← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K201768

# Voluson E10 (K201768)

_GE Healthcare · IYN · Oct 22, 2020 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K201768

## Device Facts

- **Applicant:** GE Healthcare
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Oct 22, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Intended Use

The device is a general purpose ultrasound system intended for use by qualified and trained healthcare professionals. Specific clinical applications remain the same as previously cleared: Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatric; Small Organ (breast, testes, thyroid etc.); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Musculo-skeletal Conventional and Superficial; Peripheral Vascular; Transvaginal (including GYN); Transrectal Modes of operation include: B, M, PW Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode and Combined modes: B/M, B/Color, B/Color/PWD, B/Power/PWD. The Voluson E10 is intended to be used in a hospital or medical clinic.

## Device Story

Mobile ultrasound console with keyboard, touch panel, and display; utilizes linear, curved linear, and matrix phased array transducers for real-time 3D/4D imaging. Inputs: acoustic signals from patient anatomy. Processing: beamforming, signal processing, and image reconstruction for B-mode, M-mode, Doppler, and harmonic imaging. Outputs: real-time diagnostic images and measurements on integrated display. Used in hospitals/clinics by trained clinicians. Features include SonoBiometry for automated measurements, Scan Assistant for workflow guidance, Fetal Heart Screening, and e-PAT for probe element integrity checks. Clinicians use output for diagnostic assessment and clinical decision-making. Benefits include high-performance imaging and workflow efficiency.

## Clinical Evidence

No clinical data required. Substantial equivalence supported by bench testing, including acoustic output measurement, biocompatibility, cleaning/disinfection effectiveness, and thermal/electrical/electromagnetic/mechanical safety testing. System verification and validation performed via unit-level testing, integration testing, and final acceptance testing.

## Technological Characteristics

Mobile ultrasound system; uses linear, curved linear, and matrix phased array transducers. Materials are biocompatible per ISO 10993-1. Connectivity includes DICOM (NEMA PS 3.1-3.20). Software features include automated measurement and workflow tools. Complies with AAMI/ANSI ES60601-1, IEC60601-1-2, and IEC60601-2-37. Energy source: electrical mains.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Voluson E6_E8_E10 Diagnostic Ultrasound System ([K192159](/device/K192159.md))
- LOGIQ E10 Diagnostic Ultrasound System ([K173555](/device/K173555.md))
- Vivid E95 Diagnostic Ultrasound System ([K181685](/device/K181685.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

October 22, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

GE Healthcare % Mr. Bryan Behn Regulatory Affairs Director 9900 Innovation Drive WAUWATOSA WI 53226

Re: K201768

Trade/Device Name: Voluson E10 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: September 16, 2020 Received: September 17, 2020

Dear Mr. Behn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K201768

Device Name Voluson E10

#### Indications for Use (Describe)

The device is a general purpose ultrasound system intended for use by qualified and trained healthcare professionals. Specific clinical applications remain the same as previously cleared: Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatric; Small Organ (breast, testes, thyroid etc.); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Musculo-skeletal Conventional and Superficial; Peripheral Vascular; Transvaginal (including GYN); Transrectal

Modes of operation include: B, M, PW Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode and Combined modes: B/M, B/Color, B/Color/PWD, B/Power/PWD. The Voluson E10 is intended to be used in a hospital or medical clinic.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# K201768

GE Healthcare 510(k) Premarket Notification Submission

### 510(k) Summary

| In accordance with 21 CFR 807.92 the following summary of information is provided: |                                                                                                                                                                                                |
|------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date:                                                                              | June 24, 2020                                                                                                                                                                                  |
| Submitter:                                                                         | GE Healthcare [GE Healthcare Austria GmbH & Co OG]<br>Tiefenbach 15<br>Zipf, Austria 4871                                                                                                      |
| Primary Contact<br>Person:                                                         | Bryan Behn<br>Regulatory Affairs Director<br>GE Healthcare<br>9900 Innovation Dr<br>Wauwatosa, WI 53226<br>T:(262)247-5502<br>F:(414)918-8275                                                  |
| Secondary Contact<br>Person:                                                       | Roland Kuntscher<br>Regulatory Affairs Specialist<br>GE Healthcare Austria GmbH & Co OG<br>Tiefenbach 15<br>Zipf, Austria 4871<br>T:(+43)7682-3800-660<br>F:(+43)7682 3800-47                  |
| Device:<br>Trade Name:                                                             | Voluson E10<br>Models: Voluson E10                                                                                                                                                             |
| Common/Usual<br>Name:                                                              | Ultrasound system                                                                                                                                                                              |
| Classification Names:                                                              | Class II                                                                                                                                                                                       |
| Product Code:                                                                      | Ultrasonic Pulsed Doppler Imaging System. 21CFR 892.1550, 90-IYN<br>Ultrasonic Pulsed Echo Imaging System, 21CFR 892.1560, 90-IYO<br>Diagnostic Ultrasound Transducer, 21 CFR 892.1570, 90-ITX |
| Primary Predicate<br>Device(s):                                                    | K192159 Voluson E6_E8_E10 Diagnostic Ultrasound System                                                                                                                                         |
| Secondary Predicate<br>Device(s):                                                  | K173555 LOGIQ E10 Diagnostic Ultrasound System                                                                                                                                                 |

K181685 Vivid E95 Diagnostic Ultrasound System

Image /page/3/Picture/5 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a circular border. The color of the logo is a medium blue.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'g' and 'e' intertwined in a stylized script, enclosed within a blue circle. There are decorative white swirls around the outer edge of the circle, adding a touch of elegance to the design.

## GE Healthcare 510(k) Premarket Notification Submission

- Device Description: The systems are full-featured Track 3 ultrasound systems, primarily for general radiology use and specialized for OB/GYN with particular features for real-time 3D/4D acquisition. They consist of a mobile console with keyboard control panel; color LCD/TFT touch panel, color video display and optional image storage and printing devices. They provide high performance ultrasound imaging and analysis and have comprehensive networking and DICOM capability. They utilize a variety of linear, curved linear, matrix phased array transducers including mechanical and electronic scanning transducers, which provide highly accurate real-time three-dimensional imaging supporting all standard acquisition modes.
	- The device is a general-purpose ultrasound system intended for use by Intended Use: qualified and trained healthcare professionals. Specific clinical applications remain the same as previously cleared: Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatric; Small Organ (breast, testes, thyroid etc.); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Musculoskeletal Conventional and Superficial; Peripheral Vascular; Transvaginal (including GYN); Transrectal

Modes of operation include: B, M, PW Doppler, Color Doppler, Color M Doppler, Power Doppler, Harmonic Imaging, Coded Pulse, 3D/4D Imaging mode and Combined modes: B/M, B/Color, B/PWD, B/Color/PWD, B/Power/PWD. The Voluson E10 is intended to be used in a hospital or medical clinic.

- The Voluson E Series (Voluson E10) employs the same fundamental Technology: scientific technology as its predicate devices.
#### Determination of Comparison to Predicates Substantial Equivalence:

- The proposed Voluson E10 is substantially equivalent to the predicate devices with regards to intended use, imaging capabilities, technological characteristics and safety and effectiveness.
	- The systems are all intended for diagnostic ultrasound imaging and fluid flow analysis.
	- The proposed Voluson E10 and predicate Voluson E Series systems have the same clinical intended use.
	- The proposed Voluson E10 and predicate Voluson E Series systems have the same imaging modes.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized monogram, enclosed within a blue circle. The circle has small, white, teardrop-shaped accents around its perimeter, giving it a slightly textured or dynamic appearance.

# GE Healthcare

510(k) Premarket Notification Submission

- The proposed Voluson E10 and predicate Voluson E Series . systems transducers are equivalent. Two new transducers 12S-D and eM6CG3 were added and 10 transducers removed to the proposed system.
- There is no change to the system indications for use. ●
- . The systems are manufactured with materials which have been evaluated and found to be safe for the intended use of the device.
- The systems have acoustic power levels which are below . the applicable FDA limits.
- The proposed Voluson E10 and predicate Voluson E Series systems have similar capability in terms of performing measurements, capturing digital images, reviewing and reporting studies.
- . The proposed Voluson E10 and predicate systems have been designed in compliance with approved electrical and physical safety standards.
- . There proposed Voluson E10 and predicate Voluson E series Software Features are equivalent. Some minor improvements to the existing Software features have been implemented into the proposed system.
- The proposed Voluson E10 adds additional measurements of ● existing software feature SonoBiometry (workflow improvement).
- The proposed Voluson E10 adds Image Analysis of existing ● software feature Scan Assistant (workflow improvement).
- The proposed Voluson E10 adds Fetal Heart Screening (workflow improvement).
- . The proposed Voluson E10 adds functionality check of probe crystals/elements to existing e-PAT (electronic probe assessment tool) acc. FDA guideline (modified for customer access add.to previous service tool).

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters 'GE' intertwined within a blue circle. There are small white droplets or swirls around the perimeter of the circle, giving it a dynamic and fluid appearance.

GE Healthcare 510(k) Premarket Notification Submission

### Summary of Non-Clinical Tests:

The device has been evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness as well as thermal, electrical, electromagnetic, and mechanical safety, and has been found to conform to applicable medical device safety standards. The Voluson E Series and its applications comply with voluntary standards:

- AAMI/ANSI ES60601-1, Medical Electrical Equipment Part o 1: General Requirements for Safety, 2005/(R)2012 And A1:2012
- IEC60601-1-2 Medical Electrical Equipment Part 1-2: ● General Requirements for Safety - Collateral Standard: Electromagnetic Compatibility Requirements and Tests, 2014
- IEC60601-2-37, Medical Electrical Equipment - Part 2-37: Particular Requirements for the Safety of Ultrasonic Medical Diagnostic and Monitoring Equipment, 2015
- ISO10993-1, Biological Evaluation of Medical Devices- Part 1: . Evaluation and Testing- Third Edition, 2009
- . ISO14971, Application of risk management to medical devices: Second edition 2007
- NEMA PS 3.1 3.20 (2016), Digital Imaging and ● Communications in Medicine (DICOM) Set. (Radiology)

The following quality assurance measures are applied to the development of the system:

- Risk Analysis
- Requirements Reviews ●
- Design Reviews ●
- Testing on unit level (Module verification) ●
- Integration testing (System verification) ●
- . Performance testing (Verification)
- Safety testing (Verification)
- Final Acceptance Testing (Validation) .

Transducer materials and other patient contact materials are biocompatible.

### Summary of Clinical Tests:

The subject of this premarket submission, Voluson E10 did not require

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined in a stylized script, enclosed within a blue circle. The circle has decorative white swirls around the edges, giving it a classic and recognizable appearance.

GE Healthcare 510(k) Premarket Notification Submission

clinical studies to support substantial equivalence.

Conclusion: GE Healthcare considers the Voluson E10 to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K201768](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K201768)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
