← Product Code [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN) · K151451

# Aplio 500/400/300 Diagnostic Ultrasound System V6.0 (K151451)

_Toshibamedical Systems Corporation · IYN · Jul 9, 2015 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K151451

## Device Facts

- **Applicant:** Toshibamedical Systems Corporation
- **Product Code:** [IYN](/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN.md)
- **Decision Date:** Jul 9, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The Diagnostic Ultrasound System Aplio 500 Model TUS-A500, Aplio 400 Model TUS-A400 And Aplio 300 Model TUS-A300 is indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diagnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small organs, trans-vaginal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophageal, musculoskeletal (both conventional and superficial) and laparoscopic.

## Device Story

Mobile diagnostic ultrasound system; uses wide array of probes (linear, convex, sector) with 2-12 MHz frequency range. Inputs ultrasound signals; transforms into diagnostic images/data. Used in clinical settings by trained professionals. Features include Shear Wave, Smart Navigation, SMI, Auto-NT, and Smart 3D. Output displayed on system monitor for clinician diagnosis. Enhances diagnostic accuracy for structures and dynamic processes; supports clinical decision-making through improved imaging and automated measurement tools.

## Clinical Evidence

Two clinical evaluations conducted. One study obtained representative clinical images of volunteers to confirm system provides acceptable images and numerical Shear wave results in vivo. Second study validated Auto-NT feature performance by comparing automated Nuchal Translucency thickness measurements against manual measurements.

## Technological Characteristics

Mobile ultrasound system; linear, convex, and sector array transducers (2-12 MHz). Conforms to IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, IEC 62304, and ISO 10993-1. Software-based image processing and analysis.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- Aplio 500/400/300 Diagnostic Ultrasound System, V5.0 ([K141459](/device/K141459.md))
- Aplio XG Diagnostic Ultrasound System, Model SSA-790A version 2.0 ([K072000](/device/K072000.md))
- Voluson i/e ([K131937](/device/K131937.md))
- LOGIQ E9 Diagnostic Ultrasound System ([K142160](/device/K142160.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 9, 2015

Toshiba Medical Systems Corporation % Mr. Orlando Tadeo Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780

Re: K151451

Trade/Device Name: Aplio 500/400/300 Diagnostic Ultrasound System, V6.0 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: May 26, 2015 Received: May 29, 2015

Dear Mr. Tadeo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D'Hara

For

Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

## 510(k) Number (if known)

K151451

### Device Name

Aplio 500/400/300 Diagnostic Ultrasound System, V6.0

### Indications for Use (Describe)

The Diaqnostic Ultrasound System Aplio 500 Model TUS-A500, Aplio 400 Model TUS-A400 And Aplio 300 Model TUS-A300 is indicated for the visualization of structures, and dynamic processes with the human body using ultrasound and to provide image information for diaqnosis in the following clinical applications: fetal, abdominal, intra-operative (abdominal), pediatric, small orqans, trans-vaqinal, trans-rectal, neonatal cephalic, adult cephalic, cardiac (both adult and pediatric), peripheral vascular, transesophaqeal, musculo-skeletal (both conventional and superficial) and laparoscopic.

Type of Use (Select one or both, as applicable)

ك Prescription Use (Part 21 CFR 801 Subpart D)

- Over-The-Counter Use (21 CFR 801 Subpart C)

## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1

## System: Aplio 500, Aplio 400, Aplio 300 V6.0

Transducer:

------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Clinical Application Mode of Oneration

| Clinical Application<br>Specific<br>(Tracks 3) | B | M | PW<br>D | CW<br>D | Color<br>Doppler | Combined<br>(Specify) * | THI | Dynamic<br>Flow | Power | CHI<br>2D | 4D | Other<br>[Note]                            |
|------------------------------------------------|---|---|---------|---------|------------------|-------------------------|-----|-----------------|-------|-----------|----|--------------------------------------------|
| Ophthalmic                                     |   |   |         |         |                  |                         |     |                 |       |           |    |                                            |
| Fetal                                          | P | P | P       | P       | P                | 2                       | P   | P               | P     |           | P  | 5,7,8, 9,10,14,18,19                       |
| Abdominal                                      | P | P | P       | P       | P                | 2,3                     | P   | P               | P     |           | P  | 5,7,8, 9,10,11,12,14,<br>15,16,18,19       |
| Intra-operative (Abdominal)                    | P | P | P       | P       | P                | 2                       | P   | P               | P     |           |    | 4,5,7,11,19                                |
| Intra-operative (Neuro)                        |   |   |         |         |                  |                         |     |                 |       |           |    |                                            |
| Laparoscopic                                   | P | P | P       | P       | P                | 2                       | P   | P               | P     |           |    | 5,7                                        |
| Pediatric                                      | P | P | P       | P       | P                | 2,3                     | P   | P               | P     |           | P  | 5,7,8,9,10,12,14,15,<br>18,19              |
| Small Organ (Note 1)                           | P | P | P       | P       | P                | 2                       | P   | P               | P     |           |    | 4,5,6,7,8,9,10,11,14,<br>15,16(N),17,18,19 |
| Neonatal Cephalic                              | P | P | P       | P       | P                | 3                       | P   | P               | P     |           |    |                                            |
| Adult Cephalic                                 | P | P | P       | P       | P                | 3                       | P   | P               | P     |           |    |                                            |
| Trans-rectal                                   | P | P | P       | P       | P                | 2                       | P   | P               | P     |           | P  | 4,5,7,11,12,18                             |
| Trans-vaginal                                  | P | P | P       | P       | P                | 2                       | P   | P               | P     |           | P  | 4,5,7,11,12,18                             |
| Trans-urethral                                 |   |   |         |         |                  |                         |     |                 |       |           |    |                                            |
| Trans-esoph. (non-Card.)                       |   |   |         |         |                  |                         |     |                 |       |           |    |                                            |
| Musculo-skeletal (Conventional)                | P | P | P       | P       | P                | 2                       | P   | P               | P     |           |    | 4,5,6,7,8,9,10,11,14,<br>15,17,18,19       |
| Musculo-skeletal (Superficial)                 | P | P | P       | P       | P                | 2                       | P   | P               | P     |           |    | 4,5,6,7,8,9,10,11,14,<br>15,17,18,19       |
| Intravascular                                  |   |   |         |         |                  |                         |     |                 |       |           |    |                                            |
| Other (Specify)                                |   |   |         |         |                  |                         |     |                 |       |           |    |                                            |
| Cardiac Adult                                  | P | P | P       | P       | P                | 3                       | P   | P               | P     | P         |    | 4,13                                       |
| Cardiac Pediatric                              | P | P | P       | P       | P                | 3                       | P   | P               | P     | P         |    | 4,13                                       |
| Intravascular (Cardiac)                        |   |   |         |         |                  |                         |     |                 |       |           |    |                                            |
| Trans-esoph. (Cardiac)                         | P | P | P       | P       | P                | 3                       | P   |                 |       |           |    | 4,13                                       |
| Intra-cardiac                                  |   |   |         |         |                  |                         |     |                 |       |           |    |                                            |
| Other (Specify)                                |   |   |         |         |                  |                         |     |                 |       |           |    |                                            |
| Peripheral vessel                              | P | P | P       | P       | P                | 2                       | P   | P               | P     |           |    | 4,5,6,7,8,9,10,11,14,<br>15,17,18,19       |
| Other (Specify)                                |   |   |         |         |                  |                         |     |                 |       |           |    |                                            |

N = new indication; P = previously cleared by FDA; E = added under this appendix

Previous 510(k) of the transducers: K141459, K133761, K123992, K121422, K103629

| Note 1 Small organ includes thyroid, breast and testicle                                    | Note 7 Precision Imaging       | Note 13 2D WMT           |
|---------------------------------------------------------------------------------------------|--------------------------------|--------------------------|
| Note 2 Combined mode includes B/M; B/PWD; BDF/PWD;<br>BDF/MDF; BDF/MDF/PWD                  | Note 8 STIC                    | Note 14 Boost            |
| Note 3 Combined mode includes B/M; B/PWD; BDF/PWD;<br>BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD | Note 9 3D Color (Volume Color) | Note 15 SMI              |
| Note 4 TDI                                                                                  | Note 10 STIC Color             | Note 16 Shear wave       |
| Note 5 ApliPure                                                                             | Note 11 Elastography           | Note 17 BEAM             |
| Note 6 MicroPure                                                                            | Note 12 Fusion                 | Note 18 Smart Navigation |
| Note 19 Smart 3D                                                                            |                                |                          |

{4}------------------------------------------------

### System:

Transducer: PST-25BT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application            | Mode of Operation |   |     |     |                  |                            |     |                 |       |           |      |       |        |
|---------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------|-------|-----------|------|-------|--------|
| Specific<br>(Tracks 3)          | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Dynamic<br>Flow | Power | CHI<br>2D | 4D   | Other | [Note] |
| Ophthalmic                      |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Fetal                           |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Abdominal                       | P                 | P | P   | P   | P                | 3                          | P   | P               | P     |           | 11   |       |        |
| Intra-operative (Abdominal)     |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Intra-operative (Neuro)         |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Laparoscopic                    |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Pediatric                       | P                 | P | P   | P   | P                | 3                          | P   | P               | P     |           |      |       |        |
| Small Organ (Specify) (1)       |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Neonatal Cephalic               | P                 | P | P   | P   | P                | 3                          | P   | P               | P     |           |      |       |        |
| Adult Cephalic                  | P                 | P | P   | P   | P                | 3                          | P   | P               | P     |           |      |       |        |
| Trans-rectal                    |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Trans-vaginal                   |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Trans-urethral                  |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Trans-esoph. (non-Card.)        |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Musculo-skeletal (Conventional) |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Musculo-skeletal (Superficial)  |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Intravascular                   |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Other (Specify)                 |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Cardiac Adult                   | P                 | P | P   | P   | P                | 3                          | P   | P               | P     | P         | 4,13 |       |        |
| Cardiac Pediatric               | P                 | P | P   | P   | P                | 3                          | P   | P               | P     | P         | 4,13 |       |        |
| Intravascular (Cardiac)         |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Trans-esoph. (Cardiac)          |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Intra-cardiac                   |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Other (Specify)                 |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Peripheral vessel               |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |
| Other (Specify)                 |                   |   |     |     |                  |                            |     |                 |       |           |      |       |        |

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K141459

Note 7 Precision Imaging Note 13 2D WMT Note 1 Small organ includes thyroid, breast and testicle Note 2 Combined mode includes B/M; B/PWD; BDF/PWD; Note 8 STIC Note 14 Boost BDF/MDF; BDF/MDF/PWD Note 3 Combined mode includes B/M; B/PWD; BDF/PWD; Note 9 3D Color (Volume Color) Note 15 SMI BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD Note 4 TDI Note 10 STIC Color Note 16 Shear wave Note 5 ApliPure Note 11 Elastography Note 17 BEAM Note 12 Fusion Note 18 Smart Navigation Note 6 MicroPure Note 19 Smart 3D

{5}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Specific<br>(Tracks 3) | Mode of Operation | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Dynamic<br>Flow | Power | CHI<br>2D | 4D | Other | [Note] |
|------------------------------------------------|-------------------|---|---|-----|-----|------------------|----------------------------|-----|-----------------|-------|-----------|----|-------|--------|
| Ophthalmic                                     |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Fetal                                          |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Abdominal                                      |                   | P | P | P   | P   | P                | 3                          | P   | P               | P     |           |    | 11    |        |
| Intra-operative (Abdominal)                    |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Intra-operative (Neuro)                        |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Laparoscopic                                   |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Pediatric                                      |                   | P | P | P   | P   | P                | 3                          | P   | P               | P     |           |    |       |        |
| Small Organ (Specify) (1)                      |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Neonatal Cephalic                              |                   | P | P | P   | P   | P                | 3                          | P   | P               | P     |           |    |       |        |
| Adult Cephalic                                 |                   | P | P | P   | P   | P                | 3                          | P   | P               | P     |           |    |       |        |
| Trans-rectal                                   |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Trans-vaginal                                  |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Trans-urethral                                 |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Trans-esoph. (non-Card.)                       |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Musculo-skeletal (Conventional)                |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Musculo-skeletal (Superficial)                 |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Intravascular                                  |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Other (Specify)                                |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Cardiac Adult                                  |                   | P | P | P   | P   | P                | 3                          | P   | P               | P     | P         |    | 4,13  |        |
| Cardiac Pediatric                              |                   | P | P | P   | P   | P                | 3                          | P   | P               | P     | P         |    | 4,13  |        |
| Intravascular (Cardiac)                        |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Trans-esoph. (Cardiac)                         |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Intra-cardiac                                  |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Other (Specify)                                |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Peripheral vessel                              |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Other (Specify)                                |                   |   |   |     |     |                  |                            |     |                 |       |           |    |       |        |

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K141459

| Note 1  | Small organ includes thyroid, breast and testicle                                 |
|---------|-----------------------------------------------------------------------------------|
| Note 2  | Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD                  |
| Note 3  | Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD |
| Note 4  | TDI                                                                               |
| Note 5  | ApliPure                                                                          |
| Note 6  | MicroPure                                                                         |
| Note 7  | Precision Imaging                                                                 |
| Note 8  | STIC                                                                              |
| Note 9  | 3D Color (Volume Color)                                                           |
| Note 10 | STIC Color                                                                        |
| Note 11 | Elastography                                                                      |
| Note 12 | Fusion                                                                            |
| Note 13 | 2D WMT                                                                            |
| Note 14 | Boost                                                                             |
| Note 15 | SMI                                                                               |
| Note 16 | Shear wave                                                                        |
| Note 17 | BEAM                                                                              |
| Note 18 | Smart Navigation                                                                  |
| Note 19 | Smart 3D                                                                          |

{6}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application            | Mode of Operation |   |     |     |                  |                            |     |                 |       |           |    |       |        |
|---------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------|-------|-----------|----|-------|--------|
| Specific<br>(Tracks 3)          | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Dynamic<br>Flow | Power | CHI<br>2D | 4D | Other | [Note] |
| Ophthalmic                      |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Fetal                           |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Abdominal                       | P                 | P | P   | P   | P                | 3                          | P   | P               | P     |           |    | 11    |        |
| Intra-operative (Abdominal)     |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Intra-operative (Neuro)         |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Laparoscopic                    |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Pediatric                       | P                 | P | P   | P   | P                | 3                          | P   | P               | P     |           |    |       |        |
| Small Organ (Specify) (1)       |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Neonatal Cephalic               | P                 | P | P   | P   | P                | 3                          | P   | P               | P     |           |    |       |        |
| Adult Cephalic                  | P                 | P | P   | P   | P                | 3                          | P   | P               | P     |           |    |       |        |
| Trans-rectal                    |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Trans-vaginal                   |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Trans-urethral                  |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Trans-esoph. (non-Card.)        |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Musculo-skeletal (Conventional) |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Musculo-skeletal (Superficial)  |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Intravascular                   |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Other (Specify)                 |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Cardiac Adult                   | P                 | P | P   | P   | P                | 3                          | P   | P               | P     | P         |    | 4,13  |        |
| Cardiac Pediatric               | P                 | P | P   | P   | P                | 3                          | P   | P               | P     | P         |    | 4,13  |        |
| Intravascular (Cardiac)         |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Trans-esoph. (Cardiac)          |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Intra-cardiac                   |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Other (Specify)                 |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Peripheral vessel               |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Other (Specify)                 |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K141459

| Note 1  | Small organ includes thyroid, breast and testicle                                 |
|---------|-----------------------------------------------------------------------------------|
| Note 2  | Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD                  |
| Note 3  | Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD |
| Note 4  | TDI                                                                               |
| Note 5  | ApliPure                                                                          |
| Note 6  | MicroPure                                                                         |
| Note 7  | Precision Imaging                                                                 |
| Note 8  | STIC                                                                              |
| Note 9  | 3D Color (Volume Color)                                                           |
| Note 10 | STIC Color                                                                        |
| Note 11 | Elastography                                                                      |
| Note 12 | Fusion                                                                            |
| Note 13 | 2D WMT                                                                            |
| Note 14 | Boost                                                                             |
| Note 15 | SMI                                                                               |
| Note 16 | Shear wave                                                                        |
| Note 17 | BEAM                                                                              |
| Note 18 | Smart Navigation                                                                  |
| Note 19 | Smart 3D                                                                          |

{7}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application            | Mode of Operation |   |     |     |                  |                        |     |                 |       |           |    |       |        |
|---------------------------------|-------------------|---|-----|-----|------------------|------------------------|-----|-----------------|-------|-----------|----|-------|--------|
| Specific<br>(Tracks 3)          | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)* | THI | Dynamic<br>Flow | Power | CHI<br>2D | 4D | Other | [Note] |
| Ophthalmic                      |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Fetal                           |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Abdominal                       | P                 | P | P   | P   | P                | 3                      | P   | P               | P     |           |    | 11    |        |
| Intra-operative (Abdominal)     |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Intra-operative (Neuro)         |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Laparoscopic                    |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Pediatric                       | P                 | P | P   | P   | P                | 3                      | P   | P               | P     |           |    |       |        |
| Small Organ (Specify) (1)       |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Neonatal Cephalic               | P                 | P | P   | P   | P                | 3                      | P   | P               | P     |           |    |       |        |
| Adult Cephalic                  | P                 | P | P   | P   | P                | 3                      | P   | P               | P     |           |    |       |        |
| Trans-rectal                    |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Trans-vaginal                   |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Trans-urethral                  |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Trans-esoph. (non-Card.)        |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Musculo-skeletal (Conventional) |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Musculo-skeletal (Superficial)  |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Intravascular                   |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Other (Specify)                 |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Cardiac Adult                   | P                 | P | P   | P   | P                | 3                      | P   | P               | P     | P         |    | 4,13  |        |
| Cardiac Pediatric               | P                 | P | P   | P   | P                | 3                      | P   | P               | P     | P         |    | 4,13  |        |
| Intravascular (Cardiac)         |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Trans-esoph. (Cardiac)          |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Intra-cardiac                   |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Other (Specify)                 |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Peripheral vessel               |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |
| Other (Specify)                 |                   |   |     |     |                  |                        |     |                 |       |           |    |       |        |

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K141459

| Note 1  | Small organ includes thyroid, breast and testicle                                 |
|---------|-----------------------------------------------------------------------------------|
| Note 2  | Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD                  |
| Note 3  | Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD |
| Note 4  | TDI                                                                               |
| Note 5  | ApliPure                                                                          |
| Note 6  | MicroPure                                                                         |
| Note 7  | Precision Imaging                                                                 |
| Note 8  | STIC                                                                              |
| Note 9  | 3D Color (Volume Color)                                                           |
| Note 10 | STIC Color                                                                        |
| Note 11 | Elastography                                                                      |
| Note 12 | Fusion                                                                            |
| Note 13 | 2D WMT                                                                            |
| Note 14 | Boost                                                                             |
| Note 15 | SMI                                                                               |
| Note 16 | Shear wave                                                                        |
| Note 17 | BEAM                                                                              |
| Note 18 | Smart Navigation                                                                  |
| Note 19 | Smart 3D                                                                          |

{8}------------------------------------------------

### System: _Aplio 500, Aplio 400, Aplio 300 V6.0

Transducer:___PVT-375BT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application            | Mode of Operation |   |     |     |                  |                            |     |                 |       | Other [Note] |           |                       |
|---------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------|-------|--------------|-----------|-----------------------|
| Specific<br>(Tracks 3)          | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Dynamic<br>Flow | Power |              | CHI<br>2D | 4D                    |
| Ophthalmic                      |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Fetal                           | P                 | P | P   | P   | P                | 2                          | P   | P               | P     |              |           | 5,7,19                |
| Abdominal                       | P                 | P | P   | P   | P                | 2                          | P   | P               | P     |              |           | 5,7,11,12,15,16,18,19 |
| Intra-operative (Abdominal)     |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Intra-operative (Neuro)         |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Laparoscopic                    |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Pediatric                       | P                 | P | P   | P   | P                | 2                          | P   | P               | P     |              |           | 5, 7,12,15,18,19      |
| Small Organ (Specify) (1)       |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Neonatal Cephalic               |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Adult Cephalic                  |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Trans-rectal                    |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Trans-vaginal                   |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Trans-urethral                  |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Trans-esoph. (non-Card.)        |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Musculo-skeletal (Conventional) |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Musculo-skeletal (Superficial)  |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Intravascular                   |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Other (Specify)                 |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Cardiac Adult                   |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Cardiac Pediatric               |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Intravascular (Cardiac)         |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Trans-esoph. (Cardiac)          |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Intra-cardiac                   |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Other (Specify)                 |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Peripheral vessel               |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |
| Other (Specify)                 |                   |   |     |     |                  |                            |     |                 |       |              |           |                       |

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K141459

| Note 1 | Small organ includes thyroid, breast and testicle |
|--------|---------------------------------------------------|
|--------|---------------------------------------------------|

| Note 2 | Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD |
|--------|------------------------------------------------------------------|
|--------|------------------------------------------------------------------|

| Note 3 | Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD |
|--------|-----------------------------------------------------------------------------------|
|--------|-----------------------------------------------------------------------------------|

| Note 4 | TDI |
|--------|-----|
|--------|-----|

| Note 5 | ApliPure |
|--------|----------|
|--------|----------|

| Note 6 | MicroPure |
|--------|-----------|
|--------|-----------|

| Note 7 | Precision Imaging |
|--------|-------------------|
|--------|-------------------|

| Note 8 | STIC |
|--------|------|
|--------|------|

| Note 9 | 3D Color (Volume Color) |
|--------|-------------------------|
|--------|-------------------------|

| Note 10 | STIC Color |
|---------|------------|
|---------|------------|

| Note 11 | Elastography |
|---------|--------------|
|---------|--------------|

| Note 12 | Fusion |
|---------|--------|
|---------|--------|

| Note 13 | 2D WMT |
|---------|--------|
|---------|--------|

| Note 14 | Boost |
|---------|-------|
|---------|-------|

| Note 15 | SMI |
|---------|-----|
|---------|-----|

| Note 16 | Shear wave |
|---------|------------|
|---------|------------|

| Note 17 | BEAM |
|---------|------|
|---------|------|

| Note 18 | Smart Navigation |
|---------|------------------|
|---------|------------------|

| Note 19 | Smart 3D |
|---------|----------|
|---------|----------|

{9}------------------------------------------------

### System: _Aplio 500, Aplio 400, Aplio 300 V6.0

Transducer: _ PVT-375SC

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Specific<br>(Tracks 3) | Mode of Operation |   |     |     |                  |                            |     |                 |             | Other [Note] |    |                       |
|------------------------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------|-------------|--------------|----|-----------------------|
|                                                | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Dynamic<br>Flow | Power<br>2D | CHI          | 4D |                       |
| Ophthalmic                                     |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Fetal                                          | P                 | P | P   |     | P                | 2                          | P   | P               | P           |              |    | 5,7,19                |
| Abdominal                                      | P                 | P | P   |     | P                | 2                          | P   | P               | P           |              |    | 5,7,11,12,15,16,18,19 |
| Intra-operative (Abdominal)                    |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Intra-operative (Neuro)                        |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Laparoscopic                                   |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Pediatric                                      | P                 | P | P   |     | P                | 2                          | P   | P               | P           |              |    | 5,7,12,15,18,19       |
| Small Organ (Specify) (1)                      |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Neonatal Cephalic                              |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Adult Cephalic                                 |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Trans-rectal                                   |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Trans-vaginal                                  |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Trans-urethral                                 |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Trans-esoph. (non-Card.)                       |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Musculo-skeletal<br>(Conventional)             |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Musculo-skeletal (Superficial)                 |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Intravascular                                  |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Other (Specify)                                |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Cardiac Adult                                  |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Cardiac Pediatric                              |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Intravascular (Cardiac)                        |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Trans-esoph. (Cardiac)                         |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Intra-cardiac                                  |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Other (Specify)                                |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Peripheral vessel                              |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |
| Other (Specify)                                |                   |   |     |     |                  |                            |     |                 |             |              |    |                       |

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K141459

| Note 1  | Small organ includes thyroid, breast and testicle                                 |
|---------|-----------------------------------------------------------------------------------|
| Note 2  | Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD                  |
| Note 3  | Combined mode includes B/M; B/PWD; BDF/PWD; BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD |
| Note 4  | TDI                                                                               |
| Note 5  | ApliPure                                                                          |
| Note 6  | MicroPure                                                                         |
| Note 7  | Precision Imaging                                                                 |
| Note 8  | STIC                                                                              |
| Note 9  | 3D Color (Volume Color)                                                           |
| Note 10 | STIC Color                                                                        |
| Note 11 | Elastography                                                                      |
| Note 12 | Fusion                                                                            |
| Note 13 | 2D WMT                                                                            |
| Note 14 | Boost                                                                             |
| Note 15 | SMI                                                                               |
| Note 16 | Shear wave                                                                        |
| Note 17 | BEAM                                                                              |
| Note 18 | Smart Navigation                                                                  |
| Note 19 | Smart 3D                                                                          |

{10}------------------------------------------------

### System:

Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application<br>Specific<br>(Tracks 3) | Mode of Operation | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Dynamic<br>Flow | Power | CHI<br>2D | 4D | Other [Note] |
|------------------------------------------------|-------------------|---|---|-----|-----|------------------|----------------------------|-----|-----------------|-------|-----------|----|--------------|
| Ophthalmic                                     |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Fetal                                          |                   | P | P | P   |     | P                | 2                          | P   | P               | P     |           | P  | 5,7,8,9,10   |
| Abdominal                                      |                   | P | P | P   |     | P                | 2                          | P   | P               | P     |           | P  | 5,7,8,9,10   |
| Intra-operative (Abdominal)                    |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Intra-operative (Neuro)                        |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Laparoscopic                                   |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Pediatric                                      |                   | P | P | P   |     | P                | 2                          | P   | P               | P     |           | P  | 5,7,8,9,10   |
| Small Organ (Specify) (1)                      |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Neonatal Cephalic                              |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Adult Cephalic                                 |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Trans-rectal                                   |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Trans-vaginal                                  |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Trans-urethral                                 |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Trans-esoph. (non-Card.)                       |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Musculo-skeletal (Conventional)                |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Musculo-skeletal (Superficial)                 |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Intravascular                                  |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Other (Specify)                                |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Cardiac Adult                                  |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Cardiac Pediatric                              |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Intravascular (Cardiac)                        |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Trans-esoph. (Cardiac)                         |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Intra-cardiac                                  |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Other (Specify)                                |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Peripheral vessel                              |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |
| Other (Specify)                                |                   |   |   |     |     |                  |                            |     |                 |       |           |    |              |

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K141459

| Note 1 Small organ includes thyroid, breast and testicle                                    | Note 7 Precision Imaging       | Note 13 2D WMT           |
|---------------------------------------------------------------------------------------------|--------------------------------|--------------------------|
| Note 2 Combined mode includes B/M; B/PWD; BDF/PWD;<br>BDF/MDF; BDF/MDF/PWD                  | Note 8 STIC                    | Note 14 Boost            |
| Note 3 Combined mode includes B/M; B/PWD; BDF/PWD;<br>BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD | Note 9 3D Color (Volume Color) | Note 15 SMI              |
| Note 4 TDI                                                                                  | Note 10 STIC Color             | Note 16 Shear wave       |
| Note 5 ApliPure                                                                             | Note 11 Elastography           | Note 17 BEAM             |
| Note 6 MicroPure                                                                            | Note 12 Fusion                 | Note 18 Smart Navigation |
| Note 19 Smart 3D                                                                            |                                |                          |

{11}------------------------------------------------

### System: _Aplio 500, Aplio 400, Aplio 300 V6.0

Transducer:___PVT-382BT

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application            | Mode of Operation |   |     |     |                  |                            |     |                 |       |           |    | Other | [Note]        |
|---------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------|-------|-----------|----|-------|---------------|
| Specific<br>(Tracks 3)          | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Dynamic<br>Flow | Power | CHI<br>2D | 4D |       |               |
| Ophthalmic                      |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Fetal                           | P                 | P | P   |     | P                | 2                          | P   | P               | P     |           |    |       | 5, 7,18,19    |
| Abdominal                       | P                 | P | P   |     | P                | 2                          | P   | P               | P     |           |    |       | 5, 7,12,18,19 |
| Intra-operative (Abdominal)     |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Intra-operative (Neuro)         |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Laparoscopic                    |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Pediatric                       | P                 | P | P   |     | P                | 2                          | P   | P               | P     |           |    |       | 5, 7,12,18,19 |
| Small Organ (Specify) (1)       |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Neonatal Cephalic               |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Adult Cephalic                  |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Trans-rectal                    |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Trans-vaginal                   |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Trans-urethral                  |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Trans-esoph. (non-Card.)        |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Musculo-skeletal (Conventional) |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Musculo-skeletal (Superficial)  |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Intravascular                   |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Other (Specify)                 |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Cardiac Adult                   |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Cardiac Pediatric               |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Intravascular (Cardiac)         |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Trans-esoph. (Cardiac)          |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Intra-cardiac                   |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Other (Specify)                 |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Peripheral vessel               |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |
| Other (Specify)                 |                   |   |     |     |                  |                            |     |                 |       |           |    |       |               |

N = new indication; P = previously cleared by FDA; E = added under this appendix Previous 510(k) of the transducer: K141459

| Note 1 Small organ includes thyroid, breast and testicle                                    | Note 7 Precision Imaging       | Note 13 2D WMT           |
|---------------------------------------------------------------------------------------------|--------------------------------|--------------------------|
| Note 2 Combined mode includes B/M; B/PWD; BDF/PWD;<br>BDF/MDF; BDF/MDF/PWD                  | Note 8 STIC                    | Note 14 Boost            |
| Note 3 Combined mode includes B/M; B/PWD; BDF/PWD;<br>BDF/MDF; BDF/MDF/PWD; 2D/CWD; BDF/CWD | Note 9 3D Color (Volume Color) | Note 15 SMI              |
| Note 4 TDI                                                                                  | Note 10 STIC Color             | Note 16 Shear wave       |
| Note 5 ApliPure                                                                             | Note 11 Elastography           | Note 17 BEAM             |
| Note 6 MicroPure                                                                            | Note 12 Fusion                 | Note 18 Smart Navigation |
| Note 19 Smart 3D                                                                            |                                |                          |

{12}------------------------------------------------

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application            | Mode of Operation |   |     |     |                  |                            |     |                 |       |           |    |       |        |
|---------------------------------|-------------------|---|-----|-----|------------------|----------------------------|-----|-----------------|-------|-----------|----|-------|--------|
| Specific<br>(Tracks 3)          | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(Specify)<br>* | THI | Dynamic<br>Flow | Power | CHI<br>2D | 4D | Other | [Note] |
| Ophthalmic                      |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Fetal                           | P                 | P | P   |     | P                | 2                          | P   | P               | P     |           | P  |       | 5,7,9  |
| Abdominal                       | P                 | P | P   |     | P                | 2                          | P   | P               | P     |           | P  |       | 5,7,9  |
| Intra-operative (Abdominal)     |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Intra-operative (Neuro)         |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Laparoscopic                    |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Pediatric                       | P                 | P | P   |     | P                | 2                          | P   | P               | P     |           | P  |       | 5,7,9  |
| Small Organ (Specify) (1)       |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Neonatal Cephalic               |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Adult Cephalic                  |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Trans-rectal                    |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Trans-vaginal                   |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Trans-urethral                  |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Trans-esoph. (non-Card.)        |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Musculo-skeletal (Conventional) |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Musculo-skeletal (Superficial)  |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Intravascular                   |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Other (Specify)                 |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Cardiac Adult                   |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Cardiac Pediatric               |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Intravascular (Cardiac)         |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Trans-esoph. (Cardiac)          |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Intra-cardiac                   |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Other (Specify)                 |                   |   |     |     |                  |                            |     |                 |       |           |    |       |        |
| Peripheral vessel               |                   |   |     |…

---

**Source:** [https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K151451](https://fda.innolitics.com/submissions/RA/subpart-b%E2%80%94diagnostic-devices/IYN/K151451)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
