The Teratech Corporation Terason™ uSmart3200T Plus is a prescription-only, general-purpose Ultrasound System intended for use by a qualified physician or accredited sonographer for evaluation by ultrasound imaging or fluid flow analysis of the human body. The device is intended for use in a clinic, hospital or medical-emergency setting. Specific clinical applications and exam types include: Ophthalmic, Fetal, Abdominal, Intra-operative (Spec. and Neuro.), Pediatrics, Small Organ (Thyroid, Breast, Testes); Neonatal and Adult Cephalic, Trans-rectal, Trans-vaginal, Trans-esophageal (non-cardiac), Musculo-skeletal (Conventional and Superficial), Cardiac (Adult & Pediatric), Trans-esophageal (cardiac), and Peripheral Vascular.
Device Story
Portable tablet-style diagnostic ultrasound system; acquires/displays real-time ultrasound data. Proprietary ultrasound engine controls transducer acoustic output and processes return echoes; data transferred via USB3 to tablet for image generation/display. Used in clinics, hospitals, or emergency settings by physicians or sonographers. System includes 15.6" backlit touchscreen, integrated Lithium-Polymer battery, and docking station. Supports multiple imaging modes: B-Mode, Tissue Harmonic, M-Mode, Anatomical M-Mode, Color M-Mode, Color Power Doppler, Velocity Color Doppler, PW Doppler, and TeraVision II post-processing. Output viewed on tablet screen; assists clinicians in diagnostic evaluation of human body structures and fluid flow. Benefits include high-resolution imaging in a portable form factor.
Clinical Evidence
Bench testing only. No clinical trials were referenced or conducted. Performance validated via acoustic output testing (IEC 60601-2-37, NEMA UD 2/3), B-Mode/Doppler accuracy and sensitivity testing (AIUM standards), biocompatibility testing (ISO 10993), electromagnetic compatibility (IEC 60601-1-2), and software regression testing.
Technological Characteristics
Portable tablet (15.6" touchscreen, 9.8 lbs) running Windows 10. Proprietary internal ultrasound engine; USB3 connectivity. Transducers use piezoelectric elements. Patient contact materials include silicone. Compliance with AAMI/ANSI ES 60601-1, IEC 60601-1-2, IEC 60601-1-6, and IEC 60601-2-37. Supports 384 transmit/receive channels. Sterilization/reprocessing per AAMI TIR No. 12.
Indications for Use
Indicated for ophthalmic, fetal, abdominal, intra-operative, pediatric, small organ (thyroid, breast, testes), neonatal and adult cephalic, trans-rectal, trans-vaginal, trans-esophageal (cardiac and non-cardiac), musculo-skeletal, cardiac, and peripheral vascular ultrasound imaging and fluid flow analysis in clinical, hospital, or emergency settings.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
K193510 — Terason uSmart3200T Ultrasound System · Teratech Corporation · Jan 28, 2020
K150533 — Terason uSmart3200T Ultrasound System · Teratech Corp. · May 9, 2015
K152983 — FUJIFILM SonoSite iViz Ultrasound System · FUJIFILM Sonosite, Inc. · Nov 12, 2015
K113663 — SONIXTABLET ULTRASOUND SCANNER · Ultrasonix Medical Corporation · Jan 25, 2012
K113156 — SONOSITE EDGE ULTRASOUND SYSTEM · Sonosite,Inc. · Nov 15, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
July 2, 2020
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Teratech Corporation % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite 510k SAINT PAUL MN 55114
## Re: K201633
Trade/Device Name: Terason™ uSmart 3200T Plus Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: June 12, 2020 Received: June 16, 2020
Dear Prithul Bom:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201633
Device Name
Terason uSmart3200T Plus Ultrasound System
### Indications for Use (Describe)
The Teratech Corporation Terason uSmart32007 Plus is a prescription-only, general-purpose Ultrasound System intended for use by a qualified physician or accredited sonographer for evaluation by ultrasound imaging of the human body. The device is intended for use in a clinic, hospital or medical-emergency setting. Specific clinical applications and exam types include: Ophthalmic, Fetal, Abdominal, Intra-Operative (Spec. and Neuro.), Pediatrics, Small Organ (Thyroid, Breast, Testes); Neonatal and Adult Cephalic; Trans-vaginal, Trans-esophageal (noncardiac), Musculo-skeletal (Conventional); Cardiac (Adult & Pediatric), Trans-esophageal (cardiac), and Peripheral Vascular.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) Summary
### K201633
## Teratech Corporation
## Terason uSmart3200T Plus Ultrasound System
### 1. Sponsor:
Teratech Corporation 77-79 Terrace Hall Ave. Burlington, MA 01803
Contact Person: Ben Chiampa Director of Quality Assurance Telephone: 781-270-4143
Date Prepared: May 20, 2020
## 2. Device Name
Proprietary Name: Terason uSmart3200T Plus Ultrasound System Common / Usual Name: Diaqnostic Ultrasound System Classification Name: Class II
Ultrasonic Pulsed Doppler Imaging System (21 CFR 892.1550, 90-IYN) Ultrasonic Pulsed Echo Imaging System (21 CFR 892.1560, 90-IYO) Diagnostic Ultrasonic Transducer (21 CFR 892.1570, 90-ITX)
## 3. Predicate Device
Terason™ uSmart3200T Ultrasound System (K193510)
## 4. Intended Use
The Teratech Corporation Terason™ uSmart3200T Plus is a prescription-only, general-purpose Ultrasound System intended for use by a qualified physician or accredited sonographer for evaluation by ultrasound imaging or fluid flow
{4}------------------------------------------------
analysis of the human body. The device is intended for use in a clinic, hospital or medical-emergency setting. Specific clinical applications and exam types include: Ophthalmic, Fetal, Abdominal, Intra-operative (Spec. and Neuro.), Pediatrics, Small Organ (Thyroid, Breast, Testes); Neonatal and Adult Cephalic, Trans-rectal, Trans-vaginal, Trans-esophageal (non-cardiac), Musculo-skeletal (Conventional and Superficial), Cardiac (Adult & Pediatric), Trans-esophageal (cardiac), and Peripheral Vascular.
## 5. Clinical Testing
Teratech Corporation certifies that the requirements of 42 U.S.C. 282(i), Section 402(i) of the Public Health Service Act, including 42 CFR part 11, do not apply because the application/submission which this 510(k) filing accompanies does not reference any clinical trial.
## 6. Device Description
The Terason uSmart3200T Plus ultrasound system is a portable tablet-style, fullfeature, general purpose diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through multiple imaging modes. The Terason uSmart3200T Plus Ultrasound System is equivalent to the previously cleared versions of the uSmart3200T Ultrasound Systems. The modification includes the addition of 3 transducers: 14L3, 10BP4 and 15XL4.
The Terason™ uSmart3200T ultrasound system was the previously cleared on the date January 29, 2020 as described in the 510(k) submission (K193510). This system contains a proprietary ultrasound engine for controlling the acoustic output of the transducer and processing the return echoes in real time. These data are then transferred to the tablet computer over a Universal Serial Bus (USB3) connection for further processing and generation/display of the ultrasound image.
The Terason™ uSmart3200T Plus ultrasound tablet weighs 9.8 pounds (4.2 Kq) and has an 15.6" backlit touch screen. The tablet dimensions (10.6"(H) x 15.6"(W) x 1.7"(D)) are chosen to allow portability. A Lithium-Polymer battery (integrated into the tablet) provides 2 hours of continuous ultrasound scanning. The tablet includes a docking station (for charging) that uses a medical-grade power supply. The ultrasound transducer connector is identical to that used in the
{5}------------------------------------------------
Terason™ predicate device, the uSmart3200T Plus. Optional accessories include a cart and printer.
## 7. Technology Characteristics
The design and construction of the Terason uSmart3200T Plus is the same as the Terason uSmart3200T Plus Ultrasound system which was cleared in May 2015. This system utilizes a portable computer running Windows 10 to run the ultrasound application and a custom hardware designed engine for control of the acoustic array and processing of the return echoes. The engine is housed in a compartment that is inside the tablet.
The uSmart3200T Plus system contains the same ultrasound engine as the predicate device Terason uSmart3200T Plus ultrasound system for controlling the acoustic output of the transducer and processing the return echoes in real time. These data are then transferred to the tablet computer over a USB3 connection for further processing, and generation and display of the ultrasound image
The differences between the Terason uSmart3200T Plus and the previous Terason uSmart3200T Plus Ultrasound System (the predicate device) include the following:
- . Three transducers have been added to the system: 14L3, 10BP4, and 15XL4. The software has been modified to control these transducers and ensure compliance to the standards controlling acoustic and thermal power.
- Added support for the 14L3, 10BP4, and 15XL4 transducers .
- o Confirmed transducer id numbers and names
- o Confirmed transducer geometries and characteristic parameters
- o Confirmed 14L3, 10BP4, and 15XL4 acoustic tables
- o Added 14L3, 10BP4, and 15XL4 to the table of allowed transducers
- o Added imaging presets for 14L3, 10BP4, and 15XL4 transducers.
{6}------------------------------------------------
# 8. Table of Similarities and Differences Compared to the Predicate Devices
Terason uSmart3200T Plus System and Transducers Comparison and Discussion New Transducers: 14L3, 10BP4, 15XL4 Previously cleared transducers (5V1A, 12L5A, 5C2A, 8EC4A, 16HL7, 15L4, 8L2, 8V3A, 9MC3, 8TE3, PDOF, 15L4A, 16L5, 15WL4, XY-Biplane, 10EC4) (K193510)
| | Subject Device Model | Comparable Predicate Device | Same or Different |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------|
| | Terason uSmart3200T Plus<br>(This Submission) | Terason uSmart3200T<br>K193510 | |
| Intended Use | Diagnostic Ultrasound imaging or fluid flow analysis of the human body | Diagnostic Ultrasound imaging or fluid flow analysis of the human body | Same. |
| Indication for Use | Ophthalmic, Fetal,<br>Abdominal, Intra-operative<br>(Spec.), Pediatric, Small Organ<br>(Thyroid, Breast, Testes, etc.),<br>Neonatal Cephalic, Adult<br>Cephalic, Trans-rectal, Trans-<br>vaginal, Trans-esophageal<br>(non-Cardiac), Musculo-skel.<br>(Convent.), Musculo-skel.<br>(Superfic), Cardiac Adult,<br>Cardiac Pediatric, Trans-<br>esophageal (Cardiac),<br>Peripheral vessel | Ophthalmic, Fetal,<br>Abdominal, Intra-operative<br>(Spec.), Pediatric, Small Organ<br>(Thyroid, Breast, Testes, etc.),<br>Neonatal Cephalic, Adult<br>Cephalic, Trans-rectal, Trans-<br>vaginal, Trans-esophageal<br>(non-Cardiac), Musculo-skel.<br>(Convent.), Musculo-skel.<br>(Superfic), Cardiac Adult,<br>Cardiac Pediatric, Trans-<br>esophageal (Cardiac),<br>Peripheral vessel | Same. |
| Transducer Types | Linear Array<br>Curved Array<br>Phased Array<br>Endocavity - curved array<br>Hockey Stick - Linear<br>Trans-esophageal<br>PDOF<br>192-Element<br>384-element<br>XY-Biplane<br>Endocavity-Biplane | Linear Array<br>Curved Array<br>Phased Array<br>Endocavity - curved array<br>Hockey Stick - Linear<br>Trans-esophageal<br>PDOF<br>192-Element<br>XY-Biplane | Different: Support for 384-<br>elements and biplane<br>transducers. |
| Acoustic Output and FDA<br>Limits | Display Features for High<br>Outputs | Display Features for High<br>Outputs | Same. |
| Global Maximum<br>Outputs/Worst Case Setting | ISPTA.3: 619 mW/cm² (10BP4C)<br>TI Type: TIC (14L3)<br>TI Value: 4.2 (14L3)<br>MI: 1.76 (15XL4)<br>IPA.3@MI Max: 776 W/cm²<br>(14L3) | ISPTA.3: 652.9 mW/cm² (4V2A)<br>TI Type: TIC (15L4)<br>TI Value: 5.8 (15L4)<br>MI: 1.78 (8EC4A)<br>IPA.3@MI Max: 827 W/cm²<br>(1514) | Different. Within Guideline<br>limits. |
| Modes of Operation | B-Mode Grayscale Imaging<br>Tissue Harmonic Imaging<br>M-Mode (motion)<br>Anatomical M-Mode<br>Color M-Mode<br>Color Power Doppler<br>Velocity Color Doppler<br>Duplex/Triplex – Doppler<br>imaging<br>Pulsed Wave (PW) Doppler<br>TeraVision II Postprocessing | B-Mode Grayscale Imaging<br>Tissue Harmonic Imaging<br>M-Mode (motion)<br>Anatomical M-Mode<br>Color M-Mode<br>Color Power Doppler<br>Velocity Color Doppler<br>Duplex/Triplex – Doppler<br>imaging<br>Pulsed Wave (PW) Doppler<br>TeraVision II Postprocessing | Same. |
| PW Doppler | Available for all transducers<br>Triplex Mode<br>B-Mode and PW Doppler<br>High PRF | Available for all transducers<br>Triplex Mode<br>B-Mode and PW Doppler<br>High PRF | Same. |
| Transducer Frequency | 2.0 – 15.0 MHz | 2.0 – 15.0 MHz | Same. |
| #Transmit Channels | 384 Channels | 128 Channels | Different. Supports 384<br>elements. |
| # Receive Channels | 384 Channels | 128 Channels | Different. Supports 384<br>elements. |
| Acoustic Output<br>Measurement Standard | IEC 60601-2-37 | NEMA UD 2-2004<br>NEMA UD 3-2004 | Same. |
| DICOM | DICOM 3.0 Structured<br>Reporting, Worklist - Image<br>Viewer | DICOM 3.0 Structured<br>Reporting, Worklist - Image<br>Viewer | Same. |
| Product Safety Certification | AAMI/ANSI ES 60601-<br>1:2005/(R)2012 and A1:2012<br>Medical electrical equipment<br>– Part 1: General<br>requirements for safety and<br>essential performance<br>IEC60601-1<br>IEC60601-1-2<br>IEC60601-1-6<br>IEC60601-2-37 | AAMI/ANSI ES 60601-<br>1:2005/(R)2012 and A1:2012<br>Medical electrical equipment<br>– Part 1: General<br>requirements for safety and<br>essential performance<br>IEC60601-1<br>IEC60601-1-2<br>IEC60601-1-6<br>IEC60601-2-37 | Same. |
| EMC | IEC60601-1-2<br>CISPR11 Class B | IEC60601-1-2<br>CISPR11 Class B | Same. |
| System<br>Characteristics | uSmart3200T Plus:<br>tablet computer<br>weighs 9.8 lbs (4.2 Kg)<br>18" backlit touch screen.<br>Tablet dimensions (10.6"(H) x<br>15.6"(W) x 1.7"(D)).<br>A Lithium-Polymer battery<br>(integrated into the tablet) | uSmart3200T:<br>tablet computer<br>weighs 4.9 lbs (2.21 Kg)<br>11.5" backlit touch screen.<br>Tablet dimensions (8.82"(H) x<br>12.64"(W) x 1.25"(D)).<br>A Lithium-Polymer battery<br>(integrated into the tablet) | Different.<br>Larger screen |
| continuous ultrasound<br>scanning<br>Docking station (for charging)<br>that uses a medical-grade<br>power supply<br>Data transferred to the tablet<br>computer over a USB3<br>connection | continuous ultrasound<br>scanning<br>Docking station (for charging)<br>that uses a medical-grade<br>power supply<br>Data transferred to the tablet<br>computer over a USB3<br>connection | | |
#### Terason uSmart3200T Plus Tablet Computer
{7}------------------------------------------------
{8}------------------------------------------------
Three transducers have been added to the Terason uSmart3200T Plus in this submission: 14L3, 10BP4, and 15XL4.
{9}------------------------------------------------
### SUMMARY OF NEW AND ASSOCIATED PREDICATE TRANSDUCERS
| New Subject Transducer | Comparable Predicate Transducer | Predicate Approvals and Systems |
|------------------------|---------------------------------|---------------------------------|
| Terason 14L3 | Terason 15L4 | K193510 (Terason uSmart3200T) |
| Terason 10BP4 | Terason 10EC4 | K193510 (Terason uSmart3200T) |
| Terason 15XL4 | Terason 15WL4 | K193510 (Terason uSmart3200T) |
### TRANSDUCER PERFORMANCE SUMMARY
| Transducer | Indications | Mode | Global maximum output | 510(K)<br>control number |
|------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|--------------------------------------------------------|--------------------------|
| 12L5A | Ophthalmic, Abdominal, Pediatric,<br>Small Organ (breast, testes, thyroid),<br>Neonatal and Adult Cephalic,<br>Musculo-sketetal (Conventional and<br>Superficial), and Peripheral Vascular | B, M,<br>PWD,<br>Color Doppler,<br>Combined | ISTPA.3 = 597(mW/cm²)<br>TI = 3.3<br>MI = 1.8 | K193510 |
| 5V1A | Fetal, Abdominal, Pediatric, Neonatal<br>and Adult Cephalic, Cardiac (adult<br>and pediatric) | B, M,<br>PWD,<br>Color Doppler,<br>CWD,<br>Combined | ISTPA.3 = 603 (mW/cm²)<br>TI = 5.6<br>MI = 1.5 | K193510 |
| 5C2A | Fetal, Abdominal, Pediatric, Small<br>Organ (Thyroid, Breast, Testes, etc.);<br>Musculo-sketetal (Conventional and<br>Superficial), Cardiac (adult and<br>pediatric) and Peripheral Vascular | B, M,<br>PWD,<br>Color Doppler,<br>Combined | ISTPA.3 = 660 (mW/cm²)<br>TI = 4.7<br>MI = 0.7 | K193510 |
| 8EC4A | Fetal, Trans-rectal, Trans-vaginal | B, M,<br>PWD,<br>Color Doppler,<br>Combined | ISTPA.3 = 633(mW/cm²)<br>TI = 2.29<br>MI = 1.8 | K193510 |
| 16HL7 | Intra-Operative (abdominal, organs<br>and vascular), Small Organ (Thyroid,<br>Breast, Testes); Musculo-skeletal<br>(Conventional and Superficial);<br>Peripheral Vascular. | B, M,<br>PWD,<br>Color Doppler,<br>Combined | ISTPA.3 = 554(mW/cm²)<br>TI = 1.22<br>MI = 1.6 | K193510 |
| 15L4 | Ophthalmic, Abdominal, Pediatric,<br>Small Organ (Thyroid, Breast,<br>Testes); Musculo-skeletal<br>(Conventional and Superficial);<br>Peripheral Vascular. | B, M,<br>PWD,<br>Color Doppler,<br>Combined | ISTPA.3 = 563(mW/cm²)<br>TI = 5.8<br>MI = 1.7 | K193510 |
| 8L2 | Abdominal, Pediatric, Musculo-<br>skeletal, Peripheral Vascular | B, M,<br>PWD,<br>Color Doppler,<br>Combined | ISTPA.3 = 598(mW/cm²)<br>TI = 2.8<br>MI = 1.7 | K193510 |
| 8V3A | Fetal, Abdominal, Pediatric, Cephalic,<br>and Cardiac | B, M,<br>PWD,<br>Color Doppler,<br>CWD,<br>Combined | ISTPA.3 = 560(mW/cm²)<br>TI = 4.7<br>MI = 1.7 | K193510 |
| 9MC3 | fetal, pediatric, small organs, cephalic<br>(neonatal and adult), cardiac and<br>peripheral vessels | B, M,<br>PWD,<br>Color Doppler,<br>CWD,<br>Combined | ISTPA.3 = 577(mW/cm²)<br>TI = 2.8<br>MI = 1.3 | K193510 |
| | | Combined | | |
| 8TE3 | Trans-esophageal (non-cardiac and<br>cardiac) | B, M,<br>PWD,<br>Color Doppler,<br>CWD,<br>Combined | $ISTPA.3 = 245(mW/cm2)$<br>$TI = 1.0$<br>$MI = 1.3$ | K193510 |
| PDOF | Cardiac | CWD | $ISTPA.3 = 506(mW/cm2)$<br>$TI = 4.2$<br>$MI = 0.1$ | K193510 |
| 15L4A | Ophthalmic, Abdominal, Pediatric,<br>Small Organ (Thyroid, Breast,<br>Testes); Musculo-skeletal<br>(Conventional and Superficial);<br>Peripheral Vascular | B, M,<br>PWD,<br>Color Doppler,<br>Combined | $ISTPA.3 = 620 (mW/cm2)$<br>$TI = 1.8$<br>$MI = 1.7$ | K193510 |
| 16L5 | Ophthalmic, Abdominal, Pediatric,<br>Small Organ (Thyroid, Breast,<br>Testes); Musculo-skeletal<br>(Conventional and Superficial);<br>Peripheral Vascular | B, M,<br>PWD,<br>Color Doppler,<br>Combined | $ISTPA.3 = 635 (mW/cm2)$<br>$TI = 1.6$<br>$MI = 1.7$ | K193510 |
| 15WL4 | Ophthalmic, Abdominal, Pediatric,<br>Small Organ (Thyroid, Breast,<br>Testes); Musculo-skeletal<br>(Conventional and Superficial);<br>Peripheral Vascular | B, M,<br>PWD,<br>Color Doppler,<br>Combined | $ISTPA.3 = 624 (mW/cm2)$<br>$TI = 3.3$<br>$MI = 1.8$ | K193510 |
| XY-Biplane | Fetal, Abdominal, Pediatric, Neonatal<br>and Adult Cephalic, Cardiac (adult<br>and pediatric) | B, M,<br>PWD,<br>Color Doppler,<br>CWD,<br>Combined | $ISTPA.3 = 650 (mW/cm2)$<br>$TI = 5.7$<br>$MI = 1.7$ | K193510 |
| 10EC4 | Fetal, Trans-rectal, Trans-vaginal | B, M,<br>PWD,<br>Color Doppler,<br>Combined | $ISTPA.3 = 616 (mW/cm2)$<br>$TI = 2.56$<br>$MI = 1.56$ | K193510 |
| 14L3 | Ophthalmic, Abdominal, Pediatric,<br>Small Organ (Thyroid, Breast,<br>Testes); Musculo-skeletal<br>(Conventional and Superficial);<br>Peripheral Vascular. | B, M,<br>PWD,<br>Color Doppler,<br>Combined | $ISTPA.3 = 618 (mW/cm2)$<br>$TI = 4.2$<br>$MI = 1.75$ | New |
| 10BP4 | Fetal, Trans-rectal, Trans-vaginal | B, M,<br>PWD,<br>Color Doppler,<br>Combined | $ISTPA.3 = 606 (mW/cm2)$<br>$TI = 0.99$<br>$MI = 1.75$ | New |
| 15XL4 | Abdominal, Pediatric, Small Organ<br>(Thyroid, Breast, Testes); Musculo-<br>skeletal (Conventional and<br>Superficial); Peripheral Vascular | B, M,<br>PWD,<br>Color Doppler,<br>Combined | $ISTPA.3 = 620 (mW/cm2)$<br>$TI = 3.3$<br>$MI = 1.8$ | New |
{10}------------------------------------------------
{11}------------------------------------------------
#### 14L3 Transducer
| Key Features | Subject Device Model | Comparable Predicate Device | Same or Different |
|----------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Terason uSmart3200T Plus 14L3<br>Transducer | Terason 15L4 Transducer | |
| Device<br>Classification | Class II | Class II | Same |
| 510(k) Number | KXXXXX | K193510 | n/a |
| Indications for<br>Use | The transducer is intended to be<br>used with a conventional<br>ultrasound system (Terason<br>uSmart3200T Plus) to image<br>ophthalmic, abdomen, small<br>parts, musculo-skel, peripheral<br>vascular regions. | The transducer is intended to be<br>used with a conventional<br>ultrasound system (Terason<br>uSmart3200T) to image<br>ophthalmic, abdomen, small<br>parts, musculo-skel, peripheral<br>vascular regions. | Same. The proposed transducer<br>and the predicate transducer have<br>the identical claim of imaging<br>similar regions in the human body. |
| Acoustic Array<br>Technology: | Piezoelectric elements | Piezoelectric elements | Same.<br>Regarding Safety: Both arrays<br>allow focused transmission and<br>reception of ultrasound energy to<br>enhance image quality within the<br>region of interest. |
| Acoustic Array<br>Style: | Linear Transducer | Linear Transducer | |
| Acoustic Array<br>Characteristics:<br>Element count...<br>Center<br>frequency<br>Element size<br>(diameter)... | 128<br>5 MHz<br>0.3mm X 4.5mm | 128<br>7.5 MHz<br>0.3mm X 4mm | Different.<br>The new transducer has the same<br>number of elements operating a<br>lower frequency for deeper<br>imaging. |
| Elevation focus... | 20mm | 16mm | |
| Acoustic Array | The transducer performance has<br>been evaluated in an acoustic<br>tank. | The transducer performance has<br>been evaluated in the previous<br>510(k) filings (K193510). | Same. As the predicate device and<br>therefore has same acoustic<br>characteristics. To ensure proper<br>safety guidelines are met, acoustic<br>testing was performed per the<br>IEC60601-2-37 standard. |
| Patient Contact<br>Material | Silicone (proprietary) | Silicone R1001 | Different.<br>The 14L3 has passed cytotoxicity,<br>sensitization, and irritation<br>biocompatibility tests. |
#### Discussion:
There 14L3 has the same number of elements operating at a lower frequency than the predicate device used in this comparison. The transducer has been added to the uSmart3200T Plus Ultrasound system.
Based on the identical indications for use, technological characteristics and performance testing, Teratech Corporation, Inc. believes the Terason 14L3 transducer is substantially equivalent to the predicate 15L4 transducer with respect to safety and effectiveness.
{12}------------------------------------------------
#### 15XL4 Transducer
| Key Features | Subject Device Model<br>Terason 15XL4 Transducer | Comparable Predicate<br>Device<br>Terason 15WL4 Transducer | Same or Different |
|------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Classification | Class II | Class II | Same |
| 510(k) Number | K1XXXXX | K193510 | n/a |
| Indications for Use | The transducer is intended<br>to be used with a<br>conventional ultrasound<br>system (Terason<br>uSmart3200T Plus) to image<br>Abdominal, Pediatric,<br>Musculo-skeletal, Peripheral<br>Vascular. | The transducer is intended to<br>be used with a conventional<br>ultrasound system (Terason<br>uSmart3200T) to image<br>Abdominal, Pediatric,<br>Musculo-skeletal, Peripheral<br>Vascular. | Same. |
| Acoustic Array Technology: | Piezoelectric elements | Piezoelectric elements | Same. |
| Acoustic Array Style: | Linear Array | Linear Array | Same. Regarding Safety: This<br>array allows focused<br>transmission and reception of<br>ultrasound energy to enhance<br>image quality within the<br>region of interest. |
| Acoustic Array<br>Characteristics:<br>Element count... | 384 | 192 | Different.<br>Additional 192 elements<br>operating at the same<br>frequency. |
| Center frequency... | 6.3 MHz | 6.3 MHz | |
| Element size (pitch x<br>elevation)... | 0.286mm X 4mm | 0.286mm X 4mm | |
| Elevation focus... | 20mm | 16mm | |
| Acoustic Array | The transducer imaging<br>performance has been<br>evaluated in an acoustic<br>tank. | The transducer performance<br>has been evaluated in the<br>previous 510(k) filing<br>(K193510). | Same. As the predicate<br>device and therefore has the<br>same acoustic characteristics.<br>To ensure proper safety<br>guidelines are met, acoustic<br>testing was performed per<br>the IEC60601-2-37 standard. |
| Patient Contact Material | SIM R1001 | SIM R1001 | Same. |
| Key Features | Subject Device Model | Comparable Predicate Devices | Same or Different |
| | Terason 10BP4 Transducer | Terason 10EC4 Transducer<br>Terason 15L4 Transducer | |
| Device Classification | Class II | Class II | Same |
| 510(k) Number | K1XXXXX | K193510 | n/a |
| Indications for Use | The transducer is intended<br>to be used with a<br>conventional ultrasound<br>system (Terason<br>uSmart3200T Plus) to image<br>Fetal, Trans-rectal, Trans-<br>vaginal. | The 10EC4 transducer is<br>intended to be used with a<br>conventional ultrasound<br>system (Terason<br>uSmart3200T) to image<br>Fetal, Trans-rectal, Trans-<br>vaginal. | Same.<br>Indications for Use. |
| Acoustic Array Technology: | Piezoelectric elements | Piezoelectric elements | Same.<br>Elements and shape are the<br>same. |
| Acoustic Array Style: | Linear and Micro-convex | Micro-convex (10EC4)<br>Linear (15L4) | |
| Manufacturer: | Vermon | Vermon | Same:<br>Same manufacturer. |
| Acoustic Array<br>Characteristics:<br>Element count...<br>Center frequency...<br>Element size (pitch x<br>elevation)...<br>ROC<br>Elevation focus... | 192 Curved and Linear<br>7.0 MHz<br>0.18mm X 4.2mm Curved<br>0.37mm X 5mm Linear<br>10 mm<br>20mm | 192 (10EC4) and 128 (15L4)<br>7.2 MHz (10EC4) 7.5 MHz<br>(15L4)<br>0.205mm X 5mm (10EC4)<br>0.3mm X 4mm (15L4),<br>8.8 mm (10EC4)<br>35 mm (10EC4), 16mm<br>(15L4) | Different.<br>The new 10BP4 provides a<br>slightly lower frequency<br>response. The 192 elements<br>have similar pitch and<br>elevation apertures compared<br>to the predicates. |
| Acoustic Array | The transducer imaging<br>performance has been<br>evaluated in an acoustic<br>tank. | The transducer performance<br>has been evaluated in the<br>previous 510(k) filing<br>(K193510). | Same.<br>To ensure proper safety<br>guidelines are met, acoustic<br>testing was performed per<br>the IEC60601-2-37 standard. |
| Patient Contact Material | SIM R1002 (Silicone) | 15L4 transducer<br>SIM R1002 | Same.<br>The Silicone lens is identical<br>to the previously cleared<br>(K193510) 15L4 transducer<br>lens material |
The 15XL4 has an additional 192 elements that operate at the same frequency as the predicate device. To ensure proper safety guidelines are met, acoustic testing was performed per the IEC60601-2-37 standard. The 15XL4 transducer used with the uSmart3200T Plus consists of the same patient contact materials as the predicate device.
Based on the test results, technological characteristics and performance testing, Teratech Corporation, Inc. believes the Terason 15XL4 transducer meets safety and effectiveness guidelines.
{13}------------------------------------------------
### 10BP4 Transducer
{14}------------------------------------------------
The 10BP4 transducer is manufactured by Vermon that also makes other Terason transducers. To ensure proper safety guidelines are met, acoustic testing was performed per the IEC60601-2-37 standard. The 10BP4 transducer used with the uSmart3200T Plus consists of different contact materials when compared with the predicate device.
Based on the test results, technological characteristics and performance testing, Teratech Corporation, Inc. believes the Terason 10BP4 transducer meets safety and effectiveness guidelines.
#### Previously filed transducers:
The following transducers are identical to those previously cleared under FDA 510(k) K193510, May 9, 2015 for the Terason uSmart3200T:
- . 4V2
- . 5C2A
- 8EC4A ●
- . 12L5A
- 15L4
- . 16HL7
- 8L2
- . 8TE3
- 8V3A
- 9MC3
- PDOF
- . 15L4A
- રાજાર
- . 15WL4
- XY-Biplane
- . 10EC4.
#### Conclusion:
The intended uses and features are consistent with the traditional clinical practices and FDA guidance of Diagnostic ultrasound systems and transducers. The uSmart3200T Plus and predicate device both conform to applicable electric safety medical device standards with compliance verified through independent evaluation. The uSmart3200T Plus and predicate device both meet FDA requirements for Track 3 devices, indications for use, biocompatibility similarities, and are manufactured using FDA GMPs and ISO-13485 quality systems. Teratech Corporation believes that the uSmart3200T Plus ultrasound system is substantially equivalent with regards to safety and effectiveness to the predicate device.
{15}------------------------------------------------
## 9. Summary of Bench Tests and Non-Clinical Tests
The Terason uSmart3200T Plus system has been tested for compliance to the following standards (with the corresponding report referenced for each standard).
- AAMI/ANSI ES 60601-1:2005/(R)2012 and A1:2012 Medical electrical . equipment – Part 1: General requirements for safety and essential performance
- IEC60601-1-6, Medical Electrical Equipment Part 1-6: General . requirements for safety- Collateral standard: Usability
- IEC 60601-1-2:2007, Medical Electrical Equipment Part 1-2; General . requirements for basic safety and essential performance – Collateral Standard: Electromagnetic compatibility - Requirements and tests
- EMC Test Report for the Terason uSmart3200T Plus .
- . IEC 60601-2-37 / EN60601-2-37 Medical Electrical Equipment Part 2: Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment.
- Biocompatibility Tests, ISO 10993 Part 5, Part 10 and Part 12 ● ം Biocompatibility reports for the new transducers
- AAMI TIR No. 12:210. Designing, Testing and Labeling Reusable . Medical Devices for Reprocessing in Health Care Facilities: A Guide for Device Manufacturers.
The ultrasound system acoustic output was tested in accordance with the following:
- IEC 61157. Ed. 2 2007-2008. Standard Means for the Reporting of the . Acoustic Output of Medical Diagnostic Ultrasonic Equipment
- NEMA UD 2, Acoustic Output Measurement Standard for Diagnostic . Ultrasound Equipment, Revision 3
- NEMA UD 3, Standard for Real-Time Display of Thermal and Mechanical . Acoustic Output Indices on Diagnostic Ultrasound Equipment, Revision 2.
The ultrasound system B-Mode Accuracy, Doppler Accuracy and Doppler Sensitivity for each Terason transducer have been evaluated according to the following:
{16}------------------------------------------------
- American Institute of Ultrasound in Medicine (AIUM) Quality Assurance . Manual for Gray-Scale Ultrasound Scanners
- AIUM Methods for Measuring Performance of Pulse-Echo Ultrasound . Imaging Equipment, Part II: Digital Methods
- AIUM Performance Criteria and Measurements for Doppler Ultrasound . Devices.
The Terason uSmart3200T Plus Software has undergone Quality Assurance testing consistent with IEC 62304, Software Life Cycle Process, and IEC 62366, Application for Usability.
The performance data used to validate the Terason uSmart3200T Plus and new transducers includes the following:
- Acoustic output testing .
- B-Mode accuracy, and Doppler accuracy and sensitivity .
- General requirements for safety testing ●
- Electromagnetic compatibility testing .
- . Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment testing
- Biocompatibility testing of patient contact materials .
- Burn-in testing ●
- Software performance and regression testing. ●
## 10. Summary of Conclusions
The predicate system (uSmart3200T, K193510) and associated predicate transducers are legally marketed. The new system and associated transducers have the same intended use as the predicate system and devices. The Terason uSmart3200T Plus and associated transducers represent an improved implementation of familiar technology and therefore possess new technological characteristics that are validated in this filing.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.