← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K192107

# Clarius Ultrasound Scanner (K192107)

_Clarius Mobile Health Corp. · IYN · Aug 29, 2019 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K192107

## Device Facts

- **Applicant:** Clarius Mobile Health Corp.
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Aug 29, 2019
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Intended Use

The Clarius Ultrasound Scanner is a software-based ultrasound imaging system and accessories, intended for diagnostic imaging. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: ophthalmic, fetal, abdominal, intra-operative (non-neurological), pediatric, small organ, cephalic (adult), trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), urology, gynecology, cardiac (adult, pediatric), peripheral vessel, carotid, and procedural guidance of needles into the body. The system is a transportable ultrasound system intended for use in environments where is provided by trained healthcare professionals.

## Device Story

Clarius Ultrasound Scanner is a portable, software-controlled diagnostic ultrasound system. It acquires and displays real-time ultrasound data via wireless transducers (convex, linear, phased array, intracavity) communicating with off-the-shelf iOS or Android tablets/smartphones via Wi-Fi/Bluetooth. The system is used by trained healthcare professionals (doctors, nurses, technicians) in clinical and EMS environments. It provides diagnostic imaging and fluid flow analysis, including needle enhancement in B-mode. The healthcare provider views images on the mobile device display to assist in clinical decision-making and procedural guidance. The system benefits patients by providing a portable, high-performance, low-cost diagnostic imaging platform.

## Clinical Evidence

No clinical data was required to establish substantial equivalence. Bench testing only.

## Technological Characteristics

Portable ultrasound system with wireless transducers (convex, linear, phased array, intracavity). Connectivity via Wi-Fi/Bluetooth to iOS/Android devices. Power source: Li-ion battery. Materials comply with ISO 10993-5, -10, -11. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-12, IEC 60601-2-37, IEC 62304, UD 2-2004.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Clarius Ultrasound Scanner ([K180799](/device/K180799.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Clarius Mobile Health Corp. % Prithul Bom Responsible Third Party Official Regulatory Technology Services, LLC 1000 Westgate Drive, Suite #510k SAINT PAUL MN 55114

August 29, 2019

Re: K192107

Trade/Device Name: Clarius Ultrasound Scanner Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: August 2, 2019 Received: August 5, 2019

Dear Prithul Bom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known)

K192107

Device Name Clarius Ultrasound Scanner

#### Indications for Use (Describe)

The Clarius Ultrasound Scanner is a software-based ultrasound imaging system and accessories, intended for diagnostic imaging. It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: ophthalmic, fetal, abdominal, intra-operative (non-neurological), pediatric, small organ, cephalic (adult), trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), urology, cardiac (adult, pediatric), peripheral vessel, carotid, and procedural guidance of needles into the body.

The system is a transportable ultrasound system intended for use in environments where is provided by trained healthcare professionals.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

### Clarius Ultrasound Scanner

### System: Clarius Ultrasound Scanner

### Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                                       | Mode of Operation |   |                  |                  |               |                        |        |
|---------------------------|-------------------------------------------------------|-------------------|---|------------------|------------------|---------------|------------------------|--------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)                            | B                 | M | Color<br>Doppler | Power<br>Doppler | PW<br>Doppler | Combined (Specify)     | Other* |
| Ophthalmic                | Ophthalmic                                            | P                 |   |                  |                  |               |                        |        |
| Fetal Imaging &<br>Other  | Fetal                                                 | P                 | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Abdominal                                             | P                 | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Intra-operative<br>(Abdominal organs & vascular)      | P                 | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Laparoscopic                                          |                   |   |                  |                  |               |                        |        |
|                           | Pediatric                                             | P                 | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Small Organ<br>(Thyroid, Prostate, Scrotum, Breast)   | P                 | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Neonatal Cephalic                                     |                   |   |                  |                  |               |                        |        |
|                           | Adult Cephalic                                        | P                 | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Trans-rectal                                          | P                 | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Trans-vaginal                                         | P                 | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Trans-urethral                                        |                   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (non-Cardiac)                        |                   |   |                  |                  |               |                        |        |
|                           | Musculo-skeletal (Conventional)                       | P                 | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Musculo-skeletal (Superficial)                        | P                 | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Intravascular                                         |                   |   |                  |                  |               |                        |        |
|                           | Other (Urology, Gynecology)                           | P                 | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
| Cardiac                   | Cardiac Adult                                         | P                 | P | P                | N                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Cardiac Pediatric                                     | P                 | P | P                | N                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Intravascular (Cardiac)                               |                   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (Cardiac)                            |                   |   |                  |                  |               |                        |        |
|                           | Intra-cardiac                                         |                   |   |                  |                  |               |                        |        |
|                           | Other (specify)                                       |                   |   |                  |                  |               |                        |        |
| Peripheral Vessel         | Peripheral Vessel                                     | P                 | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Other (Carotid)                                       | P                 | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | N = new indication; P = previously cleared by K180799 |                   |   |                  |                  |               |                        |        |
|                           | Note 1: Needle Enhancement in B-Mode.                 |                   |   |                  |                  |               |                        |        |

{4}------------------------------------------------

#### Clarius Scanner C3 HD

### DEVICE NAME: CLARIUS SCANNER C3 HD

### Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                                     | Mode of Operation |   |                  |                  |               |                        |        |
|---------------------------|-----------------------------------------------------|-------------------|---|------------------|------------------|---------------|------------------------|--------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)                          | B                 | M | Color<br>Doppler | Power<br>Doppler | PW<br>Doppler | Combined (Specify)     | Other* |
| Ophthalmic                | Ophthalmic                                          |                   |   |                  |                  |               |                        |        |
| Fetal Imaging &<br>Other  | Fetal                                               | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Abdominal                                           | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Intra-operative<br>(Abdominal organs & vascular)    | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Laparoscopic                                        |                   |   |                  |                  |               |                        |        |
|                           | Pediatric                                           | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Small Organ<br>(Thyroid, Prostate, Scrotum, Breast) |                   |   |                  |                  |               |                        |        |
|                           | Neonatal Cephalic                                   |                   |   |                  |                  |               |                        |        |
|                           | Adult Cephalic                                      | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Trans-rectal                                        |                   |   |                  |                  |               |                        |        |
|                           | Trans-vaginal                                       |                   |   |                  |                  |               |                        |        |
|                           | Trans-urethral                                      |                   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (non-Cardiac)                      |                   |   |                  |                  |               |                        |        |
|                           | Musculo-skeletal (Conventional)                     | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Musculo-skeletal (Superficial)                      |                   |   |                  |                  |               |                        |        |
|                           | Intravascular                                       |                   |   |                  |                  |               |                        |        |
|                           | Other (Urology, Gynecology)                         | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
| Cardiac                   | Cardiac Adult                                       | N                 | N | N                |                  | N             | B+M; B+CD; B+PWD       |        |
|                           | Cardiac Pediatric                                   | N                 | N | N                |                  | N             | B+M; B+CD; B+PWD       |        |
|                           | Intravascular (Cardiac)                             |                   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (Cardiac)                          |                   |   |                  |                  |               |                        |        |
|                           | Intra-cardiac                                       |                   |   |                  |                  |               |                        |        |
|                           | Other (specify)                                     |                   |   |                  |                  |               |                        |        |
| Peripheral Vessel         | Peripheral Vessel                                   | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Other (Carotid)                                     |                   |   |                  |                  |               |                        |        |

{5}------------------------------------------------

#### Clarius Scanner C7 HD

### DEVICE NAME: CLARIUS SCANNER C7 HD

### Intended Use: Diagnostic ultrasound Imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                                       | Mode of Operation |   |                  |                  |               |                        |        |
|---------------------------|-------------------------------------------------------|-------------------|---|------------------|------------------|---------------|------------------------|--------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)                            | B                 | M | Color<br>Doppler | Power<br>Doppler | PW<br>Doppler | Combined (Specify)     | Other* |
| Ophthalmic                | Ophthalmic                                            |                   |   |                  |                  |               |                        |        |
| Fetal Imaging &<br>Other  | Fetal                                                 | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Abdominal                                             | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Intra-operative<br>(Abdominal organs & vascular)      | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Laparoscopic                                          |                   |   |                  |                  |               |                        |        |
|                           | Pediatric                                             | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Small Organ<br>(Thyroid, Prostate, Scrotum, Breast)   | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Neonatal Cephalic                                     |                   |   |                  |                  |               |                        |        |
|                           | Adult Cephalic                                        |                   |   |                  |                  |               |                        |        |
|                           | Trans-rectal                                          |                   |   |                  |                  |               |                        |        |
|                           | Trans-vaginal                                         |                   |   |                  |                  |               |                        |        |
|                           | Trans-urethral                                        |                   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (non-Cardiac)                        |                   |   |                  |                  |               |                        |        |
|                           | Musculo-skeletal (Conventional)                       | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Musculo-skeletal (Superficial)                        |                   |   |                  |                  |               |                        |        |
|                           | Intravascular                                         |                   |   |                  |                  |               |                        |        |
|                           | Other (Urology, Gynecology)                           | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
| Cardiac                   | Cardiac Adult                                         | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Cardiac Pediatric                                     | N                 | N | N                |                  | N             | B+M; B+CD; B+PWD       |        |
|                           | Intravascular (Cardiac)                               |                   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (Cardiac)                            |                   |   |                  |                  |               |                        |        |
|                           | Intra-cardiac                                         |                   |   |                  |                  |               |                        |        |
|                           | Other (specify)                                       |                   |   |                  |                  |               |                        |        |
| Peripheral Vessel         | Peripheral Vessel                                     | N                 | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Other (Carotid)                                       |                   |   |                  |                  |               |                        |        |
|                           | N = new indication; P = previously cleared by K180799 |                   |   |                  |                  |               |                        |        |
|                           | Note 1: Needle Enhancement in B-Mode.                 |                   |   |                  |                  |               |                        |        |

{6}------------------------------------------------

#### Clarius Scanner EC7 HD

### Device Name: Clarius Scanner EC7 HD

|                           | Clinical Application<br>Mode of Operation             |   |   |                  |                  |               |                        |        |
|---------------------------|-------------------------------------------------------|---|---|------------------|------------------|---------------|------------------------|--------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)                            | B | M | Color<br>Doppler | Power<br>Doppler | PW<br>Doppler | Combined (Specify)     | Other* |
| Ophthalmic                | Ophthalmic                                            |   |   |                  |                  |               |                        |        |
| Fetal Imaging &<br>Other  | Fetal                                                 | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Abdominal                                             | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Intra-operative<br>(Abdominal organs & vascular)      |   |   |                  |                  |               |                        |        |
|                           | Laparoscopic                                          |   |   |                  |                  |               |                        |        |
|                           | Pediatric                                             |   |   |                  |                  |               |                        |        |
|                           | Small Organ<br>(Thyroid, Prostate, Scrotum, Breast)   | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Neonatal Cephalic                                     |   |   |                  |                  |               |                        |        |
|                           | Adult Cephalic                                        |   |   |                  |                  |               |                        |        |
|                           | Trans-rectal                                          | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Trans-vaginal                                         | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Trans-urethral                                        |   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (non-Cardiac)                        |   |   |                  |                  |               |                        |        |
|                           | Musculo-skeletal (Conventional)                       |   |   |                  |                  |               |                        |        |
|                           | Musculo-skeletal (Superficial)                        |   |   |                  |                  |               |                        |        |
|                           | Intravascular                                         |   |   |                  |                  |               |                        |        |
|                           | Other (Urology, Gynecology)                           | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD |        |
| Cardiac                   | Cardiac Adult                                         |   |   |                  |                  |               |                        |        |
|                           | Cardiac Pediatric                                     |   |   |                  |                  |               |                        |        |
|                           | Intravascular (Cardiac)                               |   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (Cardiac)                            |   |   |                  |                  |               |                        |        |
|                           | Intra-cardiac                                         |   |   |                  |                  |               |                        |        |
|                           | Other (specify)                                       |   |   |                  |                  |               |                        |        |
| Peripheral Vessel         | Peripheral Vessel                                     |   |   |                  |                  |               |                        |        |
|                           | Other (Carotid)                                       |   |   |                  |                  |               |                        |        |
|                           | N = new indication; P = previously cleared by K180799 |   |   |                  |                  |               |                        |        |
|                           | Note 1: Needle Enhancement in B-Mode.                 |   |   |                  |                  |               |                        |        |

{7}------------------------------------------------

#### Clarius Scanner L7 HD

| Clinical Application      | Mode of Operation                                   |   |   |                  |                  |               |                        |        |
|---------------------------|-----------------------------------------------------|---|---|------------------|------------------|---------------|------------------------|--------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)                          | B | M | Color<br>Doppler | Power<br>Doppler | PW<br>Doppler | Combined (Specify)     | Other* |
| Ophthalmic                | Ophthalmic                                          |   | N |                  |                  |               |                        |        |
| Fetal Imaging &<br>Other  | Fetal                                               |   |   |                  |                  |               |                        |        |
|                           | Abdominal                                           | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Intra-operative<br>(Abdominal organs & vascular)    | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Laparoscopic                                        |   |   |                  |                  |               |                        |        |
|                           | Pediatric                                           | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Small Organ<br>(Thyroid, Prostate, Scrotum, Breast) | N |   | N                | N                | N             | B+CD; B+PD; B+PWD      | Note 1 |
|                           | Neonatal Cephalic                                   |   |   |                  |                  |               |                        |        |
|                           | Adult Cephalic                                      |   |   |                  |                  |               |                        |        |
|                           | Trans-rectal                                        |   |   |                  |                  |               |                        |        |
|                           | Trans-vaginal                                       |   |   |                  |                  |               |                        |        |
|                           | Trans-urethral                                      |   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (non-Cardiac)                      |   |   |                  |                  |               |                        |        |
|                           | Musculo-skeletal (Conventional)                     | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Musculo-skeletal (Superficial)                      | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Intravascular                                       |   |   |                  |                  |               |                        |        |
|                           | Other (Urology, Gynecology)                         |   |   |                  |                  |               |                        |        |
| Cardiac                   | Cardiac Adult                                       |   |   |                  |                  |               |                        |        |
|                           | Cardiac Pediatric                                   |   |   |                  |                  |               |                        |        |
|                           | Intravascular (Cardiac)                             |   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (Cardiac)                          |   |   |                  |                  |               |                        |        |
|                           | Intra-cardiac                                       |   |   |                  |                  |               |                        |        |
|                           | Other (specify)                                     |   |   |                  |                  |               |                        |        |
| Peripheral Vessel         | Peripheral Vessel                                   | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |

N | N | N

### DEVICE NAME: CLARIUS SCANNER L7 HD INTENDED USE: DIAGNOSTIC ULTRASOUND IMAGING OR FLUID FLOW ANALYSIS OF THE HUMAN BODY AS FOLLOWS:

Other (Carotid) N = new indication; P = previously cleared by K180799 Note 1: Needle Enhancement in B-Mode.

N

B+M; B+CD; B+PD; B+PWD

Note 1

N

{8}------------------------------------------------

#### Clarius Scanner L15 HD

### Device Name: Clarius Scanner L15 HD

| Clinical Application      | Mode of Operation                                   |   |   |                  |                  |               |                        |        |
|---------------------------|-----------------------------------------------------|---|---|------------------|------------------|---------------|------------------------|--------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)                          | B | M | Color<br>Doppler | Power<br>Doppler | PW<br>Doppler | Combined (Specify)     | Other* |
| Ophthalmic                | Ophthalmic                                          |   |   |                  |                  |               |                        |        |
| Fetal Imaging &<br>Other  | Fetal                                               |   |   |                  |                  |               |                        |        |
|                           | Abdominal                                           | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Intra-operative<br>(Abdominal organs & vascular)    | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Laparoscopic                                        |   |   |                  |                  |               |                        |        |
|                           | Pediatric                                           | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Small Organ<br>(Thyroid, Prostate, Scrotum, Breast) | N | N | N                | N                | N             | B+CD; B+PD; B+PWD      | Note 1 |
|                           | Neonatal Cephalic                                   |   |   |                  |                  |               |                        |        |
|                           | Adult Cephalic                                      |   |   |                  |                  |               |                        |        |
|                           | Trans-rectal                                        |   |   |                  |                  |               |                        |        |
|                           | Trans-vaginal                                       |   |   |                  |                  |               |                        |        |
|                           | Trans-urethral                                      |   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (non-Cardiac)                      |   |   |                  |                  |               |                        |        |
|                           | Musculo-skeletal (Conventional)                     | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Musculo-skeletal (Superficial)                      | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Intravascular                                       |   |   |                  |                  |               |                        |        |
|                           | Other (Urology, Gynecology)                         |   |   |                  |                  |               |                        |        |
| Cardiac                   | Cardiac Adult                                       |   |   |                  |                  |               |                        |        |
|                           | Cardiac Pediatric                                   |   |   |                  |                  |               |                        |        |
|                           | Intravascular (Cardiac)                             |   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (Cardiac)                          |   |   |                  |                  |               |                        |        |
|                           | Intra-cardiac                                       |   |   |                  |                  |               |                        |        |
|                           | Other (specify)                                     |   |   |                  |                  |               |                        |        |
| Peripheral Vessel         | Peripheral Vessel                                   | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Other (Carotid)                                     | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |

{9}------------------------------------------------

#### Clarius Scanner PA HD

### DEVICE NAME: CLARIUS SCANNER PA HD

| Clinical Application      | Mode of Operation                                     |   |   |                  |                  |               |                        |        |
|---------------------------|-------------------------------------------------------|---|---|------------------|------------------|---------------|------------------------|--------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)                            | B | M | Color<br>Doppler | Power<br>Doppler | PW<br>Doppler | Combined (Specify)     | Other* |
| Ophthalmic                | Ophthalmic                                            |   |   |                  |                  |               |                        |        |
| Fetal Imaging &<br>Other  | Fetal                                                 | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Abdominal                                             | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Intra-operative<br>(Abdominal organs & vascular)      | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Laparoscopic                                          |   |   |                  |                  |               |                        |        |
|                           | Pediatric                                             | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Small Organ<br>(Thyroid, Prostate, Scrotum, Breast)   |   |   |                  |                  |               |                        |        |
|                           | Neonatal Cephalic                                     |   |   |                  |                  |               |                        |        |
|                           | Adult Cephalic                                        | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Trans-rectal                                          |   |   |                  |                  |               |                        |        |
|                           | Trans-vaginal                                         |   |   |                  |                  |               |                        |        |
|                           | Trans-urethral                                        |   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (non-Cardiac)                        |   |   |                  |                  |               |                        |        |
|                           | Musculo-skeletal (Conventional)                       |   |   |                  |                  |               |                        |        |
|                           | Musculo-skeletal (Superficial)                        |   |   |                  |                  |               |                        |        |
|                           | Intravascular                                         |   |   |                  |                  |               |                        |        |
|                           | Other (Urology, Gynecology)                           |   |   |                  |                  |               |                        |        |
| Cardiac                   | Cardiac Adult                                         | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Cardiac Pediatric                                     | N | N | N                | N                | N             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Intravascular (Cardiac)                               |   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (Cardiac)                            |   |   |                  |                  |               |                        |        |
|                           | Intra-cardiac                                         |   |   |                  |                  |               |                        |        |
|                           | Other (specify)                                       |   |   |                  |                  |               |                        |        |
| Peripheral Vessel         | Peripheral Vessel                                     |   |   |                  |                  |               |                        |        |
|                           | Other (Carotid)                                       |   |   |                  |                  |               |                        |        |
|                           | N = new indication; P = previously cleared by K180799 |   |   |                  |                  |               |                        |        |
|                           | Note 1: Needle Enhancement in B-Mode.                 |   |   |                  |                  |               |                        |        |

{10}------------------------------------------------

#### Clarius Scanner C3

### Device Name: Clarius Scanner C3

### Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      | Mode of Operation                                                                              |   |   |                  |                  |               |                        |        |  |
|---------------------------|------------------------------------------------------------------------------------------------|---|---|------------------|------------------|---------------|------------------------|--------|--|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)                                                                     | B | M | Color<br>Doppler | Power<br>Doppler | PW<br>Doppler | Combined (Specify)     | Other* |  |
| Ophthalmic                | Ophthalmic                                                                                     |   |   |                  |                  |               |                        |        |  |
| Fetal Imaging &<br>Other  | Fetal                                                                                          | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |  |
|                           | Abdominal                                                                                      | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |  |
| Other                     | Intra-operative<br>(Abdominal organs & vascular)                                               | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |  |
|                           | Laparoscopic                                                                                   |   |   |                  |                  |               |                        |        |  |
|                           | Pediatric                                                                                      | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |  |
|                           | Small Organ<br>(Thyroid, Prostate, Scrotum, Breast)                                            |   |   |                  |                  |               |                        |        |  |
|                           | Neonatal Cephalic                                                                              |   |   |                  |                  |               |                        |        |  |
|                           | Adult Cephalic                                                                                 | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |  |
|                           | Trans-rectal                                                                                   |   |   |                  |                  |               |                        |        |  |
|                           | Trans-vaginal                                                                                  |   |   |                  |                  |               |                        |        |  |
|                           | Trans-urethral                                                                                 |   |   |                  |                  |               |                        |        |  |
|                           | Trans-esophageal (non-Cardiac)                                                                 |   |   |                  |                  |               |                        |        |  |
|                           | Musculo-skeletal (Conventional)                                                                | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |  |
|                           | Musculo-skeletal (Superficial)                                                                 |   |   |                  |                  |               |                        |        |  |
|                           | Intravascular                                                                                  |   |   |                  |                  |               |                        |        |  |
|                           | Other (Urology, Gynecology)                                                                    | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |  |
| Cardiac                   | Cardiac Adult                                                                                  | P | P | P                |                  | P             | B+M; B+CD; B+PWD       |        |  |
|                           | Cardiac Pediatric                                                                              | P | P | P                |                  | P             | B+M; B+CD; B+PWD       |        |  |
|                           | Intravascular (Cardiac)                                                                        |   |   |                  |                  |               |                        |        |  |
|                           | Trans-esophageal (Cardiac)                                                                     |   |   |                  |                  |               |                        |        |  |
|                           | Intra-cardiac                                                                                  |   |   |                  |                  |               |                        |        |  |
|                           | Other (specify)                                                                                |   |   |                  |                  |               |                        |        |  |
| Peripheral Vessel         | Peripheral Vessel                                                                              | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |  |
|                           | Other (Carotid)                                                                                |   |   |                  |                  |               |                        |        |  |
|                           | N = new indication; P = previously cleared by K180799<br>Note 1: Needle Enhancement in B-Mode. |   |   |                  |                  |               |                        |        |  |

{11}------------------------------------------------

#### Clarius Scanner C7

### DEVICE NAME: CLARIUS SCANNER C7

| Clinical Application      | Mode of Operation                                     |   |   |                  |                  |               |                        |        |
|---------------------------|-------------------------------------------------------|---|---|------------------|------------------|---------------|------------------------|--------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)                            | B | M | Color<br>Doppler | Power<br>Doppler | PW<br>Doppler | Combined (Specify)     | Other* |
| Ophthalmic                | Ophthalmic                                            |   |   |                  |                  |               |                        |        |
| Fetal Imaging &<br>Other  | Fetal                                                 | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Abdominal                                             | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Intra-operative<br>(Abdominal organs & vascular)      | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Laparoscopic                                          |   |   |                  |                  |               |                        |        |
|                           | Pediatric                                             | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Small Organ<br>(Thyroid, Prostate, Scrotum, Breast)   | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Neonatal Cephalic                                     |   |   |                  |                  |               |                        |        |
|                           | Adult Cephalic                                        |   |   |                  |                  |               |                        |        |
|                           | Trans-rectal                                          |   |   |                  |                  |               |                        |        |
|                           | Trans-vaginal                                         |   |   |                  |                  |               |                        |        |
|                           | Trans-urethral                                        |   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (non-Cardiac)                        |   |   |                  |                  |               |                        |        |
|                           | Musculo-skeletal (Conventional)                       | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Musculo-skeletal (Superficial)                        |   |   |                  |                  |               |                        |        |
|                           | Intravascular                                         |   |   |                  |                  |               |                        |        |
|                           | Other (Urology, Gynecology)                           | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |
| Cardiac                   | Cardiac Adult                                         | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Cardiac Pediatric                                     | P | P | P                |                  | P             | B+M; B+CD; B+PWD       |        |
|                           | Intravascular (Cardiac)                               |   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (Cardiac)                            |   |   |                  |                  |               |                        |        |
|                           | Intra-cardiac                                         |   |   |                  |                  |               |                        |        |
|                           | Other (specify)                                       |   |   |                  |                  |               |                        |        |
| Peripheral Vessel         | Peripheral Vessel                                     | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD |        |
|                           | Other (Carotid)                                       |   |   |                  |                  |               |                        |        |
|                           | N = new indication; P = previously cleared by K180799 |   |   |                  |                  |               |                        |        |
|                           | Note 1: Needle Enhancement in B-Mode.                 |   |   |                  |                  |               |                        |        |

{12}------------------------------------------------

### Clarius Scanner EC7

### Device Name: Clarius Scanner EC7

| Clinical Application                                                                           |                                                     |   |   | Mode of Operation |                  |               |                        |        |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------|---|---|-------------------|------------------|---------------|------------------------|--------|
| General<br>(Track 1 Only)                                                                      | Specific<br>(Tracks 1 & 3)                          | B | M | Color<br>Doppler  | Power<br>Doppler | PW<br>Doppler | Combined (Specify)     | Other* |
| Ophthalmic                                                                                     | Ophthalmic                                          |   |   |                   |                  |               |                        |        |
| Fetal Imaging &<br>Other                                                                       | Fetal                                               | P | P | P                 | P                | P             | B+M; B+CD; B+PD; B+PWD |        |
|                                                                                                | Abdominal                                           | P | P | P                 | P                | P             | B+M; B+CD; B+PD; B+PWD |        |
|                                                                                                | Intra-operative<br>(Abdominal organs & vascular)    |   |   |                   |                  |               |                        |        |
|                                                                                                | Laparoscopic                                        |   |   |                   |                  |               |                        |        |
|                                                                                                | Pediatric                                           |   |   |                   |                  |               |                        |        |
|                                                                                                | Small Organ<br>(Thyroid, Prostate, Scrotum, Breast) | P | P | P                 | P                | P             | B+M; B+CD; B+PD; B+PWD |        |
|                                                                                                | Neonatal Cephalic                                   |   |   |                   |                  |               |                        |        |
|                                                                                                | Adult Cephalic                                      |   |   |                   |                  |               |                        |        |
|                                                                                                | Trans-rectal                                        | P | P | P                 | P                | P             | B+M; B+CD; B+PD; B+PWD |        |
|                                                                                                | Trans-vaginal                                       | P | P | P                 | P                | P             | B+M; B+CD; B+PD; B+PWD |        |
|                                                                                                | Trans-urethral                                      |   |   |                   |                  |               |                        |        |
|                                                                                                | Trans-esophageal (non-Cardiac)                      |   |   |                   |                  |               |                        |        |
|                                                                                                | Musculo-skeletal (Conventional)                     |   |   |                   |                  |               |                        |        |
|                                                                                                | Musculo-skeletal (Superficial)                      |   |   |                   |                  |               |                        |        |
|                                                                                                | Intravascular                                       |   |   |                   |                  |               |                        |        |
|                                                                                                | Other (Urology, Gynecology)                         | P | P | P                 | P                | P             | B+M; B+CD; B+PD; B+PWD |        |
| Cardiac                                                                                        | Cardiac Adult                                       |   |   |                   |                  |               |                        |        |
|                                                                                                | Cardiac Pediatric                                   |   |   |                   |                  |               |                        |        |
|                                                                                                | Intravascular (Cardiac)                             |   |   |                   |                  |               |                        |        |
|                                                                                                | Trans-esophageal (Cardiac)                          |   |   |                   |                  |               |                        |        |
|                                                                                                | Intra-cardiac                                       |   |   |                   |                  |               |                        |        |
|                                                                                                | Other (specify)                                     |   |   |                   |                  |               |                        |        |
| Peripheral Vessel                                                                              | Peripheral Vessel                                   |   |   |                   |                  |               |                        |        |
|                                                                                                | Other (Carotid)                                     |   |   |                   |                  |               |                        |        |
| N = new indication; P = previously cleared by K180799<br>Note 1: Needle Enhancement in B-Mode. |                                                     |   |   |                   |                  |               |                        |        |

{13}------------------------------------------------

#### Clarius Scanner L7

### Device Name: Clarius Scanner L7

| Clinical Application      | Mode of Operation                                                                              |   |   |                  |                  |               |                        |        |
|---------------------------|------------------------------------------------------------------------------------------------|---|---|------------------|------------------|---------------|------------------------|--------|
| General<br>(Track 1 Only) | Specific<br>(Tracks 1 & 3)                                                                     | B | M | Color<br>Doppler | Power<br>Doppler | PW<br>Doppler | Combined (Specify)     | Other* |
| Ophthalmic                | Ophthalmic                                                                                     | P |   |                  |                  |               |                        |        |
| Fetal Imaging &<br>Other  | Fetal                                                                                          | P |   |                  |                  |               |                        |        |
|                           | Abdominal                                                                                      | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Intra-operative<br>(Abdominal organs & vascular)                                               | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Laparoscopic                                                                                   |   |   |                  |                  |               |                        |        |
|                           | Pediatric                                                                                      | P | P | P                | P                | P             | B+M; B+CD; B+PD        | Note 1 |
|                           | Small Organ<br>(Thyroid, Prostate, Scrotum, Breast)                                            | P | P | P                | P                | P             | B+CD; B+PD; B+PWD      | Note 1 |
|                           | Neonatal Cephalic                                                                              |   |   |                  |                  |               |                        |        |
|                           | Adult Cephalic                                                                                 |   |   |                  |                  |               |                        |        |
|                           | Trans-rectal                                                                                   |   |   |                  |                  |               |                        |        |
|                           | Trans-vaginal                                                                                  |   |   |                  |                  |               |                        |        |
|                           | Trans-urethral                                                                                 |   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (non-Cardiac)                                                                 |   |   |                  |                  |               |                        |        |
|                           | Musculo-skeletal (Conventional)                                                                | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Musculo-skeletal (Superficial)                                                                 | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Intravascular                                                                                  |   |   |                  |                  |               |                        |        |
|                           | Other (Urology, Gynecology)                                                                    |   |   |                  |                  |               |                        |        |
| Cardiac                   | Cardiac Adult                                                                                  |   |   |                  |                  |               |                        |        |
|                           | Cardiac Pediatric                                                                              |   |   |                  |                  |               |                        |        |
|                           | Intravascular (Cardiac)                                                                        |   |   |                  |                  |               |                        |        |
|                           | Trans-esophageal (Cardiac)                                                                     |   |   |                  |                  |               |                        |        |
|                           | Intra-cardiac                                                                                  |   |   |                  |                  |               |                        |        |
|                           | Other (specify)                                                                                |   |   |                  |                  |               |                        |        |
| Peripheral Vessel         | Peripheral Vessel                                                                              | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | Other (Carotid)                                                                                | P | P | P                | P                | P             | B+M; B+CD; B+PD; B+PWD | Note 1 |
|                           | N = new indication; P = previously cleared by K180799<br>Note 1: Needle Enhancement in B-Mode. |   |   |                  |                  |               |                        |        |

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the Clarius logo. The logo consists of an orange symbol that looks like three letter C's nested inside each other. To the right of the symbol is the word "clarius" in dark teal, followed by the words "ultrasound anywhere" stacked on top of each other in the same color.

K192107

# 510(k) Summary

1. Submitter's name, address, telephone number, and contact person:

| Submitter:     | Brendan Seward                                         |
|----------------|--------------------------------------------------------|
|                | QA/RA Manager                                          |
|                | Clarius Mobile Health Corp.                            |
|                | 350 – 3605 Gilmore Way, Burnaby, B.C., V5G 4X5, Canada |
|                | Email: brendan.seward@clarius.com                      |
|                | Tel: (+1) 778-800-9975                                 |
|                | Fax: N/A                                               |
| Date Prepared: | May 13, 2019                                           |

- 2. Name of the device, including the trade or proprietary name, if applicable, the common or usual name, and the classification, if known:

| <b>Device Name:</b>          | Clarius Ultrasound Scanner                   |
|------------------------------|----------------------------------------------|
| <b>Common Name:</b>          | Diagnostic Ultrasound System and Accessories |
| <b>Classification:</b>       | Class II                                     |
| <b>Classification Names:</b> |                                              |

| 21 CFR Section | Classification Name                      | Product Code |
|----------------|------------------------------------------|--------------|
| 892.1550       | Ultrasonic Pulsed Doppler Imaging System | 90 IYN       |
| 892.1560       | Ultrasonic Pulsed Echo Imaging System    | 90 IYO       |
| 892.1570       | Diagnostic Ultrasound Transducer         | 90 ITX       |

### 3. Substantially Equivalent Device:

Device Name: Clarius Ultrasound Scanner 510(k) Number: K180799

### Device Description

The Clarius Ultrasound Scanner is a portable, general-purpose, software-controlled, diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through an off-the-shelf (OTS) iOS or Android device. The Clarius Ultrasound Scanner comprises a series of wireless transducers employing Bluetooth and Wi-Fi-based technology to communicate with traditional tablet/smartphone devices via direct Wi-Fi. This allows the user to export ultrasound images and display them across a range of portable personal devices.

The system is a transportable ultrasound system intended for use in environments where is provided by trained healthcare professionals, including the emergency medical services (EMS) environment.

Users will be trained medical professionals (e.g., doctors, nurses, technicians) with previous training in ultrasound.

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the Clarius logo. The logo consists of an orange symbol that looks like three curved lines forming a C shape. Next to the symbol is the word "clarius" in dark teal, followed by the words "ultrasound anywhere" stacked on top of each other, also in dark teal.

The Clarius Ultrasound Scanner product/package components include:

### Non-HD Packages

- 1. Software:
	- The Clarius Ultrasound App (Clarius App) for iOS; and
	- . The Clarius Ultrasound App (Clarius App) for Android.
- 2. Transducers/Scanners:
	- Clarius Scanner C3;
	- Clarius Scanner EC7; ●
	- . Clarius Scanner C7; and
	- Clarius Scanner L7;
- 3. Accessories:
	- Clarius-Built: ●
		- o Battery Pack (Li-ion);
		- Battery Charger; and o
		- O Clarius Fan.
	- OEM/Off-the-Shelf Product(s): ●
		- Medical Power Supply (Off-the-shelf power adaptor from SL Power Electronics; Model o Number ME20A1203B02; approved in the US).

### HD Packages

- 1. Software:
	- The Clarius Ultrasound App (Clarius App) for iOS; and
	- . The Clarius Ultrasound App (Clarius App) for Android.
- 2. Transducers/Scanners:
	- Clarius Scanner C3 HD
	- . Clarius Scanner L7 HD;
	- Clarius Scanner C7 HD;
	- Clarius Scanner EC7 HD; ●
	- Clarius Scanner L15 HD; and
	- Clarius Scanner PA HD. ●
- Accessories: 4.
	- . Clarius-Built:
		- Clarius Battery HD; o
		- Clarius Charger HD; and O
		- Clarius Fan HD. O
	- OEM/Off-the-Shelf Product(s): ●
		- Medical Power Supply (Off-the-shelf power adaptor from GlobTek, Inc.; Model Number o WR9QA3200USBNMEDR6B; approved in the US).

The concept of the Clarius Ultrasound Scanner transducers and software is primarily to provide an easy to use, high-performance, low-cost, ultrasound platform for teaching and clinical applications. The Clarius Ultrasound Scanner is intended for use in professional healthcare environments where is provided by trained medical professionals.

{16}------------------------------------------------

Image /page/16/Picture/0 description: The image shows the Clarius logo. On the left is an orange icon that looks like a stylized letter C made of three curved lines. To the right of the icon is the word "clarius" in dark teal, followed by the words "ultrasound anywhere" stacked on top of each other, also in dark teal. The logo is simple and modern, with a focus on the company name and its association with ultrasound technology.

# Changes Implemented

The subject change of this 510(k) is the addition of a new set of transducer models to the Clarius Ultrasound Scanner family. The new set of scanners comprises the Clarius Scanner L7 HD (L7 HD), Clarius Scanner C3 HD (C3 HD), Clarius Scanner C7 HD (C7 HD), Clarius Scanner EC7 HD (EC7 HD), Clarius Scanner L15 HD (L15 HD), and Clarius Scanner PA HD (PA HD). An overview of the changes implemented are provided below:

| Changed Specification    | Clarius Ultrasound Scanner (Subject 510(k))                                                                                                                                                                 | Clarius Ultrasound Scanner (K180799)                                                                                                                                    |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Models                   | C3, C7, L7, EC7, C3 HD, L7 HD, C7 HD, EC7 HD, L15 HD, PA HD                                                                                                                                                 | C3, C7, L7,…

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K192107](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K192107)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
