← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K172082

# S60 Series Digital Color Doppler Ultrasound System (K172082)

_Sonoscape Medical Corp. · IYN · Sep 21, 2017 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K172082

## Device Facts

- **Applicant:** Sonoscape Medical Corp.
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Sep 21, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Intended Use

The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

## Device Story

The S60 Series is a mobile, software-controlled, color diagnostic ultrasound system. It transmits ultrasonic energy into the body and processes received echoes to generate real-time images of anatomical structures and fluid flow. The system supports various imaging modes: B-Mode (including Tissue Harmonic Imaging), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continuous Wave Doppler, Power Doppler, Elastography, and 3D/4D imaging. It utilizes a wide array of transducers (linear, convex, phased) with frequencies from 1.0 to 25.0 MHz. Operated by qualified physicians in clinical settings, the system displays images on a color LCD monitor via a console with touch screen and keyboard controls. The output assists clinicians in diagnostic evaluation of various body systems, potentially improving patient management through non-invasive visualization.

## Clinical Evidence

No clinical data. Bench testing only, including phantom studies, electrical safety, electromagnetic compatibility, and acoustic output measurements.

## Technological Characteristics

Mobile console with color LCD, touch screen, and keyboard. Transducers: linear, convex, phased arrays (1.0-25.0 MHz). Modes: B, M, TDI, Color/Power/PWD/CWD Doppler, Elastography, 3D/4D. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-5/10, AIUM/NEMA UD 2/3.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- SonoScape S45 Digital Color Doppler Ultrasound System ([K160283](/device/K160283.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 21, 2017

SONOSCAPE MEDICAL CORP. % Ms. Toki Wu Regulatory Affairs Manager 4/f, 5/f, 8/f, 9/f & 10/f Yizhe Building, Yuquan Road, Nanshan Shenzhen 518051, GuangDong CHINA

Re: K172082

Trade/Device Name: S60 Series Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: July 4, 2017 Received: July 17, 2017

Dear Ms. Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D. O'Hara For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

#### K172082

Device Name

S60 Series Digital Color Doppler Ultrasound System

#### Indications for Use (Describe)

The Digital Color Doppler Ultrasound System is a general-purpose ultrasonic instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology,

Type of Use (Select one or both, as applicable):

| <span style="font-size:100%;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span>  |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SonoScape S60 Exp/S60/S60 Pro/S60 VO/S60 Maso/S59 System: Diagnostic Ultrasound Pulsed Echo System

Diagnostic Ultrasound Pulsed Doppler Imaging System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                    |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|--------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify  |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                    |
|                              | Fetal                              | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4,5        |
|                              | Abdominal                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4,5        |
| Fetal<br>Imaging&<br>Other   | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                    |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                    |
|                              | Laparoscopic                       | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4          |
|                              | Pediatric                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4          |
|                              | Small Organ (specify)              | P                 | P | P   |     | P                | P                               | Note 1             | Notes<br>2,4,5,6,7 |
|                              | Neonatal Cephalic                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4        |
|                              | Adult Cephalic                     | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4        |
|                              | Trans-rectal                       | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4,5        |
|                              | Trans-vaginal                      | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4,5        |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                    |
|                              | Trans-esoph.(non-Card)             |                   |   |     |     |                  |                                 |                    |                    |
|                              | Musculo-skeletal<br>(Conventional) | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4          |
|                              | Musculo-skeletal<br>(Superficial)  | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4          |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                    |
|                              | Other (Ob/GYN)                     | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4,5        |
|                              | Other (Urology)                    | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4,5        |
| Cardiac                      | Cardiac Adult                      | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4        |
|                              | Cardiac Pediatric                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4        |
|                              | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                    |
|                              | Trans-esoph.(Cardiac)              | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4        |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                    |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                    |
| Peripheral<br>Vessel         | Peripheral vessel                  | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4          |
|                              | Cerebral vascular                  |                   |   | P   |     |                  |                                 |                    |                    |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{4}------------------------------------------------

### Transducer: C322 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    |   | Mode of Operation |     |     |                  |                                 |                    |                   |  |
|------------------------------|------------------------------------|---|-------------------|-----|-----|------------------|---------------------------------|--------------------|-------------------|--|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B | M                 | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |  |
| Ophthalmic                   | Ophthalmic                         |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Fetal                              | P | P                 | P   |     | P                | P                               | Note 1             | Notes 2,4         |  |
|                              | Abdominal                          | P | P                 | P   |     | P                | P                               | Note 1             | Notes 2,4         |  |
|                              | Intra-operative Specify            |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-operative Neuro              |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Laparoscopic                       |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Pediatric                          |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Small Organ (specify)              |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Neonatal Cephalic                  |   |                   |     |     |                  |                                 |                    |                   |  |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-rectal                       |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-vaginal                      |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-urethral                     |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-esoph.(non-Card)             |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Musculo-skeletal<br>(Conventional) |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Musculo-skeletal<br>(Superficial)  |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Intravascular                      |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Other (Ob/GYN)                     | P | P                 | P   |     | P                | P                               | Note 1             | Notes 2,4         |  |
|                              | Other (Urology)                    |   |                   |     |     |                  |                                 |                    |                   |  |
| Cardiac                      | Cardiac Adult                      |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Cardiac Pediatric                  |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Intravascular(Cardiac)             |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-esoph. (Cardiac)             |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-cardiac                      |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Other (specify)                    |   |                   |     |     |                  |                                 |                    |                   |  |
| Peripheral<br>Vessel         | Peripheral vessel                  |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Cerebral vascular                  |   |                   |     |     |                  |                                 |                    |                   |  |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{5}------------------------------------------------

### Transducer: 3C-A Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |
|                              | Fetal                              | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Abdominal                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Pediatric                          |                   |   |     |     |                  |                                 |                    |                   |
|                              | Small Organ (specify)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Neonatal Cephalic                  |                   |   |     |     |                  |                                 |                    |                   |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph.(non-Card)             |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Ob/GYN)                     | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Adult                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Pediatric                  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |
| Cardiac                      | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |
| Peripheral                   | Peripheral vessel                  |                   |   |     |     |                  |                                 |                    |                   |
| Vessel                       | Cerebral vascular                  |                   |   |     |     |                  |                                 |                    |                   |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{6}------------------------------------------------

### Transducer: C1-6 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |
|                              | Fetal                              | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Abdominal                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Pediatric                          |                   |   |     |     |                  |                                 |                    |                   |
|                              | Small Organ (specify)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Neonatal Cephalic                  |                   |   |     |     |                  |                                 |                    |                   |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph.(non-Card)             |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Ob/GYN)                     | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |
| Cardiac                      | Cardiac Adult                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Pediatric                  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                                 |                    |                   |
| Vessel                       | Cerebral vascular                  |                   |   |     |     |                  |                                 |                    |                   |

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{7}------------------------------------------------

### Transducer: C613 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |
|                              | Fetal                              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Abdominal                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Pediatric                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Small Organ (specify)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Neonatal Cephalic                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph.(non-Card)             |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Ob/GYN)                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Adult                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Pediatric                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
|                              | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |
| Cardiac                      | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cerebral vascular                  |                   |   |     |     |                  |                                 |                    |                   |

P = previously cleared by FDA; E = added under this appendix N = new indication;

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{8}------------------------------------------------

### Transducer: 8P1 Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |
|                              | Fetal                              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Abdominal                          |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Pediatric                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Small Organ (specify)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Neonatal Cephalic                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
| Fetal                        | Adult Cephalic                     |                   |   |     |     |                  |                                 |                    |                   |
| Imaging&                     | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |
| Other                        | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph. (non-Card)            |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Ob/GYN)                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Adult                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Pediatric                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
|                              | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |
| Cardiac                      | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |
| Peripheral                   | Peripheral vessel                  |                   |   |     |     |                  |                                 |                    |                   |
| Vessel                       | Cerebral vascular                  |                   |   |     |     |                  |                                 |                    |                   |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{9}------------------------------------------------

### Transducer: 4P-A Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
|                              |                                    |                   |   |     |     |                  |                                 |                    |                   |
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |
|                              | Fetal                              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Abdominal                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Pediatric                          |                   |   |     |     |                  |                                 |                    |                   |
|                              | Small Organ (specify)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Neonatal Cephalic                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
|                              | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph.(non-Card)             |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Ob/GYN)                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Adult                      | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
| Cardiac                      | Cardiac Pediatric                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
|                              | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cerebral vascular                  |                   |   |     |     |                  |                                 |                    |                   |

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{10}------------------------------------------------

### Transducer: S1-5 Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |
|                              | Fetal                              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Abdominal                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Pediatric                          |                   |   |     |     |                  |                                 |                    |                   |
|                              | Small Organ (specify)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Neonatal Cephalic                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
|                              | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph.(non-Card)             |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Ob/GYN)                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Adult                      | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
| Cardiac                      | Cardiac Pediatric                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
|                              | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cerebral vascular                  |                   |   |     |     |                  |                                 |                    |                   |

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{11}------------------------------------------------

### Transducer: 7P-A Phased Array

### Diagnostic Ultrasound Transducer

### Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |  |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|--|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |  |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Fetal                              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Abdominal                          |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Pediatric                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |  |
|                              | Small Organ (specify)              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Neonatal Cephalic                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |  |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-esoph.(non-Card)             |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Other (Ob/GYN)                     |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Cardiac Adult                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Cardiac Pediatric                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |  |
|                              | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |  |
| Cardiac                      | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |  |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Cerebral vascular                  |                   |   |     |     |                  |                                 |                    |                   |  |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{12}------------------------------------------------

Transducer: 1012 Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         | Specific<br>(TRACKS 1 & 3)         | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
|------------------------------|------------------------------------|---|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Ophthalmic                         |   |   |     |     |                  |                                 |                    |                   |
| Ophthalmic                   | Fetal                              |   |   |     |     |                  |                                 |                    |                   |
|                              | Abdominal                          |   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-operative Specify            |   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-operative Neuro              |   |   |     |     |                  |                                 |                    |                   |
|                              | Laparoscopic                       |   |   |     |     |                  |                                 |                    |                   |
|                              | Pediatric                          |   |   |     |     |                  |                                 |                    |                   |
|                              | Small Organ (specify)              | P | P | P   |     | P                | P                               | Note 1             | Notes 2,4,6       |
|                              | Neonatal Cephalic                  |   |   |     |     |                  |                                 |                    |                   |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     |   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-rectal                       |   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-vaginal                      |   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-urethral                     |   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph.(non-Card)             |   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Conventional) | P | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Musculo-skeletal<br>(Superficial)  |   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular                      |   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Ob/GYN)                     |   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Urology)                    |   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Adult                      |   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Pediatric                  |   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular(Cardiac)             |   |   |     |     |                  |                                 |                    |                   |
| Cardiac                      | Trans-esoph.(Cardiac)              |   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-cardiac                      |   |   |     |     |                  |                                 |                    |                   |
|                              | Other (specify)                    |   |   |     |     |                  |                                 |                    |                   |
|                              | Peripheral vessel                  | P | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Cerebral vascular                  |   |   |     |     |                  |                                 |                    |                   |
| Peripheral<br>Vessel         |                                    |   |   |     |     |                  |                                 |                    |                   |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 3: TDI Note 4: 3D

Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{13}------------------------------------------------

### Transducer: 9L-A Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |  |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|--|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |  |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Fetal                              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Abdominal                          |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Pediatric                          |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Small Organ (specify)              | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4,6       |  |
|                              | Neonatal Cephalic                  |                   |   |     |     |                  |                                 |                    |                   |  |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-esoph.(non-Card)             |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Musculo-skeletal<br>(Conventional) | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |  |
|                              | Musculo-skeletal<br>(Superficial)  | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |  |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Other (Ob/GYN)                     |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Cardiac Adult                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Cardiac Pediatric                  |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |  |
| Cardiac                      | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |  |
| Peripheral<br>Vessel         | Peripheral vessel                  | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |  |
|                              | Cerebral vascular                  |                   |   |     |     |                  |                                 |                    |                   |  |

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{14}------------------------------------------------

### Transducer: 12L-A Linear Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |
|                              | Fetal                              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Abdominal                          |                   |   |     |     |                  |…

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K172082](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K172082)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
