Browse hierarchy Radiology (RA) Subpart B — Diagnostic Devices 21 CFR 892.1550 Product Code IYN K170277 — DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System
DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System
K170277 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · IYN · Mar 23, 2017 · Radiology
Device Facts
Record ID K170277
Device Name DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System
Applicant Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Product Code IYN · Radiology
Decision Date Mar 23, 2017
Decision SESE
Submission Type Traditional
Regulation 21 CFR 892.1550
Device Class Class 2
Attributes Pediatric
Intended Use
The DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), cardiac adult, cardiac pediatric, peripheral vessel, urology and transesophageal (Cardiac) exams.
Device Story
The DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S is a mobile, software-controlled diagnostic ultrasound system. It acquires ultrasound signals via various transducers (linear, convex, phased array) and processes them to generate images in B-mode, M-mode, PW/CW Doppler, Color Doppler, Color M-mode, Power/Dirpower, TDI, 3D/4D, and Elastography. Used in clinical settings by healthcare professionals, the system provides real-time visualization of anatomic structures and fluid flow. Output is displayed on-screen for clinical assessment, supporting diagnostic decision-making and patient management. The system includes advanced features like Smart MSP and Smart FLC for automated measurements and contrast imaging capabilities. It benefits patients by providing non-invasive, real-time diagnostic imaging across a wide range of clinical applications.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including acoustic output measurements, biocompatibility (ISO 10993-1), and compliance with electrical/mechanical safety standards (AAMI/ANSI ES60601-1, IEC 60601-2-37, IEC 62304).
Technological Characteristics
Mobile ultrasound system; linear, convex, and phased array transducers; frequency range 3-10 MHz. Operating modes: B, M, PW, CW, Color, Power/Dirpower, TDI, 3D/4D, Elastography. Software-controlled; compliant with IEC 62304 (software lifecycle) and IEC 62366 (usability). Biocompatible patient-contact materials per ISO 10993-1. Electrical safety per AAMI/ANSI ES60601-1.
Indications for Use
Indicated for adults, pregnant women, pediatric patients, and neonates for diagnostic ultrasound imaging and fluid flow analysis in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal/adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal, cardiac (adult/pediatric/transesophageal), peripheral vessel, and urology exams.
Regulatory Classification
Identification An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Related Devices
K150080 — DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Jun 12, 2015
K173471 — DC-80/DC-80 PRO/DC-80 EXP/DC-80S/DC-85 Diagnostic · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Jan 8, 2018
K190206 — DC-90/DC-90S/DC-90Q/DC-95/DC-95S/DC-88/DC-88S/DC-80A/DC-80A Exp/DC-80A Pro/DC-8X/DC-8Q/DC-81/DC-82 Diagnostic Ultrasound System · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · Mar 8, 2019
K202785 — Resona R9,Resona R9 Exp,Resona R9 Pro,Resona R9S, Nuewa R9,Nuewa R9 Exp,Nuewa R9 Pro,Nuewa R9S,Resona 7,Resona 7CV,Resona 7EXP,Resona 7S, Resona 7OB,Resona 7PRO,Imagyn 7,Resona Y Diagnostic Ultrasound System · Shenzhen Mindray Bio-Meidcal Electronics Co., Ltd. · Apr 8, 2021
K063580 — SONOACE X8 DIAGNOSTIC ULTRASOUND SYSTEM · Medison Co., Ltd. · Dec 14, 2006
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, with a flowing, ribbon-like design connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 23, 2017
Shenzhen Mindray Bio-Medical Electronics Co., Ltd. % Mr. Yang Zhaohui Engineer of Technical Regulation Mindray Building, Keji 12th Road South Hi-tech Industrial Park Nanshan, Shenzhen 518057 P.R. CHINA
CHINA
Re: K170277
Trade/Device Name: DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX, LLZ Dated: January 16, 2017 Received: January 30, 2017
Dear Mr. Zhaohui:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K170277
Device Name
DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System
Indications for Use (Describe)
The DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), neonatal cephalic, adult cephalic, trans-raginal, musculo-skeletal (conventional, superficial), cardiac adult, cardiac pediatric, peripheral vessel, urology and transesophageal (Cardiac) exams.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
PSC Publishing Services (301) 443-6740
BF
{3}------------------------------------------------
## Diagnostic Ultrasound Indications For Use Format
| System: | DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System | | | | | | | | |
|---------------------------|--------------------------------------------------------------------------------------------|---|---|-----|-----|------------------|----------------------|-----------------------|----------------------|
| Transducer: | N/A | | | | | | | | |
| 510(k) Number: | | | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | |
| Clinical Application | Mode of Operation | | | | | | | | |
| General (Track 1<br>Only) | Specific (Track 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other (specify) |
| Ophthalmic | Ophthalmic | | | | | | | | |
| | Fetal | P | P | P | | P | P | P | Note 1, 2,3, 4,6,7 |
| | Abdominal | P | P | P | P | P | P | P | Note 1, 2,3, 4,6,7,9 |
| | Intra-operative (Specify*) | P | P | P | | P | P | P | Note 1, 2, 4 |
| | Intra-operative (Neuro) | | | | | | | | |
| | Laparoscopic | | | | | | | | |
| | Pediatric | P | P | P | P | P | P | P | Note 1, 2,3, 4,6,7 |
| | Small Organ (Specify**) | P | P | P | | P | P | P | Note 1,2, 4,6,7,8 |
| Fetal Imaging &<br>Other | Neonatal Cephalic | P | P | P | P | P | P | P | Note 1, 2,4,6,7 |
| | Adult Cephalic | P | P | P | P | P | P | P | Note 1, 2,4,6,7 |
| | Trans-rectal | P | P | P | | P | P | P | Note 1, 2,3, 4,6,7 |
| | Trans-vaginal | P | P | P | | P | P | P | Note 1, 2,3, 4,6,7,8 |
| | Trans-urethral | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | |
| | Musculo-skeletal (Conventional) | P | P | P | P | P | P | P | Note 1, 2, 4,6,7 |
| | Musculo-skeletal (Superficial) | P | P | P | | P | P | P | Note 1, 2, 4,6,7 |
| | Intravascular | | | | | | | | |
| | Cardiac Adult | P | P | P | P | P | P | P | Note 1, 2,4,5,6,7,10 |
| | Cardiac Pediatric | P | P | P | P | P | P | P | Note 1, 2,4,5,6,7 |
| Cardiac | Intravascular (Cardiac) | | | | | | | | |
| | Trans-esoph. (Cardiac) | P | P | P | P | P | P | P | Note 1, 5,6 |
| | Intra-cardiac | | | | | | | | |
| Peripheral vessel | Peripheral vessel | P | P | P | P | P | P | P | Note 1, 2, 4,6,7 |
| | Other (Specify***) | P | P | P | | P | P | P | Note 1, 2, 4,6,7 |
| | N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | |
| | Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW+B. | | | | | | | | |
| | *Intraoperative includes abdominal, thoracic, and vascular. | | | | | | | | |
| | **Small organ-breast, thyroid, testes. | | | | | | | | |
| | ***Other use includes Urology. | | | | | | | | |
| | Note 1: Tissue Harmonic Imaging. | | | | | | | | |
| | Note 2: Smart3D | | | | | | | | |
| Note 3: 4D(Real-time 3D) | | | | | | | | | |
| Note 4: iScape | | | | | | | | | |
| Note 5: TDI | | | | | | | | | |
| Note 6: Color M | | | | | | | | | |
| Note 7: Biopsy Guidance | | | | | | | | | |
| | Note 8: Elastography | | | | | | | | |
| | Note 9: Contrast imaging - liver only | | | | | | | | |
| | Note10: Contrast imaging - LVO only | | | | | | | | |
{4}------------------------------------------------
#### System: DC-8/DC-8 PRO/DC-8 CV/DC-8 EXP/DC-8S Diagnostic Ultrasound System Transducer: C5-2E
510(k) Number:
Intended Use:
| Clinical Application | Mode of Operation | | | | | | | | | |
|----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|-------------------|-----|-----|-------------------|----------------------|-----------------------|--------------------|--|
| General (Track 1<br>Only) | Specific (Track 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| | Fetal | P | P | P | | P | P | P | Note 1, 2, 4,6,7 | |
| | Abdominal | P | P | P | | P | P | P | Note 1, 2, 4,6,7,9 | |
| Fetal Imaging &<br>Other | Intra-operative (Specify*) | | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | P | P | P | | P | P | P | Note 1, 2, 4,6,7 | |
| | Small Organ (Specify**) | | | | | | | | | |
| | Neonatal Cephalic | | | | | | | | | |
| | Adult Cephalic | | | | | | | | | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | | |
| | Musculo-skeletal (Conventional) | P | P | P | | P | P | P | Note 1, 2, 4,6,7 | |
| | Musculo-skeletal (Superficial) | | | | | | | | | |
| Cardiac | Intravascular | | | | | | | | | |
| | Cardiac Adult | | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | | |
| | Intravascular (Cardiac) | | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | | |
| Peripheral vessel | Intra-cardiac | | | | | | | | | |
| | Peripheral vessel | P | P | P | | P | P | P | Note 1, 2, 4,6,7 | |
| | Other (Specify***) | | | | | | | | | |
| N=new indication; P=previously cleared by FDA(K150080); E=added under Appendix E | | | | | | | | | | |
| Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW+B. | | | | | | | | | | |
| *Intraoperative includes abdominal, thoracic, and vascular. | | | | | | | | | | |
| **Small organ-breast, thyroid, testes. | | | | | | | | | | |
| ***Other use includes Urology. | | | | | | | | | | |
| Note 1: Tissue Harmonic Imaging. | | | | | | | | | | |
| Note 2: Smart3D | | | | | | | | | | |
| Note 3: 4D(Real-time 3D) | | | | | | | | | | |
| Note 4: iScape | | | | | | | | | | |
| Note 5: TDI | | | | | | | | | | |
| Note 6: Color M | | | | | | | | | | |
| Note 7: Biopsy Guidance | | | | | | | | | | |
| Note 8: Elastography | | | | | | | | | | |
| Note 9: Contrast imaging - liver only | | | | | | | | | | |
| Note10: Contrast imaging - LVO only | | | | | | | | | | |
| Clinical Application | | | Mode of Operation | | | | | | | |
| General (Track 1<br>Only) | Specific (Track 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| | Fetal | P | P | P | | P | P | P | Note 1, 2, 4,6,7 | |
| | Abdominal | P | P | P | | P | P | P | Note 1, 2, 4,6,7,9 | |
| | Intra-operative (Specify*) | | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | P | P | P | | P | P | P | Note 1, 2, 4,6,7 | |
| | Small Organ (Specify**) | | | | | | | | | |
| Fetal Imaging & | Neonatal Cephalic | | | | | | | | | |
| Other | Adult Cephalic | | | | | | | | | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | | |
| | Musculo-skeletal (Conventional) | | | | | | | | | |
| | Musculo-skeletal (Superficial) | | | | | | | | | |
| | Intravascular | | | | | | | | | |
| | Cardiac Adult | | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | | |
| Cardiac | Intravascular (Cardiac) | | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | | |
| | Intra-cardiac | | | | | | | | | |
| Peripheral vessel | Peripheral vessel | P | P | P | | P | P | P | Note 1, 2, 4,6,7 | |
| | Other (Specify***) | | | | | | | | | |
| | N=new indication; P=previously cleared by FDA(K150080); E=added under Appendix E | | | | | | | | | |
| | Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+ B、Power + PW +B. | | | | | | | | | |
| | *Intraoperative includes abdominal, thoracic, and vascular. | | | | | | | | | |
| | **Small organ-breast, thyroid, testes. | | | | | | | | | |
| | ***Other use includes Urology. | | | | | | | | | |
| | Note 1: Tissue Harmonic Imaging. | | | | | | | | | |
| Note 2: Smart3D | | | | | | | | | | |
| | Note 3: 4D(Real-time 3D) | | | | | | | | | |
| Note 4: iScape | | | | | | | | | | |
| Note 5: TDI | | | | | | | | | | |
| | Note 6: Color M | | | | | | | | | |
| Note 7: Biopsy Guidance | | | | | | | | | | |
| | Note 8: Elastography | | | | | | | | | |
| | Note 9: Contrast imaging - liver only | | | | | | | | | |
| | Note10: Contrast imaging — LVO only | | | | | | | | | |
| | Concurrence of CDRH, Office of Device Evaluation(ODE) | | | | | | | | | |
| Clinical Application | | | Mode of Operation | | | | | | | |
| General (Track 1<br>Only) | Specific (Track 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| | Fetal | | | | | | | | | |
| | Abdominal | P | P | P | | P | P | P | Note 1, 2, 4,6,7 | |
| | Intra-operative (Specify*) | | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | P | P | P | | P | P | P | Note 1, 2, 4,6,7 | |
| | Small Organ (Specify**) | P | P | P | | P | P | P | Note 1,2, 4,6,7,8 | |
| Fetal Imaging &<br>Other | Neonatal Cephalic | P | P | P | | P | P | P | Note 1, 2, 4,6,7 | |
| | Adult Cephalic | | | | | | | | | |
| | Trans-rectal | | | | | | | | | |
| | Trans-vaginal | | | | | | | | | |
| | Trans-urethral | | | | | | | | | |
| | Trans-esoph. (non-Card.) | | | | | | | | | |
| | Musculo-skeletal (Conventional) | P | P | P | | P | P | P | Note 1, 2, 4,6,7 | |
| | Musculo-skeletal (Superficial) | P | P | P | | P | P | P | Note 1, 2, 4,6,7 | |
| | Intravascular | | | | | | | | | |
| | Cardiac Adult | | | | | | | | | |
| | Cardiac Pediatric | | | | | | | | | |
| Cardiac | Intravascular (Cardiac) | | | | | | | | | |
| | Trans-esoph. (Cardiac) | | | | | | | | | |
| | Intra-cardiac | | | | | | | | | |
| Peripheral vessel | Peripheral vessel | P | P | P | | P | P | P | Note 1, 2, 4,6,7 | |
| | Other (Specify***) | | | | | | | | | |
| | N=new indication; P=previously cleared by FDA(K150080); E=added under Appendix E | | | | | | | | | |
| | Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW+B. | | | | | | | | | |
| | * Intraoperative includes abdominal, thoracic, and vascular. | | | | | | | | | |
| | **Small organ-breast, thyroid, testes. | | | | | | | | | |
| | ***Other use includes Urology. | | | | | | | | | |
| | Note 1: Tissue Harmonic Imaging. | | | | | | | | | |
| | Note 2: Smart3D | | | | | | | | | |
| | Note 3: 4D(Real-time 3D) | | | | | | | | | |
| Note 4: iScape | | | | | | | | | | |
| Note 5: TDI | | | | | | | | | | |
| Note 6: Color M | | | | | | | | | | |
| Note 7: Biopsy Guidance | | | | | | | | | | |
| Note 8: Elastography | | | | | | | | | | |
| | Note 9: Contrast imaging - liver only | | | | | | | | | |
| | Note10: Contrast imaging - LVO only | | | | | | | | | |
| Concurrence of CDRH, Office of Device Evaluation(ODE) | | | | | | | | | | |
| Clinical Application | | Mode of Operation | | | | | | | | |
| General (Track 1<br>Only) | Specific (Track 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other (specify) | |
| Ophthalmic | Ophthalmic | | | | | | | | | |
| | Fetal | | | | | | | | | |
| | Abdominal | P | P | P | | P | P | P | Note 1,2, 4,6,7 | |
| | Intra-operative (Specify*) | | | | | | | | | |
| | Intra-operative (Neuro) | | | | | | | | | |
| | Laparoscopic | | | | | | | | | |
| | Pediatric | P | P | P | | P | P | P | Note 1,2, 4,6,7 | |
| | Small Organ (Specify**) | P | P | P | | P | P | P | Note 1,2, 4,6,7,8 | |
| Fetal Imaging &<br>Other | Neonatal Cephalic | P | P | P | | P | P | P | Note 1,2, 4,6,7 | |
| | Adult Cephalic…