ArtUs

K211248 · Telemed · IYN · Feb 11, 2022 · Radiology

Device Facts

Record IDK211248
Device NameArtUs
ApplicantTelemed
Product CodeIYN · Radiology
Decision DateFeb 11, 2022
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric

Intended Use

ArtUs is a general purpose diagnostic ultrasound imaging system intended for use by qualified and appropriately trained healthcare professionals to conduct ultrasound scan process or fluid flow analysis of the human body. It is intended to be used for applications in Fetal, Abdominal, Pediatric, Small Organ (breast, thyroid, testicles), Musculo-Skeletal Conventional and Superficial, Cardiac (adult and pediatric), Adult Cephalic, and Peripheral Vascular. Modes of operation include B, M, Pulse Wave Doppler (CFM), Power Doppler (CFM), Power Doppler (PDI), Directional Power Doppler (DPDI), Combined modes (B+B, B+M, 4B, B+PWD (Duplex), B+CFM/PDI/DPDI+PWD (Triplex)), Tissue Harmonic Imaging (THI) and Inverted Tissue Harmonic Imaging (ITHI). The clinical environments where the system can be used include physician offices, clinics, hospitals, and clinical point-of-care for diagnosis of patients.

Device Story

ArtUs is a transportable, software-controlled diagnostic ultrasound system. It acquires ultrasound echo data via electronic linear, convex, or phased array transducers (2-15 MHz). The system hardware/firmware collects and pre-processes raw data, transmitting it via USB 3.0 to a Windows-based PC running Echo Wave II software. The software processes the data to display real-time B-mode, M-mode, or Doppler images (CFM, PDI, DPDI, PWD) and performs measurements. Used in clinics, hospitals, and point-of-care settings by trained professionals. The output provides visual anatomical and fluid flow information to assist clinicians in diagnosis. Benefits include real-time diagnostic imaging in a compact, transportable form factor.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including acoustic output measurements, biocompatibility testing (ISO 10993-1, -5, -10), and compliance with electrical safety and electromagnetic compatibility standards (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37).

Technological Characteristics

Transportable ultrasound system; electronic linear, convex, and phased array scanning (2-15 MHz). Materials: Silicone elastomer and plastic (acoustic lens/housing). Connectivity: USB 3.0 to host PC. Software: Echo Wave II (Windows-based). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993, IEC 62304, IEC 62366-1, ISO 14971.

Indications for Use

Indicated for qualified healthcare professionals to perform diagnostic ultrasound scans and fluid flow analysis in fetal, abdominal, pediatric, small organ (breast, thyroid, testicles), musculoskeletal, cardiac, adult cephalic, and peripheral vascular applications.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. Telemed % Yury Sokolov QA/RA Manager Highway Business Centre Savanoriu pr. 178A Vilnius, LT-03154 LITHUANIA February 11, 2022 Re: K211248 Trade/Device Name: ArtUs Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: December 30, 2021 Received: January 10, 2022 Dear Yury Sokolov: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for {1}------------------------------------------------ devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. For Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K211248 Device Name ArtUs #### Indications for Use (Describe) ArtUs is a general purpose diagnostic ultrasound imaging system intended for use by qualified and appropriately trained healthcare professionals to conduct ultrasound scan process or fluid flow analysis of the human body. It is intended to be used for applications in Fetal, Abdominal, Pediatric, Small Organ (breast, thyroid, testicles), Musculo-Skeletal Conventional and Superficial, Cardiac (adult and pediatric), Adult Cephalic, and Peripheral Vascular. Modes of operation include B, M, Pulse Wave Doppler (CFM), Power Doppler (CFM), Power Doppler (PDI), Directional Power Doppler (DPDI), Combined modes (B+B, B+M, 4B, B+PWD (Duplex), B+CFM/PDI/DPDI+PWD (Triplex)), Tissue Harmonic Imaging (THI) and Inverted Tissue Harmonic Imaging (ITHI). The clinical environments where the system can be used include physician offices, clinics, hospitals, and clinical point-ofcare for diagnosis of patients. Type of Use (Select one or both, as applicable) | <span> <b> </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |-----------------------------------------------------------------------------| | <span> <b> </b> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ In accordance with 21 CFR 807.92 the following summary of information is provided: # K211248 -510k Summary #### 1. Identifying information Manufacturer TELEMED Device Trade Name ArtUs Highway Business Centre Savanoriu pr. 178A Address Vilnius, LT-03154, Lithuania Telephone +370-5 2106272 +370-5 2106273 +370-5 2306733 Fax Web http://www.pcultrasound.com/ www.telemed.lt info@telemed.lt E-mail Yury Sokolov / QA Manager PRRC yury@pcultrasound.com Contact Person Date 510(k) Summary Prepared December 30, 2021 #### 2. Class and Predicate Information | Classification Name | Regulation Number | Product Code | | |------------------------------------------|------------------------------|-----------------------|--------------------| | Ultrasonic Pulsed Doppler Imaging System | 892.1550 | IYN | | | Ultrasonic Pulsed Echo Imaging System | 892.1560 | IYO | | | Diagnostic Ultrasonic Transducer | 892.1570 | ITX | | | Common Name | Diagnostic Ultrasound System | | | | Propietary Name | ArtUs | | | | Classification | Regulatory Class II | | | | Predicate Device | TELEMED<br>SAMSUNG MEDISON | SmartUs<br>SONOACE R7 | K163121<br>K112646 | #### 3. LIST OF CONSENSUS STANDARDS The proposed device ArtUs is in conformity with the requirements of the following consensus standards: [Rec. # 19-4] ANSI/AAMI ES60601-1 Medical electrical equipment -Part 1: General requirements for safety and essential performance, [Rec. # 19-8] IEC 60601-1-2 Medical electrical equipment --Part 1-2 General requirements for safety and essential performance -Collateral standard: Electromagnetic compatibility -Requirements and tests [Rec. # 12-293] IEC 60601-2-37 Medical electrical equipment -Part 2-37; Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment [Rec. # 13-79] IEC 62304 Medical device software - Software life cycle processes [Rec. # 5-114] IEC 62366-1 Medical devices - Part 1: Application of usability engineering to medical devices [Rec. # 5-125] ISO 14971 Medical devices - Application of risk management to medical devices [Rec. # 5-117] ISO 15223-1 Medical devices - Symbols to be used with medical device labels labelling and information to be supplied - Part 1: General requirements [Rec. # 2-220] ISO 10993-1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process [Rec. # 2-245] ISO 10993-5 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity {4}------------------------------------------------ [Rec. # 12-293] ISO 10993-10 Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization #### 4. Indication for Use ArtUs is a general purpose diagnostic ultrasound imaging system intended for use by qualified and appropriately trained healthcare professionals to conduct ultrasound scan process or fluid flow analysis of the human body. It is intended to be used for applications in Fetal, Abdominal, Pediatric, Small Organ (breast, thyroid, testicles), Musculo-Skeletal Conventional and Superficial, Cardiac (adult and pediatric), Adult Cephalic, and Peripheral Vascular. Modes of operation include B. M. Pulse Wave Doppler (PWD). Color Doppler (CFM). Power Doppler (PDI), Directional Power Doppler (DPDI), Combined modes (B+B, B+M, 4B, B+PWD (Duplex), B+CFM/PDI/DPDI+PWD (Triplex)), Tissue Harmonic Imaging (THI) and Inverted Tissue Harmonic Imaging (ITHI) The clinical environments where the system can be used include physician offices, clinics, hospitals, and clinical point-of-care for diagnosis of patients. #### 5. Device Description ArtUs system is intended for the multipurpose ultrasound examinations, based on electronic linear and convex scanning. ArtUs system is a combination of proprietary hardware and software that has been designed for real-time imaging and is intended to be a basic diagnostic tool. Its basic function is to acquire ultrasound echo data and to display the image in ultrasound B-Mode or combined modes. The system is designed for imaging with transducer ranges of 2 to 15 MHz. The devices referenced in this submission represent a transportable, software-controlled, diagnostic ultrasound system with accessories. This submission does not include technology or control feature changes nor deviations from indications for use different from those demonstrated in previously cleared devices operating in ultrasound B-Mode, M-Mode or combined modes, inclusive of the predicate devices so claimed. The ArtUs only contains the hardware and firmware, everything else (e.g. ultrasound software, database) is located on a standard PC that is connected to the ArtUs via USB 3.0. Minimum requirements are given for the PC. The Echo Wave II software was especially designed for the TELEMED devices. Software able to reside in a Windows-based PC. The device variant is: - ArtUs EXT-1H ultrasound system utilizing as hardware an ultrasound engine contained in a small standalone enclosure for connection to a host PC via a USB port with external power supply; The ArtUs can be used together with the appropriate transducers for the entire ultrasound diagnostic (2MHz to 15MHz probes). {5}------------------------------------------------ - . Transducer L15-7H40-A5, linear array, at a central frequency of approx. 12 MHz - . Transducer C5-2H60-A5, convex array at a central frequency of approx. 3.5 MHz; - Transducer P5-1S15-A6, phased array at a central ultrasonic frequency of approx. 4 MHz . #### 6. General Safety and Effectiveness. The ArtUs ultrasound system and its accessories are designed for compliance to all applicable medical devices safety standards, as referenced in Declaration of Conformity. Prior release for manufacturing, all such devices, so designed, are tested and determined to be in full compliance with acoustic output, biocompatibility, cleaning and disinfection effectiveness. - . No additional clinical testing is required, as the indications for use are not a novel indication as shown by the predicate devices. - . Maximum acoustic output level is under by the FDA recommended limit and power level is displayed all the time. - . The system's acoustic output is in accordance with ALARA principle (As Low As Reasonably Achievable) #### 7. Patient Contact Materials The parts / materials of the transducer, coming in contact with patient are: - . Acoustic Lens, Transducer Housing / Silicone Elastomer, Plastic Standards for the biological evaluation: - . ISO-10993-1:2009, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing within a risk management process - ISO-10993-5:2009, Biological Evaluation of Medical Devices Part 5: Tests for in vitro ● cytotoxicity - ISO-10993-10:2010, Biological Evaluation of Medical Devices Part 10: Tests for irritation ● and skin sensitization #### 8. Software The ArtUs system contains the hardware and software which collect and pro-processes 'rough" data and send it via USB 3.0 connection to a Windows® based PC. The main application software is Echo Wave II software running on the PC, it is receiving data, processing and showing image/data on the screen. The main user interface shows an ultrasound image, controls and drop-out menus. The ultrasound images and calculated/measured data can be stored in memory. ### 9. Determination of Substantial Equivalence The ArtUs color Doppler ultrasound system is similar to currently distributed ultrasonic pulsed echo imaging systems SmartUs (TELEMED) and SONOACE R7 (SAMSUNG MEDISON) in terms of both the intended use and technological characteristics. The ArtUs (subject device) uses the same {6}------------------------------------------------ fundamental scientific technology as the predicate devices. All systems transmit ultrasonic energy into patients, then perform post processing of received echoes to generate on-screen display of anatomic structures and fluid flow within the body. This device ArtUS is a successor of an existing licensed device SmartUs (cleared via K163121) using technologies that exist on the market as of the date of this submission. In cosmetic design, the subject device (ArtUs) has a smaller and lighter structure and appearance than SmartUs due to modern electronic components. The PCBs in the system are minimized in their size and layout, while the overall circuit operation principle is not changed. The connection of new ultrasound transducers is easier and more reliable using the new generation of connectors. The connection to the PC by USB 3.0 (comparing USB 2.0 connection for SmartUs) is increasing the dataflow performance between the device and PC. The subject device ArtUs meets FDA requirements for Track 3 devices, have biosafety equivalence, and conform to applicable electromedical devices safety standards. All the safety and performance tests of the device meet the essential requirements. Therefore, the system has the same classification and is substantially equivalent to predicate devices. Any differences between the predicate and the new device have no impact on safety or efficacy of the new device and do not raise any new potential or increased safety risks, and the new device is equivalent in performance to existing legally marketed devices. #### 10. Conclusion In accordance with the FDA and based on the information provided in this Premarket notification, TELEMED considers the ArtUs to be as safe as effective and performance is substantially equivalent to the predicate devices described herein.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...