simultaneous B-mode) N = new indication: P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes: Note 1: PWD/Color Doppler, PWD/Power Doppler Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 5: Abdominal organs and peripheral vessel Note 6: 3D Imaging **Concurrence of CDRH, Office of Device Evaluation (ODE)** Prescription Use (Per 21 CFR 801.109) David h. Segner (Division Sign-Off) Division of Reproductive, Abdominal, EN and Radiological D 510(k) Number K990970 {10}------------------------------------------------ 510(k) Number: Device Name: SonoAce 6000C / Ultramark 400C Ultrasound System C2-4/30R, 3.0MHz/30R/60D, Curved Linear Array Probe Transducer: Indications for Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | Mode of Operation (*includes simultaneous B-mode) | |--|---------------------------------------------------| |--|---------------------------------------------------| | Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) | |-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------------|------------------------|--------------------| | Ophthalmic | | | | | | | | | | | | Fetal (See Note 3) | | P | P | N | | N | N | | Note 1 | Note 6 | | Abdominal | | P | P | N | | N | N | | Note 1 | Note 6 | | Intra-Operative | | | | | | | | | | | | (See Note 5) | | | | | | | | | | | | Intra-Operative<br>Neurological | | N | N | N | | N | N | | Note 1 | Note 6 | | Pediatric | | | | | | | | | | | | Small Organ | | | | | | | | | | | | (See Note 4) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | N | N | N | | N | N | | Note 1 | Note 6 | | Cardiac | | P | P | N | | N | N | | Note 1 | Note 6 | | Transesophageal | | | | | | | | | | | | Trans-Rectal | | | | | | | | | | | | Trans-Vaginal | | | | | | | | | | | | Trans-Urethral | | | | | | | | | | | | Intra-Vascular | | | | | | | | | | | | Peripheral -Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Muscular-Skeletal<br>Conventional | | | | | | | | | | | | Muscular-Skeletal<br>Superficial | | | | | | | | | | | | Others(Specify) | | | | | | | | | | | N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes: Note 1: PWD/Color Doppler, PWD/Power Doppler Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 5: Abdominal organs and peripheral vessel Note 6: 3D Imaging Concurrence of CDRH, Office of Device Evaluation (ODE) [ Prescription Use (Per 21 CFR 801.109) David C. Stagner (Division Sign-Off) Division of Reproductive, Abdominal, El and Radiological Devi 510(k) Number K990970 {11}------------------------------------------------ 510(k) Number: SonoAce 6000C / Ultramark 400C Ultrasound System Device Name: MC2-5, 3.5MHz/40R/89D, Curved Linear Array Probe Transducer: Indications for Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) | |-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------------|------------------------|--------------------| | Ophthalmic | | | | | | | | | | | | Fetal (See Note 3) | P | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Abdominal | P | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Intra-Operative<br>(See Note 6) | | | | | | | | | | | | Intra-Operative<br>Neurological | | | | | | | | | | | | Pediatric | P | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Small Organ<br>(See Note 5) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Trans-Rectal | | | | | | | | | | | | Trans-Vaginal | | | | | | | | | | | | Trans-Urethral | | | | | | | | | | | | Intra-Vascular | | | | | | | | | | | | Peripheral -Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Muscular-Skeletal<br>Conventional | | | | | | | | | | | | Muscular-Skeletal<br>Superficial | | | | | | | | | | | | Others(Specify) | | | | | | | | | | | #### Mode of Operation (*includes simultaneous B-mode) N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes: Note 1: PWD/Color Doppler, PWD/Power Doppler Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 5: Abdominal organs and peripheral vessel Note 6: 3D Imaging Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) Daniel L. Segmen productive, Abdominal, and Rac {12}------------------------------------------------ 510(k) Number: SonoAce 6000C / Ultramark 400C Ultrasound System Device Name: HC3-6, 3.5MHz/60R/60D, Curved Linear Array Probe Transducer: Indications for Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) | |-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------------|------------------------|--------------------| | Ophthalmic | | | | | | | | | | | | Fetal (See Note 3) | | N | N | N | | N | N | | Note 1 | Notes 2, 6 | | Abdominal | | N | N | N | | N | N | | Note 1 | Notes 2, 6 | | Intra-Operative<br>(See Note 5) | | | | | | | | | | | | Intra-Operative<br>Neurological | | | | | | | | | | | | Pediatric | | N | N | N | | N | N | | Note 1 | Notes 2, 6 | | Small Organ<br>(See Note 4) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Trans-Rectal | | | | | | | | | | | | Trans-Vaginal | | | | | | | | | | | | Trans-Urethral | | | | | | | | | | | | Intra-Vascular | | | | | | | | | | | | Peripheral -Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Muscular-Skeletal<br>Conventional | | | | | | | | | | | | Muscular-Skeletal<br>Superficial | | | | | | | | | | | | Others(Specify) | | | | | | | | | | | Mode of Operation (*includes simultaneous B-mode) N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes: Note 1: PWD/Color Doppler, PWD/Power Doppler Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 5: Abdominal organs and peripheral vessel Note 6: 3D Imaging Concurrence of CDRH, Office of Device Evaluation (ODE) | / Prescription Use (Per 21 CFR 801.109) David A. Syum Reproducti and Radiological {13}------------------------------------------------ 510(k) Number: SonoAce 6000C / Ultramark 400C Ultrasound System Device Name: C4-7, 5.0MHz/40R/60D, Curved Linear Array Probe Transducer: Indications for Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Mode of Operation (*includes simultaneous B-mode) | Clinical Application | A | B | M* | PWD* | CWD* | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) | |-----------------------------------|---|---|----|------|------|-------------------|----------------------------|------------------------------|------------------------|--------------------| | Ophthalmic | | | | | | | | | | | | Fetal (See Note 3) | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Abdominal | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Intra-Operative<br>(See Note 5) | | | | | | | | | | | | Intra-Operative<br>Neurological | | | | | | | | | | | | Pediatric | | N | N | N | | N | N | | Note 1 | Notes 2, 6 | | Small Organ<br>(See Note 4) | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Trans-Rectal | | | | | | | | | | | | Trans-Vaginal | | | | | | | | | | | | Trans-Urethral | | | | | | | | | | | | Intra-Vascular | | | | | | | | | | | | Peripheral -Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Muscular-Skeletal<br>Conventional | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Muscular-Skeletal<br>Superficial | | P | P | | | | | | | | | Others(Specify) | | | | | | | | | | | N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes: Note 1: PWD/Color Doppler, PWD/Power Doppler Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 5: Abdominal organs and peripheral vessel Note 6: 3D Imaging Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) David le. Seyton Division Sign-Off) Division of Reproductive, Abdominal, E. and Radiological De 510(k) Number K9902 {14}------------------------------------------------ | 510(k) Number: | | |----------------------|------------------------------------------------------------------------------------| | Device Name: | SonoAce 6000C / Ultramark 400C Ultrasound System | | Transducer: | C4-9/10R, 6.5MHz/10R/140D, Curved Linear Array Probe | | Indications for Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | Mode of Operation (*includes simultaneous B-mode) | The fulles A B M* PWD* CWD* Color . Power Color Combined Clinical Application | Clinical Application | A | B | M* | PWD* | CWD* | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) | |-----------------------------------|---|---|----|------|------|-------------------|----------------------------|------------------------------|------------------------|--------------------| | Ophthalmic | | | | | | | | | | | | Fetal (See Note 3) | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Abdominal | | | | | | | | | | | | Intra-Operative<br>(See Note 5) | | | | | | | | | | | | Intra-Operative<br>Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ<br>(See Note 4) | | P | P | N | | N | N | | Note 1 | Note 6 | | Neonatal Cephalic | | P | P | N | | N | N | | Note 1 | Note 6 | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Trans-Rectal | | N | N | N | | N | N | | Note 1 | Notes 2, 6 | | Trans-Vaginal | | N | N | N | | N | N | | Note 1 | Notes 2, 6 | | Trans-Urethral | | | | | | | | | | | | Intra-Vascular | | | | | | | | | | | | Peripheral -Vascular | | P | P | N | | N | N | | Note 1 | Note 6 | | Laparoscopic | | | | | | | | | | | | Muscular-Skeletal<br>Conventional | | | | | | | | | | | | Muscular-Skeletal<br>Superficial | | | | | | | | | | | | Others(Specify) | | | | | | | | | | | N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes Note 1: PWD/Color Doppler, PWD/Power Doppler Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 5: Abdominal organs and peripheral vessel Note 6: 3D Imaging Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) David R. Segerson --- (Division Sign Off) (Division Sign-Off) Division of Reproductive, Abdominal, E. and Radiological Devices Radiological Devices 510(k) Number K990970 {15}------------------------------------------------ 510(k) Number: SonoAce 6000C / Ultramark 400C Ultrasound System Device Name: EC 4-9/10R, 6.5MHz/10R/140D (Endocavity), Curved Linear Array Probe Transducer: Indications for Use:Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Mode of Operation (*includes simultaneous B-mode) | Clinical Application | A | B | M* | PWD* | CWD | Color<br>Doppler* | Power<br>(Amp)<br>Doppler* | Color<br>Velocity<br>Imaging | Combined<br>(Specify)* | Other<br>(Specify) | |-----------------------------------|---|---|----|------|-----|-------------------|----------------------------|------------------------------|------------------------|--------------------| | Ophthalmic | | | | | | | | | | | | Fetal (See Note 3) | | N | N | N | | N | N | | Note 1 | Notes 2, 6 | | Abdominal | | | | | | | | | | | | Intra-Operative<br>(See Note 5) | | | | | | | | | | | | Intra-Operative<br>Neurological | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ<br>(See Note 4) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Transesophageal | | | | | | | | | | | | Trans-Rectal | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Trans-Vaginal | | P | P | N | | N | N | | Note 1 | Notes 2, 6 | | Trans-Urethral | | | | | | | | | | | | Intra-Vascular | | | | | | | | | | | | Peripheral -Vascular | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Muscular-Skeletal<br>Conventional | | | | | | | | | | | | Muscular-Skeletal<br>Superficial | | | | | | | | | | | | Others(Specify) | | | | | | | | | | | N = new indication; P = previously cleared in K981510, SA 6000/ UM 400 system; E = added under Appendix E Other Indications or Modes Note 1: PWD/Color Doppler, PWD/Power Doppler Note 2: Includes imaging for guidance of biopsy Note 3: Includes infertility monitoring of follicle development Note 4: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients Note 5: Abdominal organs and peripheral vessel Note 6: 3D Imaging Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) Toril A. Sogn Reproductive, Abdominal, olnoical D and Rad 510(k) Number
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...