Resona 7/Resona 7CV/Resona 7EXP/Resona 7S/Resona 7OB Diagnostic Ultrasound System

K152763 · Shenzhen Mindray Bio-Medical Electronics Co., Ltd. · IYN · Oct 21, 2015 · Radiology

Device Facts

Record IDK152763
Device NameResona 7/Resona 7CV/Resona 7EXP/Resona 7S/Resona 7OB Diagnostic Ultrasound System
ApplicantShenzhen Mindray Bio-Medical Electronics Co., Ltd.
Product CodeIYN · Radiology
Decision DateOct 21, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Intended Use

The Resona 7/Resona 7CV/Resona 7EXP/Resona 7S/Resona 70B diagnostic ultrasound system is applicable for adults, pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), adult cephalic, trans-rectal, trans-vaginal, musculo-skeletal (conventional, superficial), cardiac adult, cardiac pediatric, peripheral vessel and urology exams.

Device Story

General-purpose, mobile, software-controlled diagnostic ultrasound system; acquires and displays ultrasound images. Inputs: ultrasonic echoes received via linear, convex, or phased array probes (3-13.5 MHz). Processing: system transmits ultrasonic energy into patient; performs post-processing of received echoes to generate anatomic and fluid flow displays. Outputs: B-mode, M-mode, PW/CW Doppler, Color/Power Doppler, TDI, 3D/4D, Color M-mode, iScape, Elastography, LVQ, Ultrasound Fusion Imaging, and V Flow. Used in clinical settings (e.g., hospitals, clinics) by trained healthcare professionals (physicians, sonographers). Output viewed on system display for diagnostic assessment; assists in clinical decision-making by providing anatomical and physiological visualization. Benefits: non-invasive diagnostic imaging for various patient populations and clinical applications.

Clinical Evidence

No clinical data. Substantial equivalence based on non-clinical bench testing, including acoustic output measurements, biocompatibility, cleaning/disinfection effectiveness, and thermal, electrical, and mechanical safety testing in compliance with recognized standards (AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, UD 2, UD 3).

Technological Characteristics

Mobile diagnostic ultrasound system; linear, convex, and phased array transducers (3-13.5 MHz). Imaging modes: B, M, PW, CW, Color, Power/Dirpower, TDI, 3D/4D, Color M, iScape, Elastography, LVQ, Fusion, V Flow. Compliance: AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-37, UD 2, UD 3. Biocompatibility: ISO 10993-1. Software: IEC 62304. Risk management: ISO 14971.

Indications for Use

Indicated for adults, pregnant women, pediatric patients, and neonates for fetal, abdominal, pediatric, small organ (breast, thyroid, testes), adult cephalic, trans-rectal, trans-vaginal, musculoskeletal (conventional, superficial), cardiac (adult/pediatric), peripheral vessel, and urology ultrasound exams.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a wavy line below them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street, NW BUFFALO MN 55313 October 21, 2015 Re: K152763 Trade/Device Name: Resona 7CV/Resona 7CV/Resona 7S/Resona 7S/Resona 7OB Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: September 23, 2015 Received: September 24, 2015 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Robert Ocks Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. #### 510(k) Number (if known) # K152763 Device Name Resona 7/Resona 7CV/Resona 7EXP/Resona 7S/Resona 70B Indications for Use (Describe) The Resona 7/Resona 7CV/Resona 7S/Resona 7S/Resona 7OB diagnostic ultrasound system is applicable for adults. pregnant women, pediatric patients and neonates. It is intended for use in fetal, abdominal, pediatric, small organ (breast, thyroid, testes), adult cephalic, trans-vaginal,musculo-skeletal (conventional, superficial), cardiac adult, cardiac pediatric, peripheral vessel and urology exams. Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ | System: | Resona 7/Resona 7CV/Resona 7EXP/Resona 7S/Resona 7OB Diagnostic Ultrasound System | | | | | | | | | | | |--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-------------------|---|-----|-----|------------------|----------------------|-----------------------|-----------------------|--------------------------|--| | Transducer: | N/A | | | | | | | | | | | | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | | | | | Clinical Application | Mode of Operation | | | | | | | | | | | | General<br>(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | | Other<br>(Specify) | | | Ophthalmic | Ophthalmic | | | | | | | | | | | | Fetal Imaging<br>& Other | Fetal | N | N | N | N | N | N | N | | Note 1,2,3,4,6,7,10 | | | | Abdominal | N | N | N | N | N | N | N | | Note 1,2,3,4,5,6,7,10 | | | | Intra-operative<br>(Specify*) | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Pediatric | N | N | N | | N | N | N | | Note 1,2,4,7 | | | | Small Organ (Specify**) | N | N | N | | N | N | N | | Note 1,2,4,7,8 | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | N | N | N | N | N | N | N | | Note 1, 2, 4, 5, 6, 7,10 | | | | Trans-rectal | N | N | N | N | N | N | N | | Note 1, 2, 3, 4, 6, 7 | | | | Trans-vaginal | N | N | N | N | N | N | N | | Note 1, 2, 3, 4, 6, 7 | | | | Trans-urethral | | | | | | | | | | | | | Trans-esoph. (non-<br>Card.) | | | | | | | | | | | | | Musculo-skeletal<br>(Conventional) | N | N | N | | N | N | N | | Note 1, 2, 4, 6, 7,10 | | | | Musculo-skeletal<br>(Superficial) | N | N | N | | N | N | N | | Note 1, 2, 4, 7 | | | | Intravascular | | | | | | | | | | | | Cardiac | Cardiac Adult | N | N | N | N | N | N | N | | Note 1,2,4,5,6,7,9,10 | | | | Cardiac Pediatric | N | N | N | N | N | N | N | | Note 1, 2, 4, 5, 6, 7,10 | | | | Intravascular (Cardiac) | | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | Peripheral<br>vessel | Peripheral vessel | N | N | N | | N | N | N | | Note 1, 2, 4, 6, 7,10,11 | | | | Other (Specify***) | N | N | N | | N | N | N | | Note 1, 2, 4, 6, 7 | | | | N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | | | | | | Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW +Color+ B、Power + PW +B. | | | | | | | | | | | | | *Intraoperative includes abdominal, thoracic, and vascular. | | | | | | | | | | | | | **Small organ-breast, thyroid, testes. | | | | | | | | | | | | | ***Other use includes Urology. | | | | | | | | | | | | | Note 1: Tissue Harmonic Imaging. | | | | | | | | | | | | | Note 2: Smart3D | | | | | | | | | | | | | Note 3:Static 3D/4D | | | | | | | | | | | | | Note 4: iScape | | | | | | | | | | | | | Note 5: TDI | | | | | | | | | | | | | Note 6: Color M | | | | | | | | | | | | | Note 7: Biopsy Guidance | | | | | | | | | | | | | Note 8: Elastography | | | | | | | | | | | | | Note 9: Contrast imaging (Contrast agent for LVO) | | | | | | | | | | | | | Note 10: Ultrasound Fusion Imaging | | | | | | | | | | | | | Note 11: V Flow | | | | | | | | | | | | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | | | | | | | | | | | | | Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) | | | | | | | | | | | | | | | | | | | | | | | | | Transducer: | C5-1U | | | | | | | | | | | | | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | | | | Clinical Application | | Mode of Operation | | | | | | | | | | | General<br>(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other<br>(Specify) | | | | Ophthalmic | Ophthalmic | | | | | | | | | | | | Fetal Imaging<br>& Other | Fetal | N | N | N | | N | N | N | Note 1, 2, 4, 6, 7,10 | | | | | Abdominal | N | N | N | | N | N | N | Note 1, 2, 4, 6, 7,10 | | | | | Intra-operative<br>(Specify*) | | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | Small Organ (Specify**) | | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | | Trans-esoph. (non-<br>Card.) | | | | | | | | | | | | | Musculo-skeletal<br>(Conventional) | N | N | N | | N | N | N | Note 1, 2, 4, 6, 7,10 | | | | | Musculo-skeletal<br>(Superficial) | | | | | | | | | | | | | Intravascular | | | | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | | Peripheral<br>vessel | Peripheral vessel | N | N | N | | N | N | N | Note 1, 2, 4, 6, 7,10 | | | | | Other (Specify***) | | | | | | | | | | | | | N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | | | | | | Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW +B. | | | | | | | | | | | | | *Intraoperative includes abdominal, thoracic, and vascular. | | | | | | | | | | | | | **Small organ-breast, thyroid, testes. | | | | | | | | | | | | | ***Other use includes Urology. | | | | | | | | | | | | | Note 1: Tissue Harmonic Imaging. | | | | | | | | | | | | | Note 2: Smart3D | | | | | | | | | | | | | Note 3:4D(Real-time 3D) | | | | | | | | | | | | | Note 4: iScape | | | | | | | | | | | | | Note 5: TDI | | | | | | | | | | | | | Note 6: Color M | | | | | | | | | | | | | Note 7: Biopsy Guidance | | | | | | | | | | | | | Note 8: Elastography | | | | | | | | | | | | | Note 9: Contrast imaging (Contrast agent for LVO) | | | | | | | | | | | | | Note 10: Ultrasound Fusion Imaging | | | | | | | | | | | | | Note 11: V Flow | | | | | | | | | | | | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | | | | | | | | | | | | | Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIP) | | | | | | | | | | | Diagnostic Ultrasound Indications For Use Format {4}------------------------------------------------ {5}------------------------------------------------ | Transducer: | SC8-2U | | | | | | | | | | |--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|-------------------|-------------------|-----|-----|--------------------------|----------------------|-----------------------|----------------------------|--| | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | | | | Clinical Application | Mode of Operation | | | | | | | | | | | General<br>(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other<br>(Specify) | | | Ophthalmic | Ophthalmic | | | | | | | | | | | Fetal Imaging<br>& Other | Fetal | N | N | N | | N | N | N | Note 1, 2, 4, 6,10 | | | | Abdominal | N | N | N | | N | N | N | Note 1, 2, 4, 6,10 | | | | Intra-operative<br>(Specify*) | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ (Specify**) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | Trans-vaginal | | | | | | | | | | | | Trans-urethral | | | | | | | | | | | | Trans-esoph. (non-Card.) | | | | | | | | | | | | Musculo-skeletal<br>(Conventional) | | | | | | | | | | | | Musculo-skeletal<br>(Superficial) | | | | | | | | | | | | Intravascular | | | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | Peripheral<br>vessel | Peripheral vessel | | | | | | | | | | | | Other (Specify***) | | | | | | | | | | | N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | | | | | Additional comments: Combined modes--B+M、PW+B、Color + B、Power+B、PW+Color+B、Power + PW +B. | | | | | | | | | | | | *Intraoperative includes abdominal, thoracic, and vascular. | | | | | | | | | | | | ** Small organ-breast, thyroid, testes. | | | | | | | | | | | | ***Other use includes Urology. | | | | | | | | | | | | Note 1: Tissue Harmonic Imaging. | | | | | | | | | | | | Note 2: Smart3D | | | | | | | | | | | | Note 3:4D(Real-time 3D) | | | | | | | | | | | | Note 4: iScape | | | | | | | | | | | | Note 5: TDI | | | | | | | | | | | | Note 6: Color M | | | | | | | | | | | | Note 7: Biopsy Guidance | | | | | | | | | | | | Note 8: Elastography | | | | | | | | | | | | Note 9: Contrast imaging (Contrast agent for LVO) | | | | | | | | | | | | Note 10: Ultrasound Fusion Imaging | | | | | | | | | | | | Note 11: V Flow | | | | | | | | | | | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | | | | | | | | | | | | Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) | | | | | | | | | | | | System: | Resona 7/Resona 7CV/Resona 7EXP/Resona 7S/Resona 70B Diagnostic Ultrasound System | | | | | | | | | | | Transducer: | V11-3HU | | | | | | | | | | | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | | | | Clinical Application | Mode of Operation | | | | | | | | | | | General<br>(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other<br>(Specify) | | | Ophthalmic | Ophthalmic | | | | | | | | | | | Fetal Imaging<br>& Other | Fetal | N | N | N | N | N | N | N | Note 1, 2, 4, 6, 7 | | | | Abdominal | | | | | | | | | | | | Intra-operative<br>(Specify*) | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ (Specify**) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Trans-rectal | N | N | N | N | N | N | N | Note 1, 2, 4, 6, 7 | | | | Trans-vaginal | N | N | N | N | N | N | N | Note 1, 2, 4, 6, 7 | | | | Trans-urethral | | | | | | | | | | | | Trans-esoph. (non-<br>Card.) | | | | | | | | | | | | Musculo-skeletal<br>(Conventional) | | | | | | | | | | | | Musculo-skeletal<br>(Superficial) | | | | | | | | | | | | Intravascular | | | | | | | | | | | Cardiac | Cardiac Adult | | | | | | | | | | | | Cardiac Pediatric | | | | | | | | | | | | Intravascular (Cardiac) | | | | | | | | | | | | Trans-esoph. (Cardiac) | | | | | | | | | | | | Intra-cardiac | | | | | | | | | | | Peripheral<br>vessel | Peripheral vessel | | | | | | | | | | | | Other (Specify***) | N | N | N | N | N | N | N | Note 1, 2, 4, 6, 7 | | | | N=new indication; P=previously cleared by FDA; E=added under Appendix E | | | | | | | | | | | | Additional comments: Combined modes--B+M、PW+B、Color + B、Power + B、PW+Color+B、Power + PW+B. | | | | | | | | | | | | *Intraoperative includes abdominal, thoracic, and vascular. | | | | | | | | | | | | **Small organ-breast, thyroid, testes. | | | | | | | | | | | | ***Other use includes Urology. | | | | | | | | | | | | Note 1: Tissue Harmonic Imaging. | | | | | | | | | | | | Note 2: Smart3D | | | | | | | | | | | | Note 3:4D(Real-time 3D) | | | | | | | | | | | | Note 4: iScape | | | | | | | | | | | | Note 5: TDI | | | | | | | | | | | | Note 6: Color M | | | | | | | | | | | | Note 7: Biopsy Guidance | | | | | | | | | | | | Note 8: Elastography | | | | | | | | | | | | Note 9: Contrast imaging (Contrast agent for LVO) | | | | | | | | | | | | Note 10: Ultrasound Fusion Imaging | | | | | | | | | | | | Note 11: V Flow | | | | | | | | | | | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE NEEDED) | | | | | | | | | | | | Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR) | | | | | | | | | | | System: | Resona 7/Resona 7CV/Resona 7EXP/Resona 7S/Resona 7OB Diagnostic Ultrasound System | | | | | | | | | | | Transducer: | D8-4U | | | | | | | | | | | | Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: | | | | | | | | | | | | Clinical Application<br>Mode of Operation | | | | | | | | | | | General<br>(Track 1 | Specific (Track 1 & 3) | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Combined<br>(specify) | Other<br>(Specify) | | | Ophthalmic | Ophthalmic | | | | | | | | | | | Fetal Imaging<br>& Other | Fetal | N | N | N | N | N | N | N | Note 1, 3, 4, 6 | | | & Other | Abdominal | N | N | N | N | N | N | N | Note 1, 3, 4, 6 | | | | Intra-operative<br>(Specify*) | | | | | | | | | | | | Intra-operative (Neuro) | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Pediatric | | | | | | | | | | | | Small Organ (Specify**) | | | | | | | | | | | | Neonatal Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Trans-rectal | | | | | | | | | | | | Trans-vaginal | | | | | | | |…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...