← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K192030

# RS80 EVO Diagnostic Ultrasound System (K192030)

_Samsung Medison Co., Ltd. · IYN · Sep 19, 2019 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K192030

## Device Facts

- **Applicant:** Samsung Medison Co., Ltd.
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Sep 19, 2019
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** AI/ML, Pediatric

## Intended Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative, Pediatric, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (Cardiac) and Peripheral vessel.

## Device Story

RS80 EVO is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound data via various transducers (linear, curved, phased, endocavity, static) and displays it in modes including 2D, M-mode, Color/Power Doppler, PW/CW Spectral Doppler, Tissue Harmonic Imaging, Tissue Doppler, 3D/4D, and Elastoscan. Used in clinical settings by healthcare professionals to visualize anatomical structures and analyze body fluids. The system provides real-time acoustic output display (mechanical/thermal indices). It includes analysis packages to aid diagnosis. The device is used for guidance of procedures like biopsies and monitoring of follicle development. It benefits patients by providing non-invasive diagnostic imaging and fluid flow analysis across a wide range of clinical applications.

## Clinical Evidence

No clinical studies were required to support substantial equivalence. Evidence is based on non-clinical bench testing, including acoustic output, biocompatibility, cleaning/disinfection validation, and thermal/electromagnetic/mechanical safety testing in accordance with IEC 60601-1, IEC 60601-2-37, and ISO 10993-1 standards.

## Technological Characteristics

Mobile cart-based ultrasound system; 192 transmit/receive channels; 23/23.8-inch LCD monitor; 1.0-20.0 MHz frequency range. Modes: B, M, PW/CW Doppler, Color/Power Doppler, 3D/4D, Elastoscan, Tissue Harmonic Imaging. Biocompatibility per ISO 10993-1. Electrical safety per ANSI AAMI ES60601-1. EMC per IEC 60601-1-2. Acoustic output per NEMA UD-2/UD-3 and IEC 60601-2-37.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- RS85 Diagnostic Ultrasound System ([K191115](/device/K191115.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

September 19, 2019

Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

Samsung Medison Co., Ltd. % Ji Yea Lee Regulatory Affairs Specialist 3366, Hanseo-ro, Nam-myeon, Hongcheon-gun, Gangwon-do 25108 REPUBLIC OF KOREA

# Re: K192030

Trade/Device Name: RS80 EVO Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, ITX Dated: August 14, 2019 Received: August 15, 2019

Dear Ji Yea Lee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known) K192030

Device Name RS80 EVO Diagnostic Ultrasound System

#### Indications for Use (Describe)

The ultrasound diagnostic system and probes are designed to obtain ultrasound images and analyze body fluids. The clinical applications include: Fetal/Obstetrics, Abdominal, Gynecology, Intraoperative, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-vaginal, Muscular-Skeletal (Conventional, Superficial), Urology, Cardiac Adult, Cardiac Pediatric, Trans-esophageal (Cardiac) and Peripheral vessel.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.

{3}------------------------------------------------

# 510(k) No.: K192030 Device Name: RS80 EVO Diagnostic Ultrasound System Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                                           |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-------------------------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)                          |
| Ophthalmic                | Ophthalmic                    | P                                                 | P |     |     |                   |                      |                                           |
| Fetal Imaging<br>& Other  | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 3, 4, 7, 8, 9, 11, 14            |
|                           | Abdominal (See Note 10)       | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 4, 5, 6, 7, 8, 9, 11, 12, 14, 15 |
|                           | Intra-operative (See Note 6)  | P                                                 | P | P   |     | P                 | Note 1               | Notes 7, 9, 11                            |
|                           | Intra-operative (Neuro.)      | P                                                 | P | P   |     | P                 | Note 1               | Notes 7, 9, 11                            |
|                           | Laparoscopic                  |                                                   |   |     |     |                   |                      |                                           |
|                           | Pediatric                     | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9, 11                      |
|                           | Small Organ (See Note 5)      | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 5, 6, 7, 8, 9, 11, 12, 14        |
|                           | Neonatal Cephalic             | P                                                 | P | P   |     | P                 | Note 1               | Notes 8, 9, 11                            |
|                           | Adult Cephalic                | P                                                 | P | P   | P   | P                 | Note 1               | Notes 7                                   |
|                           | Trans-rectal                  | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7, 8, 9, 11, 12                  |
|                           | Trans-vaginal                 | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7, 8, 9, 11, 12                  |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                                           |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                                           |
|                           | Musculo-skel. (Convent.)      | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 5, 6, 7, 8, 9, 11, 14            |
|                           | Musculo-skel. (Superfic.)     | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 5, 6, 7, 8, 9, 11, 14            |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                                           |
|                           | Other (See Note 13)           | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7, 8, 9, 12                      |
| Cardiac                   | Cardiac Adult                 | P                                                 | P | P   | P   | P                 | Note 1               | Notes 4, 7                                |
|                           | Cardiac Pediatric             | P                                                 | P | P   | P   | P                 | Note 1               | Notes 4, 7                                |
|                           | Trans-esophageal (Cardiac)    | P                                                 | P | P   | P   | P                 | Note 1               | Notes 4, 7                                |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                                           |
| Peripheral<br>Vessel      | Peripheral vessel             | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 5, 6, 7, 8, 9, 11, 14            |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                                           |

N= new indication; P= previously cleared by; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

- Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion

{4}------------------------------------------------

510(k) No.: K192030

#### Device Name: CA1-7A for use with RS80 EVO

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                          |
|----------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|--------------------------|
| General              | Specific                      | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)         |
| (Track I only)       | (Tracks I & III)              |   |                                                   |     |     |                   |                      |                          |
| Ophthalmic           | Ophthalmic                    |   |                                                   |     |     |                   |                      |                          |
|                      | Fetal/Obstetrics (See Note 3) | P | P                                                 | P   | P   | P                 | Note 1               | Notes 2,4,7,8,9,11,14    |
|                      | Abdominal (See Note 10)       | P | P                                                 | P   | P   | P                 | Note 1               | Notes 2,6,7,8,9,11,14,15 |
|                      | Intra-operative (See Note 6)  |   |                                                   |     |     |                   |                      |                          |
|                      | Intra-operative (Neuro.)      |   |                                                   |     |     |                   |                      |                          |
| Fetal Imaging        | Laparoscopic                  |   |                                                   |     |     |                   |                      |                          |
| &<br>Other           | Pediatric                     | P | P                                                 | P   | P   | P                 | Note 1               | Notes 2,6,7,8,9,11,14    |
|                      | Small Organ (See Note 5)      |   |                                                   |     |     |                   |                      |                          |
|                      | Neonatal Cephalic             |   |                                                   |     |     |                   |                      |                          |
|                      | Adult Cephalic                |   |                                                   |     |     |                   |                      |                          |
|                      | Trans-rectal                  |   |                                                   |     |     |                   |                      |                          |
|                      | Trans-vaginal                 |   |                                                   |     |     |                   |                      |                          |
|                      | Trans-urethral                |   |                                                   |     |     |                   |                      |                          |
|                      | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |                   |                      |                          |
|                      | Musculo-skel. (Convent.)      | P | P                                                 | P   | P   | P                 | Note 1               | Notes 2,6,7,8,9,11,14    |
|                      | Musculo-skel. (Superfic.)     |   |                                                   |     |     |                   |                      |                          |
|                      | Intra-luminal                 |   |                                                   |     |     |                   |                      |                          |
|                      | Other (See Note 13)           | P | P                                                 | P   | P   | P                 | Note 1               | Notes 2,6,7,8,9,11,14    |
|                      | Cardiac Adult                 |   |                                                   |     |     |                   |                      |                          |
| Cardiac              | Cardiac Pediatric             |   |                                                   |     |     |                   |                      |                          |
|                      | Trans-esophageal (Cardiac)    |   |                                                   |     |     |                   |                      |                          |
|                      | Other (spec.)                 |   |                                                   |     |     |                   |                      |                          |
| Peripheral           | Peripheral vessel             | P | P                                                 | P   | P   | P                 | Note 1               | Notes 2,6,7,8,9,11,14    |
| Vessel               | Other (spec.)                 |   |                                                   |     |     |                   |                      |                          |

N= new indication; P= previously cleared by FDA K191115; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPI, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

- Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

{5}------------------------------------------------

### 510(k) No.: K192030 Device Name: CF4-9 for use with RS80 EVO Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                   |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)  |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                   |
|                           | Fetal/Obstetrics (See Note 3) |                                                   |   |     |     |                   |                      |                   |
|                           | Abdominal (See Note 10)       | P                                                 | P | P   |     | P                 | Note 1               | Notes 7, 8, 9, 11 |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                   |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                   |
| Fetal Imaging             | Laparoscopic                  |                                                   |   |     |     |                   |                      |                   |
| &<br>Other                | Pediatric                     | P                                                 | P | P   |     | P                 | Note 1               | Notes 7, 8, 9, 11 |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                   |
|                           | Neonatal Cephalic             | P                                                 | P | P   |     | P                 | Note 1               | Notes 7, 8, 9, 11 |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                   |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                   |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                   |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                   |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                   |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                   |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                   |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                   |
|                           | Other (See Note 13)           |                                                   |   |     |     |                   |                      |                   |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                   |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                   |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                   |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                   |
| Peripheral                | Peripheral vessel             | P                                                 | P | P   |     | P                 | Note 1               | Notes 8, 9, 11    |
| Vessel                    | Other (spec.)                 |                                                   |   |     |     |                   |                      |                   |

N= new indication; P= previously cleared by FDA K191115; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPL, B+CW, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

- Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

{6}------------------------------------------------

510(k) No.: K192030

### Device Name: CA2-8A for use with RS80 EVO

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                         |  |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|-------------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)        |  |
| Ophthalmic                | Ophthalmic                    |   |                                                   |     |     |                   |                      |                         |  |
|                           | Fetal/Obstetrics (See Note 3) | P | P                                                 | P   | P   | P                 | Note 1               | Notes 2, 4, 7, 8, 9, 11 |  |
|                           | Abdominal (See Note 10)       | P | P                                                 | P   | P   | P                 | Note 1               | Notes 2, 4, 7, 8, 9, 11 |  |
|                           | Intra-operative (See Note 6)  |   |                                                   |     |     |                   |                      |                         |  |
|                           | Intra-operative (Neuro.)      |   |                                                   |     |     |                   |                      |                         |  |
| Fetal Imaging<br>& Other  | Laparoscopic                  |   |                                                   |     |     |                   |                      |                         |  |
|                           | Pediatric                     |   |                                                   |     |     |                   |                      |                         |  |
|                           | Small Organ (See Note 5)      |   |                                                   |     |     |                   |                      |                         |  |
|                           | Neonatal Cephalic             |   |                                                   |     |     |                   |                      |                         |  |
|                           | Adult Cephalic                |   |                                                   |     |     |                   |                      |                         |  |
|                           | Trans-rectal                  |   |                                                   |     |     |                   |                      |                         |  |
|                           | Trans-vaginal                 |   |                                                   |     |     |                   |                      |                         |  |
|                           | Trans-urethral                |   |                                                   |     |     |                   |                      |                         |  |
|                           | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |                   |                      |                         |  |
|                           | Musculo-skel. (Convent.)      |   |                                                   |     |     |                   |                      |                         |  |
|                           | Musculo-skel. (Superfic.)     |   |                                                   |     |     |                   |                      |                         |  |
|                           | Intra-luminal                 |   |                                                   |     |     |                   |                      |                         |  |
|                           | Other (See Note 13)           |   |                                                   |     |     |                   |                      |                         |  |
|                           | Cardiac Adult                 |   |                                                   |     |     |                   |                      |                         |  |
| Cardiac                   | Cardiac Pediatric             |   |                                                   |     |     |                   |                      |                         |  |
|                           | Trans-esophageal (Cardiac)    |   |                                                   |     |     |                   |                      |                         |  |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                         |  |
| Peripheral                | Peripheral vessel             |   |                                                   |     |     |                   |                      |                         |  |
| Vessel                    | Other (spec.)                 |   |                                                   |     |     |                   |                      |                         |  |

N= new indication; P= previously cleared by FDA K191115; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+PPI, B+CW, B+C+PW, B+PD+PW, B+PPI+PW, B+PY+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

- Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development

- Note 4: Color M-mode
Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: S-Fusion Note 15: S-Shearwave

{7}------------------------------------------------

### 510(k) No.: K192030 Device Name: CA3-10A for use with RS80 EVO Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                         |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)        |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                         |
|                           | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 3, 7, 8, 9, 11 |
|                           | Abdominal (See Note 10)       | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7, 8, 9, 11    |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                         |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                         |
| Fetal Imaging             | Laparoscopic                  |                                                   |   |     |     |                   |                      |                         |
| &<br>Other                | Pediatric                     | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7, 8, 9, 11    |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                         |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                         |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                         |
|                           | Musculo-skel. (Convent.)      | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7, 8, 9, 11    |
|                           | Musculo-skel. (Superfic.)     | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7, 8, 9, 11    |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                         |
|                           | Other (See Note 13)           |                                                   |   |     |     |                   |                      |                         |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                         |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                         |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                         |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                         |
| Peripheral                | Peripheral vessel             | P                                                 | P | P   |     | P                 | Note 1               | Notes 2, 7, 8, 9, 11    |
| Vessel                    | Other (spec.)                 |                                                   |   |     |     |                   |                      |                         |

N= new indication; P= previously cleared by FDA K191115; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+TD, B+C+PW, B+PD+PW, B+PPF+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

- Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development

- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

- Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)

- Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)

- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

{8}------------------------------------------------

### 510(k) No.: K192030 Device Name: E3-12A for use with RS80 EVO Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               |  | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                      |  |
|---------------------------|-------------------------------|--|---------------------------------------------------|---|-----|-----|-------------------|----------------------|----------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  |  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)     |  |
| Ophthalmic                | Ophthalmic                    |  |                                                   |   |     |     |                   |                      |                      |  |
|                           | Fetal/Obstetrics (See Note 3) |  | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9     |  |
|                           | Abdominal (See Note 10)       |  | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9, 12 |  |
|                           | Intra-operative (See Note 6)  |  |                                                   |   |     |     |                   |                      |                      |  |
|                           | Intra-operative (Neuro.)      |  |                                                   |   |     |     |                   |                      |                      |  |
| Fetal Imaging<br>& Other  | Laparoscopic                  |  |                                                   |   |     |     |                   |                      |                      |  |
|                           | Pediatric                     |  |                                                   |   |     |     |                   |                      |                      |  |
|                           | Small Organ (See Note 5)      |  |                                                   |   |     |     |                   |                      |                      |  |
|                           | Neonatal Cephalic             |  |                                                   |   |     |     |                   |                      |                      |  |
|                           | Adult Cephalic                |  |                                                   |   |     |     |                   |                      |                      |  |
|                           | Trans-rectal                  |  | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9, 12 |  |
|                           | Trans-vaginal                 |  | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9, 12 |  |
|                           | Trans-urethral                |  |                                                   |   |     |     |                   |                      |                      |  |
|                           | Trans-esoph. (non-Cardiac)    |  |                                                   |   |     |     |                   |                      |                      |  |
|                           | Musculo-skel. (Convent.)      |  |                                                   |   |     |     |                   |                      |                      |  |
|                           | Musculo-skel. (Superfic.)     |  |                                                   |   |     |     |                   |                      |                      |  |
|                           | Intra-luminal                 |  |                                                   |   |     |     |                   |                      |                      |  |
|                           | Other (See Note 13)           |  | P                                                 | P | P   | P   | P                 | Note 1               | Notes 2, 7, 8, 9, 12 |  |
|                           | Cardiac Adult                 |  |                                                   |   |     |     |                   |                      |                      |  |
| Cardiac                   | Cardiac Pediatric             |  |                                                   |   |     |     |                   |                      |                      |  |
|                           | Trans-esophageal (Cardiac)    |  |                                                   |   |     |     |                   |                      |                      |  |
|                           | Other (spec.)                 |  |                                                   |   |     |     |                   |                      |                      |  |
|                           |                               |  |                                                   |   |     |     |                   |                      |                      |  |
| Peripheral<br>Vessel      | Peripheral vessel             |  |                                                   |   |     |     |                   |                      |                      |  |
|                           | Other (spec.)                 |  |                                                   |   |     |     |                   |                      |                      |  |

N= new indication; P= previously cleared by FDA K191115; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPL, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion

Note 2: Includes imaging for guidance of biopsy

{9}------------------------------------------------

510(k) No.: K192030

### Device Name: L3-12A for use with RS80 EVO

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                                   |  |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-----------------------------------|--|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)                  |  |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                                   |  |
|                           | Fetal/Obstetrics (See Note 3) |                                                   |   |     |     |                   |                      |                                   |  |
|                           | Abdominal (See Note 10)       | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11, 14     |  |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                                   |  |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                                   |  |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                                   |  |
|                           | Pediatric                     |                                                   |   |     |     |                   |                      |                                   |  |
|                           | Small Organ (See Note 5)      | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11, 12, 14 |  |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                                   |  |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                                   |  |
|                           | Trans-rectal                  |                                                   |   |     |     |                   |                      |                                   |  |
|                           | Trans-vaginal                 |                                                   |   |     |     |                   |                      |                                   |  |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                                   |  |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                                   |  |
|                           | Musculo-skel. (Convent.)      | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11, 14     |  |
|                           | Musculo-skel. (Superfic.)     | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11, 14     |  |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                                   |  |
|                           | Other (See Note 13)           |                                                   |   |     |     |                   |                      |                                   |  |
|                           | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                                   |  |
| Cardiac                   | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                                   |  |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                                   |  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                                   |  |
| Peripheral<br>Vessel      | Peripheral vessel             | P                                                 | P | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11, 14     |  |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                                   |  |

N= new indication; P= previously cleared by FDA K191115; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPI, B+CW, B+CW, B+C+PW, B+PD+PW, B+PPH+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion

Note 2: Includes imaging for guidance of biopsy

{10}------------------------------------------------

#### 510(k) No.: K192030

#### Device Name: EA2-11B for use with RS80 EVO

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                        |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)       |
| Ophthalmic                | Ophthalmic                    |                                                   |   |     |     |                   |                      |                        |
|                           | Fetal/Obstetrics (See Note 3) | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 8, 9        |
|                           | Abdominal (See Note 12)       | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 6, 7, 8, 9, 12 |
|                           | Intra-operative (See Note 6)  |                                                   |   |     |     |                   |                      |                        |
|                           | Intra-operative (Neuro.)      |                                                   |   |     |     |                   |                      |                        |
| Fetal Imaging<br>& Other  | Laparoscopic                  |                                                   |   |     |     |                   |                      |                        |
|                           | Pediatric                     |                                                   |   |     |     |                   |                      |                        |
|                           | Small Organ (See Note 5)      |                                                   |   |     |     |                   |                      |                        |
|                           | Neonatal Cephalic             |                                                   |   |     |     |                   |                      |                        |
|                           | Adult Cephalic                |                                                   |   |     |     |                   |                      |                        |
|                           | Trans-rectal                  | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 8, 9, 12    |
|                           | Trans-vaginal                 | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 8, 9, 12    |
|                           | Trans-urethral                |                                                   |   |     |     |                   |                      |                        |
|                           | Trans-esoph. (non-Cardiac)    |                                                   |   |     |     |                   |                      |                        |
|                           | Musculo-skel. (Convent.)      |                                                   |   |     |     |                   |                      |                        |
|                           | Musculo-skel. (Superfic.)     |                                                   |   |     |     |                   |                      |                        |
|                           | Intra-luminal                 |                                                   |   |     |     |                   |                      |                        |
|                           | Other (See Note 13)           | P                                                 | P | P   | P   | P                 | Note 1               | Note 2, 7, 8, 9, 12    |
| Cardiac                   | Cardiac Adult                 |                                                   |   |     |     |                   |                      |                        |
|                           | Cardiac Pediatric             |                                                   |   |     |     |                   |                      |                        |
|                           | Trans-esophageal (Cardiac)    |                                                   |   |     |     |                   |                      |                        |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                        |
| Peripheral<br>Vessel      | Peripheral vessel             |                                                   |   |     |     |                   |                      |                        |
|                           | Other (spec.)                 |                                                   |   |     |     |                   |                      |                        |

N= new indication: P= previously cleared by FDA K191115; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPI, B+TD, B+CW, B+C+PW, B+C+PW, B+PD+PW, B+PPH+PW, B+TD+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

- Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development

- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: S-Fusion
- Note 15: S-Shearwave

{11}------------------------------------------------

### 510(k) No.: K192030 Device Name: LA2-9A for use with RS80 EVO Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                                 |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|---------------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)                |
| Ophthalmic                | Ophthalmic                    |   |                                                   |     |     |                   |                      |                                 |
|                           | Fetal/Obstetrics (See Note 3) |   |                                                   |     |     |                   |                      |                                 |
|                           | Abdominal (See Note 10)       | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11       |
|                           | Intra-operative (See Note 6)  |   |                                                   |     |     |                   |                      |                                 |
|                           | Intra-operative (Neuro.)      |   |                                                   |     |     |                   |                      |                                 |
| Fetal Imaging<br>& Other  | Laparoscopic                  |   |                                                   |     |     |                   |                      |                                 |
|                           | Pediatric                     |   |                                                   |     |     |                   |                      |                                 |
|                           | Small Organ (See Note 5)      | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11,12,15 |
|                           | Neonatal Cephalic             |   |                                                   |     |     |                   |                      |                                 |
|                           | Adult Cephalic                |   |                                                   |     |     |                   |                      |                                 |
|                           | Trans-rectal                  |   |                                                   |     |     |                   |                      |                                 |
|                           | Trans-vaginal                 |   |                                                   |     |     |                   |                      |                                 |
|                           | Trans-urethral                |   |                                                   |     |     |                   |                      |                                 |
|                           | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |                   |                      |                                 |
|                           | Musculo-skel. (Convent.)      | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11,12,15 |
|                           | Musculo-skel. (Superfic.)     | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11,12,15 |
|                           | Intra-luminal                 |   |                                                   |     |     |                   |                      |                                 |
|                           | Other (See Note 13)           |   |                                                   |     |     |                   |                      |                                 |
|                           | Cardiac Adult                 |   |                                                   |     |     |                   |                      |                                 |
| Cardiac                   | Cardiac Pediatric             |   |                                                   |     |     |                   |                      |                                 |
|                           | Trans-esophageal (Cardiac)    |   |                                                   |     |     |                   |                      |                                 |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                                 |
| Peripheral<br>Vessel      | Peripheral vessel             | P | P                                                 | P   |     | P                 | Note 1               | Note 2, 5, 6, 7, 8, 9, 11       |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                                 |

N= new indication; P= previously cleared by FDA K191115; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+PPL, B+CW, B+CW, B+C+PW, B+PD+PW, B+PD+PW, B+PPH+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

- Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development

Note 4: Color M-mode

Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

Note 6: Abdominal organs and peripheral vessel

Note 7: Tissue Harmonic Imaging (THI)

Note 8: 3D imaging

Note 9: Spatial Compound Imaging

Note 10: Includes Renal, Gynecology/Pelvis

Note 11: Panoramic imaging

Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion

{12}------------------------------------------------

### 510(k) No.: K192030 Device Name: LA3-16A for use with RS80 EVO Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                            |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|----------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)           |
| Ophthalmic                | Ophthalmic                    |   |                                                   |     |     |                   |                      |                            |
| Fetal Imaging<br>& Other  | Fetal/Obstetrics (See Note 3) |   |                                                   |     |     |                   |                      |                            |
|                           | Abdominal (See Note 10)       |   |                                                   |     |     |                   |                      |                            |
|                           | Intra-operative (See Note 6)  |   |                                                   |     |     |                   |                      |                            |
|                           | Intra-operative (Neuro.)      |   |                                                   |     |     |                   |                      |                            |
|                           | Laparoscopic                  |   |                                                   |     |     |                   |                      |                            |
|                           | Pediatric                     |   |                                                   |     |     |                   |                      |                            |
|                           | Small Organ (See Note 5)      | P | P                                                 | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11, 12 |
|                           | Neonatal Cephalic             |   |                                                   |     |     |                   |                      |                            |
|                           | Adult Cephalic                |   |                                                   |     |     |                   |                      |                            |
|                           | Trans-rectal                  |   |                                                   |     |     |                   |                      |                            |
|                           | Trans-vaginal                 |   |                                                   |     |     |                   |                      |                            |
|                           | Trans-urethral                |   |                                                   |     |     |                   |                      |                            |
|                           | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |                   |                      |                            |
|                           | Musculo-skel. (Convent.)      | P | P                                                 | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |
|                           | Musculo-skel. (Superfic.)     | P | P                                                 | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |
|                           | Intra-luminal                 |   |                                                   |     |     |                   |                      |                            |
|                           | Other (See Note 13)           |   |                                                   |     |     |                   |                      |                            |
| Cardiac                   | Cardiac Adult                 |   |                                                   |     |     |                   |                      |                            |
|                           | Cardiac Pediatric             |   |                                                   |     |     |                   |                      |                            |
|                           | Trans-esophageal (Cardiac)    |   |                                                   |     |     |                   |                      |                            |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                            |
| Peripheral<br>Vessel      | Peripheral vessel             | P | P                                                 | P   | P   | P                 | Note 1               | Note 2, 5, 6, 7, 9, 11     |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                            |

N= new indication; P= previously cleared by FDA K191115; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+TD, B+C+PW, B+PD+PW, B+PPF+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

- Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development

- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

- Note 6: Abdominal organs and peripheral vessel
- Note 7: Tissue Harmonic Imaging (THI)
- Note 8: 3D imaging
- Note 9: MultiVision (Spatial Compound Imaging)
- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan
- Note 13: Includes Urology/Prostate
- Note 14: S-Fusion Note 15: S-Shearwave

{13}------------------------------------------------

#### 510(k) No.: K192030 Device Name: LA3-16AI for use with RS80 EVO Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               |   | Mode of Operation (*includes simultaneous B-mode) |     |     |                   |                      |                  |
|---------------------------|-------------------------------|---|---------------------------------------------------|-----|-----|-------------------|----------------------|------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B | M                                                 | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                | Ophthalmic                    |   |                                                   |     |     |                   |                      |                  |
|                           | Fetal/Obstetrics (See Note 3) |   |                                                   |     |     |                   |                      |                  |
|                           | Abdominal (See Note 10)       |   |                                                   |     |     |                   |                      |                  |
|                           | Intra-operative (See Note 6)  | P | P                                                 | P   | P   | P                 | Note 1               | Note 7, 9, 11    |
|                           | Intra-operative (Neuro.)      | P | P                                                 | P   | P   | P                 | Note 1               | Note 7, 9, 11    |
| Fetal Imaging<br>& Other  | Laparoscopic                  |   |                                                   |     |     |                   |                      |                  |
|                           | Pediatric                     |   |                                                   |     |     |                   |                      |                  |
|                           | Small Organ (See Note 5)      |   |                                                   |     |     |                   |                      |                  |
|                           | Neonatal Cephalic             |   |                                                   |     |     |                   |                      |                  |
|                           | Adult Cephalic                |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-rectal                  |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-vaginal                 |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-urethral                |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-esoph. (non-Cardiac)    |   |                                                   |     |     |                   |                      |                  |
|                           | Musculo-skel. (Convent.)      | P | P                                                 | P   | P   | P                 | Note 1               | Note 7, 9, 11    |
|                           | Musculo-skel. (Superfic.)     | P | P                                                 | P   | P   | P                 | Note 1               | Note 7, 9, 11    |
|                           | Intra-luminal                 |   |                                                   |     |     |                   |                      |                  |
|                           | Other (See Note 13)           |   |                                                   |     |     |                   |                      |                  |
|                           | Cardiac Adult                 |   |                                                   |     |     |                   |                      |                  |
| Cardiac                   | Cardiac Pediatric             |   |                                                   |     |     |                   |                      |                  |
|                           | Trans-esophageal (Cardiac)    |   |                                                   |     |     |                   |                      |                  |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                  |
| Peripheral<br>Vessel      | Peripheral vessel             | P | P                                                 | P   | P   | P                 | Note 1               | Note 7, 9, 11    |
|                           | Other (spec.)                 |   |                                                   |     |     |                   |                      |                  |

N= new indication; P= previously cleared by FDA K191115; E= added under Appendix E

#### Additional Comments:

Color Doppler includes Power (Amplitude) Doppler

Note 1: B+M, B+PW, B+C, B+PD, B+DPD, B+PPI, B+TD, B+C+PW, B+PD+PW, B+PPF+PW, B+PPI+PW, B+C+M, Dual/Quad, B+C+CW, B+PD+CW, B+E, B+B/C

- Note 2: Includes imaging for guidance of biopsy
Note 3: Includes infertility monitoring of follicle development

- Note 4: Color M-mode
Note 5: For example: thyroid, parathyroid, breast, scrotum and penis in adult, pediatric and neonatal patients

- Note 6: Abdominal organs and peripheral vessel
Note 7: Tissue Harmonic Imaging (THI)

- Note 8: 3D imaging
Note 9: MultiVision (Spatial Compound Imaging)

- Note 10: Includes Renal, Gynecology/Pelvis
- Note 11: Panoramic imaging
- Note 12: ElastoScan

Note 13: Includes Urology/Prostate

Note 14: S-Fusion Note 15: S-Shearwave

Concurrence of Center for Devices and Radiological Health (CDRH) Prescription Use (Per 21 CFR 801.109)

{14}------------------------------------------------

### 510(k) No.: K192030 Device Name: LA4-18B for use with RS80 EVO Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                               | Mode of Operation (*includes simultaneous B-mode) |   |     |     |                   |                      |                                   |
|---------------------------|-------------------------------|---------------------------------------------------|---|-----|-----|-------------------|----------------------|-----------------------------------|
| General<br>(Track I only) | Specific<br>(Tracks I & III)  | B                                                 | M | PWD | CWD | Color<br>Doppler* | Combined*<br>(Spec.) | Other<br>(Spec.)                  |
| Ophthalmic                | Ophthalmic                    |…

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K192030](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K192030)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com