← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K162269

# Voluson E6, Voluson E8, Voluson E10 (K162269)

_GE Healthcare · IYN · Oct 3, 2016 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K162269

## Device Facts

- **Applicant:** GE Healthcare
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Oct 3, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Intended Use

Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatric; Small Organ (breast, testes, thyroid etc.); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Musculo-skeletal Conventional and Superficial; Peripheral Vascular; Transvaginal (including GYN); Transrectal

## Device Story

The Voluson E6, E8, and E10 are full-featured, mobile console-based ultrasound systems for general radiology and specialized OB/GYN use. They utilize various linear, curved linear, and matrix phased array transducers to acquire ultrasound signals. The system transforms these signals into real-time 2D, 3D, and 4D images and performs fluid flow analysis. Operators (clinicians) interact with the system via a keyboard, touch panel, and display. The device includes networking and DICOM capabilities for image storage and reporting. It features semi-automated analysis tools, such as SonoVCAD Heart, to assist in cardiac assessment. By providing high-performance imaging and automated measurement tools, the device aids clinicians in diagnostic decision-making, potentially improving diagnostic accuracy and workflow efficiency in obstetric and general imaging contexts.

## Clinical Evidence

No clinical data was required to support substantial equivalence. The device evaluation relied on bench testing, including acoustic output, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, electromagnetic, and mechanical safety testing in accordance with standards such as IEC 60601-1, IEC 60601-2-37, and NEMA UD 2/UD 3.

## Technological Characteristics

Mobile console with keyboard, color LCD/TFT touch panel, and display. Utilizes linear, curved linear, and matrix phased array transducers. Supports B, M, PW/CW Doppler, Color/Power Doppler, 3D/4D imaging, and harmonic imaging. Complies with AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, NEMA UD 2, NEMA UD 3, and ISO 10993-1. Connectivity via DICOM. Software includes semi-automated analysis features.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Voluson E6_E8_E10 Diagnostic Ultrasound System ([K152567](/device/K152567.md))
- LOGIQ E9 Diagnostic Ultrasound System ([K152309](/device/K152309.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized design featuring three human profiles facing to the right, with flowing lines above them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 3, 2016

GE Healthcare % Mr. Bryan Behn Regulatory Affairs Director 9900 Innovation Drive, Milwaukee County Research Park WAUWATOSA WI 53226

Re: K162269

Trade/Device Name: Voluson E6, Voluson E8, Voluson E10 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: September 8, 2016 Received: September 9, 2016

Dear Mr. Behn:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert Ocko

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K162269

Device Name Voluson E6, Voluson E8, Voluson E10

#### Indications for Use (Describe)

The device is a general purpose ultrasound system. Specific clinical applications remain the same as previously cleared: Fetal/OB; Abdominal (including GYN, pelvic and infertility monitoring/follicle development); Pediatic; Small Organ (breast, testes, thyroid etc.); Neonatal and Adult Cephalic; Cardiac (adult and pediatric); Musculo-skeletal Conventional and Superficial; Peripheral Vascular; Transvaginal (including GYN); Transrectal

Type of Use (Select one or both, as applicable)

| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
|-------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label>             |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized script, enclosed within a blue circle. The blue circle has a textured or slightly shaded appearance, giving it a three-dimensional effect.

Indications for Use Forms

The following forms represent indications with clinical applications and exam types along with the modes of operation for the Voluson E Series system and for all of its probe/mode combinations. Combinations identified as "P" represents those previously cleared with another GE Ultrasound system. This modification does not add to indications to the previously cleared system level or transducer indications or clinical applications.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative swirls around the perimeter, giving it a classic and recognizable appearance. The logo is simple, yet distinctive, representing the brand identity of General Electric.

### Diagnostic Ultrasound Indications for Use Form GE Voluson E Series (Voluson E6 / Voluson E8 / Voluson E10)

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                   |                    |                  |                    |                     |                |                  |
|----------------------------------------------------|-------------------|---|---------------|---------------|-------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color#<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic                                         |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Fetal / Obstetrics[7]                              | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Abdominal[1]                                       | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Pediatric                                          | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Small Organ[2]                                     | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Neonatal Cephalic                                  | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5 ]            |
| Adult Cephalic                                     | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              |                  |
| Cardiac[3]                                         | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5 ]            |
| Peripheral Vascular                                | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Musculo-skeletal Conventional                      | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Musculo-skeletal Superficial                       | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Other                                              |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Exam Type, Means of Access                         |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Transesophageal                                    |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Transrectal[8]                                     | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Transvaginal[10]                                   | P                 | P | P             | P             | P                 | P                  | P                | P                  | P                   | P              | [ 5,6,9]         |
| Transuretheral                                     |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Intraoperative                                     |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Intraoperative Neurological                        |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Intravascular                                      |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |
| Laparoscopic                                       |                   |   |               |               |                   |                    |                  |                    |                     |                |                  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

[1] Abdominal includes renal, GYN/Pelvic, Urology Notes:

[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients

[3] Cardiac is Adult and Pediatric

[5] 3D/4D Imaging Mode

[6] Includes imaging of guidance of biopsy

[7] Includes infertility monitoring of follicle development

[8] Includes urology/prostate

[9] Elastography imaging- Elasticity

[10] Transvaginal includes GYN

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

[*] 4D color Doppler (only with eM6C probe)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)

Prescription Use (Per 21 CFR 801.109)

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three swirls around the outside of the circle. The logo is simple and recognizable.

### Diagnostic Ultrasound Indications for Use Form

#### GE Voluson E6 / Voluson E8 / Voluson E10 with RAB2-5-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                               | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |
|-------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Clinical Application          | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Anatomy/Region of Interest    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Ophthalmic                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Fetal / Obstetrics[7]         | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [ 5,6]           |
| Abdominal[1]                  | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [ 5,6]           |
| Pediatric                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Small Organ[2]                |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Neonatal Cephalic             |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Adult Cephalic                |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Cardiac[3]                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Peripheral Vascular           |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Musculo-skeletal Conventional | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [ 5,6]           |
| Musculo-skeletal Superficial  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Other                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Exam Type, Means of Access    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transesophageal               |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transrectal                   |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transvaginal                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transuretheral                |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative                |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative Neurological   |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intravascular                 |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Laparoscopic                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes renal, GYN/Pelvic, Urology

[5] 3D/4D Imaging Mode

[6] Includes imaging of guidance of biopsy (3D/4D)

[7] Includes infertility monitoring of follicle development

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. There are three swirls around the outside of the circle. The logo is simple and recognizable.

### Diagnostic Ultrasound Indications for Use Form

#### GE Voluson E6 / Voluson E8 / Voluson E10 with RIC5-9-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                           | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |
|-----------------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Clinical Application<br><i>Anatomy/Region of Interest</i> | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic                                                |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Fetal / Obstetrics[7]                                     | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [5,6,9]          |
| Abdominal[1]                                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Pediatric                                                 |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Small Organ[2]                                            |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Neonatal Cephalic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Adult Cephalic                                            |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Cardiac[3]                                                |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Peripheral Vascular                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Musculo-skeletal Conventional                             |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Musculo-skeletal Superficial                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Other                                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| <i>Exam Type, Means of Access</i>                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transesophageal                                           |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transrectal[8]                                            | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [5,6,9]          |
| Transvaginal[10]                                          | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [5,6,9]          |
| Transuretheral                                            |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative                                            |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative Neurological                               |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intravascular                                             |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Laparoscopic                                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [5] 3D/4D Imaging Mode

[6] Includes imaging of guidance of biopsy (3D/4D)

[7] Includes infertility monitoring of follicle development

[8] Includes urology/prostate

[9] Elastography Imaging- Elasticity

[10] Transvaginal includes GYN

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative swirls or flourishes around the perimeter, giving it a vintage or classic appearance.

### Diagnostic Ultrasound Indications for Use Form GE Voluson E6 / Voluson E8 / Voluson E10 with 4C-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Fetal / Obstetrics[7]                              | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |
| Abdominal[1]                                       | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |
| Pediatric                                          |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Small Organ[2]                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Cardiac[3]                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Peripheral Vascular                                | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |
| Musculo-skeletal Conventional                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Other                                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes renal, GYN/Pelvic, Urology

[6] Includes imaging of guidance of biopsy (2D)

[7] Includes infertility monitoring of follicle development

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. The circle has decorative swirls or flourishes around the perimeter, giving it a classic or vintage appearance.

### Diagnostic Ultrasound Indications for Use Form GE Voluson E6 / Voluson E8 / Voluson E10 with IC5-9-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                   | Mode of Operation |   |         |         |         |         |         |                    |                     |                |                  |
|-----------------------------------|-------------------|---|---------|---------|---------|---------|---------|--------------------|---------------------|----------------|------------------|
| Clinical Application              | B                 | M | PW      | CW      | Color   | Color M | Power   | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |
| <i>Anatomy/Region of Interest</i> |                   |   | Doppler | Doppler | Doppler | Doppler | Doppler |                    |                     |                |                  |
| Ophthalmic                        |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Fetal / Obstetrics[7]             | P                 | P | P       |         | P       | P       | P       | P                  | P                   | P              | [ 6,9]           |
| Abdominal[1]                      |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Pediatric                         |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Small Organ[2]                    |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Neonatal Cephalic                 |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Adult Cephalic                    |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Cardiac[3]                        |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Peripheral Vascular               |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Musculo-skeletal Conventional     |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Musculo-skeletal Superficial      |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Other                             |                   |   |         |         |         |         |         |                    |                     |                |                  |
| <i>Exam Type, Means of Access</i> |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Transesophageal                   |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Transrectal[8]                    | P                 | P | P       |         | P       | P       | P       | P                  | P                   | P              | [ 6,9]           |
| Transvaginal[10]                  | P                 | P | P       |         | P       | P       | P       | P                  | P                   | P              | [ 6,9]           |
| Transurethral                     |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Intraoperative                    |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Intraoperative Neurological       |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Intravascular                     |                   |   |         |         |         |         |         |                    |                     |                |                  |
| Laparoscopic                      |                   |   |         |         |         |         |         |                    |                     |                |                  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [6] Includes imaging of guidance of biopsy (2D)

[7] Includes infertility monitoring of follicle development

[8] Includes urology/prostate

[9] Elastography Imaging- Elasticity

[10] Transvaginal includes GYN

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The letters are enclosed within a circular border, and the entire logo is presented in a blue color. 

### Diagnostic Ultrasound Indications for Use Form

#### GE Voluson E6 / Voluson E8 / Voluson E10 with RSP6-16-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Fetal / Obstetrics[7]                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Abdominal[1]                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Pediatric                                          | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [ 5,6]           |  |
| Small Organ[2]                                     | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [ 5,6]           |  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Cardiac[3]                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Peripheral Vascular                                | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [ 5,6]           |  |
| Musculo-skeletal Conventional                      | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [ 5,6]           |  |
| Musculo-skeletal Superficial                       | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [ 5,6]           |  |
| Other                                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transrectal[8]                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients Notes:

[5] 3D/4D Imaging Mode

[6] Includes imaging of guidance of biopsy (3D/4D)

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' written in a stylized, cursive font, enclosed within a circular shape. The circle has a decorative, swirling pattern around the letters, and the entire logo is colored in a shade of blue.

### Diagnostic Ultrasound Indications for Use Form

#### GE Voluson E6 / Voluson E8 / Voluson E10 with RIC6-12-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                               | Mode of Operation |   |         |         |         |         |         |        |                   |       |         |  |
|-------------------------------|-------------------|---|---------|---------|---------|---------|---------|--------|-------------------|-------|---------|--|
| Clinical Application          | B                 | M | PW      | CW      | Color   | Color M | Power   |        | Combined Harmonic | Coded | Other   |  |
| Anatomy/Region of Interest    |                   |   | Doppler | Doppler | Doppler | Doppler | Doppler | Modes* | Imaging           | Pulse | [Notes] |  |
| Ophthalmic                    |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Fetal / Obstetrics[7]         | P                 | P | P       |         | P       | P       | P       | P      | P                 | P     | [ 5,6]  |  |
| Abdominal[1]                  |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Pediatric                     |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Small Organ[2]                |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Neonatal Cephalic             |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Adult Cephalic                |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Cardiac[3]                    |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Peripheral Vascular           |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Musculo-skeletal Conventional |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Musculo-skeletal Superficial  |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Other                         |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Exam Type, Means of Access    |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Transesophageal               |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Transrectal[8]                | P                 | P | P       |         | P       | P       | P       | P      | P                 | P     | [ 5,6]  |  |
| Transvaginal[10]              | P                 | P | P       |         | P       | P       | P       | P      | P                 | P     | [ 5,6]  |  |
| Transuretheral                |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Intraoperative                |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Intraoperative Neurological   |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Intravascular                 |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Laparoscopic                  |                   |   |         |         |         |         |         |        |                   |       |         |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [5] 3D/4D Imaging Mode

[6] Includes imaging of guidance of biopsy (3D/4D)

[7] Includes infertility monitoring of follicle development

[8] Includes urology/prostate

[10] Transvaginal includes GYN

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a circular border. The logo is colored in blue, with the letters and border in a lighter shade of blue and the background in a darker shade.

### Diagnostic Ultrasound Indications for Use Form

### GE Voluson E6 / Voluson E8 / Voluson E10 with 9L-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Fetal / Obstetrics[7]                              | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |  |
| Abdominal[1]                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |  |
| Small Organ[2]                                     | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Cardiac[3]                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Peripheral Vascular                                | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |  |
| Musculo-skeletal Conventional                      | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [6]              |  |
| Musculo-skeletal Superficial                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Other                                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

[2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients Notes:

[6] Includes imaging of guidance of biopsy (2D)

[7] Includes infertility monitoring of follicle development

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{12}------------------------------------------------

Image /page/12/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. The circle has decorative swirls or flourishes around the perimeter, giving it a classic or vintage appearance.

### Diagnostic Ultrasound Indications for Use Form GE Voluson E6 / Voluson E8 / Voluson E10 with 11L-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Fetal / Obstetrics[7]                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Abdominal[1]                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Pediatric                                          | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [6,9]            |  |
| Small Organ[2]                                     | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [6,9]            |  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Adult Cephalic                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Cardiac[3]                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Peripheral Vascular                                | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [6,9]            |  |
| Musculo-skeletal Conventional                      | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [6,9]            |  |
| Musculo-skeletal Superficial                       | P                 | P | P             |               | P                | P                  | P                | P                  | P                   | P              | [6,9]            |  |
| Other                                              |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Exam Type, Means of Access                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transesophageal                                    |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transrectal                                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transvaginal                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Transuretheral                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intraoperative Neurological                        |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Intravascular                                      |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Laparoscopic                                       |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients

[6] Includes imaging of guidance of biopsy (2D)

[9] Elastography Imaging- Elasticity

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{13}------------------------------------------------

Image /page/13/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" written in a stylized script, enclosed within a blue circle. The circle has decorative swirls or flourishes around the perimeter, giving it a vintage or classic appearance.

### Diagnostic Ultrasound Indications for Use Form GE Voluson E6 / Voluson E8 / Voluson E10 with C1-5-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                               | Mode of Operation |   |         |         |         |         |         |          |          |       |         |  |
|-------------------------------|-------------------|---|---------|---------|---------|---------|---------|----------|----------|-------|---------|--|
| Clinical Application          | B                 | M | PW      | CW      | Color   | Color M | Power   | Combined | Harmonic | Coded | Other   |  |
| Anatomy/Region of Interest    |                   |   | Doppler | Doppler | Doppler | Doppler | Doppler | Modes*   | Imaging  | Pulse | [Notes] |  |
| Ophthalmic                    |                   |   |         |         |         |         |         |          |          |       |         |  |
| Fetal / Obstetrics[7]         | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [6]     |  |
| Abdominal[1]                  | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [6]     |  |
| Pediatric                     |                   |   |         |         |         |         |         |          |          |       |         |  |
| Small Organ[2]                |                   |   |         |         |         |         |         |          |          |       |         |  |
| Neonatal Cephalic             |                   |   |         |         |         |         |         |          |          |       |         |  |
| Adult Cephalic                |                   |   |         |         |         |         |         |          |          |       |         |  |
| Cardiac[3]                    |                   |   |         |         |         |         |         |          |          |       |         |  |
| Peripheral Vascular           | P                 | P | P       | P       | P       | P       | P       | P        | P        | P     | [6]     |  |
| Musculo-skeletal Conventional |                   |   |         |         |         |         |         |          |          |       |         |  |
| Musculo-skeletal Superficial  |                   |   |         |         |         |         |         |          |          |       |         |  |
| Other                         |                   |   |         |         |         |         |         |          |          |       |         |  |
| Exam Type, Means of Access    |                   |   |         |         |         |         |         |          |          |       |         |  |
| Transesophageal               |                   |   |         |         |         |         |         |          |          |       |         |  |
| Transrectal                   |                   |   |         |         |         |         |         |          |          |       |         |  |
| Transvaginal                  |                   |   |         |         |         |         |         |          |          |       |         |  |
| Transuretheral                |                   |   |         |         |         |         |         |          |          |       |         |  |
| Intraoperative                |                   |   |         |         |         |         |         |          |          |       |         |  |
| Intraoperative Neurological   |                   |   |         |         |         |         |         |          |          |       |         |  |
| Intravascular                 |                   |   |         |         |         |         |         |          |          |       |         |  |
| Laparoscopic                  |                   |   |         |         |         |         |         |          |          |       |         |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [1] Abdominal includes renal, GYN/Pelvic, Urology

[6] Includes imaging of guidance of biopsy (2D)

[7] Includes infertility monitoring of follicle development

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{14}------------------------------------------------

Image /page/14/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. The circle has decorative swirls or flourishes around the perimeter, giving it a classic or vintage appearance.

### Diagnostic Ultrasound Indications for Use Form GE Voluson E6 / Voluson E8 / Voluson E10 with ML6-15-D Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                               | Mode of Operation |   |         |         |         |         |         |        |                   |       |         |  |
|-------------------------------|-------------------|---|---------|---------|---------|---------|---------|--------|-------------------|-------|---------|--|
| Clinical Application          | B                 | M | PW      | CW      | Color   | Color M | Power   |        | Combined Harmonic | Coded | Other   |  |
| Anatomy/Region of Interest    |                   |   | Doppler | Doppler | Doppler | Doppler | Doppler | Modes* | Imaging           | Pulse | [Notes] |  |
| Ophthalmic                    |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Fetal / Obstetrics[7]         |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Abdominal[1]                  |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Pediatric                     | P                 | P | P       | P       | P       | P       | P       | P      | P                 | P     | [6,9]   |  |
| Small Organ[2]                | P                 | P | P       | P       | P       | P       | P       | P      | P                 | P     | [ 6,9]  |  |
| Neonatal Cephalic             |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Adult Cephalic                |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Cardiac[3]                    |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Peripheral Vascular           | P                 | P | P       | P       | P       | P       | P       | P      | P                 | P     | [6,9]   |  |
| Musculo-skeletal Conventional | P                 | P | P       | P       | P       | P       | P       | P      | P                 | P     | [ 6,9]  |  |
| Musculo-skeletal Superficial  | P                 | P | P       | P       | P       | P       | P       | P      | P                 | P     | [ 6,9]  |  |
| Other                         |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Exam Type, Means of Access    |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Transesophageal               |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Transrectal                   |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Transvaginal                  |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Transuretheral                |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Intraoperative                |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Intraoperative Neurological   |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Intravascular                 |                   |   |         |         |         |         |         |        |                   |       |         |  |
| Laparoscopic                  |                   |   |         |         |         |         |         |        |                   |       |         |  |

N = new indication; P = previously cleared by FDA; E = added under Appendix E

Notes: [2] Small organ includes breast, testes, thyroid, salivary gland, lymph nodes, pediatric and neonatal patients

[6] Includes imaging of guidance of biopsy (2D)

[9] Elastography Imaging-Elasticity

[*] Combined modes are B/M, B/Color M, B/PWD or CWD, B/Color/PWD or CWD, B/Power/PWD

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

{15}------------------------------------------------

Image /page/15/Picture/0 description: The image shows the logo for General Electric (GE). The logo consists of the letters "GE" in a stylized script, enclosed within a blue circle. The circle has decorative swirls or flourishes around the perimeter, giving it a classic or vintage appearance.

### Diagnostic Ultrasound Indications for Use Form GE Voluson E6 / Voluson E8 / Voluson E10 with RM6C Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

|                                                    | Mode of Operation |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
|----------------------------------------------------|-------------------|---|---------------|---------------|------------------|--------------------|------------------|--------------------|---------------------|----------------|------------------|--|
| Clinical Application<br>Anatomy/Region of Interest | B                 | M | PW<br>Doppler | CW<br>Doppler | Color<br>Doppler | Color M<br>Doppler | Power<br>Doppler | Combined<br>Modes* | Harmonic<br>Imaging | Coded<br>Pulse | Other<br>[Notes] |  |
| Ophthalmic                                         |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Fetal / Obstetrics[7]                              | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [5,6]            |  |
| Abdominal[1]                                       | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [5,6]            |  |
| Pediatric                                          | P                 | P | P             | P             | P                | P                  | P                | P                  | P                   | P              | [5,6]            |  |
| Small Organ[2]                                     |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Neonatal Cephalic                                  |                   |   |               |               |                  |                    |                  |                    |                     |                |                  |  |
| Adult Cephalic                                     |                   |   |               |               |…

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K162269](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K162269)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
