← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K251853

# Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System (K251853)

_Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · IYN · Nov 7, 2025 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K251853

## Device Facts

- **Applicant:** Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Nov 7, 2025
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Intended Use

The Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System is applicable for for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional), musculo-skeletal (superficial), thoracic/pleural, cardiac adult, cardiac pediatric, peripheral vessel and urology, intra-operative (abdominal, thoracic, and vascular) exams. This device is a general purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid, which is intended to be used in a hospital or medical clinic. Modes of operation include: B, M, PWD, CWD, Color Doppler, Amplitude Doppler, Combined mode (B+M, PW+B, Color+B, Power+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, iScape View, TDI, Color M, Smart 3D, 4D (Real-time 3D), Strain Elastography, Contrast imaging (Contrast agent for LVO), Contrast imaging (Contrast agent for Liver).

## Device Story

General-purpose, mobile, software-controlled diagnostic ultrasound system; acquires and displays ultrasound images. Inputs: ultrasonic energy transmitted into patients; received echoes processed to generate anatomical/fluid flow images. Outputs: onscreen display of structures/flow; anatomical measurements; analysis packages. Used in hospitals/clinics by trained healthcare professionals. Healthcare providers use output for diagnosis and clinical decision-making. Benefits: non-invasive visualization of internal structures and fluid flow to aid clinical diagnosis.

## Clinical Evidence

Bench testing only. No clinical data required. Evidence includes acoustic output measurements (NEMA UD 2), biocompatibility (ISO 10993-1), and electrical/mechanical safety testing (IEC 60601 series).

## Technological Characteristics

Mobile ultrasound system; ultrasonic pulsed Doppler/echo imaging. Materials: patient-contact parts (transducers/brackets) compliant with ISO 10993-1. Connectivity: standard ultrasound interfaces. Software: IEC 62304 compliant. Sterilization: cleaning/disinfection per manufacturer instructions. Energy: ultrasonic energy.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Consona N9 ([K221300](/device/K221300.md))

## Reference Devices

- Resona I9 ([K240684](/device/K240684.md))
- Resona I9 ([K210699](/device/K210699.md))
- MX6 ([K250110](/device/K250110.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA U.S. FOOD &amp; DRUG ADMINISTRATION

November 7, 2025

Shenzhen Mindray Bio-medical Electronics Co., LTD.
Lu Chen
Engineer, Technical Regulation Department
Mindray Building, Keji 12th Road South,
Hi-Tech Industrial Park, Nanshan
Shenzhen, Guangdong 518057
China

Re: K251853

Trade/Device Name: Diagnostic Ultrasound System (Consona N5); Diagnostic Ultrasound System (Consona N5 Pro); Diagnostic Ultrasound System (Consona N5 Super); Diagnostic Ultrasound System (Consona N5 Exp); Diagnostic Ultrasound System (Consona ER); Diagnostic Ultrasound System (Consona N5S); Diagnostic Ultrasound System (Consona EI); Diagnostic Ultrasound System (Consona EF); Diagnostic Ultrasound System (Consona N5 Elite); Diagnostic Ultrasound System (Consona N5 Ultra); Diagnostic Ultrasound System (Consona N3); Diagnostic Ultrasound System (Consona N3 Exp)

Regulation Number: 21 CFR 892.1550
Regulation Name: Ultrasonic Pulsed Doppler Imaging System
Regulatory Class: Class II
Product Code: IYN, IYO, ITX
Dated: October 22, 2025
Received: October 22, 2025

Dear Lu Chen:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

U.S. Food &amp; Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

{1}

K251853 - Lu Chen
Page 2

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-

{2}

K251853 - Lu Chen
Page 3

devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

**YANNA S. KANG -S**

Yanna Kang, Ph.D.
Assistant Director
Mammography and Ultrasound Team
DHT8C: Division of Radiological Imaging and Radiation Therapy Devices
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

{3}

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.

Submission Number (if known)

K251853

Device Name

Diagnostic Ultrasound System (Consona N5);
Diagnostic Ultrasound System (Consona N5 Pro);
Diagnostic Ultrasound System (Consona N5 Super);
Diagnostic Ultrasound System (Consona N5 Exp);
Diagnostic Ultrasound System (Consona ER);
Diagnostic Ultrasound System (Consona N5S);
Diagnostic Ultrasound System (Consona EI);
Diagnostic Ultrasound System (Consona EF);
Diagnostic Ultrasound System (Consona N5 Elite);
Diagnostic Ultrasound System (Consona N5 Ultra);
Diagnostic Ultrasound System (Consona N3);
Diagnostic Ultrasound System (Consona N3 Exp)

Indications for Use (Describe)

The Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ(breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional), musculo-skeletal (superficial), thoracic/ pleural, cardiac adult, cardiac pediatric, peripheral vessel and urology, intra-operative (abdominal, thoracic, and vascular) exams.

This device is a general purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid, which is intended to be used in a hospital or medical clinic.

Modes of operation include: B, M, PWD, CWD, Color Doppler, Amplitude Doppler, Combined mode (B+M, PW+B, Color+B, Power+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, iScape View, TDI, Color M, Smart 3D, 4D(Real-time 3D), Strain Elastography, Contrast imaging (Contrast agent for LVO), Contrast imaging (Contrast agent for Liver).

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov

“An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.”

1/1

{4}

510(K) SUMMARY

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92(c).

The assigned 510(k) number: K251853

1. **Submitter**

Shenzhen Mindray Bio-medical Electronics Co., LTD
Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Tel: +86 755 8188 5655
Fax: +86 755 2658 2680

**Contact Person**

Chen Lu
Shenzhen Mindray Bio-medical Electronics Co., LTD
Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
**Date Prepared**: May 16, 2025

2. **Device Name**

Diagnostic Ultrasound System (Consona N5);
Diagnostic Ultrasound System (Consona N5 Pro);
Diagnostic Ultrasound System (Consona N5 Super);
Diagnostic Ultrasound System (Consona N5 Exp);
Diagnostic Ultrasound System (Consona ER);
Diagnostic Ultrasound System (Consona N5S);
Diagnostic Ultrasound System (Consona EI);
Diagnostic Ultrasound System (Consona EF);
Diagnostic Ultrasound System (Consona N5 Elite);
Diagnostic Ultrasound System (Consona N5 Ultra);
Diagnostic Ultrasound System (Consona N3);
Diagnostic Ultrasound System (Consona N3 Exp)

**Classification**

Regulatory Class: II

{5}

Review Category: Tier II

21 CFR 892.1550 Ultrasonic Pulsed Doppler Imaging System (IYN)
21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (IYO)
21 CFR 892.1570 Diagnostic Ultrasound Transducer (ITX)

## 3. Predicate devices

Consona N5/N3 series Diagnostic Ultrasound System is comparable with and substantially equivalent to the predicate listed below. Consona N9 is the main predicate device.

|  Device | Manufacturer | Model | Device Class | Product Code | 510K Number  |
| --- | --- | --- | --- | --- | --- |
|  1. Main predicate device | Mindray | Consona N9 | II | IYN, IYO, ITX | K221300  |
|  2. Reference device | Mindray | Resona I9 | II | IYN, IYO, ITX | K240684  |
|  3. Reference device | Mindray | Resona I9 | II | IYN, IYO, ITX | K210699  |
|  4. Reference device | Mindray | MX6 | II | IYN, IYO, ITX | K250110  |

The result shows the conformance of subject device to the predicate.

Regulation name and code

21 CFR 892.1550 Ultrasonic Pulsed Dopple Imaging System(IYN)
21 CFR 892.1560 Ultrasonic Pulsed Echo Imaging System (IYO)
21 CFR 892.1570 Diagnostic Ultrasound Transducer (ITX)

## 4. Device Description:

The Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System is a general purpose, mobile, software controlled, ultrasonic diagnostic system. Its function is to acquire and display ultrasound images.

Modes of operation include: B, M, PWD, CWD, Color Doppler, Amplitude Doppler, Combined mode (B+M, PW+B, Color+B, Power+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, iScape View, TDI, Color M, Smart 3D, 4D(Real-time 3D), Strain Elastography, Contrast imaging (Contrast agent for LVO), Contrast imaging (Contrast agent for Liver).

The Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System can also measure anatomical structures and offer analysis packages to provide information based on which the competent health care professionals can make the diagnosis.

{6}

5. Intended Use:

The Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System is applicable for for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional), musculo-skeletal (superficial), thoracic/pleural, cardiac adult, cardiac pediatric, peripheral vessel and urology, intra-operative (abdominal, thoracic, and vascular) exams.

This device is a general purpose diagnostic ultrasound system intended for use by qualified and trained healthcare professionals for ultrasound imaging, measurement, display and analysis of the human body and fluid, which is intended to be used in a hospital or medical clinic.

Modes of operation include: B, M, PWD, CWD, Color Doppler, Amplitude Doppler, Combined mode (B+M, PW+B, Color+B, Power+B, PW+Color+B, Power+PW+B), Tissue Harmonic Imaging, iScape View, TDI, Color M, Smart 3D, 4D (Real-time 3D), Strain Elastography, Contrast imaging (Contrast agent for LVO), Contrast imaging (Contrast agent for Liver).

6. Comparison with Predicate Devices:

Subject device Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System is comparable with and substantially equivalent to the predicate device with regards to intended use, imaging modes, features and functions and technological characteristics.

- All systems transmit ultrasonic energy into patients, perform post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body. All systems allow for specialized measurements of structures and flow, as well as calculations.
- Subject device Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System has the same intended uses as the predicate Consona N9 (K221300).
- The patient contact materials of the transducers and needle-guided brackets of

{7}

subject device Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System are the same to the predicate or tested under ISO 10993-1.

- The acoustic power levels of Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System are levels specified in the FDA guidance document, which are the same as the predicate device Consona N9 (K221300).
- Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System is designed in compliance with the FDA recognized electrical and physical safety standards, which are the same as the predicate device Consona N9 (K221300).
- Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System has the equivalent features and functions as the predicate device Consona N9 (K221300).

The main differences are between the subject and predicate

|   | Subject Device ConsonaN5/N3 series | Main predicate device Consona N9 (K221300) | Reference device Resona I9 (K240684) | Reference device Resona I9 (K210699) | Reference device MX6 (K250110)  |
| --- | --- | --- | --- | --- | --- |
|  Indications for use | It is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ(breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional), | It is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, Intra-operative (includes abdominal, thoracic, and vascular), Laparoscopic, pediatric, small organ (breast, thyroid, testes), | It is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, Intra-operative (includes abdominal, thoracic, and vascular), Laparoscopic, pediatric, small organ (breast, thyroid, testes), neonatal and adult cephalic, trans-rectal, | It is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in ophthalmic, fetal, abdominal, pediatric, small organ(breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal(conventi | It is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in ophthalmic, fetal, abdominal, pediatric, small organ(breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal(conventi  |
|   | recent, and in the case of children, and the child is not a child. It is a child, and the child is not a child. It is a child, and the child is not a child. | recent, and in the case of children, and the child is not a child. It is a child, and the child is not a child. It is a child, and the child is not a child. | recent, and in the case of children, and the child is not a child. It is a child, and the child is not a child. | recent, and in the case of children, and the child is not a child. | recent, and in the case of children, and the child is not a child.  |

{8}

|   | musculo-skeletal (superficial), thoracic/pleural, cardiac adult, cardiac pediatric, peripheral vessel, and urology, intra-operative (abdominal, thoracic, and vascular) exams. | vaginal, musculo-skeletal(conventional), Musculo skeletal(superficial), Thoracic/Pleural, cardiac adult, cardiac pediatric, Trans-esoph., peripheral vessel and urology exams. | neonatal and adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), Thoracic/Pleural, adult and pediatric cardiac, trans-esoph. (Cardiac), peripheral vessel, urology and Thoracic/Pleural exams. | trans-vaginal, musculo-skeletal (conventional, superficial), adult and pediatric cardiac, trans-esoph. (Cardiac), peripheral vessel, urology and Thoracic/Pleural exams. | onal), musculoskeletal(superficial), thoracic/pleural, cardiac adult, cardiac pediatric, peripheral vessel and urology, intra-operative(abdominal, thoracic, and vascular), trans-esoph(Cardiac) exams.  |
| --- | --- | --- | --- | --- | --- |
|  Difference Description | All intended uses of the subject device have already been cleared on the main predicate device. The predicate device has more functions than the subject device.  |   |   |   |   |
|  |   |   |   |   |   |
|   | Subject Device Consona N5/N3 series | Main predicate device Consona N9 (K221300) | Reference device Resona I9 (K240684) | Reference device Resona I9 (K210699) | Reference device MX6 (K250110)  |
|  Options hardware | Printer fixed bracket | / | / | / | /  |
|  Difference Description | Most of the hardware has been cleared on the main predicate device (K221300) The Ultrasound gel warmer has been cleared on the reference device (K210699) The Probe Adapter PCM-CWm01 has been cleared on the reference device (K250110) The subject device includes 1 newly added hardware, which is: Printer Fixed Bracket: Used for securely mounting the printer to the system, enhancing stability during operation.  |   |   |   |   |
|  |   |   |   |   |   |
|   | Subject Device Consona N5/N3 series | Main predicate device Consona N9 (K221300) | Reference device Resona I9 (K240684) | Reference device Resona I9 (K210699) | Reference device MX6 (K250110)  |
|  Options software | / | / | / | / | /  |
|  Difference Description | All the software of the subject device has been cleared in the main predicate device(K221300).  |   |   |   |   |
|  |   |   |   |   |   |

{9}

|  Items | Subject Device Consona N5/N3 series | Main predicate device Consona N9 (K221300) | Reference device Resona I9 (K240684) | Reference device Resona I9 (K210699) | Reference device MX6 (K250110)  |
| --- | --- | --- | --- | --- | --- |
|  Transducers | / | / | / | / | /  |
|  Difference Description | All the probes have been cleared on the reference device MX6 (K250110).  |   |   |   |   |
|  |   |   |   |   |   |
|  Items | Subject Device Consona N5/N3 series | Main predicate device Consona N9 (K221300) | Reference device Resona I9 (K240684) | Reference device Resona I9 (K210699) | Reference device MX6 (K250110)  |
|  Needle-guided brackets | / | / | / | / | /  |
|  Difference Description | All the Needle-guided brackets have been on the reference device MX6 (K250110).  |   |   |   |   |
|  |   |   |   |   |   |
|   | Subject Device Consona N5/N3 series | Main predicate device Consona N9 (K221300) | Reference device Resona I9 (K240684) | Reference device Resona I9 (K210699) | Reference device MX6 (K250110)  |
|  Accessories | Multilingual control panel overlay | / | / | / | /  |
|  Difference Description | All accessories of the subject device have already been cleared on the main predicate device (K221300) except the Multilingual control panel overlay, which is a layer that displays control labels in different languages.  |   |   |   |   |

## 7. Non-clinical Tests:

Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical and mechanical safety, and this device has been designed to conform with applicable medical safety standards.

This device has been tested and evaluated under the following standards:

{10}

- NEMA UD 2-2004 (R2009), acoustic output measurement standard for diagnostic ultrasound equipment revision 3.
- ANSI AAMI ES60601-1:2005/(R)2012 &amp; A1:2012, C1:2009/(R)2012 &amp; A2:2010/(R)2012 (Cons. Text) [Incl. AMD2:2021] Medical electrical equipment - Part 1: General requirements for basic safety and essential performance (IEC 60601-1:2005, MOD) [Including Amendment 2 (2021)].
- IEC 60601-1-2 Edition 4.1 2020-09 CONSOLIDATED VERSION, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests.
- IEC TS 60601-4-2 Edition 1.0 2024-03 Medical electrical equipment - Part 4-2: Guidance and interpretation - Electromagnetic immunity: performance of medical electrical equipment and medical electrical systems.
- IEC 60601-1-6 Edition 3.2 2020-07 CONSOLIDATED VERSION, Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability.
- IEC 60601-2-37 Edition 2.1 2015, Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment.
- ISO 14971 Third Edition 2019-12, Medical devices - Application of risk management to medical devices.
- ISO 10993-1 Fifth edition 2018-08, biological evaluation of medical devices - part 1: evaluation and testing within a risk management process.
- IEC 62304 Edition 1.1 2015-06, medical device software - software life cycle processes.
- IEC 62366-1 Edition 1.1 2020-06 CONSOLIDATED VERSION, Medical devices - Part 1: Application of usability engineering to medical devices.

These non-clinical tests relied on in this premarket notification submission can support the determination of substantial equivalence of the subject device.

## 8. Clinical Studies

Not applicable. The subject of this submission, Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System, does not require clinical studies to support substantial equivalence.

## 9. Summary

{11}

Based on the performance data as documented in the study, Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System was found to have a safety and effectiveness profile that is similar to the predicate device.

## 10. Conclusion:

Intended uses and other key features are consistent with traditional clinical practices, FDA guidelines, and established methods of patient examination. The design, development and quality process of the manufacturer conforms with 21 CFR 820, ISO 9001 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards. Therefore, the Consona N5, Consona N5 Pro, Consona N5 Super, Consona N5 Exp, Consona ER, Consona N5S, Consona EI, Consona EF, Consona N5 Elite, Consona N5 Ultra, Consona N3, Consona N3 Exp Diagnostic Ultrasound System is substantially equivalent with respect to safety and effectiveness to the predicate device.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K251853](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K251853)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com