← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K211299

# X-CUBE 70 (K211299)

_Alpinion Medical Systems Co., Ltd. · IYN · Aug 9, 2021 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K211299

## Device Facts

- **Applicant:** Alpinion Medical Systems Co., Ltd.
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Aug 9, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The X-CUBE 70 diagnostic ultrasound system is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Trans-rectal; Trans-vaginal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiac(adult& pediatric); Peripheral Vessel(PV); and Urology(including prostate). And, in the imaging modes of 2D(B) mode; Harnonic mode(HAR); M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode; High PRF Doppler mode; Tissue Doppler Imaging(TDI) Mode; 3D/4D mode. The X-CUBE 70 is intended to be used in a hospital or medical clinic.

## Device Story

X-CUBE 70 is a diagnostic ultrasound system for medical imaging and fluid flow analysis. It processes acoustic signals from various transducers to produce 2D, M-mode, Doppler (Color, Power, PWD, CWD, TDI), and 3D/4D images. Operated by physicians in hospitals or clinics, the system features an ergonomic control panel and flat-panel display. It includes advanced imaging functions like Xpeed, Full SRI, Spatial Compounding, and specialized tools (e.g., Cube Strain, Live HQ, Needle Vision, Elastography, Auto EF). Output is used by clinicians to evaluate soft tissue and blood flow, aiding in diagnosis and clinical decision-making. The system supports connectivity via wireless LAN and integrates with peripherals like printers and foot switches.

## Clinical Evidence

No clinical studies were required to support substantial equivalence. The device was evaluated through bench testing, including biocompatibility, acoustic output, thermal, electrical, electromagnetic, and mechanical safety testing, confirming compliance with applicable standards (e.g., IEC 60601 series, ISO 10993, NEMA UD 2/3).

## Technological Characteristics

Ultrasound imaging system utilizing pulsed Doppler and echo imaging. Features include 2D, M-mode, Doppler modes, and 3D/4D imaging. Connectivity via Wireless LAN. Safety compliance: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 2, NEMA UD 3. Biocompatible patient-contact materials per ISO 10993.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- X-CUBE 70 Diagnostic Ultrasound System ([K200450](/device/K200450.md))

## Reference Devices

- E-CUBE 12 Diagnostic Ultrasound System ([K181277](/device/K181277.md))
- E-CUBE 15 Diagnostic Ultrasound System ([K150773](/device/K150773.md))
- E-CUBE 8 Diagnostic Ultrasound System ([K181617](/device/K181617.md))
- E-CUBE 11 Diagnostic Ultrasound System ([K161439](/device/K161439.md))
- HS70A Diagnostic Ultrasound System ([K173713](/device/K173713.md))
- TOMTEC-ARENA ([K201632](/device/K201632.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

August 9, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Alpinion Medical Systems Co., Ltd. % Boyeon Cho Quality Management Representative 5FL, I dong, 77, heungan-daero 81 beon-gil dongan-gu Anyang-si. Gyeonggi-do 14117 REPUBLIC OF KOREA

Re: K211299

Trade/Device Name: X-CUBE 70 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: June 10, 2021 Received: June 11, 2021

Dear Boyeon Cho:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR

{1}------------------------------------------------

803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.

510(k) Number (if known) K211299

Device Name X-CUBE 70

#### Indications for Use (Describe)

The X-CUBE 70 diagnostic ultrasound system is intended for use by, or by the order of, and under the supervision of, a licensed physician who is qualified for the evaluation of soft tissue and blood flow in the clinical applications of Fetal; Abdominal(renal & GYN/pelvic); Pediatric; Small Organ(breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Trans-rectal; Trans-vaginal; Musculo-skeletal(Conventional); Musculo-skeletal(Superficial); Cardiac(adult& pediatric); Peripheral Vessel(PV); and Urology(including prostate).

And, in the imaging modes of 2D(B) mode; Harnonic mode(HAR); M mode; Anatomical M mode; Color Flow Doppler(CF) Mode; Power Doppler(PD) Mode; Directional PD mode; Pulsed Wave Doppler(PWD) Mode; Continuous Wave Doppler(CWD) Mode; High PRF Doppler mode; Tissue Doppler Imaging(TDI) Mode; 3D/4D mode. The X-CUBE 70 is intended to be used in a hospital or medical clinic.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# X-CUBE 70 Ultrasound System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                          | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| Clinical Application                     | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic                               |                   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                                    | P                 | P | P   |     | P                | P                | P                             | P                      | P                    |
| Abdominal                                | P                 | P | P   |     | P                | P                | P                             | P                      | P                    |
| Intra-operative (Specify)                |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                             |                   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                                | P                 | P | P   |     | P                | P                | P                             | P                      | P                    |
| Small Organ<br>(breast, testes, thyroid) | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Neonatal Cephalic                        | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Adult Cephalic                           | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Trans-rectal                             | P                 | P | P   |     | P                | P                | P                             | P                      | P                    |
| Trans-vaginal                            | P                 | P | P   |     | P                | P                | P                             | P                      | P                    |
| Trans-urethral                           |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal<br>(Conventional)       | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Musculo-skeletal<br>(Superficial)        | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Intravascular                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Adult                            | P                 | P | P   | P   | P                | P                | P                             | P                      |                      |
| Cardiac Pediatric                        | P                 | P | P   | P   | P                | P                | P                             | P                      |                      |
| Intravascular (Cardiac)                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (Cardiac)                   |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-cardiac                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Peripheral vessel                        | P                 | P | P   | P   | P                | P                | P                             | P                      |                      |
| Urology (including prostate)             | P                 | P | P   |     | P                | P                | P                             | P                      | P                    |

N = new indication; P = previously cleared by FDA K200450; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

{4}------------------------------------------------

# X-CUBE 70 with L3-8H Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application                     | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
|------------------------------------------|---|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| Ophthalmic                               |   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                                    |   |   |     |     |                  |                  |                               |                        |                      |
| Abdominal                                |   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Specify)                |   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)                  |   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                             |   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                                | P | P | P   |     | P                | P                | P                             | P                      |                      |
| Small Organ<br>(breast, testes, thyroid) | P | P | P   |     | P                | P                | P                             | P                      |                      |
| Neonatal Cephalic                        |   |   |     |     |                  |                  |                               |                        |                      |
| Adult Cephalic                           |   |   |     |     |                  |                  |                               |                        |                      |
| Trans-rectal                             |   |   |     |     |                  |                  |                               |                        |                      |
| Trans-vaginal                            |   |   |     |     |                  |                  |                               |                        |                      |
| Trans-urethral                           |   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)                 |   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal<br>(Conventional)       | P | P | P   |     | P                | P                | P                             | P                      |                      |
| Musculo-skeletal<br>(Superficial)        | P | P | P   |     | P                | P                | P                             | P                      |                      |
| Intravascular                            |   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Adult                            |   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Pediatric                        |   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular (Cardiac)                  |   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (Cardiac)                   |   |   |     |     |                  |                  |                               |                        |                      |
| Intra-cardiac                            |   |   |     |     |                  |                  |                               |                        |                      |
| Peripheral vessel                        | P | P | P   |     | P                | P                | P                             | P                      |                      |
| Urology (including prostate)             |   |   |     |     |                  |                  |                               |                        |                      |

N = new indication; P = previously cleared by FDA K200450; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

{5}------------------------------------------------

# X-CUBE 70 with L3-12X Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                          | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| Clinical Application                     | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic                               |                   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                                    |                   |   |     |     |                  |                  |                               |                        |                      |
| Abdominal                                |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Specify)                |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                             |                   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                                | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Small Organ<br>(breast, testes, thyroid) | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Neonatal Cephalic                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Adult Cephalic                           |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-rectal                             |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-vaginal                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-urethral                           |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal<br>(Conventional)       | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Musculo-skeletal<br>(Superficial)        | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Intravascular                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Adult                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Pediatric                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular (Cardiac)                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (Cardiac)                   |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-cardiac                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Peripheral vessel                        | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Urology (including prostate)             | P                 | P | P   |     | P                | P                | P                             | P                      |                      |

N = new indication; P = previously cleared by FDA K200450; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

{6}------------------------------------------------

# X-CUBE 70 with L3-15H Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                              | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |
|------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| Clinical Application         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic                   |                   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Abdominal                    |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Specify)    |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)      |                   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                    |                   |   |     |     |                  |                  |                               |                        |                      |
| Small Organ                  | N                 | N | N   |     | N                | N                | N                             | N                      |                      |
| Neonatal Cephalic            |                   |   |     |     |                  |                  |                               |                        |                      |
| Adult Cephalic               |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-rectal                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-vaginal                |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-urethral               |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)     |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal             | N                 | N | N   |     | N                | N                | N                             | N                      |                      |
| Musculo-skeletal             | N                 | N | N   |     | N                | N                | N                             | N                      |                      |
| Intravascular                |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Adult                |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Pediatric            |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular (Cardiac)      |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (Cardiac)       |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-cardiac                |                   |   |     |     |                  |                  |                               |                        |                      |
| Peripheral vessel            | N                 | N | N   |     | N                | N                | N                             | N                      |                      |
| Urology (including prostate) |                   |   |     |     |                  |                  |                               |                        |                      |

N = new indication; P = previously cleared by FDA; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

{7}------------------------------------------------

### X-CUBE 70 with SL3-19H Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                              | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |
|------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| Clinical Application         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic                   |                   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                        | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Abdominal                    | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Intra-operative (Specify)    |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)      |                   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                    | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Small Organ                  | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| (breast, testes, thyroid)    |                   |   |     |     |                  |                  |                               |                        |                      |
| Neonatal Cephalic            | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Adult Cephalic               |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-rectal                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-vaginal                |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-urethral               |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)     |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal             | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| (Conventional)               |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal             | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| (Superficial)                |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular                |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Adult                |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Pediatric            |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular (Cardiac)      |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (Cardiac)       |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-cardiac                |                   |   |     |     |                  |                  |                               |                        |                      |
| Peripheral vessel            | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Urology (including prostate) |                   |   |     |     |                  |                  |                               |                        |                      |

N = new indication; P = previously cleared by FDA K200450; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

{8}------------------------------------------------

# X-CUBE 70 with SL3-19X Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                              | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |
|------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| Clinical Application         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic                   |                   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                        | N                 | N | N   |     | N                | N                | N                             | N                      |                      |
| Abdominal                    | N                 | N | N   |     | N                | N                | N                             | N                      |                      |
| Intra-operative (Specify)    |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)      |                   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                    | N                 | N | N   |     | N                | N                | N                             | N                      |                      |
| Small Organ                  | N                 | N | N   |     | N                | N                | N                             | N                      |                      |
| Neonatal Cephalic            | N                 | N | N   |     | N                | N                | N                             | N                      |                      |
| Adult Cephalic               |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-rectal                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-vaginal                |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-urethral               |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)     |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal             | N                 | N | N   |     | N                | N                | N                             | N                      |                      |
| Musculo-skeletal             | N                 | N | N   |     | N                | N                | N                             | N                      |                      |
| Intravascular                |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Adult                |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Pediatric            |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular (Cardiac)      |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (Cardiac)       |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-cardiac                |                   |   |     |     |                  |                  |                               |                        |                      |
| Peripheral vessel            | N                 | N | N   |     | N                | N                | N                             | N                      |                      |
| Urology (including prostate) |                   |   |     |     |                  |                  |                               |                        |                      |

N = new indication; P = previously cleared by FDA; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

{9}------------------------------------------------

# X-CUBE 70 with L10-25H Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                          | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| Clinical Application                     | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic                               |                   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                                    |                   |   |     |     |                  |                  |                               |                        |                      |
| Abdominal                                |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Specify)                |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                             |                   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                                |                   |   |     |     |                  |                  |                               |                        |                      |
| Small Organ<br>(breast, testes, thyroid) | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Neonatal Cephalic                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Adult Cephalic                           |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-rectal                             |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-vaginal                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-urethral                           |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal<br>(Conventional)       | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Musculo-skeletal<br>(Superficial)        | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Intravascular                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Adult                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Pediatric                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular (Cardiac)                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (Cardiac)                   |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-cardiac                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Peripheral vessel                        | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Urology (including prostate)             |                   |   |     |     |                  |                  |                               |                        |                      |

N = new indication; P = previously cleared by FDA K200450; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

{10}------------------------------------------------

# X-CUBE 70 with IO7-18 Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                          | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| Clinical Application                     | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic                               |                   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                                    |                   |   |     |     |                  |                  |                               |                        |                      |
| Abdominal                                |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Specify)                |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                             |                   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                                |                   |   |     |     |                  |                  |                               |                        |                      |
| Small Organ<br>(breast, testes, thyroid) | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Neonatal Cephalic                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Adult Cephalic                           |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-rectal                             |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-vaginal                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-urethral                           |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal<br>(Conventional)       | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Musculo-skeletal<br>(Superficial)        | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Intravascular                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Adult                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Pediatric                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular (Cardiac)                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (Cardiac)                   |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-cardiac                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Peripheral vessel                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Urology (including prostate)             |                   |   |     |     |                  |                  |                               |                        |                      |

N = new indication; P = previously cleared by FDA K200450; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

{11}------------------------------------------------

# X-CUBE 70 with C5-8NT Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                          | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |  |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|--|
| Clinical Application                     | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |  |
| Ophthalmic                               |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Fetal                                    |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Abdominal                                | P                 | P | P   |     | P                | P                | P                             | P                      |                      |  |
| Intra-operative (Specify)                |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Intra-operative (Neuro)                  |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Laparoscopic                             |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Pediatric                                | P                 | P | P   |     | P                | P                | P                             | P                      |                      |  |
| Small Organ<br>(breast, testes, thyroid) |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Neonatal Cephalic                        | P                 | P | P   |     | P                | P                | P                             | P                      |                      |  |
| Adult Cephalic                           |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Trans-rectal                             |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Trans-vaginal                            |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Trans-urethral                           |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Trans-esoph. (non-Card.)                 |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Musculo-skeletal<br>(Conventional)       |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Musculo-skeletal<br>(Superficial)        |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Intravascular                            |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Cardiac Adult                            |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Cardiac Pediatric                        | P                 | P | P   |     | P                | P                | P                             | P                      |                      |  |
| Intravascular (Cardiac)                  |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Trans-esoph. (Cardiac)                   |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Intra-cardiac                            |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Peripheral vessel                        |                   |   |     |     |                  |                  |                               |                        |                      |  |
| Urology (including prostate)             |                   |   |     |     |                  |                  |                               |                        |                      |  |

N = new indication; P = previously cleared by FDA K200450; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

{12}------------------------------------------------

# X-CUBE 70 with SC1-7H Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                              | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |
|------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| Clinical Application         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic                   |                   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                        | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Abdominal                    | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Intra-operative (Specify)    |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)      |                   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                    | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Small Organ                  |                   |   |     |     |                  |                  |                               |                        |                      |
| (breast, testes, thyroid)    |                   |   |     |     |                  |                  |                               |                        |                      |
| Neonatal Cephalic            |                   |   |     |     |                  |                  |                               |                        |                      |
| Adult Cephalic               |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-rectal                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-vaginal                |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-urethral               |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)     |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal             |                   |   |     |     |                  |                  |                               |                        |                      |
| (Conventional)               |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal             |                   |   |     |     |                  |                  |                               |                        |                      |
| (Superficial)                |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular                |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Adult                |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Pediatric            |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular (Cardiac)      |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (Cardiac)       |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-cardiac                |                   |   |     |     |                  |                  |                               |                        |                      |
| Peripheral vessel            |                   |   |     |     |                  |                  |                               |                        |                      |
| Urology (including prostate) | P                 | P | P   |     | P                | P                | P                             | P                      |                      |

N = new indication; P = previously cleared by FDA K200450; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

{13}------------------------------------------------

# X-CUBE 70 with SC2-9H Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         | B | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
|------------------------------|---|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| Ophthalmic                   |   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                        | P | P | P   |     | P                | P                | P                             | P                      |                      |
| Abdominal                    | P | P | P   |     | P                | P                | P                             | P                      |                      |
| Intra-operative (Specify)    |   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)      |   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                 |   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                    | P | P | P   |     | P                | P                | P                             | P                      |                      |
| Small Organ                  |   |   |     |     |                  |                  |                               |                        |                      |
| (breast, testes, thyroid)    |   |   |     |     |                  |                  |                               |                        |                      |
| Neonatal Cephalic            |   |   |     |     |                  |                  |                               |                        |                      |
| Adult Cephalic               |   |   |     |     |                  |                  |                               |                        |                      |
| Trans-rectal                 |   |   |     |     |                  |                  |                               |                        |                      |
| Trans-vaginal                |   |   |     |     |                  |                  |                               |                        |                      |
| Trans-urethral               |   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)     |   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal             |   |   |     |     |                  |                  |                               |                        |                      |
| (Conventional)               |   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal             |   |   |     |     |                  |                  |                               |                        |                      |
| (Superficial)                |   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular                |   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Adult                |   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Pediatric            |   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular (Cardiac)      |   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (Cardiac)       |   |   |     |     |                  |                  |                               |                        |                      |
| Intra-cardiac                |   |   |     |     |                  |                  |                               |                        |                      |
| Peripheral vessel            |   |   |     |     |                  |                  |                               |                        |                      |
| Urology (including prostate) | P | P | P   |     | P                | P                | P                             | P                      |                      |

N = new indication; P = previously cleared by FDA K200450; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

{14}------------------------------------------------

# X-CUBE 70 with SC2-11H Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                              | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |
|------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| Clinical Application         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic                   |                   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Abdominal                    | N                 | N | N   | N   | N                | N                | N                             | N                      |                      |
| Intra-operative (Specify)    |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)      |                   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                    | N                 | N | N   | N   | N                | N                | N                             | N                      |                      |
| Small Organ                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Neonatal Cephalic            |                   |   |     |     |                  |                  |                               |                        |                      |
| Adult Cephalic               |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-rectal                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-vaginal                |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-urethral               |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)     |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal             |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal             |                   |   |     |     |                  |                  |                               |                        |                      |
| Intravascular                |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Adult                |                   |   |     |     |                  |                  |                               |                        |                      |
| Cardiac Pediatric            | N                 | N | N   | N   | N                | N                | N                             | N                      |                      |
| Intravascular (Cardiac)      |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (Cardiac)       |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-cardiac                |                   |   |     |     |                  |                  |                               |                        |                      |
| Peripheral vessel            |                   |   |     |     |                  |                  |                               |                        |                      |
| Urology (including prostate) |                   |   |     |     |                  |                  |                               |                        |                      |

N = new indication; P = previously cleared by FDA; E = added under appendix

* Combined: B/Color Doppler, B/PWD, B/Color Doppler/PWD; **Other: 3D, 4D

### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostic Devices (OIVD)

{15}------------------------------------------------

# X-CUBE 70 with MP1-5X Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                          | Mode of Operation |   |     |     |                  |                  |                               |                        |                      |
|------------------------------------------|-------------------|---|-----|-----|------------------|------------------|-------------------------------|------------------------|----------------------|
| Clinical Application                     | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>Doppler | Tissue<br>Harmonic<br>Imaging | Combined*<br>(Specify) | Other**<br>(Specify) |
| Ophthalmic                               |                   |   |     |     |                  |                  |                               |                        |                      |
| Fetal                                    |                   |   |     |     |                  |                  |                               |                        |                      |
| Abdominal                                | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Intra-operative (Specify)                |                   |   |     |     |                  |                  |                               |                        |                      |
| Intra-operative (Neuro)                  |                   |   |     |     |                  |                  |                               |                        |                      |
| Laparoscopic                             |                   |   |     |     |                  |                  |                               |                        |                      |
| Pediatric                                | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Small Organ<br>(breast, testes, thyroid) |                   |   |     |     |                  |                  |                               |                        |                      |
| Neonatal Cephalic                        |                   |   |     |     |                  |                  |                               |                        |                      |
| Adult Cephalic                           | P                 | P | P   |     | P                | P                | P                             | P                      |                      |
| Trans-rectal                             |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-vaginal                            |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-urethral                           |                   |   |     |     |                  |                  |                               |                        |                      |
| Trans-esoph. (non-Card.)                 |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skeletal<br>(Conventional)       |                   |   |     |     |                  |                  |                               |                        |                      |
| Musculo-skele…

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K211299](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K211299)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
