← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K160283

# S45 Digital Color Doppler Ultrasound System (K160283)

_Sonoscape Medical Corp. · IYN · Apr 4, 2016 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K160283

## Device Facts

- **Applicant:** Sonoscape Medical Corp.
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Apr 4, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The SonoScape S45 system is a general-purpose ultrasonic intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Transrectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

## Device Story

Mobile, software-controlled, color diagnostic ultrasound system; transmits ultrasonic energy into patient body; implements post-processing of received echoes; generates real-time display of anatomic structures and fluid flow. Used in clinical settings by qualified physicians. Displays B-Mode (including Tissue Harmonic Imaging), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continuous Wave Doppler, Power Doppler, Elastography, and 3D/4D imaging. Output viewed on integrated color LCD monitor; assists physicians in diagnostic evaluation of various body systems; provides non-invasive diagnostic imaging benefits.

## Clinical Evidence

No clinical data. Bench testing only, including electrical, mechanical, thermal, and electromagnetic compatibility safety, biocompatibility, and acoustic output measurements. Conformance to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-5, ISO 10993-10, and AIUM/NEMA UD 2/UD 3 standards verified.

## Technological Characteristics

Mobile console with touch screen, keyboard, power supply, and color LCD monitor. Supports linear, convex, and phased array transducers. Modes: B, M, TDI, Color-Flow, PWD, CWD, Power Doppler, Elastography, 3D/4D. Software-controlled. Biocompatibility per ISO 10993. Acoustic output per AIUM/NEMA UD 2/3.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- SonoScape S50 Digital Color Doppler Ultrasound System ([K152854](/device/K152854.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing to the right. The profiles are connected and form a single, flowing shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 4, 2016

SONOSCAPE MEDICAL CORP. Toki Wu 4/f, 5/f, 8/f, 9/f & 10/f, Yizhe Building, Yuquan Road Nanshan, Shenzhen 518051, Guangdong CHINA

Re: K160283 Trade/Device Name: S45 Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: January 22, 2016 Received: February 1, 2016

Dear Ms. Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Michael D'Hara

For

Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known) K160283

#### Device Name

S45 Digital Color Doppler Ultrasound System

Indications for Use (Describe)

The SonoScape S45 system is a general-purpose ultrasonic intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Transrectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (pediatric and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

| Type of Use (Select one or both, as applicable)                                                        |                                             |
|--------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <div><span style="font-family: sans-serif;"> </span>Prescription Use (Part 21 CFR 801 Subpart D)</div> | Over-The-Counter Use (21 CFR 801 Subpart C) |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services
  
Food and Drug Administration
  
Office of Chief Information Officer
  
Paperwork Reduction Act (PRA) Staff
  
PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

SonoScape S45 System:

Diagnostic Ultrasound Pulsed Echo System

Diagnostic Ultrasound Pulsed Doppler Imaging System

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    |   | Mode of Operation |     |     |                  |                                 |                    |                    |
|------------------------------|------------------------------------|---|-------------------|-----|-----|------------------|---------------------------------|--------------------|--------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B | M                 | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify  |
| Ophthalmic                   | Ophthalmic                         |   |                   |     |     |                  |                                 |                    |                    |
|                              | Fetal                              | N | N                 | N   |     | N                | N                               | Note 1             | Notes 2,4,5        |
|                              | Abdominal                          | N | N                 | N   |     | N                | N                               | Note 1             | Notes 2,4,5        |
|                              | Intra-operative Specify            |   |                   |     |     |                  |                                 |                    |                    |
|                              | Intra-operative Neuro              |   |                   |     |     |                  |                                 |                    |                    |
|                              | Laparoscopic                       | N | N                 | N   |     | N                | N                               | Note 1             | Notes 2,4          |
|                              | Pediatric                          | N | N                 | N   |     | N                | N                               | Note 1             | Notes 2,4          |
|                              | Small Organ (specify)              | N | N                 | N   |     | N                | N                               | Note 1             | Notes<br>2,4,5,6,7 |
|                              | Neonatal Cephalic                  | N | N                 | N   | N   | N                | N                               | Note 1             | Notes 2,3,4        |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     | N | N                 | N   | N   | N                | N                               | Note 1             | Notes 2,3,4        |
|                              | Trans-rectal                       | N | N                 | N   |     | N                | N                               | Note 1             | Notes 2,4,5        |
|                              | Trans-vaginal                      | N | N                 | N   |     | N                | N                               | Note 1             | Notes 2,4,5        |
|                              | Trans-urethral                     |   |                   |     |     |                  |                                 |                    |                    |
|                              | Trans-esoph.(non-Card)             |   |                   |     |     |                  |                                 |                    |                    |
|                              | Musculo-skeletal<br>(Conventional) | N | N                 | N   |     | N                | N                               | Note 1             | Notes 2,4          |
|                              | Musculo-skeletal<br>(Superficial)  | N | N                 | N   |     | N                | N                               | Note 1             | Notes 2,4          |
|                              | Intravascular                      |   |                   |     |     |                  |                                 |                    |                    |
|                              | Other (Ob/GYN)                     | N | N                 | N   |     | N                | N                               | Note 1             | Notes 2,4,5        |
|                              | Other (Urology)                    | N | N                 | N   |     | N                | N                               | Note 1             | Notes 2,4,5        |
|                              | Cardiac Adult                      | N | N                 | N   | N   | N                | N                               | Note 1             | Notes 2,3,4        |
|                              | Cardiac Pediatric                  | N | N                 | N   | N   | N                | N                               | Note 1             | Notes 2,3,4        |
| Cardiac                      | Intravascular(Cardiac)             |   |                   |     |     |                  |                                 |                    |                    |
|                              | Trans-esoph. (Cardiac)             | N | N                 | N   | N   | N                | N                               | Note 1             | Notes 2,3,4        |
|                              | Intra-cardiac                      |   |                   |     |     |                  |                                 |                    |                    |
|                              | Other (specify)                    |   |                   |     |     |                  |                                 |                    |                    |
| Peripheral<br>Vessel         | Peripheral vessel                  | N | N                 | N   |     | N                | N                               | Note 1             | Notes 2,4          |
|                              | Cerebral vascular                  |   |                   | N   |     |                  |                                 |                    |                    |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{4}------------------------------------------------

### Transducer: C322 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |
|                              | Fetal                              | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Abdominal                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Pediatric                          |                   |   |     |     |                  |                                 |                    |                   |
|                              | Small Organ (specify)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Neonatal Cephalic                  |                   |   |     |     |                  |                                 |                    |                   |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph.(non-Card)             |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Ob/GYN)                     | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Adult                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Pediatric                  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |
| Cardiac                      | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cerebral vascular                  |                   |   |     |     |                  |                                 |                    |                   |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{5}------------------------------------------------

### Transducer: C542 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |
|                              | Fetal                              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Abdominal                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Pediatric                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Small Organ (specify)              |                   |   |     |     |                  |                                 |                    |                   |
| Fetal<br>Imaging&<br>Other   | Neonatal Cephalic                  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Adult Cephalic                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph. (non-Card)            |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Ob/GYN)                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |
| Cardiac                      | Cardiac Adult                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Pediatric                  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cerebral vascular                  |                   |   |     |     |                  |                                 |                    |                   |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{6}------------------------------------------------

### Transducer: 3C-A Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         | Mode of Operation                  |   |   |     |     |                  |                                 |                    |                   |
|------------------------------|------------------------------------|---|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic                   | Ophthalmic                         |   |   |     |     |                  |                                 |                    |                   |
|                              | Fetal                              | P | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Abdominal                          | P | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Intra-operative Specify            |   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-operative Neuro              |   |   |     |     |                  |                                 |                    |                   |
|                              | Laparoscopic                       |   |   |     |     |                  |                                 |                    |                   |
|                              | Pediatric                          |   |   |     |     |                  |                                 |                    |                   |
|                              | Small Organ (specify)              |   |   |     |     |                  |                                 |                    |                   |
|                              | Neonatal Cephalic                  |   |   |     |     |                  |                                 |                    |                   |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     |   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-rectal                       |   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-vaginal                      |   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-urethral                     |   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph. (non-Card)            |   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Conventional) |   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Superficial)  |   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular                      |   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Ob/GYN)                     | P | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Other (Urology)                    |   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Adult                      |   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Pediatric                  |   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular(Cardiac)             |   |   |     |     |                  |                                 |                    |                   |
| Cardiac                      | Trans-esoph.(Cardiac)              |   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-cardiac                      |   |   |     |     |                  |                                 |                    |                   |
|                              | Other (specify)                    |   |   |     |     |                  |                                 |                    |                   |
| Peripheral                   | Peripheral vessel                  |   |   |     |     |                  |                                 |                    |                   |
| Vessel                       | Cerebral vascular                  |   |   |     |     |                  |                                 |                    |                   |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{7}------------------------------------------------

### Transducer: C1-6 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Mode of Operation                                    |                                    |                   |   |     |     |                  |                                 |                    |                   |
|------------------------------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| Clinical Application<br>General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic                                           | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Fetal                              | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                                                      | Abdominal                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                                                      | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Pediatric                          |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Small Organ (specify)              |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Neonatal Cephalic                  |                   |   |     |     |                  |                                 |                    |                   |
| Fetal<br>Imaging&<br>Other                           | Adult Cephalic                     |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Trans-esoph. (non-Card)            |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Other (Ob/GYN)                     | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                                                      | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Cardiac Adult                      |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Cardiac Pediatric                  |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |
| Cardiac                                              | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |
|                                                      | Peripheral<br>Vessel               | Peripheral vessel |   |     |     |                  |                                 |                    |                   |
|                                                      |                                    | Cerebral vascular |   |     |     |                  |                                 |                    |                   |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THI; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{8}------------------------------------------------

#### Transducer: VC6-2 Curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |  |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|--|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |  |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Fetal                              | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4,5       |  |
|                              | Abdominal                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4,5       |  |
|                              | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Pediatric                          |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Small Organ (specify)              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Neonatal Cephalic                  |                   |   |     |     |                  |                                 |                    |                   |  |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-esoph.(non-Card)             |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Other (Ob/GYN)                     | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4,5       |  |
|                              | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Cardiac Adult                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Cardiac Pediatric                  |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |  |
| Cardiac                      | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |  |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Cerebral vascular                  |                   |   |     |     |                  |                                 |                    |                   |  |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D Note 5: 4D

Note 6: Small Organ: breast, thyroid, testes

{9}------------------------------------------------

#### Transducer: C613 Micro-curved Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |  |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|--|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |  |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Fetal                              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Abdominal                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |  |
|                              | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Pediatric                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |  |
|                              | Small Organ (specify)              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Neonatal Cephalic                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |  |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-esoph.(non-Card)             |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Other (Ob/GYN)                     |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Cardiac Adult                      |                   |   |     |     |                  |                                 |                    |                   |  |
| Cardiac                      | Cardiac Pediatric                  | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,3,4       |  |
|                              | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |  |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Cerebral vascular                  |                   |   |     |     |                  |                                 |                    |                   |  |

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{10}------------------------------------------------

#### Transducer: 5P2 Phased Array

Diagnostic Ultrasound Transducer

#### Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |
|                              | Fetal                              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Abdominal                          |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Pediatric                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Small Organ (specify)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Neonatal Cephalic                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph.(non-Card)             |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Ob/GYN)                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Adult                      |                   |   |     |     |                  |                                 |                    |                   |
| Cardiac                      | Cardiac Pediatric                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
|                              | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cerebral vascular                  |                   |   |     |     |                  |                                 |                    |                   |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 5: 4D

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{11}------------------------------------------------

#### Transducer: 8P1 Phased Array

#### Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |  |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|--|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |  |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Fetal                              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Abdominal                          |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Pediatric                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |  |
|                              | Small Organ (specify)              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Neonatal Cephalic                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |  |
| Fetal<br>Imaging&<br>Other   | Adult Cephalic                     |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-esoph.(non-Card)             |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Other (Ob/GYN)                     |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Cardiac Adult                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Cardiac Pediatric                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |  |
|                              | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |  |
| Cardiac                      | Trans-esoph.(Cardiac)              |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |  |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                                 |                    |                   |  |
|                              | Cerebral vascular                  |                   |   |     |     |                  |                                 |                    |                   |  |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD; B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

Note 7: Elastography

Note 5: 4D

{12}------------------------------------------------

#### Transducer: 4P-A Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |
|------------------------------|------------------------------------|-------------------|---|-----|-----|------------------|---------------------------------|--------------------|-------------------|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B                 | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |
| Ophthalmic                   | Ophthalmic                         |                   |   |     |     |                  |                                 |                    |                   |
| Fetal<br>Imaging&<br>Other   | Fetal                              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Abdominal                          | P                 | P | P   |     | P                | P                               | Note 1             | Notes 2,4         |
|                              | Intra-operative Specify            |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-operative Neuro              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Laparoscopic                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Pediatric                          |                   |   |     |     |                  |                                 |                    |                   |
|                              | Small Organ (specify)              |                   |   |     |     |                  |                                 |                    |                   |
|                              | Neonatal Cephalic                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
|                              | Adult Cephalic                     | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
|                              | Trans-rectal                       |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-vaginal                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-urethral                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph. (non-Card)            |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                                 |                    |                   |
|                              | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intravascular                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Ob/GYN)                     |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (Urology)                    |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cardiac Adult                      | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
|                              | Cardiac Pediatric                  | P                 | P | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |
| Cardiac                      | Intravascular(Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |
|                              | Trans-esoph. (Cardiac)             |                   |   |     |     |                  |                                 |                    |                   |
|                              | Intra-cardiac                      |                   |   |     |     |                  |                                 |                    |                   |
|                              | Other (specify)                    |                   |   |     |     |                  |                                 |                    |                   |
| Peripheral<br>Vessel         | Peripheral vessel                  |                   |   |     |     |                  |                                 |                    |                   |
|                              | Cerebral vascular                  |                   |   |     |     |                  |                                 |                    |                   |

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{13}------------------------------------------------

### Transducer: S1-5 Phased Array

Diagnostic Ultrasound Transducer

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    |   | Mode of Operation |     |     |                  |                                 |                    |                   |  |
|------------------------------|------------------------------------|---|-------------------|-----|-----|------------------|---------------------------------|--------------------|-------------------|--|
| General<br>(TRACK 1<br>ONLY) | Specific<br>(TRACKS 1 & 3)         | B | M                 | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Other*<br>Combined | Other*<br>Specify |  |
| Ophthalmic                   | Ophthalmic                         |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Fetal                              |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Abdominal                          | P | P                 | P   |     | P                | P                               | Note 1             | Notes 2,4         |  |
|                              | Intra-operative Specify            |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-operative Neuro              |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Laparoscopic                       |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Pediatric                          |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Small Organ (specify)              |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Neonatal Cephalic                  | P | P                 | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |  |
| Fetal                        | Adult Cephalic                     | P | P                 | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |  |
| Imaging&                     | Trans-rectal                       |   |                   |     |     |                  |                                 |                    |                   |  |
| Other                        | Trans-vaginal                      |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-urethral                     |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Trans-esoph.(non-Card)             |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Musculo-skeletal<br>(Conventional) |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Musculo-skeletal<br>(Superficial)  |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Intravascular                      |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Other (Ob/GYN)                     |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Other (Urology)                    |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Cardiac Adult                      | P | P                 | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |  |
|                              | Cardiac Pediatric                  | P | P                 | P   | P   | P                | P                               | Note 1             | Notes 2,3,4       |  |
|                              | Intravascular(Cardiac)             |   |                   |     |     |                  |                                 |                    |                   |  |
| Cardiac                      | Trans-esoph. (Cardiac)             |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Intra-cardiac                      |   |                   |     |     |                  |                                 |                    |                   |  |
|                              | Other (specify)                    |   |                   |     |     |                  |                                 |                    |                   |  |
| Peripheral                   | Peripheral vessel                  |   |                   |     |     |                  |                                 |                    |                   |  |
| Vessel                       | Cerebral vascular                  |   |                   |     |     |                  |                                 |                    |                   |  |

N = new indication; P = previously cleared by FDA; E = added under this appendix Note 1: Other Combined includes: B/M; B/PWD; B/THl; M/Color Doppler; B/Color Doppler/PWD;

B/Power Doppler/PWD

Note 2: Tissue Harmonic Imaging. The feature does not use contrast agents Note 5: 4D

Note 3: TDI Note 4: 3D

Note 6: Small Organ: breast, thyroid, testes

{14}------------------------------------------------

#### Transducer: 7P-A Phased Array

Diagnostic Ultrasound Transducer

#### Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application         |                                    | Mode of Operation |   |     |     |                  |                                 |                    |                   |
|--------------------------…

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K160283](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K160283)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
