← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K213102

# ES1 System (K213102)

_E-Scopics · IYN · Jan 11, 2022 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K213102

## Device Facts

- **Applicant:** E-Scopics
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Jan 11, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Intended Use

The ES1 Ultrasound Diagnostic System is intended for general purpose pulse echo ultrasound imaging, soft tissue elasticity imaging of the human body and provides measurements of shear wave speed and tissue stiffness, ultrasound beam attenuation and estimates of speed of sound, in internal structures of the body. This device is intended to be used by trained healthcare professionals, in a healthcare environment.

## Device Story

Ultraportable ultrasound system; consists of software app on consumer off-the-shelf (COTS) host and curved array probe (e.C5-1). Probe uses mechanical vibrator to induce shear waves in tissue; 2D B-mode ultrasound tracks shear wave propagation to estimate tissue stiffness. System also performs B-mode imaging to measure ultrasound attenuation, speed of sound, and brightness ratios. Operated by trained healthcare professionals in clinical environments via multi-touch screen. Output displayed on host monitor; includes images, quantitative measurements, and printable reports. Aids clinicians in diagnosing and monitoring liver disease by providing real-time anatomical visualization alongside quantitative tissue property data.

## Clinical Evidence

Bench testing only. Accuracy and precision assessed on calibrated phantoms. Shear wave speed/stiffness wCV 0.1-6.0%, bias -15.7% to 11.7%. Attenuation precision 1.5-4.5%, bias -22.2% to 1.4%. Speed of sound precision 0.05-0.3%, bias -0.7% to 0.7%. High correlation (Pearson 0.997-0.999) with predicate device. Acoustic output (MI, TI, ISPTA3) within FDA Track 1 limits. Feasibility confirmed on limited subjects in clinical setting.

## Technological Characteristics

Curved array probe (e.C5-1) with mechanical vibrator. Diverging wave transmission/reception. B-mode imaging, 2D-Transient Elastography (2D-TE), shear wave speed, attenuation, and speed of sound estimation. COTS host computer. Software-based processing. Complies with IEC 60601-1, IEC 60601-2-37, IEC 62359, ISO 10993-1, IEC 62304.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- FibroScan Family of Products ([K203273](/device/K203273.md))

## Reference Devices

- AIXPLORER MACH 20 & 30 Ultrasound Diagnostic Systems ([K191007](/device/K191007.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The Department of Health & Human Services logo is a stylized representation of a human figure, while the FDA part includes the acronym "FDA" in a blue square and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue font.

E-Scopics Claude Cohen-Bacrie CEO 931 Chemin De La Bosque D'Antonelle Aix-en-Provence, PACA 13090 France

Re: K213102

January 25, 2022

Trade/Device Name: ES1 System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX

Dear Claude Cohen-Bacrie:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter for your device cleared on January 11, 2022. Specifically, FDA is updating this SE letter due to a typo in the clearance date, which was incorrectly dated as January 11, 2021.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jessica Lamb, OHT7: Office of In Vitro Diagnostics and Radiological Health,

Sincerely.

Jessica Lamb

Jessica Lamb Assistant Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains two logos. On the left is the seal for the Department of Health & Human Services - USA. On the right is the logo for the U.S. Food & Drug Administration (FDA). The FDA logo has the letters FDA in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.

E-Scopics % Claude Cohen-Bacrie, CEO 931 chemin de la Bosque d'Antonelle 13090 Aix-en-Provence FRANCE

Re: K213102

January 11, 2021

Trade/Device Name: ES1 System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: December 3, 2021 Received: December 9, 2021

Dear Claude Cohen-Bacrie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809 ); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{2}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jessica Lamb

Jessica Lamb, Ph.D. Assistant Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known) K213102

Device Name ES1 System

#### Indications for Use (Describe)

The ES1 Ultrasound Diagnostic System is intended for general purpose pulse echo ultrasound imaging, soft tissue elasticity imaging of the human body and provides measurements of shear wave speed and tissue stiffness, ultrasound beam attenuation and estimates of speed of sound, in internal structures of the body. This device is intended to be used by trained healthcare professionals, in a healthcare environment.

The ES1 Ultrasound Diagnostic System is indicated for imaging of anatomical structures in the abdomen and measurements of physical properties in the liver and the spleen.

In particular, the ES1 ultrasound diagnostic system is intended to provide:

- Linear distance measurements of anatomical structures,
- Measurement of shear wave speed at selected shear wave frequencies, and estimates in the liver and the spleen,

- Estimates of ultrasound attenuation coefficient in the liver at 3.5 MHz.
- Measurement of brightness ratio between structures and in particular between the liver and the kidney,

- Estimates of speed of sound in the liver.

The shear wave speed measurements, ultrasound beam attenuation, elastic properties estimates ratio may be used as an aid to the diagnosis, montoring and clinical management of adult and pediatric patients with liver disease.

Type of Use (Select one or both, as applicable)

| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)     </span> </div> |
|----------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)     </span> </div>  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

## Premarket Notification [510(k)] Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements detailed in 21 CFR 807.92

The assigned 510(k) number is: K213102

- Company: E-Scopics 931 chemin de la Bosque d'Antonelle 13090 Aix-en-Provence France Telephone:
Contact Person: Claude COHEN- BACRIE

Date Prepared: 03rd January 2022

#### Device Names:

| Trade/Proprietary Name: | ES1 System                                                                                                                                                                  |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common or Usual Name:   | Diagnostic Ultrasound System and Accessories                                                                                                                                |
| Device Class            | Class II                                                                                                                                                                    |
| Classifications:        | §892.1550 : Ultrasonic pulsed doppler imaging system (IYN)<br>§892.1560 : Ultrasonic Pulsed Echo Imaging System (IYO)<br>§892.1570 : Diagnostic Ultrasound Transducer (ITX) |

#### Substantially Equivalent/Predicate Devices:

The data and information supplied in this submission demonstrates substantial equivalence to their respective predicate devices : Predicate Device: FibroScan® Family of Products (K203273), cleared on 03/25/2021. Reference Devices: AIXPLORER® MACH 20 & 30 Ultrasound Diagnostic Systems (K191007), cleared on 10/25/2019.

{5}------------------------------------------------

#### Device description:

E-Scopics' ES1 System is an ultraportable ultrasound imaging system used to perform non-invasive diagnostic general purpose ultrasound imaging and quantitative imaging studies. The ES1 System consists of a Software App (running on a consumer off-the-shelf Selected Host) and an accessory curved array probe. The system produces images and quantifications, which are displayed on the monitor of the Selected Host. The ES1 System is operated from the Selected Host multi-touch screen. The ES1 System also allows the user to perform measurements, to capture images, and to generate printable reports.

The ES1 System is designed to perform non-invasive measurements of liver/spleen shear wave speed and to estimate tissue stiffness. The e.C5-1 probe is equipped with a mechanical vibrator to produce shear waves in tissue. When used on the liver or the spleen applications, shear waves travel through the skin and intercostal space into the abdominal organ. 2D ultrasound is used to track the shear wave and measure its speed. The system then provides an estimate of tissue stiffness. In addition, the ES1 System is designed to measure several quantitative parameters from B Mode imaging: ultrasound attenuation, speed of sound, and compute ultrasound brightness ratio between 2 regions in the image. The results of ES1 quantitative imaging modalities are displayed on the Selected Host monitor.

#### Intended Use & Indications for Use:

The ES1 Ultrasound Diagnostic System is intended for general purpose pulse echo ultrasound imaging, soft tissue elasticity imaging of the human body and provides measurements of shear wave speed and tissue stiffness, ultrasound beam attenuation and estimates of speed of sound, in internal structures of the body. This device is intended to be used by trained healthcare professionals, in a healthcare environment.

The ES1 Ultrasound Diagnostic System is indicated for imaging of anatomical structures in the abdomen and measurements of physical properties in the liver and the spleen.

In particular, the ES1 ultrasound diagnostic system is intended to provide:

- Linear distance measurements of anatomical structures. –
- Measurement of shear wave speed at selected shear wave frequencies, and estimates of tissue stiffness in the liver and the spleen,
- Estimates of ultrasound attenuation coefficient in the liver at 3.5 MHz, -
- -Measurement of brightness ratio between structures and in particular between the liver and the kidney.
- Estimates of speed of sound in the liver. -

The shear wave speed measurements, ultrasound beam attenuation, elastic properties estimates, brightness ratio may be used as an aid to the diagnosis, monitoring and clinical management of adult and pediatric patients with liver disease.

{6}------------------------------------------------

#### Substantial Equivalence discussion

The ES1 System operating modalities, principles, and features (B-Mode, 2D-Transient Elastography, estimation of shear wave speed, ultrasound attenuation & speed of sound, and associated combinations as listed in the Diagnostic Ultrasound Indications for Use Table) have been demonstrated to be substantially equivalent to operating modalities. principles, and features of the Predicate & Reference Devices.

Shear wave speed measurement and tissue stiffness estimation are available on the ES1 System with transient elastography (TE) using a mechanical vibration to induce shear waves, thereby technically equivalent to the Predicate Device. The shear wave propagation is tracked by both devices at high pulse repetition frequency, the only difference between them being that tracking is performed by 1D M-Mode with the Predicate Device & 2D B-Mode with ES1 System.

Results of these assessments provide substantially equivalent measurements serving the same intended use as per the Predicate Device, without raising any safety & effectiveness issues.

The ES1 System also provides shear wave speed maps leveraging tracking of shear waves generated by the probe vibrations using 2D imaging, displayed as images of estimated stiffness of tissue. This bidimensional stiffness map extends the effectiveness with respect to the 1D measurement provided by the Predicate Device, and as such similar display to the stiffness map provided by the Reference Device. The ES1 System 2D maps of tissue stiffness derive from different technical characteristics with respect to the Reference Device without raising any issue of safety & effectiveness.

Both the measurement and the map of shear wave speed (and estimated stiffness) are displayed while providing real time B-Mode. This simultaneous display of stiffness and B-Mode is as effective as the Reference Device displaying for the operator the anatomical region where the measurement is performed. The operator can effectively visualize the liver tissue, where the measurement should be made.

The ES1 System provides estimates of ultrasound attenuation in tissue and is substantially equivalent to the Predicate Device which uses CAP (Controlled Attenuation Parameter) technology. The technological differences between methods implemented on the ES1 System and the Predicate Device do not raise issues of safety and effectiveness. This estimation is performed simultaneously as a real time B-Mode imaging, providing the operator with visualization of the location where the measurement is performed. This visualization of anatomical structures, while measuring the attenuation, contributes to the effectiveness of the method as it is the case in the Reference Device with Att PLUS.

The ES1 System provides estimates of speed of sound in tissue in order to complement information on the tissue as performed on the Reference Device with Ssp PLUS. This measurement does not raise an issue of safety & effectiveness.

{7}------------------------------------------------

The ES1 System operates a single curvilinear array to perform all ultrasound modes, including conventional B-Mode. Although the ES1 System B-Mode imaging has different technological characteristics with respect to the Reference Device, both allow the operator to visualize where measurements are performed and do not raise issues of safety and effectiveness.

The following Table 1 provides supporting elements to the substantial equivalence discussion.

This part of the page is left intentionally blank.

.../...

{8}------------------------------------------------

| Table 1: Substantial Equivalence Comparison |                                                                                         |                                                                                                            |                                                                                                             |
|---------------------------------------------|-----------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|
| Key points                                  | Predicate Device                                                                        | Reference Device                                                                                           | New Device                                                                                                  |
| Device Name                                 | FibroScan® Family of Products<br>(Models: 502 Touch, 530 Compact,<br>430 Mini+ and 630) | AIXPLORER® MACH 30 and<br>AIXPLORER® MACH 20                                                               | ES1 System                                                                                                  |
|                                             |                                                                                         | Ultrasound Diagnostic Systems                                                                              |                                                                                                             |
| Manufactured by                             | Echosens, France                                                                        | SuperSonic Imagine, S.A., France                                                                           | E-Scopics S.A.S., France                                                                                    |
| 510(k) Clearance # and Date                 | K203273; 03/25/2021                                                                     | K191007; 10/25/2019                                                                                        | K213102 ;                                                                                                   |
| Classification Name                         | Ultrasonic pulsed echo imaging<br>system                                                | Ultrasonic pulsed doppler imaging<br>system                                                                | Ultrasonic pulsed doppler imaging<br>system                                                                 |
| Class                                       | II                                                                                      | II                                                                                                         | II                                                                                                          |
| Intended use                                | Diagnostic ultrasound imaging or<br>fluid flow analysis of the human<br>body            | Diagnostic ultrasound imaging, soft<br>tissue elasticity imaging, fluid flow<br>analysis of the human body | Diagnostic ultrasound imaging, soft<br>tissue elasticity imaging, fluid flow<br>analysis of the human body. |

| Key points                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Predicate Definition                                                                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                        | Contrast-enhanced ultrasound<br>imaging         |                                                                                                                                                                                    |                                                                                                                                         |                                                                                                                                                                                           |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| as for use<br>The FibroScan Family of Products<br>(Models: 502 Touch, 530 Compact, 430 Mini+ and 630) is intended to provide shear wave speed measurements and estimates of tissue stiffness as well as ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter) in internal structures of the body. The Shear wave speed and stiffness measurements may be used as an aid to clinical management of adult patients with liver disease. | The SuperSonic Imagine AIXPLORER® MACH range ultrasound diagnostic systems and transducers are intended for general purpose pulse echo ultrasound imaging, soft tissue viscoelasticity imaging, doppler fluid flow analysis of the human body.                                                                                                                                                                          | The ES1 Ultrasound Diagnostic System is intended for general purpose pulse echo ultrasound imaging, soft tissue elasticity imaging of the human body and provides measurements of shear wave speed and tissue stiffness, ultrasound beam attenuation and estimates of speed of sound, in internal structures of the body. This device is intended to be used by trained healthcare professionals, in a healthcare environment.                                            |                                                                                                                                                                                                                                                                                                                        | Conventional Modes                              | B-Mode (Harmonic, Fundamental)                                                                                                                                                     | B-Mode                                                                                                                                  |                                                                                                                                                                                           |
| The FibroScan Family of Products<br>(Models: 502 Touch, 530 Compact, 430 Mini+ and 630) is indicated for non-invasive measurement in the liver of 50 Hz shear wave speed and estimates of stiffness as well as determining a 3.5 MHz ultrasound coefficient of attenuation (CAP: Controlled Attenuation Parameter).                                                                                                                                          | The SuperSonic Imagine AIXPLORER® MACH ultrasound diagnostic systems are indicated for use in the following applications, for imaging and measurement of anatomical structures: Abdominal, Small Organs, Musculoskeletal, Superficial Musculoskeletal, Vascular, Peripheral Vascular, Intraoperative, OB-GYN, Pelvic, Pediatric, Urology, Transrectal, Trans-vaginal and Neonatal/Adult Cephalic, Non-invasive Cardiac. | The ES1 Ultrasound Diagnostic System is indicated for imaging and measurements of anatomical structures in the abdomen.                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                        | Ultrasound Parameter Estimation and Display     |                                                                                                                                                                                    |                                                                                                                                         |                                                                                                                                                                                           |
| The shear wave speed and stiffness, and CAP may be used as an aid to diagnosis and monitoring of adult patients with liver disease, as part of an overall assessment of the liver.                                                                                                                                                                                                                                                                           | In addition, the SuperSonic Imagine AIXPLORER® MACH ultrasound diagnostic systems and associated transducers are intended for:<br>- Measurements of abdominal anatomical structures,                                                                                                                                                                                                                                    | In particular, the ES1 ultrasound diagnostic system is intended to provide:<br>- Linear distance measurements of anatomical structures,<br>- Measurement of shear wave speed at selected shear wave frequencies, and estimates of tissue stiffness in the liver and the spleen,<br>- Estimates of ultrasound attenuation coefficient in the liver at 3.5 MHz,<br>- Measurement of brightness ratio between structures and in particular between the liver and the kidney, | B-Mode                                                                                                                                                                                                                                                                                                                 | Name: Controlled Attenuation<br>Parameter (CAP) | Name: Att PLUS                                                                                                                                                                     | Name: ATT                                                                                                                               |                                                                                                                                                                                           |
| Shear wave speed and stiffness, and CAP* may be used as an aid in the                                                                                                                                                                                                                                                                                                                                                                                        | - Measurements of broad band shear wave speed, and tissue stiffness in internal structures of the liver and the spleen,                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                        | Probe: S+, M+ and XL+ Probes<br>only for liver  | Probe: C6-1X probe only for<br>liver and abdomen applications                                                                                                                      | Probe: e.C5-1 probe                                                                                                                     |                                                                                                                                                                                           |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                              | clinical management of pediatric<br>patients with liver disease.                                                                                                                                                                                                                                                                                                                                                        | - Measurements of brightness ratio<br>between liver and kidney,<br>- Visualization of abdominal<br>vascularization,<br>microvascularization and perfusion,<br>- Quantification of abdominal<br>vascularization and perfusion.                                                                                                                                                                                                                                             | - Estimates of speed of sound in the<br>liver.<br>- The shear wave speed<br>measurements, ultrasound beam<br>attenuation, elastic properties<br>estimates, brightness ratio may be<br>used as an aid to the diagnosis,<br>monitoring and clinical<br>management of adult and pediatric<br>patients with liver disease. |                                                 | Technological Characteristics:<br>single beam transmission and<br>reception of ultrasound signals<br>using the monoelement probe<br>with<br>estimation of attenuation at<br>3,5MHz | Technological Characteristics:<br>Plane wave ultrasound<br>transmission and reception<br>Wide bandwidth attenuation<br>slope estimation | Technological characteristics:<br>Diverging wave ultrasound<br>transmission and reception<br>Wide bandwidth attenuation<br>slope estimation AND<br>Estimation of attenuation at<br>3,5Mhz |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                              | FibroScan 630 (Expert) is also<br>indicated for noninvasive<br>measurement in the spleen of 100<br>Hz shear wave speed and estimates<br>of stiffness that may be used as an<br>aid to diagnosis, monitoring and<br>clinical management of adult<br>patients with liver disease, as part<br>of an overall assessment of the<br>liver.                                                                                    | The shearwave speed, beam<br>attenuation, viscosity and stiffness<br>measurements, the brightness ratio,<br>the visualization of vascularization,<br>microvascularization and perfusion,<br>the quantification of<br>vascularization and perfusion may<br>be used as an aid to clinical<br>management of adult and pediatric<br>patients with liver disease.                                                                                                              |                                                                                                                                                                                                                                                                                                                        | Ultrasound attenuation                          | Range: 100 to 400 dB/m                                                                                                                                                             | Range:<br>0.2 dB/cm/MHz to 1.6<br>dB/cm/MHz                                                                                             | Range:<br>105 to 350 dB/m<br>0.3 dB/cm/MHz to 1<br>dB/cm/MHz                                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                              | *CAP for pediatric patients with<br>liver disease is only available with<br>SmartExam capability on<br>FibroScan Models: 530 Compact,<br>430 Mini+, and 630.                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                        |                                                 | Display:<br>Median, interquartile range<br>(IQR) and IQR/Median ratio<br>(%)                                                                                                       | Display:<br>Average value                                                                                                               | Display:<br>Median, interquartile range<br>(IQR) and IQR/Median ratio<br>(%)                                                                                                              |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                         | Modes of Operation                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                        |                                                 | Bias range (dB/m @ 3.5 MHz):<br>(6.9%) - (12.2%)                                                                                                                                   |                                                                                                                                         | Bias:<br>(dB/m @ 3.5 MHz): (-22.2%) -<br>(1.4%)                                                                                                                                           |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                         | A-Mode<br>B-Mode (Harmonic,<br>Fundamental)<br>Spatial compounding,<br>Panoramic<br>M-Mode                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                        |                                                 | Precision range (dB/m @ 3.5<br>MHz):                                                                                                                                               |                                                                                                                                         |                                                                                                                                                                                           |
| Imaging & quantification<br>modes                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                         | B-Mode (FS630E only for organ<br>localization)<br>Transient Elastography / Shear<br>Wave<br>CAPTM                                                                                                                                                                                                                                                                                                                                                                         | B-mode<br>Transient Elastography<br>2D Transient Elastography and<br>Shear Wave Imaging<br>Attenuation measurement<br>Speed of sound measurement                                                                                                                                                                       |                                                 |                                                                                                                                                                                    |                                                                                                                                         |                                                                                                                                                                                           |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                         | PW<br>CW<br>Color Doppler<br>Amplitude Doppler<br>Microvascular (Angio PL.U.S)<br>Att PLUS<br>SSP PLUS<br>Shear Wave Elastography                                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                                                                                                                                                        |                                                 |                                                                                                                                                                                    |                                                                                                                                         |                                                                                                                                                                                           |

{9}------------------------------------------------

{10}------------------------------------------------

{11}------------------------------------------------

{12}------------------------------------------------

|                                  | (0.3%) - (2.1%) |                                                                                                                                                                                                                                                                                                       | (dB/cm/MHz): (-22.2%) - (1.4%)<br>Precision:<br>(dB/m @ 3.5 MHz): (1.5%) - (4.5%)<br>(dB/cm/MHz): (1.5%) - (4.5%)<br>Name: SOS<br>Probe: e.C5-1 probe<br>Technological Characteristics:<br>Diverging wave ultrasound transmission and reception<br>Wide bandwidth frequency estimation<br>Range:<br>1450 m/s to 1600 m/s<br>Display:<br>Median, interquartile range (IQR) and IQR/Median ratio (%)<br>Bias:<br>(m/s): (-0.7%) - (0.7%)<br>Precision:<br>(m/s): (0.05%) - (0.3%)<br>Name: Transient Elastography (2D-TE) |
|----------------------------------|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                  | N/A             | Name: Speed of sound measurement (Ssp PLUS)<br>Probe: C6-1X probe only for liver and abdomen applications<br>Technological Characteristics:<br>Plane wave ultrasound transmission and reception<br>Wide bandwidth frequency estimation<br>Range:<br>1450 m/s to 1600 m/S<br>Display:<br>Average value | Name: ShearWave Elastography (SWE)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Other Modes<br>Elastography Mode | Speed of sound  |                                                                                                                                                                                                                                                                                                       | Name: Vibration-Controlled Transient Elastography (VCTE)                                                                                                                                                                                                                                                                                                                                                                                                                                                                |

{13}------------------------------------------------

| Type: 1D, quantification without<br>imaging capabilities                                                                                                                    | Type: 2D, quantification with<br>imaging capabilities                                                                                                     | Type: 2D, quantification with<br>imaging capabilities                                                                                                                                           |                                                | M+ (kPa): (-21.6%) - (1.3%)<br>M+ (m/s): (-11.5%) - (0.7%)<br>XL+ (kPa): (-25.7%) - (2.6%)<br>XL+ (m/s): (-13.9%) - (1.3%) | Precision:<br>S+ (kPa): (1.4%) - (4.0%)<br>S+ (m/s): (0.7%) - (2.0%)<br>M+ (kPa): (1.3%) - (3.8%)<br>M+ (m/s): (0.6%) - (1.9%)<br>XL+ (kPa): (0.0%) - (6.3%)<br>XL+ (m/s): (0.0%) - (3.1%)       | Precision:<br>(kPa): (0.2%) - (6.0%)<br>(m/s): (0.1%) - (2.9%)                                           |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Shear wave source: External<br>electromechanical vibrator<br>creating a punctual source of<br>vibration                                                                     | Shear wave source: Supersonic<br>acoustic radiation force                                                                                                 | Shear wave source: External<br>electromechanical vibrator<br>creating a non-punctual source<br>of vibration, spread along the<br>probe surface                                                  | Combined Modes                                 | N/A                                                                                                                        | Combined modes may include:<br>- B + Color Flow<br>- B + SWE<br>- B + PW<br>- B + WP + Color Flow<br>- B + M<br>- B + SWE + Color Flow<br>- B + CW<br>- B + M + Color Flow<br>- B + Strain + SWE | Combined modes include:<br>- B + 2D-TE or B + SWS<br>- B + ATT<br>- B + SOS<br>- B + Q.US (ATT + SOS)    |
| Ultrasound tracking: Focused<br>monoelement transmission and<br>reception                                                                                                   | Ultrasound tracking: Plane wave<br>transmission and reception                                                                                             | Ultrasound tracking: Plane wave<br>transmission and reception                                                                                                                                   | System Components                              | Product Design                                                                                                             |                                                                                                                                                                                                  |                                                                                                          |
| Range:<br>Liver: Shear wave speed<br>(0.8-5.0 m/s); Stiffness values<br>(2.0-75 kPa)<br>Spleen (FS630E only): Shear<br>wave speed (1.4-5.8 m/s);<br>Stiffness (6.0-100 kPa) | Range:<br>Elastography range displayed on<br>the screen (next to color bar)                                                                               | Range:<br>Liver: shear wave speed<br>(0.0-5.8); Stiffness values<br>(0.0-100 kPa)<br>Elastography range displayed on<br>the screen (next to color bar)<br>and depends on vibration<br>frequency |                                                | - Software<br>- Analog cable hardwired probes<br>- Cart (including computer host)                                          | - Software<br>- Analog cable hardwired probes<br>- Cart (including computer host)                                                                                                                | - Software<br>- USB digital hardwired probe<br>- Consumer off-the-shelf<br>(COTS) Selected Host computer |
| Display:<br>Liver: Median value;<br>Interquartile range (IQR);<br>IQR/Median ratio<br>Spleen: Median value;<br>Interquartile range (IQR)                                    | Display:<br>Mean value and Standard<br>Deviation<br>Median/Average values;<br>Interquartile range<br>(IQR)/Standard Deviation<br>values; IQR/Median ratio | Display:<br>Median value; Interquartile<br>range (IQR); IQR/Median ratio                                                                                                                        | Form factor (Cart based /<br>portable mobile ) | Mobile cart-based product with<br>monitor                                                                                  | Mobile cart based product with<br>control panel and monitor                                                                                                                                      | Portable product with COTS<br>selected host                                                              |
| Bias:<br>S+ (kPa): (-25.1%) - (7.6%)<br>S+ (m/s): (-13.5%) - (3.6%)                                                                                                         |                                                                                                                                                           | Bias:<br>(kPa): (-15.7%) - (11.7%)<br>(m/s): (-8.6%) - (5.7%)                                                                                                                                   | Transducers<br>Ultrasound source               | Piezoelectric ultrasound source                                                                                            | Piezoelectric ultrasound source                                                                                                                                                                  | Piezoelectric ultrasound source                                                                          |

{14}------------------------------------------------

{15}------------------------------------------------

| Transducers                                                                                                                                                                              |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Probe type                                                                                                                                                                               |
| Mono-element ultrasound<br>transducers:<br>M+-probe (3.5 MHz)<br>XL+ probe (2.5 MHz)<br>S+ probe (5 MHz)                                                                                 |
| Multi-elements ultrasound<br>transducer:<br>Convex array: ES-C5-2R60S-3<br>(3.5 MHz) (FS630E only), OEM<br>MicrUs EXT-1H (#K161968)                                                      |
| Multi-elements ultrasound<br>transducer:<br>Linear arrays: L18-5, L10-2,<br>LV16-5, LH20-6<br>Convex arrays: C6-1X, C9-2X<br>Microconvex arrays: E12-3,<br>MC12-3<br>Phased array: P5-1X |
| Multi-elements ultrasound<br>transducer:<br>Convex array: e.C5-1                                                                                                                         |

{16}------------------------------------------------

### Application of Recognized Standards

The ES1 System has been designed & developed to ensure compliance with the following requirements : biocompatibility, acoustic output, cleaning & disinfection, thermal, mechanical, electrical & electromagnetic safety performance taking into account & in accordance with the following standards :

- . IEC 60601-1:2005. AMD1:2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- . IEC 60601-1-2: Ed 4.0 : 2014-02 Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests
- . IEC 60601-2-37 Ed2.1 2015 : Medical electrical equipment - Part 2-37: Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment
- IEC 62127-1 :2007 + amd1:2013 Ultrasonics -- Hydrophones -- Part 1: Measurement and ● characterization of medical ultrasonic fields up to 40 MHz
- IEC 62359 Edition 2.1 2017-09 Ultrasonics - Field characterization - Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields
- ISO 10993-1 Fifth edition 2018-08 Biological evaluation of medical devices Part 1: ● Evaluation and testing within a risk management process
- o IEC 62304: Medical Device Software - Software Life Cycle Processes: First Edition 2006- 05, Equivalent to IEC 62304: 2006/A1:2016.
- . ISO 14971 Second: Medical Devices - Application of Risk Management To Medical Devices: Edition 2007-03-01.
- IEC 62366-1 Edition 1.1 2020-06 Medical devices Part 1: Application of usability ● engineering to medical devices

### Discussion of Performance Data:

The accuracy and precision of the device ultrasound modalities were tested on calibrated phantoms with known reference values.

For B-Mode imaging, the in-vitro performances were good. Contrast resolution enables to distinguish targets having a contrast ratio of +/- 3 dB as compared to the phantom background. The axial resolution of the system is 1 mm, and remains constant over the whole image depth. The lateral resolution ranges from 1 mm to 4 mm depending on the depth of structures. In addition, these performances were shown to not suffer from probe variability. These in vitro assessments were confirmed by the use of the ES1 System on a few subjects in a clinical setting.

The accuracy of quantitative ultrasound modes of the ES1 System was assessed in-vitro by looking at the precision and the bias obtained on a sample of calibrated phantoms. Precision (in %), also known as the within subject coefficient of variation (wCV), was defined as the standard deviation of the independent measurements normalized by the reference value. Bias (in %) was defined as the difference in the mean value measured and the nominal value of the phantom, normalized by the nominal value.

{17}------------------------------------------------

For the estimation of the Speed of Sound, the in-vitro experiments reveal a strong inter-system repeatability on the two considered phantoms. The method demonstrated its robustness against external parameters such as element sensitivity, electronic noise levels of the acquisition chain and small variations of probe geometries on a wide range of speed of sound values. Bias values range in [-0.7 %, 0.7 %] and precision range in [0.05 %, 0.3 %]. The use of Speed of Sound measurements on a few subjects in a clinical setting was also demonstrated to be feasible and safe.

In addition, the accuracy of shear wave speed/stiffness and ultrasound attenuation measurements of the ES1 System and its Predicate Device were compared looking at the precision and the bias obtained with both systems on the same sample of calibrated shear wave speed and ultrasound attenuation phantoms.

For shear wave speed measurements and liver stiffness estimation by the 2D TE method, in-vitro experiments reveal that the stiffness estimation is robust on a wide range of stiffness values and satisfies OIBA technical performance claims. Indeed, the wCV lies in the range [0.1 %, 2.9 %] for the SWS and in the range [0.2 %, 6.0 %] for the corresponding Young's modulus. Bias values lie in the range [-8.6 %, 5.7 %] for the SwS and in the range [-15.7 %, 11.7 %] for Young's modulus. It also reveals a strong inter-system repeatability.

When comparing shear wave speed measurements performed with the ES1 System and its Predicate Device, a high Pearson correlation coefficient of 0.997 was obtained between the 2 systems and methods.

For measurement of the ultrasound attenuation coefficient, the in-vitro performance tests revealed a strong inter-system repeatability. The algorithm is robust against external parameters such as element sensitivity, electronic noise levels of the acquisition chain and small variations of probe geometries. On tested phantoms, the precision lies in the range [1.5 %, 4.5 %], while bias values range in [-22.2% %, 1.4 %]. When compared to its Predicate Device on the same set of calibrated phantoms, ultrasound attenuation estimations with the ES1 System were highly correlated with FibroScan CAP method with a Pearson correlation coefficient of 0.999.

Acoustic Pressure and Intensities of the ultrasound modalities available to the user on the ES1 System were measured. Five transmit patterns (TP) were measured, each of them being used respectively in the five composed imaging modes available on the system. These test results show that the upper-limit values for MI, TI and ISPTA3 are always below the limits recommended in the FDA Guidance for applications other than for ophthalmic use, for ultrasound systems, therefore legitimating the Track 1 approach.

### Conclusions for Performance Testing :

The performance testing data conclude that the safety and effectiveness of the ES1 System is not compromised and meet all acceptance criteria, demonstrating that the system is substantially equivalent to the relevant Predicate Device.

{18}------------------------------------------------

#### Conclusion regarding Substantial Equivalence :

In conclusion, the ES1 System was demonstrated to be substantially equivalent to its Predicate Device K203273, as they have the same intended use, equivalent indications for use, and the technological differences do not raise issues of safety and effectiveness. ES1 System features equivalent to those of the Reference Device K191007 are contributing to ES1 System effectiveness without raising any safety issue as it allows a visualization of the tissue where measurement is performed.

This has been demonstrated via safety & performance testing and comparative data between the ES1 System, its Predicate Device and its Reference Device.

- End of document -

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K213102](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K213102)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
