VINNO 6PRO, VINNO 6EXP, VINNO 5PRO, VINNO 5EXP, VINNO 3, VINNO 3PRO, VINNO 3EXP

K221911 · Vinno Technology (Suzhou) Co.,Ltd · IYN · Dec 21, 2022 · Radiology

Device Facts

Record IDK221911
Device NameVINNO 6PRO, VINNO 6EXP, VINNO 5PRO, VINNO 5EXP, VINNO 3, VINNO 3PRO, VINNO 3EXP
ApplicantVinno Technology (Suzhou) Co.,Ltd
Product CodeIYN · Radiology
Decision DateDec 21, 2022
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric

Intended Use

The device is general purpose diagnostic ultrasound system for use by qualified healthcare professionals. It is applicable for adults, pregnant women, pediatric patients and neonates. The device is intended for ultrasound imaging, measurement and analysis of human body and fluid for multiple clinical applications including: abdominal (GYN and Urology), Thoracic/Pleural, Fetal/Ob, small organ (including breast, thyroid, testes), peripheral vessel, neonatal cephalic, pediatric, musculo-skeletal (conventional, superficial), transrectal, trans-vaginal, cardiac pediatric, magnetic Needle guidance and imaging guidance of interventional procedures (e.g. biopsy). This device is intended to use by, or by the order of, and under the supervision of a licensed physician qualified to direct the use of the device. The device is used in hospital, clinical point-of-care for diagnosis of patients. The operating modes supported by the device are B, M,PWD CWD, Tissue Doppler, Color M Doppler, Power Doppler, Tissue Velocity Imaging, Harmonic Imaging, 3D/4D, CBI, EI, Combine modes.

Device Story

Laptop-based digital color ultrasonic diagnostic system; transmits ultrasound waves into body tissues; displays echo images of tissues and blood flow. Operates via integrated battery or external power supply. Used in hospitals, clinics, and point-of-care settings by qualified healthcare professionals/physicians. Input: ultrasound signals via probes. Processing: digital acquisition and image processing. Output: real-time diagnostic images, measurements, and analysis. Clinical utility: aids diagnosis, measurement, and interventional guidance (e.g., biopsy). Benefits: provides non-invasive visualization of internal structures and blood flow for various clinical applications.

Clinical Evidence

Bench testing only. No clinical data provided. Performance supported by biocompatibility testing, electrical safety/EMC testing, and acoustic output measurements compliant with NEMA UD2 and IEC standards.

Technological Characteristics

Laptop-based ultrasound system. Probes consist of shell and lens materials (biocompatibility verified per ISO 10993-1). Energy source: RF radiation/ultrasound energy. Connectivity: integrated battery or external power supply. Software: moderate level of concern. Imaging modes: B, M, PWD, CWD, Tissue Doppler, Color Doppler, Power Doppler, 3D/4D, etc. Complies with IEC 60601-1, IEC 60601-2-37, and IEC 60601-1-2.

Indications for Use

Indicated for ultrasound imaging, measurement, and analysis of human body and fluid in adults, pregnant women, pediatric patients, and neonates. Clinical applications include abdominal (GYN/Urology), thoracic/pleural, fetal/OB, small organs (breast, thyroid, testes), peripheral vessels, neonatal/adult cephalic, pediatric, musculoskeletal, transrectal, transvaginal, and cardiac (adult/pediatric). Supports interventional guidance (e.g., biopsy, magnetic needle).

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ December 21, 2022 VINNO Technology (Suzhou) Co., Ltd. % Cordelia Liu Regulatory Registered Engineer 5F Building A, 4F Building C No. 27 Xinfa Rd. Suzhou Industrial Park Suzhou, Jiangsu 215123 CHINA Re: K221911 Trade/Device Name: VINNO 6PRO, VINNO 6EXP, VINNO 5PRO, VINNO 5EXP, VINNO 3, VINNO 3PRO, VINNO 3EXP Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic Pulsed Doppler Imaging System Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: November 9, 2022 Received: November 21, 2022 Dear Cordelia Liu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Yanna S. Kang -S Yanna Kang, Ph.D. Assistant Director Mammography and Ultrasound Team DHT8C: Division of Radiological Imaging and Radiation Therapy Devices OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K221911 #### Device Name VINNO 6EXP, VINNO 6PRO, VINNO 5EXP, VINNO 5PRO, VINNO 3, VINNO 3EXP, VINNO 3PRO Indications for Use (Describe) The device is general purpose diagnostic ultrasound system for use by qualified healthcare professionals. It is applicable for adults, pregnant women, pediatric patients and neonates. The device is intended for ultrasound imaging, measurement and analysis of human body and fluid for multiple clinical applications including: abdominal (GYN and Urology), Thoracic/Pleural, Fetal/Ob, small organ (including breast, thyroid, testes), peripheral vessel, neonatal cephalic, pediatric, musculo-skeletal (conventional, superficial), transrectal, trans-vaginal, cardiac pediatric, magnetic Needle guidance and imaging guidance of interventional procedures (e.g. biopsy). This device is intended to use by, or by the order of, and under the supervision of a licensed physician qualified to direct the use of the device. The device is used in hospital, clinical point-of-care for diagnosis of patients. The operating modes supported by the device are B, M,PWD CWD, Tissue Doppler, Color M Doppler, Power Doppler, Tissue Velocity Imaging, Harmonic Imaging, 3D/4D, CBI, EI, Combine modes. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| | <span style="font-size: 10pt"> <span style="font-family: Symbol">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 10pt"> <span style="font-family: Symbol">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------| |----------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------| #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### K221911 # Section 5 510(k) summary #### l Submitter Device submitter: VINNO Technology (Suzhou) Co., Ltd. 5F Building A. 4F Building C No. 27 XinFa Rd. Suzhou Industrial Park, SuZhou 215123 Jiangsu China Contact person: Cordelia.liu Regulatory Registered Engineer Phone: +86 512 62873806 Fax: +86 512 62873801 Email: cordelia.liu@vinno.com Date written: 2022-12-19 #### II Device Trade Name of Device: VINNO 6EXP,VINNO 6PRO,VINNO 5EXP,VINNO 5PRO, VINNO 3, VINNO 3EXP, VINNO 3PRO Regulation name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1550 Requlatory Class: II Product code: IYN, IYO, ITX #### III Predicate Devices Trade name: VINNO 8, VINNO 6, VINNO 5 Regulation Name: Ultrasonic pulsed doppler imaging system Regulation Number: 21 CFR 892.1550 Regulatory Class: II Product code: IYN, IYO, ITX Premarket Notification: k190120 #### IV Device description The VINNO 6EXP, VINNO 6PRO, VINNO 5EXP, VINNO 5PRO, VINNO 3, VINNO 3EXP, VINNO 3PRO ultrasound devices are laptop digital color ultrasonic diagnostic devices which transmit ultrasound waves into the body tissues and display the echo images of the tissues and blood flow accordingly. The devices are capable of digital acquisition, processing and display and operate from an integrated battery or separate power supply/charger. {4}------------------------------------------------ #### V Indications for use The device is general purpose diagnostic ultrasound system for use by qualified healthcare professionals. It is applicable for adults, pregnant women, pediatric patients and neonates. The device is intended for ultrasound imaging, measurement and analysis of human body and fluid for multiple clinical applications including: abdominal (GYN and Urology), Thoracic/ Pleural, Fetal/Ob, small organ (including breast, thyroid, testes), peripheral vessel, neonatal cephalic, adult cephalic, pediatric, musculo-skeletal (conventional, superficial), trans-rectal, trans-vaginal, cardiac adult, cardiac pediatric, magnetic Needle guidance and imaging guidance of interventional procedures (e.g. biopsy). This device is intended to use by, or by the order of, and under the supervision of a licensed physician qualified to direct the use of the device. The device is used in hospital, clinics and clinical point-of-care for diagnosis of patients. The operating modes supported by the device are B, M,PWD CWD, Tissue Doppler, Color Doppler, Color M Doppler, Power Doppler, Tissue Velocity Imaging, Harmonic Imaging, 3D/4D, CBI, El, Combine modes. ## VI Comparison of technological characteristics with the predicate devices The VINNO 6EXP, VINNO 6PRO, VINNO 5EXP, VINNO 5PRO, VINNO 3, VINNO 3EXP, VINNO 3PRO ultrasound devices have the same technological characteristics and fundamental design as the predicate devices. The VINNO 6, VINNO 6, VINNO 5 ultrasound devices and the predicate device are all lap-top general purpose ultrasound devices designed to provide real-time images for diagnosis. The differences between the VINNO 8, VINNO 6, VINNO 5 ultrasound devices and predicate devices do not alter suitability of the proposed device for its intended use. See the following table for details on the comparison of technical features with the predicate device: | Technological<br>characteristics | Subject device | Predicate device | Comparison | |----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------| | Indications for use | The device is intended for<br>ultrasound imaging,<br>measurement and analysis of<br>human body and fluid for multiple<br>clinical applications including: | The device is intended for<br>ultrasound imaging,<br>measurement and analysis of<br>human body and fluid for multiple<br>clinical applications including: | No<br>difference | | | abdominal (GYN and Urology),<br>Thoracic/Pleural, Fetal/Ob, small<br>organ (including breast, thyroid,<br>testes), peripheral vessel,<br>neonatal cephalic, adult cephalic,<br>pediatric, musculo-skeletal<br>(conventional, superficial),<br>trans-rectal, trans-vaginal,<br>cardiac adult, cardiac pediatric,<br>magnetic Needle guidance and<br>imaging guidance of<br>interventional procedures (e.g.<br>biopsy). | abdominal (GYN and Urology),<br>Thoracic/Pleural, Fetal/Ob, small<br>organ (including breast, thyroid,<br>testes), peripheral vessel,<br>neonatal cephalic, adult cephalic,<br>pediatric, musculo-skeletal<br>(conventional, superficial),<br>trans-rectal, trans-vaginal, cardiac<br>adult, cardiac pediatric, magnetic<br>Needle guidance and imaging guidance of<br>interventional procedures (e.g. biopsy). | | | Design | This device is designed for the<br>safety of the patient and operator.<br>This product is designed to fully<br>comply with the EN60601-1-2<br>(IEC60601-1-2),<br>Class A, in medical electric<br>equipment EMC regulations. | This device is designed for the<br>safety of the patient and operator.<br>This product is designed to fully<br>comply with the EN60601-1-2<br>(IEC60601-1-2),<br>Class A, in medical electric<br>equipment EMC regulations. | The subject<br>device was<br>consistent<br>with the<br>predicate<br>device in<br>terms of | | Material | The probe materials based on the biocompatibility test were all<br>composed of a shell and a lens, the probe of the predicate device<br>could cover the subject device, and the probe materials passed the<br>biocompatibility test. | | design,<br>material,<br>chemical<br>composition, | | Chemical<br>composition | NA<br>Our device does not involve the composition of chemical. | | and energy<br>source, and | | Energy source | This device can produce and use the radiation RF energy.<br>Ultrasound system utilizes ultrasound energy that must be coupled to<br>the patient by direct physical contact In order to assure optimal<br>transmission of energy between the patient and probe, a conductive<br>gel or couplant must be applied liberally to the patient where scanning<br>will be performed. | | there was<br>no<br>difference | | Imaging features | See Table 1 of the article for details | | The imaging<br>features of<br>the subject<br>device can<br>be covered<br>by<br>the<br>predicated<br>device | | Key functionalities | | | | | | Gain, Depth, Focus, TGC, B Steer,<br>2D Automatic Optimization,<br>Harmonic Imaging, L/R, U/D,<br>Frequency, PRF, Wall Filter,<br>Packet Size, Color Level, Invert,<br>Color Map, Line Density, Sync<br>Display, Persistence, Flash<br>Reduction, Base Line,<br>Transparency, Steer, Focus<br>Position, Focus Number, VFusion,<br>VSpeckle, Dynamic Range | Gain, Depth, Focus, TGC, B Steer,<br>2D Automatic Optimization,<br>Harmonic Imaging,<br>L/R, U/D, Frequency, PRF, Wall<br>Filter, Packet Size, Color Level,<br>Invert, Color Map, Line<br>Density, Sync Display, Persistence,<br>Flash Reduction, Base Line,<br>Transparency, Steer, Focus<br>Position, Focus Number, VFusion,<br>VSpeckle, Dynamic Range | No<br>difference | | Measurements | Depth, Distance, Perimeter, Area,<br>Volume, Angle, Stenosis, A/B<br>ratio, Time, Speed, Heart Rate,<br>Stenosis, A/B ratio, Acceleration,<br>PS, ED, TAMAX, TAMEAN, RI,<br>PI, PS.ED ratio, ED/PS ratio,<br>Flow volume, MaxPG, MeanPG,<br>Stroke Volume, Heart rate | Depth, Distance, Perimeter, Area,<br>Volume, Angle, Stenosis, A/B<br>ratio, Time, Speed, Heart Rate,<br>Stenosis, A/B ratio, Acceleration,<br>PS, ED, TAMAX, TAMEAN, RI,<br>PI, PS.ED ratio, ED/PS ratio,<br>Flow volume, MaxPG, MeanPG,<br>Stroke Volume, Heart rate | No<br>difference | | Probes | Convex array<br>Linear array<br>Phased array | Convex array<br>Linear array<br>Phased array | The subject<br>device was<br>consistent<br>with the<br>predicate<br>device in<br>terms of<br>probes | | Acoustic output | Comply with Track 3 limits:<br>Ispta.3 $≤$ 720mW/cm²<br>MI $≤$ 1.9 | Comply with Track 3 limits:<br>Ispta.3 $≤$ 720mW/cm²<br>MI $≤$ 1.9 | No<br>difference | {5}------------------------------------------------ {6}------------------------------------------------ ## VII Performance data The following performance data were provided in support of the substantial equivalence determination. # Biocompatibility testing Biocompatibility of the probes was evaluated in accordance with ISO 10993-1:2009. All evaluation acceptance criteria were met {7}------------------------------------------------ ## Electrical safety and electromagnetic compatibility (EMC) Electrical safety and EMC testing were conducted on the Ultrasound System. The system complies with the IEC 60601-1 and IEC 60601-2-37 for safety and the IEC 60601-1-2 for EMC. #### Software Verification and Validation Testing Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern. #### Acoustic output testing Acoustic output testing was performed according to NEMA UD2 and IEC60601-2-37. ## VIII Conclusion The VINNO 6EXP, VINNO 6PRO, VINNO 5EXP, VINNO 5PRO, VINNO 3, VINNO 3EXP, VINNO 3PRO ultrasound devices are substantially equivalent to their predicate devices. The non-clinical testing demonstrates that the device is as safe, as effective and performs as well as the legally marketed device. {8}------------------------------------------------ | Table 1 | | | | | | | | | |-------------------------------------|---------|------------|------------|------------|------------|------------|------------|---| | Product | VINNO 8 | VINNO 6EXP | VINNO 6PRO | VINNO 5EXP | VINNO 5PRO | VINNO 3EXP | VINNO 3PRO | | | 2D Imaging<br>Features | | | | | | | | | | Frequency range | 25Mhz | 25Mhz | 25Mhz | 25Mhz | 25Mhz | 25Mhz | 25Mhz | | | VFusion | X | X | X | X | X | X | X | | | VSpeckle I | X | X | X | X | X | X | X | | | VTissue | X | X | X | O | O | O | O | | | Auto optimization | X | X | X | X | X | X | X | | | Easy Compare | X | X | X | X | X | X | X | | | Duplex/Triplex | X | X | X | X | X | X | X | | | CW Mode | X | X | X | O | - | - | - | | | PW Mode | X | X | X | X | X | X | X | | | Zoom | X | X | X | X | X | X | X | | | FULL screen | X | X | X | X | X | X | X | | | B Mode | X | X | X | X | X | X | X | | | M Mode | X | X | X | X | X | X | X | | | CF Mode | X | X | X | X | X | X | X | | | PDI Mode | X | X | X | X | X | X | X | | | B+CF<br>simultaneously | X | X | X | X | X | X | X | | | TView | X | X | X | X | X | X | X | | | Pview | O | O | O | O | O | O | O | | | TD Mode | X | X | X | - | - | - | - | | | Tissue Velocity<br>Imaging (TVI) | O | O | O | - | - | - | - | | | Auto IMT | X | X | X | O | O | O | O | | | Auto EF | O | O | O | O | O | O | O | | | Auto Bladder | X | X | X | X | X | X | X | | | Elastography | O | O | O | - | - | - | - | | | Contrast Agent<br>imaging | O | - | - | - | - | - | - | | | Needle<br>enhancement | O | O | O | O | O | O | O | | | SGC (Scanline gain<br>compensation) | O | O | O | O | O | O | O | | | Stress Echo | X | X | O | O | O | O | O | | | Flow Volume | O | O | O | O | O | O | O | | | 3D/4D Imaging<br>Features | | | | | | | | | | 3D imaging | X | X | X | X | X | X | X | X | | Real-time<br>grayscale 4D | X | X | X | X | X | X | X | X | | Tomographic<br>display (Mcut) | X | X | X | O | O | O | O | O | | HQ (High Quality)<br>3D/4D | O | O | O | O | O | O | O | O | | STIC Mode | O | O | O | - | - | - | - | - | | Magic cut | X | O | O | O | O | O | O | O | | Free view | X | O | O | O | O | O | O | O | | Free 3D | O | O | O | O | O | O | O | O | | Export 3D data for<br>3D printer | X | O | O | O | O | O | O | O | {9}------------------------------------------------
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...