← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K162288

# SonoSite iViz Ultrasound System (K162288)

_FUJIFILM Sonosite, Inc. · IYN · Sep 8, 2016 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K162288

## Device Facts

- **Applicant:** FUJIFILM Sonosite, Inc.
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Sep 8, 2016
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric, 3rd-Party Reviewed

## Intended Use

The SonoSite iViz Ultrasound System is a general purpose ultrasound system and non-continuous patient monitoring platform intended in clinical care by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Fetal - OB Abdominal Pediatric Small Organ (breast, thyroid, testicles, prostate) Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Cardiac Adult Cardiac Pediatric Peripheral vessel

## Device Story

The SonoSite iViz is a handheld, battery-operated digital ultrasound tablet. It acquires real-time ultrasound data via connected transducers and displays images on a 7-inch LCD screen. The system supports B-mode, M-mode, Color Doppler, and Tissue Harmonic Imaging. It features SonoHD3 speckle reduction for image enhancement. Used in clinical settings by physicians and healthcare professionals, the device aids in diagnostic evaluation and fluid flow analysis. It supports wireless connectivity for image transfer and over-the-air software updates. The output allows clinicians to visualize internal structures and blood flow, facilitating diagnostic decision-making and patient monitoring.

## Clinical Evidence

No clinical studies were required to support the determination of substantial equivalence. The device evaluation relied on bench testing, including electrical, thermal, mechanical, and EMC safety, biocompatibility, and acoustic output measurements.

## Technological Characteristics

Handheld tablet-based ultrasound system; 7-inch 1920x1200 LCD; Android OS. Transducers use Polysulfone UDEL P1700, PVC, and silicone rubber. Imaging modes: B-mode, M-mode, Color Doppler, Tissue Harmonic Imaging, SonoHD3 speckle reduction. Connectivity: 802.11 wireless for image transfer and OTA updates. Battery-powered. Complies with ISO 10993-1, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-37, IEC 62304, IEC 62359, ISO 14971, and NEMA UD 2-2004.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- SonoSite iViz Ultrasound System ([K161119](/device/K161119.md))
- SonoSite Edge Ultrasound System ([K133454](/device/K133454.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three human profiles facing right, stacked on top of each other, with a flowing design below them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 8, 2016

FUJIFILM SonoSite, Inc. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street, NW BUFFALO, MN 55313

Re: K162288

Trade/Device Name: SonoSite iViz Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: August 11, 2016 Received: August 15, 2016

Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)
TBD K162288

Device Name SonoSite iViz Ultrasound System

### Indications for Use (Describe)

The SonoSite iViz Ultrasound System is a general purpose ultrasound system and non-continuous patient monitoring platform intended in clinical care by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include:

Fetal - OB Abdominal Pediatric Small Organ (breast, thyroid, testicles, prostate) Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Cardiac Adult Cardiac Pediatric Peripheral vessel

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Table 1.3-1: Diagnostic Ultrasound Indications for Use Form – FUJIFILM SonoSite iViz Ultrasound System

| System:                                               | FUJIFILM SonoSite iViz Ultrasound System                                              |   |     |     |                  |                     |                  |
|-------------------------------------------------------|---------------------------------------------------------------------------------------|---|-----|-----|------------------|---------------------|------------------|
| Transducer:                                           | N/A                                                                                   |   |     |     |                  |                     |                  |
| Intended Use:                                         | Diagnostic ultrasound imaging or fluid flow analysis of the human<br>body as follows: |   |     |     |                  |                     |                  |
| Clinical Application                                  | Mode of Operation                                                                     |   |     |     |                  |                     |                  |
|                                                       | B                                                                                     | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                                            |                                                                                       |   |     |     |                  |                     |                  |
| Fetal                                                 | P                                                                                     | P |     |     | P                | B+M; B+CD           | 1-3              |
| Abdominal                                             | P                                                                                     | P |     |     | P                | B+M; B+CD           | 1-3              |
| Intra-operative (Abdominal<br>organs and vascular)    |                                                                                       |   |     |     |                  |                     |                  |
| Intra-operative (Neuro.)                              |                                                                                       |   |     |     |                  |                     |                  |
| Laparoscopic                                          |                                                                                       |   |     |     |                  |                     |                  |
| Pediatric                                             | P                                                                                     | P |     |     | P                | B+M; B+CD           | 1-3              |
| Small Organ (breast, thyroid,<br>testicles, prostate) | P                                                                                     | P |     |     | P                | B+M; B+CD           | 1,3              |
| Neonatal Cephalic                                     |                                                                                       |   |     |     |                  |                     |                  |
| Adult Cephalic                                        |                                                                                       |   |     |     |                  |                     |                  |
| Trans-rectal                                          |                                                                                       |   |     |     |                  |                     |                  |
| Trans-vaginal                                         |                                                                                       |   |     |     |                  |                     |                  |
| Trans-urethral                                        |                                                                                       |   |     |     |                  |                     |                  |
| Trans-esoph. (non-Card.)                              |                                                                                       |   |     |     |                  |                     |                  |
| Musculo-skel. (Convent.)                              | P                                                                                     | P |     |     | P                | B+M; B+CD           | 1,3              |
| Musculo-skel. (Superfic.)                             | P                                                                                     | P |     |     | P                | B+M; B+CD           | 1,3              |
| Intra-luminal                                         |                                                                                       |   |     |     |                  |                     |                  |
| Other (spec.)                                         |                                                                                       |   |     |     |                  |                     |                  |
| Cardiac Adult                                         | P                                                                                     | P |     |     | P                | B+M; B+CD           | 1-3              |
| Cardiac Pediatric                                     | P                                                                                     | P |     |     | P                | B+M; B+CD           | 1-3              |
| Trans-esophageal (card.)                              |                                                                                       |   |     |     |                  |                     |                  |
| Other (spec.)                                         |                                                                                       |   |     |     |                  |                     |                  |
| Peripheral vessel                                     | N                                                                                     | N |     |     | N                | B+M; B+CD           | 1,3              |
| Other (spec.)                                         |                                                                                       |   |     |     |                  |                     |                  |

N= new indication; P= previously cleared by FDA; E= added under this appendix

#### Additional Comments:

1: Color Doppler includes Power/Velocity/Variance

2: Tissue Harmonic Imaging (THI)

3: SonoHD3 Imaging (Speckle Reduction)

All items marked "P" were previously cleared in 510(k) K161119.

Prescription Use (Per 21 CFR 801.109)

Page 2 of 4

{4}------------------------------------------------

Table 1.3-2: Diagnostic Ultrasound Indications for Use Form – L38v/10-5 MHz Transducer

| System:                       | FUJIFILM SonoSite iViz Ultrasound System                          |   |     |     |                  |                     |                  |
|-------------------------------|-------------------------------------------------------------------|---|-----|-----|------------------|---------------------|------------------|
| Transducer:                   | L38v/10-5 MHz Transducer                                          |   |     |     |                  |                     |                  |
| Intended Use:                 | Diagnostic ultrasound imaging or fluid flow analysis of the human |   |     |     |                  |                     |                  |
|                               | body as follows:                                                  |   |     |     |                  |                     |                  |
| Clinical Application          | Mode of Operation                                                 |   |     |     |                  |                     |                  |
|                               |                                                                   |   |     |     |                  |                     |                  |
|                               | B                                                                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                    |                                                                   |   |     |     |                  |                     |                  |
| Fetal                         |                                                                   |   |     |     |                  |                     |                  |
| Abdominal                     | P                                                                 | P |     |     | P                | B+M; B+CD           | 1,3              |
| Intra-operative (Abdominal    |                                                                   |   |     |     |                  |                     |                  |
| organs and vascular)          |                                                                   |   |     |     |                  |                     |                  |
| Intra-operative (Neuro.)      |                                                                   |   |     |     |                  |                     |                  |
| Laparoscopic                  |                                                                   |   |     |     |                  |                     |                  |
| Pediatric                     | P                                                                 | P |     |     | P                | B+M; B+CD           | 1,3              |
| Small Organ (breast, thyroid, | P                                                                 | P |     |     | P                | B+M; B+CD           | 1,3              |
| testicles, prostate)          |                                                                   |   |     |     |                  |                     |                  |
| Neonatal Cephalic             |                                                                   |   |     |     |                  |                     |                  |
| Adult Cephalic                |                                                                   |   |     |     |                  |                     |                  |
| Trans-rectal                  |                                                                   |   |     |     |                  |                     |                  |
| Trans-vaginal                 |                                                                   |   |     |     |                  |                     |                  |
| Trans-urethral                |                                                                   |   |     |     |                  |                     |                  |
| Trans-esoph. (non-Card.)      |                                                                   |   |     |     |                  |                     |                  |
| Musculo-skel. (Convent.)      | P                                                                 | P |     |     | P                | B+M; B+CD           | 1,3              |
| Musculo-skel. (Superfic.)     | P                                                                 | P |     |     | P                | B+M; B+CD           | 1,3              |
| Intra-luminal                 |                                                                   |   |     |     |                  |                     |                  |
| Other (spec.)                 |                                                                   |   |     |     |                  |                     |                  |
| Cardiac Adult                 |                                                                   |   |     |     |                  |                     |                  |
| Cardiac Pediatric             |                                                                   |   |     |     |                  |                     |                  |
| Trans-esophageal (card.)      |                                                                   |   |     |     |                  |                     |                  |
| Other (spec.)                 |                                                                   |   |     |     |                  |                     |                  |
| Peripheral vessel             | P                                                                 | P |     |     | P                | B+M; B+CD           | 1,3              |
| Other (snec.)                 |                                                                   |   |     |     |                  |                     |                  |

N= new indication; P= previously cleared by FDA; E= added under this appendix

#### Additional Comments:

1: Color Doppler includes Power/Velocity/Variance

2: Tissue Harmonic Imaging (THI)

3: SonoHD3 Imaging (Speckle Reduction)

All items marked "P" were previously cleared in 510(k) K133454.

Prescription Use (Per 21 CFR 801.109)

Page 3 of 4

{5}------------------------------------------------

Table 1.3-3: Diagnostic Ultrasound Indications for Use Form – P21v/5-1 MHz Transducer

| System:                                               | FUJIFILM SonoSite iViz Ultrasound System                                              |   |     |     |                  |                     |                  |
|-------------------------------------------------------|---------------------------------------------------------------------------------------|---|-----|-----|------------------|---------------------|------------------|
| Transducer:                                           | P21v/5-1 MHz Transducer                                                               |   |     |     |                  |                     |                  |
| Intended Use:                                         | Diagnostic ultrasound imaging or fluid flow analysis of the human<br>body as follows: |   |     |     |                  |                     |                  |
| Clinical Application                                  | B                                                                                     | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic                                            |                                                                                       |   |     |     |                  |                     |                  |
| Fetal                                                 | P                                                                                     | P |     |     | P                | B+M; B+CD           | 1-3              |
| Abdominal                                             | P                                                                                     | P |     |     | P                | B+M; B+CD           | 1-3              |
| Intra-operative (Abdominal<br>organs and vascular)    |                                                                                       |   |     |     |                  |                     |                  |
| Intra-operative (Neuro.)                              |                                                                                       |   |     |     |                  |                     |                  |
| Laparoscopic                                          |                                                                                       |   |     |     |                  |                     |                  |
| Pediatric                                             | P                                                                                     | P |     |     | P                | B+M; B+CD           | 1-3              |
| Small Organ (breast, thyroid,<br>testicles, prostate) |                                                                                       |   |     |     |                  |                     |                  |
| Neonatal Cephalic                                     |                                                                                       |   |     |     |                  |                     |                  |
| Adult Cephalic                                        |                                                                                       |   |     |     |                  |                     |                  |
| Trans-rectal                                          |                                                                                       |   |     |     |                  |                     |                  |
| Trans-vaginal                                         |                                                                                       |   |     |     |                  |                     |                  |
| Trans-urethral                                        |                                                                                       |   |     |     |                  |                     |                  |
| Trans-esoph. (non-Card.)                              |                                                                                       |   |     |     |                  |                     |                  |
| Musculo-skel. (Convent.)                              |                                                                                       |   |     |     |                  |                     |                  |
| Musculo-skel. (Superfic.)                             |                                                                                       |   |     |     |                  |                     |                  |
| Intra-luminal                                         |                                                                                       |   |     |     |                  |                     |                  |
| Other (spec.)                                         |                                                                                       |   |     |     |                  |                     |                  |
| Cardiac Adult                                         | P                                                                                     | P |     |     | P                | B+M; B+CD           | 1-3              |
| Cardiac Pediatric                                     | P                                                                                     | P |     |     | P                | B+M; B+CD           | 1-3              |
| Trans-esophageal (card.)                              |                                                                                       |   |     |     |                  |                     |                  |
| Other (spec.)                                         |                                                                                       |   |     |     |                  |                     |                  |
| Peripheral vessel                                     |                                                                                       |   |     |     |                  |                     |                  |
| Other (spec.)                                         |                                                                                       |   |     |     |                  |                     |                  |

N= new indication; P= previously cleared by FDA; E= added under this appendix

#### Additional Comments:

1: Color Doppler includes Power/Velocity/Variance

2: Tissue Harmonic Imaging (THI)

3: SonoHD3 Imaging (Speckle Reduction)

All items marked "P" were previously cleared in 510(k) K152983.

Prescription Use (Per 21 CFR 801.109)

Page 4 of 4

{6}------------------------------------------------

# 510(K) Summary

This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.

### 1) Submitter's name, address, telephone number, contact person:

FUJIFILM SonoSite, Inc. 21919 30th Drive SE Bothell, WA 98021-3904

| Corresponding Official: | Patricia Liau<br>Manager, Regulatory Affairs |
|-------------------------|----------------------------------------------|
| E-mail:                 | patricia.liau@fujifilm.com                   |
| Telephone:              | (425) 951-6870                               |
| Facsimile:              | (425) 951-1201                               |
| Date prepared:          | August 5, 2016                               |

### 2) Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:

Common/ Usual Name

Diagnostic Ultrasound System with Accessories

Proprietary Name

SonoSite iViz Ultrasound System (subject to change)

Classification Names

| Name                                     | FR Number | Product Code |
|------------------------------------------|-----------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550  | 90-IYN       |
| Ultrasonic Pulsed Echo Imaging System    | 892.1560  | 90-IYO       |
| Diagnostic Ultrasound Transducer         | 892.1570  | 90-ITX       |

#### 3) Identification of the predicate or legally marketed device:

| SonoSite iViz Ultrasound System | K161119 |
|---------------------------------|---------|
| SonoSite Edge Ultrasound System | K133454 |

#### 4) Device Description:

The SonoSite iViz Ultrasound System is a highly featured, general purpose, software controlled, diagnostic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through multiple imaging modes. iViz is a custom fabricated digital electronic handheld tablet that is highly portable, battery-operated, and consists of an active transducer that connects to and is controlled by the tablet. iViz supports Bluetooth and wireless network connectivity for image transfer and over-the-air (OTA) software updates.

{7}------------------------------------------------

### 5) Intended Use:

The SonoSite iViz Ultrasound System is a general purpose ultrasound system and non-continuous patient monitoring platform intended in clinical care by qualified physicians and healthcare professionals for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include:

### Fetal - OB Abdominal Pediatric Small Organ (breast, thyroid, testicles, prostate) Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Cardiac Adult Cardiac Pediatric Peripheral vessel

#### 6) Technological Characteristics:

SonoSite iViz and Edge Ultrasound Systems are Track 3 devices that employ the same fundamental scientific technology. A comparison table is provided below.

| Feature                 | SonoSite iViz Ultrasound                                                                                                                                                                                                     | SonoSite iViz Ultrasound                                                                                                                                                                                | SonoSite Edge                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                         | System<br>(This submission)                                                                                                                                                                                                  | System<br>(K161119)                                                                                                                                                                                     | Ultrasound System<br>(K133454)                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Intended Use            | Diagnostic ultrasound imaging or<br>fluid flow analysis of the human<br>body                                                                                                                                                 | Diagnostic ultrasound imaging or<br>fluid flow analysis of the human<br>body                                                                                                                            | Diagnostic ultrasound imaging or<br>fluid flow analysis of the human<br>body                                                                                                                                                                                                                                                                                                                                                                            |
| Indications for<br>Use  | Fetal - OB<br>Abdominal<br>Pediatric<br>Small Organ (breast, thyroid,<br>testicle, prostate)<br>Musculo-skeletal (Conventional)<br>Musculo-skeletal (Superficial)<br>Cardiac Adult<br>Cardiac Pediatric<br>Peripheral Vessel | Fetal - OB<br>Abdominal<br>Pediatric<br>Small Organ (breast, thyroid,<br>testicle, prostate)<br>Musculo-skeletal (Conventional)<br>Musculo-skeletal (Superficial)<br>Cardiac Adult<br>Cardiac Pediatric | Opthalmic<br>Fetal - OB/GYN<br>Abdominal<br>Intraoperative (abdominal organs<br>and vascular)<br>Intra-operative (Neuro.)<br>Pediatric<br>Small Organ (breast, thyroid,<br>testicle, prostate)<br>Neonatal Cephalic<br>Adult Cephalic<br>Trans-Rectal<br>Trans-Vaginal<br>Musculo-skeletal (Conventional)<br>Musculo-skeletal (Superficial)<br>Cardiac Adult<br>Cardiac Pediatric<br>Trans-esophageal (cardiac)<br>Peripheral Vessel<br>Needle guidance |
| Transducer<br>Types     | Linear Array<br>Phased Array                                                                                                                                                                                                 | Linear Array<br>Phased Array                                                                                                                                                                            | Linear Array<br>Curved Linear Array<br>Intracavitary<br>Phased Array<br>Static Probes<br>Trans-esophageal                                                                                                                                                                                                                                                                                                                                               |
| Transducer<br>Frequency | 1.0 - 10.0 MHz                                                                                                                                                                                                               | 1.0 - 10.0 MHz                                                                                                                                                                                          | 1.0 - 15.0 MHz                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Modes of<br>Operation   | B-mode Grayscale Imaging<br>Tissue Harmonic Imaging<br>M-mode<br>Color M-Mode<br>Color Power Doppler<br>Zoom                                                                                                                 | B-mode Grayscale Imaging<br>Tissue Harmonic Imaging<br>M-mode<br>Color M-Mode<br>Color Power Doppler<br>Zoom                                                                                            | B-mode Grayscale Imaging<br>Tissue Harmonic Imaging<br>M-mode<br>Color M-Mode<br>Color Power Doppler<br>Zoom                                                                                                                                                                                                                                                                                                                                            |

{8}------------------------------------------------

| Feature                         | SonoSite iViz Ultrasound<br>System<br>(This submission)                                                                                                                                                                                                                     | SonoSite iViz Ultrasound<br>System<br>(K161119)                                                                                                                                                                                                                             | SonoSite Edge<br>Ultrasound System<br>(K133454)                                                                                                                                                                                                                                                                                                                                                        |
|---------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                 | Combination Modes<br>SonoHD3 Noise Reduction<br>Velocity Color Doppler                                                                                                                                                                                                      | Combination Modes<br>SonoHD3 Noise Reduction<br>Velocity Color Doppler                                                                                                                                                                                                      | Combination Modes<br>Pulsed Wave (PW) Doppler<br>Continuous Wave (CW) Doppler<br>SonoHD2 Noise Reduction<br>SonoMB/MBe Image<br>Compounding<br>Steered CW Doppler<br>Velocity Color Doppler<br>Tissue Doppler Imaging (TDI)                                                                                                                                                                            |
| PW Doppler                      | Not available                                                                                                                                                                                                                                                               | Not available                                                                                                                                                                                                                                                               | Available                                                                                                                                                                                                                                                                                                                                                                                              |
| CW Doppler                      | Not available                                                                                                                                                                                                                                                               | Not available                                                                                                                                                                                                                                                               | Available                                                                                                                                                                                                                                                                                                                                                                                              |
| Patient<br>Contact<br>Materials | Transducers:<br>Polysulfone UDEL P1700<br>Poly-Vinyl-Chloride (PVC)<br>Silicone Rubber                                                                                                                                                                                      | Transducers:<br>Polysulfone UDEL P1700<br>Poly-Vinyl-Chloride (PVC)<br>Silicone Rubber                                                                                                                                                                                      | Transducers:<br>Acrylonitrile-butadien-styrene<br>(ABS)<br>Cycoloy<br>Dow Medical Adhesive, Type A<br>Epoxy paste adhesive<br>Polyethylene (PE) Ionomer<br>Polyetheretherketone (PEEK)<br>Polysulfone UDEL P1700<br>Polyurethane<br>Poly-Vinyl-Chloride (PVC)<br>Silicone RTV Adhesive<br>Silicone Rubber<br>Urethane<br>Needle Guides:<br>Acetal copolymer<br>Acrylonitrile-butadien-styrene<br>(ABS) |
| System<br>Characteristics       | iViz:<br>Handheld tablet<br>7", 1920 x 1200 pixels LCD<br>Operating system: Android<br>iViz ultrasound software running<br>as an "app" on tablet<br>System operates via battery<br>Wireless 802.11 support for image<br>transfer and over-the-air (OTA)<br>software updates | iViz:<br>Handheld tablet<br>7", 1920 x 1200 pixels LCD<br>Operating system: Android<br>iViz ultrasound software running<br>as an "app" on tablet<br>System operates via battery<br>Wireless 802.11 support for image<br>transfer and over-the-air (OTA)<br>software updates | Edge:<br>Handheld display and control<br>12.1", 800 x 600 pixels, LCD<br>Operating system: Windows CE<br>System operates via battery or AC<br>power<br>Wireless 802.11 support for image<br>transfer                                                                                                                                                                                                   |
| 510(k) Track                    | Track 3                                                                                                                                                                                                                                                                     | Track 3                                                                                                                                                                                                                                                                     | Track 3                                                                                                                                                                                                                                                                                                                                                                                                |

## 7) Determination of Substantial Equivalence:

## Summary of Non-Clinical Tests:

The iViz Ultrasound System has been evaluated for electrical, thermal, mechanical, and EMC safety. Additionally, cleaning/disinfection, biocompatibility, and acoustic output have been evaluated, and the device has been found to conform to applicable mandatory medical device safety standards. Assurance of quality was established by employing the following elements of product development but were not limited to: Design Phase Reviews, Risk Assessment, Requirements Development, and Verification and Validation.

The iViz Ultrasound System is designed to comply with the following FDA recognized standards.

{9}------------------------------------------------

| Reference No.  | Title                                                                                                                                                                                                                                             |
|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993-1    | AAMI / ANSI / ISO 10993-1:2009/(R)2013, Biological evaluation of medical devices -<br>Part 1: Evaluation and testing within a risk management process                                                                                             |
| IEC 60601-1    | AAMI / ANSI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and<br>A2:2010/(R)2012 (Consolidated Text) Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential performance (IEC 60601-1:2005, MOD)      |
| IEC 60601-1-2  | AAMI / ANSI / IEC 60601-1-2:2007(R)2012, Medical electrical equipment - Part 1-2:<br>General requirements for basic safety and essential performance - Collateral<br>standard: Electromagnetic compatibility - Requirements and tests (Edition 3) |
| IEC 60601-1-6  | IEC 60601-1-6 Edition 3.1 2013-10, Medical electrical equipment - Part 1-6: General<br>requirements for basic safety and essential performance - Collateral standard:<br>Usability                                                                |
| IEC 60601-2-37 | IEC 60601-2-37:2007 Edition 2.0 2007-08, Medical electrical equipment - Part 2-37:<br>Particular requirements for the basic safety and essential performance of ultrasonic<br>medical diagnostic and monitoring equipment                         |
| IEC 62304      | AAMI / ANSI / IEC 62304:2006, Medical device software - Software life cycle<br>processes                                                                                                                                                          |
| IEC 62359      | IEC 62359 Edition 2.0 2010-10-10, Ultrasonics - Field characterization - Test<br>methods for the determination of thermal and mechanical indices related to medical<br>diagnostic ultrasonic fields [Including: Technical corrigendum 1 (2011)]   |
| ISO 14971      | ISO 14971:2007, Medical devices - Application of risk management to medical<br>devices                                                                                                                                                            |
| NEMA UD 2-2004 | Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment                                                                                                                                                                          |

### Summary of Clinical Tests:

The iViz Ultrasound System and transducers, subject of this submission, did not require clinical studies to support the determination of substantial equivalence.

#### 8) Conclusion:

Intended uses and other key features are consistent with traditional clinical practice and FDA guidance. The iViz system and predicates meet FDA requirements for Track 3 devices, share indications for use, have biosafety equivalence, and conform to applicable electromedical device safety standards. FUJIFILM SonoSite, Inc. believes that the iViz Ultrasound System is substantially equivalent with regard to safety and effectiveness to the predicate devices.

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K162288](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K162288)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
